← Product Code [DXO](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DXO) · K141495

# TRUWAVE DISPOSABLE PRESSURE TRANSDUCER (K141495)

_Edwards Lifesciences, LLC · DXO · Sep 3, 2014 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DXO/K141495

## Device Facts

- **Applicant:** Edwards Lifesciences, LLC
- **Product Code:** [DXO](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DXO.md)
- **Decision Date:** Sep 3, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2870
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Pressure Monitoring Kit with TruWave Disposable Pressure Transducer is for use on patients requiring intravascular, intracranial, or intrauterine pressure monitoring.

## Device Story

The Pressure Monitoring Kit with TruWave Disposable Pressure Transducer is a sterile, single-use device designed for monitoring intravascular, intracranial, or intrauterine pressure. It features a straight, flow-through design where fluid passes across a pressure sensor. The kit includes a sterile cable (available in 30 cm or 120 cm lengths) that interfaces with an Edwards Lifesciences cable specifically wired for the patient monitor used to display pressure data. The device may include an integral flush device. It is intended for use in hospitals and clinical environments by healthcare professionals. The device provides real-time pressure data to the patient monitor, assisting clinicians in hemodynamic or pressure-related decision-making, thereby supporting patient management in critical care settings.

## Clinical Evidence

Bench testing only. No clinical data was required. The device underwent functional performance testing and biocompatibility testing in accordance with applicable consensus standards to demonstrate substantial equivalence.

## Technological Characteristics

Flow-through pressure transducer design; sterile, single-use; includes integral flush device option; cable lengths of 30 cm or 120 cm; biocompatible materials per ISO 10993; interfaces with patient monitors via proprietary cabling.

## Regulatory Identification

A catheter tip pressure transducer is a device incorporated into the distal end of a catheter. When placed in the bloodstream, its mechanical or electrical properties change in relation to changes in blood pressure. These changes are transmitted to accessory equipment for processing.

## Predicate Devices

- Phoenix Disposable Pressure Transducer #73-600 ([K925638](/device/K925638.md))

## Reference Devices

- VAMP ([K896819](/device/K896819.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human face in profile, composed of three overlapping silhouettes. The silhouettes are arranged in a way that suggests a sense of community and support. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the logo.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

September 3, 2014

Edwards Lifesciences, LLC. Mr. Andrew S. Mazurkiewicz, Jr., MBA/MKT Senior Associate, Regulatory Affairs Critical Care One Edwards Way, CA 92614

Re: K141495

> Trade/Device Names: Pressure Monitoring Kit with TruWave Disposable Pressure Transducers, with list of 577 models attached Regulation Number: 21 CFR 870.2870 Regulation Name: Catheter Tip Pressure Transducer Regulatory Class: Class II (two) Product Code: DXO Dated: August 22, 2014 Received: August 25, 2014

Dear Mr. Mazurkiewicz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 – Mr. Andrew S. Mazurkiewicz, Jr.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours

forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# ATTACHMENT 1

## SUPPLEMENT TO TABLE 11-2: TRUWAVE MODELS

# (Total of 577 models)

{3}------------------------------------------------

| Supplement to Table 11-2: TruWave™ Models |                                     |                                    |                       |                                                         |
|-------------------------------------------|-------------------------------------|------------------------------------|-----------------------|---------------------------------------------------------|
| Model #                                   | # of<br>Transducers<br>(P/N 500479) | Transducer<br>Flow Rate<br>(mL/hr) | Tubing Length<br>(cm) | Blood Sampling<br>Reservoir Size (cc)<br>VAMP (K896819) |
| PXMK2043                                  | 3                                   | 3                                  | 183                   | N/A                                                     |
| PX284AN                                   | 1                                   | 3                                  | 210                   | N/A                                                     |
| PXMK1940                                  | 1                                   | 3                                  | 183                   | N/A                                                     |
| PX2X2284C                                 | 2                                   | 3                                  | 211                   | N/A                                                     |
| PXMK2002                                  | 1                                   | 3                                  | 91                    | N/A                                                     |
| PXMK478                                   | 1                                   | 3                                  | 150                   | N/A                                                     |
| T494C00B                                  | 2                                   | 3                                  | 180                   | N/A                                                     |
| PXVMP260                                  | 1                                   | 3                                  | 150                   | 5                                                       |
| PXVP0861                                  | 2                                   | 3                                  | 170                   | 12                                                      |
| T001650A                                  | 2                                   | 3                                  | 150                   | N/A                                                     |
| VMP426PX                                  | 1                                   | 3                                  | 66                    | 3                                                       |
| PXMK2041                                  | 1                                   | 3                                  | 183                   | N/A                                                     |
| PXVP2284                                  | 1                                   | 3                                  | 210                   | 12                                                      |
| T001709A                                  | 2                                   | 3                                  | 210                   | N/A                                                     |
| T432105A                                  | 1                                   | 30                                 | 130                   | N/A                                                     |
| T322242B                                  | 2                                   | 3                                  | 210                   | N/A                                                     |
| T531401A                                  | 2                                   | 3                                  | 0                     | N/A                                                     |
| T434303A                                  | 2                                   | 3                                  | 210                   | 5                                                       |
| T001759A                                  | 2                                   | 3                                  | 171                   | 12                                                      |
| T001737A                                  | 2                                   | 3                                  | 180                   | 5                                                       |
| PX12N                                     | 1                                   | 30                                 | 30                    | N/A                                                     |
| T001657A                                  | 2                                   | 3                                  | 210                   | N/A                                                     |
| T001775A                                  | 1                                   | 3                                  | 195                   | N/A                                                     |
| T433803A                                  | 1                                   | 3                                  | 200                   | N/A                                                     |
| PX200                                     | 1                                   | 3                                  | 0                     | N/A                                                     |
| PX1X2                                     | 1                                   | 3                                  | 150                   | N/A                                                     |
| PXVMP2X31                                 | 2                                   | 3                                  | 160                   | 5                                                       |
| PX6001                                    | 1                                   | 3                                  | 0                     | N/A                                                     |
| PXMK1849                                  | 1                                   | 3                                  | 193                   | N/A                                                     |
| T001633A                                  | 1                                   | 3                                  | 150                   | N/A                                                     |
| T001624A                                  | 1                                   | 3                                  | 150                   | N/A                                                     |
| T321571A                                  | 2                                   | 3                                  | 165                   | N/A                                                     |
| PXMK0666                                  | 2                                   | 3                                  | 208                   | N/A                                                     |
| T001658A                                  | 2                                   | 3                                  | 180                   | N/A                                                     |
| PX2X2YB                                   | 2                                   | 3                                  | 180                   | N/A                                                     |
| T391T00A                                  | 2                                   | 3                                  | 221                   | 12                                                      |
| T001712A                                  | 2                                   | 3                                  | 195                   | N/A                                                     |
| T441T01C                                  | 1                                   | 3                                  | 225                   | N/A                                                     |
| T491Y01B                                  | 1                                   | 3                                  | 185                   | N/A                                                     |
