← Product Code [DXH](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DXH) · K143032

# eMotion Faros ECG Mobile (K143032)

_Mega Electronics , Ltd. · DXH · Mar 29, 2015 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DXH/K143032

## Device Facts

- **Applicant:** Mega Electronics , Ltd.
- **Product Code:** [DXH](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DXH.md)
- **Decision Date:** Mar 29, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2920
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The eMotion Faros ECG sensor is a wearable, externally applied, electrocardiograph recorder and transmitter for the purpose of health monitoring, biofeedback and scientific research. The eMotion Faros ECG Mobile is intended for use in clinical settings to collect and transmit health parameters to healthcare professionals for monitoring and evaluation. Health parameters are collected from a variety of commercially available, external plug-in devices such as ECG sensors, Weight Scales, Blood Pressure Meters and Pulse Oximeters. Indicated for adult patients who require clinical or non-clinical ECG monitoring in healthcare facility environment under supervision of a physician or prescript by the supervising physician to supplement data acquisition in home environment. The eMotion Faros ECG Mobile does not provide any automatic analysis or diagnosis.

## Device Story

Mobile telemetry solution for ambulatory monitoring of chronic patients; collects health data (ECG, weight, blood pressure, SpO2) from external plug-in devices via Bluetooth. Data transmitted via mobile network to 'Virtual Clinic' server; healthcare professionals review data via PC-based interface. Device includes eMotion Faros ECG sensors (90, 180, 360 models) which record ECG data to internal memory or transmit in real-time. Used in clinical settings or home environments under physician supervision. Provides no automatic analysis or diagnosis; serves as data acquisition and transmission tool to support clinical decision-making.

## Clinical Evidence

No clinical testing performed. Substantial equivalence established via non-clinical bench testing, verification, and validation against established design protocols and international standards.

## Technological Characteristics

Wearable ECG sensors (Faros 90, 180, 360) with 14-bit or 24-bit resolution; 3.7V Li-ion battery; Bluetooth connectivity; internal memory (512MB-1GB). Standards: EN 60601-1 (safety), EN 60601-1-2 (EMC), ISO 10993-1 (biocompatibility), EN 60601-2-25 (ECG performance).

## Regulatory Identification

A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.

## Predicate Devices

- eMotion ECG Mobile ([K131699](/device/K131699.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 29, 2015

Mega Electronics Ltd. Taneli Vaaraniemi Quality Engineer Pioneerinkatu 6 Kuopio, 70800 FI

Re: K143032

Trade/Device Name: Emotion Faros ECG Mobile Regulation Number: 21 CFR 870.2920 Regulation Name: Telephone Electrocardiograph Transmitter and Receiver Regulatory Class: Class II Product Code: DXH Dated: February 6, 2015 Received: February 9, 2015

Dear Taneli Vaaraniemi,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mitchell Stein

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K143032

Device Name eMotion Faros ECG Mobile

#### Indications for Use (Describe)

The eMotion Faros ECG sensor is a wearable, externally applied, electrocardiograph recorder and transmitter for the purpose of health monitoring, biofeedback and scientific research.

The eMotion Faros ECG Mobile is intended for use in clinical settings to collect and transmit health parameters to healthcare professionals for monitoring and evaluation.

Health parameters are collected from a variety of commercially available, external plug-in devices such as ECG sensors, Weight Scales, Blood Pressure Meters and Pulse Oximeters.

Indicated for adult patients who require clinical ECG monitoring in healthcare facility environment under supervision of a physician or prescript by the supervising physician to supplement data acquisition in home environment.

The eMotion Faros ECG Mobile does not provide any automatic analysis or diagnosis.

| Type of Use (Select one or both, as applicable)                                                     |                                                                                  |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> <b>      ☑     </b>     Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>     ☐ Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Mega Electronics Ltd. 510(k) Premarket Submission eMotion Faros ECG Mobile

Image /page/3/Picture/1 description: The image shows the logo for Mega. The logo consists of a square with rounded corners that is pink. Inside the square is a white silhouette of an archer aiming an arrow. To the right of the square is the word "mega" in black, lowercase letters.

