BIOZ DX HEMODYNAMIC MONITOR WITH PHILIPS 12-LEAD ECG, MODEL 5100
Applicant
Cardiodynamics International Corp.
Product Code
DSB · Cardiovascular
Decision Date
Sep 26, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The BioZ Dx device is intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The device is intended to monitor hemodynamic parameters within hospitals and other healthcare facilities providing patient care. The modified BioZ Dx device has these same intended uses.
Device Story
BioZ Dx is a noninvasive impedance cardiography (ICG) monitor; measures thoracic electrical bioimpedance via high-frequency, low-amplitude alternating current injected through sensors on neck and mid-axillary line. Device detects impedance changes over time to calculate hemodynamic parameters (e.g., stroke volume, cardiac output). Includes optional 12-lead ECG capability. Used in hospitals/healthcare facilities by trained personnel or under clinician supervision. Output displayed for clinical review to assist in hemodynamic assessment, diagnosis, and treatment decision-making. Benefits include noninvasive monitoring of cardiovascular status. Version 4.0b5 adds parameters (SVRI, LCWI, TFCI, Base Impedance, Heather Index, Q-C Interval, Stroke Index, Left Stroke Work Index) and corrects software bugs.
Clinical Evidence
Bench testing only. Quality assurance measures included risk analysis, design requirements/traceability, and unit/system-level software/firmware verification and system-level validation. Results demonstrate performance comparable to predicate devices.
Indicated for hemodynamic monitoring in hospital/healthcare settings for adult and pediatric patients. ICG function provides parameters including HR, SBP, DBP, MAP, SI, SV, CI, CO, SVR, SVRI, LCW, ACI, VI, TFC, Zo, TFCI, STR, STRI, PEP, LVET, TAC, TACI, LSWI, HI, QC, LCWI, SSRI, EDV, and EDI. 12-lead ECG function indicated for clinicians to evaluate ECG for diagnosis, treatment planning, treatment effectiveness, or symptom rule-out.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
CardioDynamics BioZ® Dx Hemodynamic Monitor with Philips 12-Lead ECG (K051228)
{0}------------------------------------------------
## Section 5 - Premarket Notification 510(k) Summary
SEF 26 2007
### GENERAL INFORMATION:
Device Name and Classification
| Product Name: | BioZ Dx Hemodynamic Monitor with 12-Lead ECG |
|-----------------------|---------------------------------------------------------|
| Classification Name: | Plethysmography, Impedance |
| Common Name: | Impedance Cardiograph (ICG) |
| Classification Panel: | Cardiovascular |
| Regulation Number: | 21 CFR 870.2770 (for ICG) and 21 CFR 870.2340 (for ECG) |
| Device Class: | II |
| Product Code: | 74 DSB (for ICG) and 74 DPS (for ECG) |
Manufacturer and Contact Person
CardioDynamics International Corporation 6175 Nancy Ridge Drive, # 300 San Diego, CA 92121 USA Phone: 858-535-0202 Fax: 858-535-0294
Paul L. Shaffer Director, Quality and Regulatory
Date of Summary Preparation: January 15, 2007
## SUBSTANTIAL EQUIVALENCE:
Predicate Devices:
K051228 - CardioDynamics BioZ® Dx Hemodynamic Monitor with Philips 12-Lead ECG K011439 - CardioDynamics BioZ .com Hemodynamic Monitor (for additional parameters) K041434 - Analogic C3TM ICG Monitor (for additional parameters) K010164 - GE Medical Systems ICG Module (for additional parameters)
Device Description:
This submission covers Version 4.0b5 of the BioZ Dx System Software for the BioZ Dx (K051228), which is a noninvasive impedance cardiography (ICG) device that provides hemodynamic parameters based on the measurement of thoracic electrical bioimpedance. The BioZ Dx measures this change in impedance by injecting a high frequency, low amplitude alternating electrical current through the thorax between a pair of sensors placed on the neck and another pair placed on the mid-axillary line at the xiphoid process level. By detecting and measuring the change in thoracic impedance as a function of time, the BioZ Dx is able to calculate stroke volume, cardiac output and many other hemodynamic parameters. The device additionally includes the capability of performing a standard 12-Lead ECG test.
The modified BioZ Dx software and firmware applies the same functions and scientific concepts as the predicate devices. Additional parameters have been added in this release that exists in the predicate devices, and a small number of routine software bugs were corrected with this release.
1/3
{1}------------------------------------------------
#### Intended Use:
The BioZ Dx device is intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The device is intended to monitor hemodynamic parameters within hospitals and other healthcare facilities providing patient care. The modified BioZ Dx device has these same intended uses.