| T581203B                                  | 1                                   | 3                                  | 210                   | N/A                                                     |
| Supplement to Table 11-2: TruWave™ Models |                                     |                                    |                       |                                                         |
| Model #                                   | # of<br>Transducers<br>(P/N 500479) | Transducer<br>Flow Rate<br>(mL/hr) | Tubing Length<br>(cm) | Blood Sampling<br>Reservoir Size (cc)<br>VAMP (K896819) |
| PXMK2043                                  | 3                                   | 3                                  | 183                   | N/A                                                     |
| T530217B                                  | 1                                   | 3                                  | 180                   | N/A                                                     |
| VMP448PX                                  | 1                                   | 3                                  | 122                   | 3                                                       |
| PXVP2284DZ                                | 1                                   | 3                                  | 210                   | 12                                                      |
| T001660A                                  | 3                                   | 3                                  | 150                   | N/A                                                     |
| T100202A                                  | 1                                   | 3                                  | 105                   | N/A                                                     |
| T001741A                                  | 2                                   | 3                                  | 180                   | 12                                                      |
| PXVP0549                                  | 3                                   | 3                                  | 221                   | 12                                                      |
| PX3X3284C                                 | 3                                   | 3                                  | 211                   | N/A                                                     |
| PX2X3                                     | 2                                   | 3                                  | 150                   | N/A                                                     |
| T005002A                                  | 1                                   | 3                                  | 180                   | N/A                                                     |
| T001745A                                  | 1                                   | 3                                  | 180                   | 12                                                      |
| PX3X3                                     | 3                                   | 3                                  | 150                   | N/A                                                     |
| PXVK0458                                  | 3                                   | 3                                  | 236                   | 5                                                       |
| PXVMP3X31                                 | 3                                   | 3                                  | 160                   | 5                                                       |
| PX600F                                    | 1                                   | 3                                  | 0                     | N/A                                                     |
| T005025A                                  | 1                                   | 3                                  | 180                   | N/A                                                     |
| PX4X4                                     | 4                                   | 3                                  | 211                   | N/A                                                     |
| PX3X3AN                                   | 3                                   | 3                                  | 211                   | N/A                                                     |
| PXMK2011                                  | 1                                   | 3                                  | 196                   | N/A                                                     |
| PXMK9146                                  | 2                                   | 3                                  | 178                   | N/A                                                     |
| T001789B                                  | 1                                   | 3                                  | 150                   | N/A                                                     |
| T325202A                                  | 1                                   | 3                                  | 150                   | N/A                                                     |
| PX3X3272                                  | 3                                   | 3                                  | 178                   | N/A                                                     |
| PXVMP172                                  | 1                                   | 3                                  | 180                   | 5                                                       |
| PX600F30                                  | 1                                   | 30                                 | 0                     | N/A                                                     |
| PXMK053                                   | 1                                   | 3                                  | 210                   | N/A                                                     |
| PXMK2008                                  | 1                                   | 3                                  | 180                   | N/A                                                     |
| PXVP2284AT3                               | 1                                   | 3                                  | 210                   | 12                                                      |
| PXVP23X3AT3                               | 3                                   | 3                                  | 231                   | 12                                                      |
| PXVP2X2AT3                                | 2                                   | 3                                  | 213                   | 12                                                      |
| PXVK0388                                  | 3                                   | 3                                  | 188                   | N/A                                                     |
| PXCMK590                                  | 2                                   | 3                                  | 208                   | N/A                                                     |
| PXVK0387                                  | 2                                   | 3                                  | 188                   | N/A                                                     |
| PX266C                                    | 1                                   | 3                                  | 168                   | N/A                                                     |
| PXMK1479                                  | 1                                   | 3                                  | 148                   | N/A                                                     |
| PXVP2272                                  | 1                                   | 3                                  | 180                   | 12                                                      |
| PX600                                     | 1                                   | 3                                  | 0                     | N/A                                                     |
| PXMK0695                                  | 1                                   | 3                                  | 130                   | N/A                                                     |
| PX284                                     | 1                                   | 3                                  | 210                   | N/A                                                     |