## 007 – 510(k) Summary

### 1. Submitter

Mega Electronics Ltd Pioneerinkatu 6 KUOPIO FI-70800 Finland Phone: +358 17 581 7700 Fax: +358 17 580 0978 Contact: Taneli Vääräniemi, Quality Engineer Email: taneli.vaaraniemi@megaemg.com

Registration Number: 1000425315

### 2. Date

March 25, 2015

### 3. Device Identification

| Modified ECG Module Names:        | a) eMotion Faros 90<br>b) eMotion Faros 180<br>c) eMotion Faros 360 |
|-----------------------------------|---------------------------------------------------------------------|
| Modified Health Gateway Interface | Virtual Clinic                                                      |
| Trade/Proprietary Name:           | eMotion Faros ECG Mobile                                            |
| Common/Usual Name:                | Digital Ambulatory Monitor                                          |
| Classification Name:              | Transmitters And Receivers, Electrocardiograph Telephone            |
| Classification Regulation:        | 21 CFR 870.2920                                                     |
| Product Code:                     | DXH                                                                 |
| Device Class:                     | Class II                                                            |
| Classification Panel:             | Cardiovascular                                                      |
| Device Classification EU:         | Class IIa                                                           |
| Medical Device Directive:         | 93/42EEC as a amended by 2007/47/EC                                 |

# 4. Substantial Equivalence, Predicate Device

eMotion ECG Mobile, 510(k) Number: K131699

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains a logo with a pink square on the left side and the word "mega" in black on the right side. The pink square contains a white silhouette of an archer aiming an arrow. The word "mega" is written in a bold, sans-serif font.

## 5. Device Description

The eMotion Faros ECG Mobile is a mobile device. PC and Internet based telemetry solution for the ambulant monitoring of the pluq-in device data of chronic patients via a mobile network. Pluq-in devices can be ECG devices, blood pressure monitors, weighing scales, etc. The device reads data from the plug-in device via a Bluetooth connection. The mobile device sends the data to a server (Health Gateway) over mobile networks using the secured connection.

Data can be viewed from the Health Gateway server using the Virtual Clinic interface. Monitoring is performed using PC application that reads data from the server over the internet using the secured connection. ECG data is possible to record on eMotion Faros ECG sensors internal memory. Data from ECG sensors can be reviewed with eMotion EDF Viewer or uploaded via USB connection.

Data acquisition is performed using eMotion Faros ECG sensors and applicable plug-in devices. Faros ECG sensors are modified versions of eMotion ECG sensor which the FDA granted clearance on November 26, 2013 (K131699).

## Intended Use

The eMotion Faros ECG sensor is a wearable, externally applied, electrocardiograph recorder and transmitter for the purpose of health monitoring, biofeedback and scientific research.

The eMotion Faros ECG Mobile is intended for use in clinical settings to collect and transmit health parameters to healthcare professionals for monitoring and evaluation.

Health parameters are collected from a variety of commercially available, external plug-in devices such as ECG sensors, Weight Scales, Blood Pressure Meters and Pulse Oximeters.

Indicated for adult patients who require clinical or non-clinical ECG monitoring in healthcare facility environment under supervision of a physician or prescript by the supervising physician to supplement data acquisition in home environment.

The eMotion Faros ECG Mobile does not provide any automatic analysis or diagnosis.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows a logo with a pink square on the left and the word "mega" in black on the right. The pink square contains a white silhouette of an archer aiming an arrow. The word "mega" is written in a simple, sans-serif font and is positioned to the right of the square.

## 6. Comparison of Technological Characteristics

The eMotion Faros ECG Mobile is modified device of the predicate device. With respect to intended use, technological characteristics and principles of operation are identical with Predicate device, eMotion ECG Mobile K131699. Comparison to Legally Marketed Device in section 009 provides more detailed information regarding the basis for the determination of substantial equivalence.