3
Additional parameter outputs are being displayed based on calculations incorporated into predicate CardioDynamic (Systemic Vascular Resistance Index and Left Cardiac Work Index), Analogic (Thoracic Fluid Content Index, Base Impedance, Heather Index, Q-C Interval & Stroke Index) and GE (Left Stroke Work Index) devices. These additional parameter calculations use measurements and/or parameters which already existed in the BioZ Dx predicate device.
#### Technology:
The modified BioZ Dx device has the same technological characteristics as the predicate BioZ Dx device. A human factors improvement has been made to the cable and electrode accessories whereby the connections with the dual element electrode and patient leadwires have unique connector fittings to further ensure proper connections. The accessory change was determined to be insignificant (not requiring a 510(k) submission), and was documented to file and implemented into the current BioZ Dx.
#### General Safety & Effectiveness Concerns:
The instructions for use for this device contain the necessary cautions and warnings to provide for safe and effective use for the device.
Risk management is an essential element of the design and development process, and hazard analysis assessments performed on the modified BioZ Dx device at the System and Software levels revealed that the enhancements present no additional risks from the predicate devices. The software hazard analysis for the modified device in included in section 16.1 and the system hazard analysis is included in section 18.1.
CardioDynamics International Corporation is presently certified to ISO13485:2003 by Nemko. No significant audit findings have been presented since initial ISO 13485 certification in May of 1998. All products are designed and manufactured under CDIC's Quality Management System, which includes Design Control and Good Manufacturing Practices.
#### Performance Testing:
There have been no hardware modifications between the modified BioZ Dx and the predicate BioZ Dx monitor device, thus the relevant electrical safety standards have not been impacted by this product change. The improvements to the cable/electrode accessories were confined to the patient sensor connector mechanism only, and there were no impacts to the electromagnetic compatibility standards when compared to the predicate BioZ Dx device.
Design Review is required, at a minimum, at the completion of each step in the design control process (i.e. Design Input, Design Output, Design Verification, Design Validation and Design Transfer). The design review process identifies the required deliverables for each step in the process and verifies the content and accuracy of each deliverable.
{2}------------------------------------------------
The following quality assurance measures were conducted for the modified BioZ Dx device and are included in this submission:
- . Risk analysis
- . Design requirements and traceability
- . Unit and system level software and firmware verifications
- . System level validations
The results of verification and validation tests concluded that the functionality and performance characteristics of the modified BioZ Dx are comparable to the currently marketed predicate devices.
Conclusion:
The results of all testing demonstrate that the modified BioZ Dx Hemodynamic Monitor with 12-Lead ECG do not raise any new issues of safety, effectiveness or performance of the device when compare to the existing predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its body and wing. The eagle faces left. Encircling the eagle is the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 6 2007
CardioDynamics International Corporation c/o Mr. Paul L. Shaffer Director, Quality and Regulatory 6175 Nancy Ridge Drive #300 San Diego, CA 92121
Re: K070156
Trade/Device Name: BioZ Dx Hemodynamic Monitor with Philips 12-lead ECG Regulation Number: 21 CFR 870.2770 Regulation Name: Impedance Plethysmograph Regulatory Class: Class II (two) Product Code: DSB and DPS Dated: September 10, 2007 Received: September 12, 2007
Dear Mr. Shaffer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Paul L. Shaffer
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Blymmerman for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: BioZ Dx HemoDynamic Monitor and Philips 12-lead ECG
Indications For Use:
For the ICG function:
The BioZ Dx Hemodynamic Monitor is intended to monitor and display a patient's hemodynamic parameters. These parameters include:
- Heart Rate (HR) Systolic Blood Pressure (SBP) Diastolic Blood Pressure (DBP) Mean Arterial Blood Pressure (MAP) Stroke Index (SI) Stroke Volume (SV) Cardiac Index (CI) Cardiac Output (CO) Systemic Vascular Resistance (SVR) Systemic Vascular Resistance Index (SVRI) Left Cardiac Work (LCW) Acceleration Index (ACI) Velocity Index or Index of Contractility (VI, IC) Thoracic Fluid Content (TFC) Base Impedance (TFI or Zo)
Thoracic Fluid Content Index (TFCI) Systolic Time Ratio (STR) Systolic Time Ratio Index (STRI) Pre-Ejection Period (PEP) Left Ventricular Ejection Time (LVET) Total Arterial Compliance (TAC) Total Arterial Compliance Index (TACI) Left Stroke Work Index (LSWI) Heather Index (HI) Q-C Interval (QC) Left Cardiac Work Index (LCWI) Electrocardiograph (ECG) Systemic Stroke Resistance Index (SSRI or SSVRI) End Diastolic Volume (EDV) End Diastolic Index (EDI)
For the optional 12-lead ECG function:
Where the clinician decides to evaluate the electrocardiogram of adult and pediatric as part of decisions regarding possible diagnosis, potential treatment, effectiveness of treatment or to ruleout causes for symptoms.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blummen
iciovi 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.