← Product Code [MNW](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MNW) · K014009

# TANITA BODY COMPOSITION ANALYZER; TANITA BODY FAT MONITOR; TANITA ULTIMATE SCALE (K014009)

_Tanita Corp. of America · MNW · Jun 24, 2002 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MNW/K014009

## Device Facts

- **Applicant:** Tanita Corp. of America
- **Product Code:** [MNW](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MNW.md)
- **Decision Date:** Jun 24, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2770
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

Intended to be used as a body fat analyzer that determines body weight and estimates body fat composition with the use of BIA (bioelectrical impedance analysis).

## Device Story

Device consists of platform scale with four stainless steel footpad electrodes; subject stands on scale to measure body weight and lower extremity impedance. System utilizes bioelectrical impedance analysis (BIA) to estimate body composition parameters. Used in professional and consumer settings; operated by clinicians or patients. Output displayed to user to assist in monitoring body composition and weight management. Benefits include non-invasive assessment of body fat and related metrics.

## Clinical Evidence

No clinical data provided; substantial equivalence based on bench testing and comparison of BIA methodology to predicate.

## Technological Characteristics

Platform scale with stainless steel footpad electrodes; utilizes 'foot-to-foot' bioelectrical impedance analysis (BIA).

## Regulatory Identification

An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- TANITA Body Fat Analyzer Model TBF-105 ([K930599](/device/K930599.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92

| Trade Name:          | TANITA Body Fat Analyzer TBF/BF/ Ultimate Scale                                                |
|----------------------|------------------------------------------------------------------------------------------------|
| Common Name:         | Body Fat Analyzer/ Body Fat Monitors/ Body Composition Analyzers/<br>Body Composition Monitors |
| Classification Name: | ANALYZER, BODY COMPOSITION<br>21 CFR § 870.2770                                                |

## Description of Applicant Device:

The TANITA Body Fat Monitor/Scales (multiple models) are designed to determine body weight and body fat composition. Each model will offer a range of features. The system consists of two subdivided stainless steel footpad electrodes mounted on a platform scale. The four-foot electrodes are designed where each foot makes contact with a heel and toe electrode. Impedance of the lower extremities and body weight is measured simultaneously while the subject stands on the scale.

## Intended Uses of Applicant Device:

Intended to be used as a body fat analyzer that determines body weight and estimates body fat composition with the use of BIA (bioelectrical impedance analysis).

#### Predicate Device:

TANITA Body Fat Analyzer Model TBF-105 K930599

## Scientific Concepts and Significant Performance Characteristics:

|                                  | Body Fat Analyzer Model TBF-<br>105<br>K930599                                                                                                             | TANITA Body Fat Analyzer<br>TBF/BF/ Ultimate Scale                                                                                                         |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| INTENDED USE:                    | A combination non-invasive<br>device, which determines weight<br>and estimates body fat<br>composition using BIA<br>(bioelectrical impedance<br>analysis). | A combination non-invasive<br>device, which determines weight<br>and estimates body fat<br>composition using BIA<br>(bioelectrical impedance<br>analysis). |
| PRODUCT DESCRIPTION:             | Body composition analyzer/scales<br>utilize a “foot-to-foot” BIA (bio-<br>electrical impedance) technology -<br>to determine internal body<br>composition. | Body composition analyzer/scales<br>utilize a “foot-to-foot” BIA (bio-<br>electrical impedance) technology<br>to determine internal body<br>composition.   |
| ANALYTICAL<br>METHOD/MEASUREMENT | • “Foot-to-Foot” BIA<br>• Patented in-house BIA and<br>DEXA reference methods                                                                              | • “Foot-to-Foot” BIA<br>• Patented in-house BIA and<br>DEXA reference methods                                                                              |

{1}------------------------------------------------

# 510(k) SUMMARY, continued

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92

Side by side comparisons of the TANITA Body Fat Analyzer TBF/ BF/ Ultimate Scale to the predicate device Body Fat Analyzer Model TBF-105 clearly demonstrate that the applicant device is substantially equivalent to the legally marketed device.

Based on the results of using the previously approved "Foot-to-Foot" BIA methodology with our patented in-house BIA it was concluded that the TANITA Body Fat Analyzer(s) TBF/BF/Ultimate Scales perform as well as the predicate device and therefore have proven its safety and efficacy.

Carol Alloian (Benson) TANITA Corporation of America Manager Requiatory Affairs Manager Phone: (847) 640-9251 ext. 116 (847) 640-7978 Fax:

November 30, 2001

{2}------------------------------------------------

Public Health Service

Image /page/2/Picture/2 description: The image shows a circular logo for the U.S. Department of Health & Human Services. The logo features the department's name around the perimeter of the circle. Inside the circle is a stylized emblem with three wave-like shapes.

JUN 2 4 2002

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Carol Alloian (Benson) Regulatory Affairs Manager TANITA Corporation of America 2625 South Clearbrook Drive ARLINGTON HEIGHTS IL 60005 Re: K014009 Trade/Device Name: Body Fat Analyzer, Professional Series: Models TBF-300A. TBF-310. TBF-410, TBF-215, and BF-350; Consumer Body Fat Monitor/Scale Series: Models TBF-611, TBF-612, TBF-621, TBF-622, BF-623, BF-625, BF-626, BF-541, BF-555, BF-542, TBF-551, BF-556, BF-558, BF-559, BF-572, TBF-521 Ultimate Scales 2000, 2001, and 2001T Regulation Number: 21 CFR 870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: 74 MNW Dated: March 25, 2002 Received: March 26, 2002

Dear Ms. Alloian:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C.Bogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K014009 510(k) Number (if known):

Tanita Body Fat Analyzer Device Name:

Indications for Use:

MODELS TBF-300A, TBF-310, TBF-410, TBF-215, BF-350:

A body composition analyzer that measures body weight and impedance and estimates body fat percent. BMI, fat mass, fat free mass, basal metabolic rate, and total body water using BIA (bioelectrical impedance analysis).

MODELS TBF-611, TBF-612, TBF-621, TBF-622, BF-623, BF-625, BF-626, BF-541, BF-555, BF-542, TBF-551, BF-556, BF-558, BF-559, BF-572, TBF-521, Ultimate Scale 2001:

A body composition analyzer that measures body weight and impedance and estimates body fat percent using BIA (bioelectrical impedance analysis).

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

|                                          |    | Concurrence of CDRH, Office of Device Evaluation (ODE)                 |
|------------------------------------------|----|------------------------------------------------------------------------|
| Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-the-Counter Use <span style="text-decoration: overline;">✓</span> |
|                                          |    | (Optional Format 1-2-96)                                               |

(Division Sign-Off)

Division of Reproductive, Abdominal,

Revised: 6/24/02

1

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MNW/K014009](https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MNW/K014009)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