The technological characteristics and principles of operation of the modified device are the same as the predicate device.

|                         | eMotion ECG<br>sensor (K131699)               | eMotion Faros 90<br>(modified device) | eMotion Faros 180<br>(modified device)        | eMotion Faros 360<br>(modified device)        |
|-------------------------|-----------------------------------------------|---------------------------------------|-----------------------------------------------|-----------------------------------------------|
| ECG<br>Channel(s)       | Online: 1-channel                             | Offline:<br>1-channel                 | Online and offline:<br>1-channel              | Online: 1-channel<br>Offline:3-channels       |
| Sensor<br>Dimensions    | 35 x 35 x 15 (mm)                             | 48 x 29 x 12 (mm)                     | 48 x 29 x 12 (mm)                             | 48 x 29 x 12 (mm)                             |
| Sensor Weight           | 16g                                           | 13g                                   | 13g                                           | 13g                                           |
| Internal Data<br>logger | no                                            | Yes                                   | yes                                           | yes                                           |
| Memory<br>Capacity      | no                                            | 512Mb                                 | 1 GB                                          | 1 GB or more                                  |
| ECG Sampling<br>Rate    | Selectable: 100,<br>125, 250, 500, 1000<br>Hz | Selectable: 100,<br>125, 250 Hz       | Selectable: 100,<br>125, 250, 500, 1000<br>Hz | Selectable: 100,<br>125, 250, 500, 1000<br>Hz |
| Resolution              | 14-bit                                        | 14-bit                                | 14-bit                                        | 24-bit                                        |
| Bluetooth               | yes                                           | no                                    | yes                                           | yes                                           |
| Battery                 | 3,7 V Li-ion battery                          | 3,7 V Li-ion battery                  | 3,7 V Li-ion battery                          | 3,7 V Li-ion battery                          |

Table 2: Technological Characteristics comparison Between ECG Sensors

|                                                                                                       | Web Monitor<br>(Predicate Device) | Virtual Clinic<br>(Subject Device) |
|-------------------------------------------------------------------------------------------------------|-----------------------------------|------------------------------------|
| Login screen                                                                                          | Yes                               | Yes                                |
| Patient list                                                                                          | Yes                               | Yes                                |
| Document list                                                                                         | Yes                               | Yes                                |
| Doctor info view                                                                                      | Yes                               | Yes                                |
| Doctor list                                                                                           | Yes                               | Yes                                |
| Measurement views<br>- ECG<br>- ECG Report and listing<br>- Blood pressure<br>- Body weight<br>- SpO2 | Yes                               | Yes                                |
| Patient management                                                                                    | Yes                               | Yes                                |
| Measurement value limits with<br>notifications                                                        | No                                | Yes                                |
| Administration Management                                                                             | Yes                               | Yes                                |

Table 3: Comparison Between UI Interfaces features

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains the logo for Mega Limited. On the left side of the image is a pink square with rounded corners. Inside the square is a white silhouette of an archer holding a bow and arrow. To the right of the square is the word "mega" in black, lowercase letters.

## 7. Performance Data Summary

The eMotion Faros ECG Mobile has been fully verified and validated against written test protocols to demonstrate that the design meets the requirements and performs as intended. The test results including product release information summary is documented as a part of the section 011 Summary of Design Control Activities.

Performance is tested and evaluated according to the applicable sections of the following standards:

- EN 60601-1:2006, Medical Electrical Equipment Part 1 General Requirements for Basic ● Safety and Essential Performance
- EN 60601-1-2: 2007, Medical Electrical Equipment Part 1-2: General Requirements for ● Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
- ISO 10993-1: 2003, Biological Evaluation of Medical Devices Part 1: Evaluation and ● Testing
- . EN 60601-2-25:1999 Medical electrical equipment - Part 2-25: Particular requirements for the basic safety and essential performance of electrocardiographs

## 8. Clinical Testing

There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device. The non-clinical testing performed supports the substantial equivalence of the device.

## 9. Statement of Substantial Equivalence

By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the modified device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.

It has been shown in this 510(k) submission that the difference between the eMotion Faros ECG Mobile and the predicate devices do not raise any questions regarding its safety and effectiveness. Performance testing and compliance with voluntary standards, demonstrate that the eMotion Faros ECG Mobile is substantially equivalent to the relevant aspects of the predicate devices in terms of design, components, materials, principals of operation, biocompatibility, performance characteristics, and intended use.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DXH/K143032](https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DXH/K143032)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
