← Product Code [JOM](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/JOM) · K022047

# PHLEBOTEST 2000 (K022047)

_Osborn Medical Corp. · JOM · Nov 4, 2002 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/JOM/K022047

## Device Facts

- **Applicant:** Osborn Medical Corp.
- **Product Code:** [JOM](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/JOM.md)
- **Decision Date:** Nov 4, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2780
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

Phlebotest 2000 is a system for peripheral vascular diagnostic investigations. Outflow phethysmography (OP) Passive Drain and Refill (PDR) Exercise Venous Plethysmography (EVP) Triple Test (TTT) Combination of OP, PDR, and EVP Either Strain Gauges or Air Pressure Cuffs Prescription device by a physician

## Device Story

Phlebotest 2000 is a peripheral vascular diagnostic system. System includes a motorized patient chair with adjustable heel/foot supports for sitting or laying positions, controlled via remote. Integrated pneumatic system (compressor, pressure tank, valves) and sensors (strain gauges or air pressure cuffs) measure vascular parameters. Hardware and software interface with a Windows-based PC to process sensor data. Software programs support specific diagnostic tests: Outflow Plethysmography (OP), Passive Drain and Refill (PDR), Exercise Venous Plethysmography (EVP), and Triple Test (TTT). Clinicians use the system to perform vascular investigations; output is displayed on the PC to assist in clinical decision-making regarding vascular health.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

System includes motorized patient chair, pneumatic compressor/tank/valve assembly, and sensors (strain gauges or air pressure cuffs). Connectivity via PC Windows-based computer. Software-controlled diagnostic system. No specific material standards or algorithm architecture details provided.

## Regulatory Identification

A hydraulic, pneumatic, or photoelectric plethysmograph is a device used to estimate blood flow in a region of the body using hydraulic, pneumatic, or photoelectric measurement techniques.

## Predicate Devices

- Varitest (Phlebotest) ([K882547](/device/K882547.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

4 2002 NOV

Image /page/0/Picture/1 description: The image shows the logo for Osborn Medical. The logo consists of a stylized human figure with arms raised above its head, followed by the text "Osborn Medical" in a serif font. The figure is black, and the text is also black.

osbornmed@osbornmedical.com

Customer Service: 800.535.5865 Fax: 507.932.5044 Telephone: 507.932.5028

100 West Main P.O. Box 324 Utica, MN 55979 USA

022047

### 510(k) SUMMARY of SAFETY and EFFECTIVENESS

### A. General Information

- 1. Submitter's Name: Osborn Medical Corporation
- 100 West Main 2. Address: Utica Minnesota 55979
- 3. Telephone: 507-932-5028
- 4. Contact Person: Bill Davis
- 5. Date Prepared: June 18, 2002
- 6. Registration Number: 2184046

### B. Device

- Phlebotest 2000 1. Name:
- 2. Trade Name: Phlebotest 2000
- 3. Common Name: Plethysmograph
- Plethysmograph 4. Classification Name:
- 5. Product Code: JOM
- 6. Class: II
- 7. Regulation Number: 870.2870

### C. Identification of Legally Marketed Devices

- l. Name: Varitest (Phlebotest) 2. K Number: K882547
w . o s b o r n m e d i c a l . e o ۱۱: | | |

{1}------------------------------------------------

# K022047 22/2

### D. Description of the Device

The PHLEBOTEST 2000 is a system designed for peripheral vascular diagnostic investigations.

The center of the system is a specially developed patient chair, which can be positioned according to the investigation performed. The patient can be in a sitting or laying position, with legs positioned in different angels using the movable heel and foot support. Patient elevation between laying and sitting positions can be carried out in various speeds according to investigational requirements.

All movements are carried out with low-voltage electrical motors and are directed by a remote hand controller. A pneumatic system consisting of a compressor, pressure tank and valve system is integrated in the system.

Different sensors can be attached to the patients for measuring the various examination possibilities. The sensors and pneumatic system are fully integrated and controlled by hardware and software in connection with a PC Windows-based computer.

A number of special developed software programs for various diagnostic tests support the system.

#### E. Intended Use Statement

- 트 Phlebotest 2000 is a system for peripheral vascular diagnostic investigations.
	- 트 Outflow phethysmography (OP)
	- 트 Passive Drain and Refill (PDR)
	- Exercise Venous Plethysmography (EVP)
	- I Triple Test (TTT) Combination of OP, PDR, and EVP
	- 프 Either Strain Gauges or Air Pressure Cuffs

#### F. Technical Characteristics Summary

The Phlebotest 2000 is substantially equivalent to the Eureka Varitest (Phlebotest) cleared on November 3, 1988.

Osborn Medical Corporation undertook extensive verifications of the software and testing to appropriate standards to verify the Phlebotest's safety and functional outcomes.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white and appears to be a scan of a document.

: 发生

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV. 4 2002

Osborn Medical, Inc. c/o Mr. Bill D. Davis President 100 West Main P.O. Box 324 Utica, Minnesota 55979

Re: K022047

Trade Name: Phlebotest 2000 Regulation Number: 21 CFR 870.2780 Regulation Name: Hydraulic, Pneumatic, or Photoelectric Plethysmograph Regulatory Class: Class II (two) Product Code: JOM Dated: October 8, 2002 Received: October 9, 2002

Dear Mr. Davis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

### Page 2 - Mr. Bill D. Davis

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Bram D. Zuckerman, M.D.
Director

Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## 510(k) Number: To be determined 长心 乙

Device Name: Phlebotest 2000

Indications for Use:

- Phlebotest 2000 is a system for peripheral vascular diagnostic investigations. ●
	- Outflow phethysmography (OP) .
	- Passive Drain and Refill (PDR) ●
	- Exercise Venous Plethysmography (EVP) .
	- Triple Test (TTT) Combination of OP, PDR, and EVP ●
	- Either Strain Gauges or Air Pressure Cuffs .
	- Prescription device by a physician

### PLEASE DO NOT WRITE BELOW THIS LINE -Continue on another Page if needed

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use X

OR

OVER-THE-COUNTER USE (optional Form 1-2-96)

Division of Cardiovascular & Respiratory Devices
510(k) Number K022047

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/JOM/K022047](https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/JOM/K022047)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
