← Product Code [DXH](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DXH) · K133703

# PHYSIOGLOVE ET SYSTEM (K133703)

_Commwell R&D , Ltd. · DXH · Aug 18, 2014 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DXH/K133703

## Device Facts

- **Applicant:** Commwell R&D , Ltd.
- **Product Code:** [DXH](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DXH.md)
- **Decision Date:** Aug 18, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2920
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The PhysioGlove ET is intended for self-testing by patients and by health care professionals at home and at medical settings. This 12-lead electrocardiogram (ECG), allows remote patients to display and transmit their ECG data to medical professionals via a mobile platform (communication device) to a remote server. Specifically, the PhysioGlove ET is intended for patients who are concerned about their heart health and have the following problems that are suggestive of abnormal heart function: - Skipped beats - Pounding heart (palpitations) - Heart racing or irregular pulse - Lightheadedness or faintness - Chest discomfort - Shortness of breath - High risk for heart problems - History of heart problems - Recipient of heart medications The system only provides waveform parameters for healthcare provider interpretation and does not provide suggested interpretations.

## Device Story

PhysioGlove ET is a 12-lead ECG system for remote cardiac monitoring. Input: ECG signals captured via Ag/AgCl electrodes embedded in a wearable glove. Processing: Data acquisition circuitry in glove transmits raw ECG data via Bluetooth to a mobile platform (smartphone/tablet); mobile software forwards data via Wi-Fi/GPRS to a remote server. Output: Waveform data presented via Mercury Telemedicine System (MTS) software at a receiving center (e.g., nurse station). Usage: Patient or healthcare professional at home or medical setting. Clinical decision-making: Medical professional interprets raw waveforms; device provides no automated diagnostic suggestions. Benefit: Enables remote cardiac monitoring for patients with symptoms of abnormal heart function.

## Clinical Evidence

Bench testing only. Testing included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), home healthcare environment safety (IEC 60601-1-11), and electrocardiograph performance (IEC 60601-2-25). Software validation and a usability study were conducted. Results demonstrated the system functions as expected.

## Technological Characteristics

12-lead ECG system; Ag/AgCl electrodes; Bluetooth connectivity to mobile platform; Wi-Fi/GPRS transmission to remote server. Components: PhysioGlove (acquisition/transmission circuitry), Mobile Platform Software, Mercury Telemedicine System (MTS) software. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-25.

## Regulatory Identification

A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.

## Predicate Devices

- PMP4 SelfCheck ECG ([K042254](/device/K042254.md))
- PhysioGlove ES-1 with ECG Analysis Program ([K103791](/device/K103791.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white seal for the Department of Health and Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three faces in profile, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

August 18, 2014

Commwell Ltd. c/o Mr. Jonathan S. Kahan Regulatory Counsel Hogan Lovells US LLP 555 Thirteenth Street NW Washington, DC 20004

Re: K133703

> Trade/Device Names: PhysioGlove ET System Regulation Number: 21 CFR 870.2920 Regulation Name: Telephone Electrocardiograph Transmitter and Receiver Regulatory Class: Class II (two) Product Code: DXH, DPS Dated: July 29, 2014 Received: July 29, 2014

Dear Mr. Kahan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 – Mr. Jonathan S. Kahan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mitchell Stein

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# INDICATIONS FOR USE

510(k) Number (if known): K133703

Device Name: PhysioGlove ET

Indications for Use:

The PhysioGlove ET is intended for self-testing by patients and by health care professionals at home and at medical settings. This 12-lead electrocardiogram (ECG), allows remote patients to display and transmit their ECG data to medical professionals via a mobile platform (communication device) to a remote server.

Specifically, the PhysioGlove ET is intended for patients who are concerned about their heart health and have the following problems that are suggestive of abnormal heart function:

- · Skipped beats
- · Pounding heart (palpitations)
- · Heart racing or irregular pulse
- · Lightheadedness or faintness
- · Chest discomfort
- Shortness of breath
- · High risk for heart problems
- · History of heart problems
- · Recipient of heart medications

The system only provides waveform parameters for healthcare provider interpretation and does not provide suggested interpretations.

| Prescription<br>(Part 21 CFR 801 Subpart D) | Use<br>√<br> | AND/OR | Over-The-Counter Use<br>(21 CFR 801 Subpart C) |
|---------------------------------------------|--------------|--------|------------------------------------------------|
|---------------------------------------------|--------------|--------|------------------------------------------------|

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

### K133703

### 510(K) SUMMARY

510(k) Number K

#### 6.1 Applicant's Name

Commwell Ltd. 4 Yad Harutzim Street Kfar Saba, 4464102, Israel Tel: +972-9-7668094 Fax: +972-9-7668099

#### 6.2 Contact Person

Jonathan S. Kahan Hogan Lovells US LLP Columbia Square 555 Thirteenth Street, NW Washington, DC 20004-1109 Tel: (202) 637-5794 Fax: (202) 637-5910 Mail: jonathan.kahan@hoqanlovells.com

and/or

Merav Yarmus, Ph.D. Biomedical Strategy (2004) Ltd. 155 Bialik Sheet, Ramat Gan, 5252346, Israel Tel: +972-3-6123281 Fax: +972-3-6123282 Mail: merav@ebms.so.il

#### Date Prepared 6.3

December 3, 2013

{4}------------------------------------------------

#### 6.4 Trade Name

PhysioGlove ET

#### 6.5 Classification Name

Transmitters and receivers, electrocardiograph, telephone

#### 6.6 Product Code

DXH

### 6.7 Device Class Class II

#### Regulation Number 6.8

21 CFR 870.2920

- 6.9 Panel Cardiovascular
## 6.10 Predicate Devices

- · PMP4 SelfCheck ECG [Card G u a r d Scientific Survival Ltd.] cleared under K042254
- · PhysioGlove ES-1 with ECG Analysis Program [Commwell Ltd.] cleared under K103791

## 6.11 Intended Use / Indication for Use

The PhysioGlove ET is intended for self-testing by patients and by healthcare professionals at home and in medical settings. This 12-lead electrocardiogram (ECG) allows remote patients to display and transmit their ECG data to medical professionals via a mobile platform (communication device) to a remote server.

