MICROVIT MT-100

K973735 · Schiller America, Inc. · DSH · Aug 25, 1998 · Cardiovascular

Device Facts

Record IDK973735
Device NameMICROVIT MT-100
ApplicantSchiller America, Inc.
Product CodeDSH · Cardiovascular
Decision DateAug 25, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2800
Device ClassClass 2

Indications for Use

The ECG Holter device. MICROVIT MI-100. Is a recording device used to record long term Electrocardlographs for the diagnostics of symptomatic and asymptomatic arrhythmlas.

Device Story

Microvit MT-100 is a 2-channel solid-state Holter recorder for 24-hour ambulatory ECG monitoring. Device captures ECG signals from patient; stores data for subsequent analysis. Unlike predicate using PCMCIA cards, MT-100 utilizes serial optical transmission to transfer data to an analysis station. Signal integrity maintained with <20 microvolt deviation. Used in clinical settings or by patients in ambulatory environments under physician supervision. Healthcare providers review recorded ECG data to diagnose arrhythmias and ischemia, evaluate treatment efficacy, or assess post-surgical recovery. Benefits include portable, long-term cardiac rhythm monitoring for improved diagnostic accuracy of transient cardiac events.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 601-1), electromagnetic compatibility (IEC 801 series, EN 55011), and environmental testing (vibration, shock, temperature) per MIL-STD-81D and AAMI EC38-1994 standards.

Technological Characteristics

2-channel solid-state Holter recorder. Input impedance >10 MOhm; DC offset ±300mVDC; CMRR >80dB; bandwidth 0.05Hz to 40Hz. Dimensions 4.3" x 2.7" x 0.7"; weight 140g. Powered by 1.5V AA alkaline batteries. Data transmission via serial optical interface. Compliant with AAMI EC38-1994, IEC 601-1, and MIL-STD-81D.

Indications for Use

Indicated for human patients of any age, sex, or race requiring long-term ECG recording for diagnosis of symptomatic/asymptomatic arrhythmias (e.g., bradycardia, tachycardia, coronary artery disease, cardiomyopathy, hypertension, long QT syndrome), evaluation of palpitations, syncope, or dizziness, and monitoring of medical therapies or post-operative status (e.g., bypass, PTCA). Includes ST segment analysis for ischemia detection.

