K180558 · St. Jude Medical (Now Part of Abbott Medical) · DXO · Mar 28, 2018 · Cardiovascular
Device Facts
Record ID
K180558
Device Name
PressureWire X
Applicant
St. Jude Medical (Now Part of Abbott Medical)
Product Code
DXO · Cardiovascular
Decision Date
Mar 28, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2870
Device Class
Class 2
Indications for Use
The guidewire is uniquely paired with a specific connection cable or with a specific transmitter. Both PressureWire™ X guidewire connection configurations connect to a diagnostic computer or a catheter laboratory hemodynamic recording system. The PressureWire™ guidewire is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a vessel. The signal output from the sensor is transmitted to associated equipment for analysis, calculations, and display of physiological parameters or indices based on pressure or temperature, e.g. Fractional Flow Reserve (FFR).
Device Story
PressureWire™ X is a guidewire with an integrated distal sensor for measuring blood pressure and temperature in heart, coronary, and peripheral vessels. Used in clinical settings (e.g., catheter labs) by physicians to steer catheters through vasculature. The device connects to a diagnostic computer or hemodynamic recording system via a cable or transmitter. The sensor output is transmitted to external equipment for analysis, calculation, and display of physiological indices like Fractional Flow Reserve (FFR). The device aids clinical decision-making by providing real-time hemodynamic data during interventional procedures.
Clinical Evidence
No clinical testing was performed or relied upon. Evidence consists of bench testing and simulated-use validation.
Technological Characteristics
Integrated sensor element at distal tip; operates at -30 to +300 mmHg pressure range and 34-42°C temperature range. Radio frequency transmission (2.4000 - 2.4835 GHz ISM-band). Available in two lengths. Connects to external hemodynamic recording systems.
Indications for Use
Indicated for directing a catheter through a blood vessel and measuring physiological parameters (blood pressure, blood temperature) in the heart and coronary/peripheral blood vessels. Contraindicated for use in cerebral vasculature.
Regulatory Classification
Identification
A catheter tip pressure transducer is a device incorporated into the distal end of a catheter. When placed in the bloodstream, its mechanical or electrical properties change in relation to changes in blood pressure. These changes are transmitted to accessory equipment for processing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 28, 2018
St. Jude Medical (now part of Abbott Medical) Josh Johnson Regulatory Affairs Specialist One St. Jude Medical Drive St. Paul, Minnesota 55117
Re: K180558
Trade/Device Name: PressureWire™ X Regulation Number: 21 CFR 870.2870 Regulation Name: Catheter tip pressure transducer Regulatory Class: Class II Product Code: DXO, DQX, DRG Dated: February 26, 2018 Received: March 1, 2018
Dear Josh Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
M.A. Hillemann
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180558
Device Name PressureWire™ X
Indications for Use (Describe)
The PressureWire™ X guidewire is indicated to direct a catheter through a blood vessel and to measure physiological parameters in the heart and in the coronary and peripheral blood vessels. Physiological parameters include blood pressure. The PressureWire™ X guidewire can also measure blood temperature.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:100%">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size:100%">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| 510(k) Number: | K180558 |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | March 27, 2018 |
| Submitter Name &<br>Address: | St. Jude Medical<br>One St. Jude Medical Dr.<br>St. Paul, MN 55117 |
| Contact Person: | Josh Johnson<br>Regulatory Affairs Specialist<br>Phone (651) 756-3767<br>Fax (651) 756-3301<br>josh.johnson@abbott.com |
| Alternative Contact<br>Person: | Marlene Peterson<br>Regulatory Affairs Manager<br>Phone (651) 756-3268<br>Fax (651) 756-3301<br>marlene.peterson@abbott.com |
| Proprietary Name: | PressureWire™ X |
| Common/Usual<br>Name: | PressureWire |
| Product Classification<br>Code: | Primary: DXO<br>Secondary: DQX & DRG |
| Product Regulation<br>Number and Name: | 870.2870, Catheter tip pressure transducer<br>870.1330 – Wire, Guide, Catheter<br>870.2910 - Transmitters and Receivers, Physiological Signal, Radiofrequency |
| Device Class: | II |
| Predicate Device: | PressureWire™ X (K161171 ) |
| Device Description: | The PressureWire™ X guidewire has an integrated sensor element at the tip to enable<br>measurements of physiological parameters. The wire is introduced into the patient's<br>blood vessel. A torque device is used to steer the wire and sensor into the required<br>position for pressure measurements according to standard clinical practice.<br>PressureWire™ X guidewire is available in two different lengths. |
| Intended Use: | The guidewire is uniquely paired with a specific connection cable or with a specific<br>transmitter. Both PressureWire™ X guidewire connection configurations connect to a<br>diagnostic computer or a catheter laboratory hemodynamic recording system.<br>The PressureWire™ guidewire is designed to fit inside a percutaneous catheter for the<br>purpose of directing the catheter through a vessel. The signal output from the sensor is<br>transmitted to associated equipment for analysis, calculations, and display of<br>physiological parameters or indices based on pressure or temperature, e.g. Fractional<br>Flow Reserve (FFR). |
| Indications for Use: | The PressureWire™ X guidewire is indicated to direct a catheter through a blood vessel<br>and to measure physiological parameters in the heart and in the coronary and peripheral<br>blood vessels. Physiological parameters include blood pressure. The PressureWire™ X<br>guidewire can also measure blood temperature. |