Specifically, the PhysioGlove ET is intended for patients who are concerned about their heart health and have the following problems that are suggestive of abnormal heart function:

{5}------------------------------------------------

- Skipped beats .
- Pounding heart (palpitations) .
- Heart racing or irregular pulse .
- Lightheadedness or faintness ●
- . Chest discomfort
- Shortness of breath .
- . High risk for heart problems
- History of heart problems
- Recipient of heart medications ●

The system only provides waveform parameters for healthcare provider interpretation and does not provide suggested interpretations.

### 6.12 Device Description

The PhysioGlove ET system is a 12-lead electrocardiograph capable of recording and transmitting standard ECGs for the purpose of cardiac monitoring and diagnosis in a remote location. The PhysioGlove ET system consists of the following 3 primary components:

The PhysioGlove - a glove that incorporates a system of Ag/AgCl electrodes for use in recording 12-lead ECGs. The PhysioGlove of the ET system is a modification of the cleared PhysioGlove ES-1 system (originally cleared under K083677 and later incorporated under K103791). In the ET system, however, the PhysioGlove encompasses also electronic elements (mainly a data acquisition circuitry and software packed into the glove, based on the cleared PhysioGlove ES-1) and a recording and transmitting circuitry to permit the transfer of the ECG data to a mobile platform.

The Mobile Platform Software - designed to be downloaded to a compatible mobile platform such as a smart cell-phone, a tablet computer, or another communication device. The proprietary Mobile Platform Software is responsible to provide the user with the PhysioGlove ET system operational instructions, as well as to receive the ECG data from the PhysioGlove and to transfer them to the receiving center. No graphical presentation of the acquired ECG is enabled on the mobile platform.

The Mercury Telemedicine System (MTS) - proprietary software designed to be installed on a computer in the PhysioGlove ET receiving center (e.g., Nurse Station). The MTS is responsible to receive the recorded ECG data, save it in a database and present it to a 

{6}------------------------------------------------

medical professional who is capable of interpreting the results. The MTS is based on the PhysioGlove ES-1 software (originally cleared under K050674 and later incorporated under K103791).

The PhysioGlove ET system operates as follows:

The PhysioGlove records the patient's ECG and sends the data via Bluetooth to a mobile platform. Then, the recorded data is automatically transferred via the Internet using Wi-Fi to connect to a modem or via GPRS cellular communication. The data is sent to a receiving center (e.g., Nurse Station) to be reviewed by a medical professional who is capable of interpreting the ECG data.

### 6.13 Substantial Equivalence

The PhysioGlove ET System is substantially equivalent to the SelfCheck and PhysioGlove ES-1 with regard to intended use and indications for use. All three devices are reusable 12-Lead ECG examination systems, intended for recording of ECG. Both the PhysioGlove ET and the SelfCheck device are designed to be used by a patient or a health care professional who are concerned about their heart health and have problems that are suggestive of abnormal heart function.

Also, all three devices, the PhysioGlove ET, the SelfCheck Device, and the PhysioGlove ES-1 System, include hardware (HW) and software (SW) elements responsible for the collection and conditioning of the ECG signal so that it can be transmitted to another location for display. In the PhysioGlove ET, and similarly in the SelfCheck Device, the ECG recording may be conducted at home while the data display is performed at a remote location such as hospital, physician's office or a monitoring center.

The PhysioGlove ET System and the SelfCheck Device both use a Mobile Platform (e.g., a smart cell-phone, a compatible generic PDA or another communication device). The platform receives the ECG signal from the ECG acquisition element (the PhysioGlove or the PMP4 SelfCheck ECG) via Bluetooth and transmits it to the receiving center via cellular communication. Both devices contain methods for error testing and retransmission of the data.

The essential functionality of the SW elements included in the PhysioGlove ET system and its two predicates is substantially similar, i.e., receive the ECG data generated from the electrodes and transfer the digitized ECG data to the receiving center. Specifically, the PhysioGlove ET SW is a modification of the cleared PhysioGlove ES-1 SW (K103791) and thus, substantially equivalent the company's cleared device.

{7}------------------------------------------------

### 6.14 Performance Testing

Performance testing was conducted in order to demonstrate the performance, safety and usability of the PhysioGlove ET system. The testing plan included electrical safety and electromagnetic compatibility testing (according to IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety, and IEC 60601-1-2, Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic Compatibility - Requirements and Tests ), compliance with performance standards (IEC 60601-1-11, General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment, and IEC 60601-2-25,

Particular requirements for the basic safety and essential performance of electrocardiographs). SW validation testing and a usability study.

Test results indicated that the PhysioGlove ET functions as expected and is substantially equivalent to the predicate devices for its intended use. Therefore, it is concluded that the PhysioGlove ET system is substantially equivalent to its predicate devices without raising any new questions of safety and/or effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DXH/K133703](https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DXH/K133703)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