{4}------------------------------------------------

{5}------------------------------------------------

| Supplement to Table 11-2: TruWave™ Models |                                     |                                    |                       |                                                         |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
|-------------------------------------------|-------------------------------------|------------------------------------|-----------------------|---------------------------------------------------------|----------|---|---|-----|-----|----------|---|---|-----|-----|----------|---|---|---|-----|----------|---|---|---|-----|----------|---|---|---|-----|----------|---|---|-----|-----|----------|---|---|-----|-----|----------|---|---|-----|-----|----------|---|---|-----|-----|----------|---|---|-----|-----|----------|---|---|-----|-----|----------|---|---|-----|-----|----------|---|---|---|-----|---------|---|---|----|-----|
| Model #                                   | # of<br>Transducers<br>(P/N 500479) | Transducer<br>Flow Rate<br>(mL/hr) | Tubing Length<br>(cm) | Blood Sampling<br>Reservoir Size (cc)<br>VAMP (K896819) |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXMK2043                                  | 3                                   | 3                                  | 183                   | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PX212                                     | 1                                   | 3                                  | 30                    | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXMK0976                                  | 2                                   | 3                                  | 147                   | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXVK0398                                  | 2                                   | 3                                  | 188                   | 5                                                       |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXVMP160                                  | 1                                   | 3                                  | 150                   | 5                                                       |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PX284R                                    | 1                                   | 3                                  | 210                   | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXMK0688                                  | 1                                   | 30                                 | 119                   | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PX260<br>(Representative<br>Model)        | 1                                   | 3                                  | 150                   | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PX272                                     | 1                                   | 3                                  | 180                   | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PX2X2                                     | 2                                   | 3                                  | 150                   | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXVMP120                                  | 1                                   | 3                                  | 50                    | 5                                                       |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXVP0897                                  | 3                                   | 3                                  | 193                   | 12                                                      |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXVP23X3                                  | 3                                   | 3                                  | 170                   | 12                                                      |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXVMP184                                  | 1                                   | 3                                  | 210                   | 5                                                       |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXVMP272                                  | 1                                   | 3                                  | 180                   | 5                                                       |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXVMP284                                  | 1                                   | 3                                  | 210                   | 5                                                       |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXVP2260                                  | 1                                   | 3                                  | 150                   | 12                                                      |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| T001631A                                  | 1                                   | 3                                  | 150                   | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| T001645A                                  | 1                                   | 3                                  | 165                   | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| T001670A                                  | 1                                   | 3                                  | 150                   | 5                                                       |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| T001671A                                  | 1                                   | 3                                  | 180                   | 5                                                       |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| T001673A                                  | 2                                   | 3                                  | 150                   | 5                                                       |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PX3X3MT                                   | 3                                   | 3                                  | 211                   | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXMK1876                                  | 3                                   | 3                                  | 213                   | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXMK1805                                  | 3                                   | 3                                  | 211                   | N/A                                                     | PXMK2012 | 4 | 3 | 196 | N/A | T100209A | 1 | 3 | 150 | N/A | PXAK1568 | 1 | 3 | 0 | N/A | PXAK2066 | 3 | 3 | 0 | N/A | PXAK2296 | 4 | 3 | 0 | N/A | PXMK0933 | 1 | 3 | 208 | N/A | PXMK1300 | 3 | 3 | 274 | N/A | PXMK1737 | 2 | 3 | 208 | N/A | PXMK1863 | 4 | 3 | 178 | N/A | PXMK2143 | 3 | 3 | 198 | N/A | PXMK2144 | 3 | 3 | 211 | N/A | PXMK2258 | 4 | 3 | 183 | N/A | PXMK2270 | 5 | 3 | 0 | N/A | PXMK560 | 2 | 3 | 30 | N/A |
| PXMK1805                                  | 3                                   | 3                                  | 211                   | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXMK2012                                  | 4                                   | 3                                  | 196                   | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| T100209A                                  | 1                                   | 3                                  | 150                   | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXAK1568                                  | 1                                   | 3                                  | 0                     | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXAK2066                                  | 3                                   | 3                                  | 0                     | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXAK2296                                  | 4                                   | 3                                  | 0                     | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXMK0933                                  | 1                                   | 3                                  | 208                   | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXMK1300                                  | 3                                   | 3                                  | 274                   | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXMK1737                                  | 2                                   | 3                                  | 208                   | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXMK1863                                  | 4                                   | 3                                  | 178                   | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXMK2143                                  | 3                                   | 3                                  | 198                   | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXMK2144                                  | 3                                   | 3                                  | 211                   | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXMK2258                                  | 4                                   | 3                                  | 183                   | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXMK2270                                  | 5                                   | 3                                  | 0                     | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |
| PXMK560                                   | 2                                   | 3                                  | 30                    | N/A                                                     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |          |   |   |   |     |          |   |   |   |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |     |     |          |   |   |   |     |         |   |   |    |     |