Regulatory Classification

Identification

A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/3 description: The image shows a date, "AUG 25 1998". The month is August, the day is the 25th, and the year is 1998. The text is in a bold, sans-serif font and is black in color. The background is white. K973735 # 510 K Summary according to 21 CFR 807.92 #### A1 Address SCHILLER America, Inc. 3002 Dow Avenue, #138 Tustin, CA 92780 Mr. Markus Maritz Contact Name: 1619) 635 6023 Tel: 26th September 1997 Date: #### Device Name A2 ﺮ | 1. Trade Name: | Microvit MT-100 | |-------------------------|-------------------------------------------| | 2. Common Name: | Solid State Holter Recorder | | 3. Classification Name: | Medical Magnetic Tape Recorder (Class II) | Classification Class: 74 DSH #### A3 Legally Marketed Device Legally Marketed Device to which this submitted device is compared: Braemar DL 700 Holter Monitor (K945130) #### A4 Description of Submitted Device The ECG Holter device, MICROVIT MT-100, is a recording device used to record long term Electrocardlographs for 24 hours, on two channels. #### A5 Intended Use The ECG Holter device. MICROVIT MI-100. Is a recording device used to record long term Electrocardlographs for the diagnostics of symptomatic and asymptomatic arrhythmlas. - 1. Patlent Population: human beings without restrictions of age, sex or race, - 2. Difference to the Predicate Device: mode of data transmission to the analysis station. Predicate Device: PCMCIA Card Submitted Device: Serial Optical Transmission. The maximum deviation between the original ECG signal and the transmitted and restored ECG signal Is less than 20 microvotts. This difference In fransmission rises no concern regarding the safety and effectiveness of the device, {1}------------------------------------------------ Image /page/1/Picture/3 description: The image shows the word "SCHILLER" in large, bold, black letters. Below the word is a horizontal line, and below that is the word "SWITZERLAND" in smaller, black letters. The text appears to be a logo or brand name. #### A6 Table of Comparison Predicate device: Braemar DL 700 Holter Monitor (K9451 30) | Dimensions<br>Weight | DL 700 | MT - 100 | |-----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|------------------------------------------------------------------| | | 6.0"x3.5"x0.95"<br>10 ounces incl. batteries | 4.3" x 2.7" x 0.7"<br>140g, 5 ounces incl.<br>batteries | | Battery:<br>Channels | 1.5V Alkaline AA<br>2 or 3 | SAME<br>2 | | Environmental Conditions<br>Operating temperature<br>Storage temperature<br>Relative humidity | 0 to 70 degree Celsius<br>-40 to +50 degree Celsius<br>10% to 90%<br>(non condensing) | +10 to +40 °C<br>-10 to +50 °C<br>25% to 95%<br>(non condensing) | | ECG Amplifier<br>Input Impedance<br>DC Offset<br>Common Mode rejection<br>Recording Bandwidth | > 5 MOhm<br>± 300mVDC max.<br>60dB<br>0.05 Hz to 50 Hz | > 10 MOhm<br>SAME<br>> 80dB<br>0.05Hz to 40 Hz | | Recording Time | 24 hours | SAME | | Electro Static Discharge /<br>Electro Magnetic<br>Compatibility<br>ESD | Fully functional below 4 kV<br>(Open Air)<br>No damage below 8kV<br>(Open Air) | SAME<br>SAME | | Radiated Emissions<br>Radiated Immunity | Less than 30 dB Microvolts<br>Less than 3 Volts per meter | SAME<br>SAME | ### Discussion of differences - 1 . Environmental conditions claimed are In accordance with MIL-STD-81D. No concern related to safety and effectiveness of the device. - 2. Input Impedance cialmed are In accordance with AAMI EC38-1994. No concern related to safety and effectiveness of the device. - 3. Common Mode Rejection claimed are In accordance with AAMI EC38-1994. No concern related to safety and effectiveness of the device. - 4. Recording Bandwidth claimed are in accordance with AAMI EC38-1994. No concern related to safety and effectiveness of the device. {2}------------------------------------------------ Image /page/2/Picture/3 description: The image shows the word "SCHILLER" in large, bold, black letters. Below the word "SCHILLER" is a horizontal line. Under the line, the word "SWITZERLAND" is written in smaller, black letters. The text appears to be a logo or brand name. #### Non-Clinical Tests B1 - 1. Electrical Safety and Reliability The Microvit MT-100 device has been tested to be in accordance with the following standards: - ANSI / AAMI, EC 38-1994: Ambulatory Electrocardlographs Type 1 Devloes - - IEC 601-1:1988: Safety of Medical Electrical Equipment part 1, General : requirements - IEC 801-1:1984, IEC 801-2:1991, IEC 801-3:1984, Electromagnetic Compatibility Test, Electrostatio Discharge, Radlo Frequency Electromagnetic Fleld. All tests are passed. - 2. Environmental Testing - The following environmental tests have been conducted: - mechanical tests (vibration, shock, fall, toppling) (IEC 601 1 , MIL-STD-81D) - climate tests (storage/operation temperatures) (MIL-STD-81D) All tests are passed. # 3. Electro Magnetic Compatibility (EMC) The following electromagnetic magnetic compatibility tests have been conducted: - EN 55011 class B Electromagnetic radiation . - IEC 801-1 Electromagnetic susceptibility - IEC 801-2 Electrostatic Discharge - IEC 801-3 Radlo Frequency Electromagnetic Field All tests are passed. The test reports are enclosed. ### 4) Data related to software quality Schiller has reviewed its software development process following the guideline * reviewer guidance for computer controlled medical devices undergoing 510 (k) review". Device software requirements, software structure chart, software development, software revision/ modification, software Identification, software verification, validation and testing are described in the data attached. #### B2 Cilnical Tests n.a. #### B3 Conclusions from Tests The fulfilment of the above standards ensures the safety and effectiveness of the submitted device. We consider the submitted device to be as safe and effective as the Predicate (Braemer) Device. LD/510 K Summary.doc {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three human profiles below three horizontal bars. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 25 1998 Mr. Markus Maritz Schiller America, Inc. 3002 Dow Avenue, Suite 138 Tustin, CA 92780 Re: K973735 Microvit MT-100 Requlatory Class: II (two) Product Code: 74 DSH Dated: May 29, 1998 Received: June 2, 1998 Dear Mr. Maritz: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ ### Page 2 - Mr. Markus Maritz This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Thomas J. Callahan Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ WELCH ALLYN 07/31/88 11:28 . 07/31/88 13:04 FAI ID:619-621-6610 2201 143 9101 AUG 04'98 FD4-0ST : DECS Page 1 of _1 810(k) Number (if known): 973735 Holter MICROVIT MT-100 Device Name:_ ## Indications For Use: The ECG Holter device, MICROVIT MT-100, is a recording device used to record long term Electrocardiographs for the diagnostics of symptomatic and asymptomatic arrhythmias, i.e. bradycardia, tachycardia, diagnostic of arrhythmias in patients with coronary artery disease, in patients after reanimation, in patients with different diseases such as cardiomyopathy, high blood pressure, long QT syndrome. A Holter ECG recording is also used for clarifying palpitations or syncopes and attacks of dizziness or for controlling medical therapies, or after operative treatments such as bypass operations or PTCA. The ST segment analysis of a Holter ECG is used for the detection of symptomatic or asymptomatic ischemias. # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRH, Office of Device Evaluation (ODE) | | |-------------------------------------------------------------------|---------| | <i>for</i> (Division Sign-Off) | | | Division of Cardiovascular, Respiratory, and Neurological Devices | | | 510(k) Number | K973735 | Prescription Use_V (Per 21 CFR 801.109) OR Over-The-Counter Use_ (Optional Pormat 1-2-90)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...