| Comparison of<br>Subject to Predicate<br>Device: | A minor design modification was made to the pressure sensor jacket component of the<br>PressureWire™ X guidewire. Specifically, the sensor jacket design changed from a one-<br>window to a three-window jacket design, and the window length was extended. The<br>sensor jacket is located near the distal tip of the PressureWire™ X, and functions to<br>protect the pressure sensor. The openings in the sensor jacket, or windows, function to<br>allow fluid flow to the sensor. There were no other changes to the device or IFU. |
{4}------------------------------------------------
The intended use, indications for use, and technological characteristics of the PressureWire™ X device that is subject to this Special 510(K) submission are the same as the predicate, and are summarized in the following table:
| | PressureWire™ X<br>(Subject Device, K180558) | PressureWire™ X<br>(Predicate Device, K161171) | |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary<br>Name: | PressureWire™ X | PressureWire™ X | |
| Common<br>Name: | PressureWire | PressureWire | |
| Product<br>Classific-<br>ation Code: | Primary: DXO<br>Secondary: DQX & DRG | Primary: DXO<br>Secondary: DQX & DRG | |
| Product<br>Regulation<br>Number &<br>Name: | Primary: 870.2870, Catheter tip<br>pressure transducer<br>Secondary: 870.1330, catheter<br>guide wire and 870.2910<br>radiofrequency physiological<br>signal transmitters and receivers | Primary: 870.2870, Catheter tip<br>pressure transducer<br>Secondary: 870.1330, catheter<br>guide wire and 870.2910<br>radiofrequency physiological<br>signal transmitters and receivers | |
| Device<br>Class: | II | II | |
| Device<br>Description: | The PressureWire™ X guidewire<br>has an integrated sensor element<br>at the tip to enable measurements<br>of physiological parameters. The<br>wire is introduced into the<br>patient's blood vessel. A torque<br>device is used to steer the wire<br>and sensor into the required<br>position for pressure<br>measurements according to<br>standard clinical practice.<br>PressureWire™ X guidewire is<br>available in two different lengths.<br>The guidewire is uniquely paired<br>with a specific connection cable<br>or with a specific transmitter.<br>Both PressureWire™ X<br>guidewire connection<br>configurations connect to a<br>diagnostic computer or a catheter<br>laboratory hemodynamic<br>recording system. | The PressureWire™ X guidewire<br>has an integrated sensor element<br>at the tip to enable measurements<br>of physiological parameters. The<br>wire is introduced into the<br>patient's blood vessel. A torque<br>device is used to steer the wire<br>and sensor into the required<br>position for pressure<br>measurements according to<br>standard clinical practice.<br>PressureWire™ X guidewire is<br>available in two different lengths.<br>The guidewire is uniquely paired<br>with a specific connection cable<br>or with a specific transmitter.<br>Both PressureWire™ X<br>guidewire connection<br>configurations connect to a<br>diagnostic computer or a catheter<br>laboratory hemodynamic<br>recording system. | |
| Intended<br>Use: | The PressureWire™ guidewire is<br>designed to fit inside a<br>percutaneous catheter for the<br>purpose of directing the catheter<br>through a vessel. The signal<br>output from the sensor is<br>transmitted to associated<br>equipment for analysis,<br>calculations, and display of | The PressureWire™ guidewire is<br>designed to fit inside a<br>percutaneous catheter for the<br>purpose of directing the catheter<br>through a vessel. The signal<br>output from the sensor is<br>transmitted to associated<br>equipment for analysis,<br>calculations, and display of | |
| | | temperature, e.g. Fractional Flow<br>Reserve (FFR). | temperature, e.g. Fractional Flow<br>Reserve (FFR). |
| | Indications<br>for Use: | The PressureWire™ X guidewire<br>is indicated to direct a catheter<br>through a blood vessel and to<br>measure physiological parameters<br>in the heart and in the coronary<br>and peripheral blood vessels.<br>Physiological parameters include<br>blood pressure. The<br>PressureWire™ X guidewire can | The PressureWire™ X guidewire<br>is indicated to direct a catheter<br>through a blood vessel and to<br>measure physiological<br>parameters in the heart and in the<br>coronary and peripheral blood<br>vessels. Physiological parameters<br>include blood pressure. The<br>PressureWire™ X guidewire can |
| | Contra-<br>indications: | also measure blood temperature.<br>This guidewire is contraindicated<br>for use in the cerebral<br>vasculature. | also measure blood temperature.<br>This guidewire is contraindicated<br>for use in the cerebral<br>vasculature. |
| | Technical<br>Specifi-<br>cations | Operating Pressure: -30 to +300<br>mmHg<br>Temp. Performance<br>Specification:<br>34 - 42°C<br>Radio Frequency Range:<br>2.4000 - 2.4835 GHz (ISM-band) | Operating Pressure: -30 to +300<br>mmHg<br>Temp. Performance<br>Specification:<br>34 - 42°C<br>Radio Frequency Range:<br>2.4000 - 2.4835 GHz (ISM-band) |
| Summary on Non-<br>Clinical Testing | Performance bench testing was completed as part of design verification to demonstrate<br>safety and effectiveness, and that the subject device performed as intended. Tests were<br>conducted to evaluate the following:<br>• Guidewire Tensile Strength<br>• Torque Strength<br>• Fatigue Resistance<br>• Fracture Resistance<br>• Signal Quality in Severe Bend<br>• Signal Quality during Pullback Procedure<br>Additionally, simulated-use testing was completed as part of design validation to<br>demonstrate the subject device met user needs and the intended use. The following test<br>was performed:<br>• Signal Drift | | |
| Summary of Clinical<br>Testing: | No new clinical testing was completed, nor relied upon, in support of this Special<br>510(k) submission. | | |
| Statement of<br>Equivalence: | The PressureWire™ X device described in this Special 510(k) submission is<br>substantially equivalent to the currently-marketed predicate device, K161171, based on<br>comparisons of the device classifications, technological characteristics, intended use,<br>and indications for use. The subject device met the pre-determined requirements, and<br>raised no new safety or effectiveness concerns. | | |
{5}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.