{6}------------------------------------------------

| Supplement to Table 11-2:  TruWave™ Models |                                     |                                    |                       |                                                         |
|--------------------------------------------|-------------------------------------|------------------------------------|-----------------------|---------------------------------------------------------|
| Model #                                    | # of<br>Transducers<br>(P/N 500479) | Transducer<br>Flow Rate<br>(mL/hr) | Tubing Length<br>(cm) | Blood Sampling<br>Reservoir Size (cc)<br>VAMP (K896819) |
| PXMK2043                                   | 3                                   | 3                                  | 183                   | N/A                                                     |
| PXVK0363                                   | 3                                   | 3                                  | 241                   | 5                                                       |
| PXVK0438                                   | 4                                   | 3                                  | 163                   | 5                                                       |
| PXVK0738                                   | 4                                   | 3                                  | 244                   | 5                                                       |
| PXVK0755                                   | 3                                   | 3                                  | 244                   | 5                                                       |
| PXVK0797                                   | 3                                   | 3                                  | 224                   | 5                                                       |
| PXVK0965                                   | 3                                   | 3                                  | 173                   | 5                                                       |
| PXVK1001                                   | 3                                   | 3                                  | 206                   | 5                                                       |
| PXVK1003                                   | 4                                   | 3                                  | 173                   | 5                                                       |
| PXVK1086                                   | 3                                   | 3                                  | 99                    | 5                                                       |
| PXVK268                                    | 3                                   | 3                                  | 203                   | 5                                                       |
| PXVP0686                                   | 2                                   | 3                                  | 220                   | 12                                                      |
| PXVP0812                                   | 3                                   | 3                                  | 155                   | 12                                                      |
| PXVP0880                                   | 3                                   | 3                                  | 231                   | 12                                                      |
| PXVP0902                                   | 3                                   | 3                                  | 206                   | 12                                                      |
| PXVP0966                                   | 3                                   | 3                                  | 231                   | 12                                                      |
| PXVP1066                                   | 3                                   | 3                                  | 177                   | 12                                                      |
| PXVP1087                                   | 4                                   | 3                                  | 199                   | 12                                                      |
| T493Y02B                                   | 3                                   | 3                                  | 195                   | N/A                                                     |
| T001619A                                   | 1                                   | 3                                  | 30                    | N/A                                                     |
| T310195A                                   | 1                                   | 30                                 | 150                   | N/A                                                     |
| T001691A                                   | 1                                   | 3                                  | 171                   | 12                                                      |
| T100214A                                   | 1                                   | 3                                  | 180                   | N/A                                                     |
| T100215A                                   | 1                                   | 3                                  | 210                   | N/A                                                     |
| T430023B                                   | 2                                   | 3                                  | 215                   | 12                                                      |
| T001702A                                   | 2                                   | 3                                  | 135                   | N/A                                                     |
| T001724A                                   | 1                                   | 3                                  | 210                   | N/A                                                     |
| T001758A                                   | 1                                   | 3                                  | 0                     | N/A                                                     |
| T001760A                                   | 2                                   | 3                                  | 171                   | 12                                                      |
| T001786A                                   | 2                                   | 3                                  | 195                   | N/A                                                     |
| T005004A                                   | 1                                   | 3                                  | 150                   | N/A                                                     |
| T005052A                                   | 1                                   | 3                                  | 210                   | N/A                                                     |
| T005074A                                   | 2                                   | 3                                  | 175                   | 12                                                      |
| T005080A                                   | 1                                   | 3                                  | 180                   | N/A                                                     |
| T100212A                                   | 1                                   | 3                                  | 150                   | N/A                                                     |
| T100500A                                   | 1                                   | 3                                  | 60                    | 5                                                       |
| T100504A                                   | 1                                   | 3                                  | 190                   | 5                                                       |
| T100671B                                   | 3                                   | 3                                  | 171                   | 12                                                      |
| T330639B                                   | 2                                   | 3                                  | 150                   | N/A                                                     |
| T330831A                                   | 1                                   | 3                                  | 210                   | N/A                                                     |
| Supplement to Table 11-2: TruWaveTM Models |                                     |                                    |                       |                                                         |
| Model #                                    | # of<br>Transducers<br>(P/N 500479) | Transducer<br>Flow Rate<br>(mL/hr) | Tubing Length<br>(cm) | Blood Sampling<br>Reservoir Size (cc)<br>VAMP (K896819) |
| PXMK2043                                   | 3                                   | 3                                  | 183                   | N/A                                                     |
| T330918A                                   | 2                                   | 3                                  | 200                   | N/A                                                     |
| T330919A                                   | 1                                   | 3                                  | 200                   | N/A                                                     |
| T330V11A                                   | 3                                   | 3                                  | 200                   | N/A                                                     |
| T331E02A                                   | 1                                   | 3                                  | 0                     | N/A                                                     |
| T331F06B                                   | 1                                   | 3                                  | 90                    | N/A                                                     |
| T331S05B                                   | 2                                   | 3                                  | 200                   | N/A                                                     |
| T332J01A                                   | 1                                   | 3                                  | 150                   | N/A                                                     |
| T333529A                                   | 2                                   | 3                                  | 150                   | N/A                                                     |
| T333709B                                   | 2                                   | 3                                  | 0                     | N/A                                                     |
| T333710C                                   | 2                                   | 3                                  | 0                     | N/A                                                     |
| T333918A                                   | 2                                   | 3                                  | 210                   | N/A                                                     |
| T333919A                                   | 1                                   | 3                                  | 150                   | N/A                                                     |
| T333C00A                                   | 1                                   | 3                                  | 180                   | N/A                                                     |
| T333D00A                                   | 1                                   | 3                                  | 150                   | N/A                                                     |
| T333F00A                                   | 1                                   | 3                                  | 210                   | N/A                                                     |
| T334041A                                   | 2                                   | 3                                  | 180                   | N/A                                                     |
| T334402A                                   | 2                                   | 3                                  | 180                   | N/A                                                     |
| T334708C                                   | 1                                   | 3                                  | 0                     | N/A                                                     |
| T336546B                                   | 2                                   | 3                                  | 197                   | 12                                                      |
| T336547B                                   | 3                                   | 3                                  | 197                   | 12                                                      |
| T336548B                                   | 1                                   | 3                                  | 197                   | 12                                                      |
| T337003A                                   | 1                                   | 3                                  | 150                   | N/A                                                     |
| T337119B                                   | 2                                   | 3                                  | 200                   | N/A                                                     |
| T337305B                                   | 3                                   | 3                                  | 240                   | N/A                                                     |
| T391T01A                                   | 3                                   | 3                                  | 221                   | 12                                                      |
| T441809C                                   | 2                                   | 3                                  | 150                   | N/A                                                     |
| T450552C                                   | 1                                   | 30                                 | 156                   | 3                                                       |
| T491U01B                                   | 1                                   | 3                                  | 180                   | N/A                                                     |
| T492011A                                   | 3                                   | 3                                  | 135                   | N/A                                                     |
| T493Y03B                                   | 1                                   | 30                                 | 156                   | N/A                                                     |
| T495009A                                   | 3                                   | 3                                  | 175                   | N/A                                                     |
| T321570A                                   | 1                                   | 3                                  | 120                   | N/A                                                     |
| T321572A                                   | 3                                   | 3                                  | 150                   | N/A                                                     |
| T321573A                                   | 1                                   | 3                                  | 165                   | N/A                                                     |
| T321575A                                   | 1                                   | 3                                  | 105                   | N/A                                                     |
| T449103B                                   | 2                                   | 3                                  | 160                   | 5                                                       |
| T100209A                                   | 1                                   | 3                                  | 150                   | N/A                                                     |
| T391314A                                   | 3                                   | 3                                  | 210                   | N/A                                                     |
| T770000A                                   | 1                                   | 3                                  | 210                   | 5                                                       |
| Supplement to Table 11-2: TruWave™ Models  |                                     |                                    |                       |                                                         |
| Model #                                    | # of<br>Transducers<br>(P/N 500479) | Transducer<br>Flow Rate<br>(mL/hr) | Tubing Length<br>(cm) | Blood Sampling<br>Reservoir Size (cc)<br>VAMP (K896819) |
| PXMK2043                                   | 3                                   | 3                                  | 183                   | N/A                                                     |
| T005086A                                   | 1                                   | 3                                  | 210                   | N/A                                                     |
| T440B26B                                   | 2                                   | 3                                  | 200                   | 5                                                       |
| T321811A                                   | 1                                   | 3                                  | 162                   | 5                                                       |
| T321811Z                                   | 1                                   | 3                                  | 162                   | 5                                                       |
| T443422A                                   | 2                                   | 3                                  | 210                   | N/A                                                     |
| T391312A                                   | 1                                   | 3                                  | 210                   | N/A                                                     |
| T320161B                                   | 1                                   | 3                                  | 150                   | N/A                                                     |
| T460061A                                   | 1                                   | 3                                  | 195                   | N/A                                                     |
| T341211A                                   | 1                                   | 3                                  | 0                     | N/A                                                     |
| T433202B                                   | 1                                   | 3                                  | 210                   | N/A                                                     |
| T460225D                                   | 3                                   | 3                                  | 205                   | N/A                                                     |
| T395705A                                   | 1                                   | 3                                  | 180                   | N/A                                                     |
| T322241A                                   | 1                                   | 3                                  | 150                   | N/A                                                     |
| T325615A                                   | 1                                   | 3                                  | 195                   | N/A                                                     |
| T398807D                                   | 1                                   | 3                                  | 58                    | 12                                                      |
| T531900A                                   | 1                                   | 3                                  | 55                    | N/A                                                     |
| T341210A                                   | 1                                   | 3                                  | 150                   | N/A                                                     |
| T397705B                                   | 2                                   | 3                                  | 180                   | N/A                                                     |
| T433802A                                   | 1                                   | 3                                  | 220                   | 5                                                       |
| T320943A                                   | 2                                   | 3                                  | 120                   | N/A                                                     |
| T320940B                                   | 3                                   | 3                                  | 120                   | N/A                                                     |
| PXVJ0441                                   | 1                                   | 30                                 | 168                   | 3                                                       |
| T005094A                                   | 1                                   | 3                                  | 150                   | N/A                                                     |
| T430181A                                   | 1                                   | 3                                  | 221                   | 5                                                       |
| PX601                                      | 1                                   | 3                                  | 0                     | N/A                                                     |
| T395606B                                   | 3                                   | 3                                  | 180                   | N/A                                                     |
| T451703A                                   | 3                                   | 3                                  | 215                   | 12                                                      |
| T451703A                                   | 3                                   | 3                                  | 215                   | 12                                                      |
| T433806A                                   | 2                                   | 3                                  | 220                   | 5                                                       |
| T310682A                                   | 3                                   | 3                                  | 150                   | N/A                                                     |
| T432703A                                   | 2                                   | 3                                  | 220                   | 5                                                       |
| T349302A                                   | 1                                   | 30                                 | 150                   | N/A                                                     |
| T530317A                                   | 1                                   | 3                                  | 30                    | N/A                                                     |
| T413405A                                   | 1                                   | 3                                  | 150                   | N/A                                                     |
| T431305A                                   | 3                                   | 3                                  | 190                   | 5                                                       |
| T320162C                                   | 2                                   | 3                                  | 60                    | N/A                                                     |
| T349301B                                   | 1                                   | 3                                  | 180                   | N/A                                                     |
| T328500A                                   | 1                                   | 3                                  | 180                   | N/A                                                     |
| PXMK0010                                   | 1                                   | 3                                  | 230                   | N/A                                                     |
| Supplement to Table 11-2:  TruWave™ Models |                                     |                                    |                       |                                                         |
| Model #                                    | # of<br>Transducers<br>(P/N 500479) | Transducer<br>Flow Rate<br>(mL/hr) | Tubing Length<br>(cm) | Blood Sampling<br>Reservoir Size (cc)<br>VAMP (K896819) |
| PXMK2043                                   | 3                                   | 3                                  | 183                   | N/A                                                     |
| PXMK0011                                   | 2                                   | 3                                  | 230                   | N/A                                                     |
| PXMK0012                                   | 3                                   | 3                                  | 230                   | N/A                                                     |
| PXMK0013                                   | 1                                   | 3                                  | 0                     | N/A                                                     |
| PXMK0014                                   | 1                                   | 3                                  | 150                   | N/A                                                     |
| T610002A                                   | 1                                   | 3                                  | 30                    | N/A                                                     |
| PXMK2329                                   | 2                                   | 3                                  | 89                    | N/A                                                     |
| PXMK2330                                   | 3                                   | 3                                  | 119                   | N/A                                                     |
| T413675A                                   | 2                                   | 3                                  | 225                   | N/A                                                     |
| T397704A                                   | 1                                   | 3                                  | 180                   | N/A                                                     |
| T001725A                                   | 1                                   | 3                                  | 150                   | N/A                                                     |
| T322238A                                   | 1                                   | 3                                  | 190                   | N/A                                                     |
| T322515A                                   | చి                                  | చి                                 | 180                   | N/A                                                     |
| T398603A                                   | 2                                   | 3                                  | 220                   | 5                                                       |
| T001744A                                   | 1                                   | 3                                  | 220                   | 5                                                       |
| T323922D                                   | 2                                   | 3                                  | 180                   | N/A                                                     |
| T434304A                                   | 1                                   | 3                                  | 190                   | 5                                                       |
| T433204B                                   | 1                                   | 3                                  | 210                   | N/A                                                     |
| PX24N                                      | 1                                   | 30                                 | 60                    | N/A                                                     |
| T323923A                                   | 1                                   | 3                                  | 30                    | N/A                                                     |
| T433801A                                   | 1                                   | 3                                  | 200                   | N/A                                                     |
| T430142A                                   | 4                                   | 3                                  | 150                   | N/A                                                     |
| T312306A                                   | 1                                   | 3                                  | 15                    | N/A…

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DXO/K141495](https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DXO/K141495)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
