DISPOSABLE ECG MONITORING ELECTRODES, MODEL 4110, 4140
K990110 · Maersk Medical A/S · DRX · Dec 6, 1999 · Cardiovascular
Device Facts
Record ID
K990110
Device Name
DISPOSABLE ECG MONITORING ELECTRODES, MODEL 4110, 4140
Applicant
Maersk Medical A/S
Product Code
DRX · Cardiovascular
Decision Date
Dec 6, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2360
Device Class
Class 2
Attributes
Pediatric
Indications for Use
ECG monitoring electrodes for short-term use (< 24 hours) in adults (Model 4110) and infants (Model 4140).
Device Story
Niko ECG Monitoring Electrodes are disposable, pregelled electrodes used to acquire electrical signals from the heart for ECG monitoring. The device consists of an Ag/AgCl sensor element (10-20 mm diameter) and an adhesive backing (20-70 mm diameter or rectangular/square). Electrodes are applied to the patient's skin by clinical staff in a hospital or clinical setting. The device acts as a conductive interface between the patient's skin and an external ECG monitoring system. By facilitating the transmission of cardiac electrical activity to the monitor, the electrodes enable healthcare providers to observe heart rhythms, aiding in the diagnosis and management of cardiac conditions. The disposable nature of the electrodes helps maintain hygiene and signal integrity during short-term monitoring sessions.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Ag/AgCl sensor element (10-20 mm diameter); adhesive backing (20-70 mm diameter or rectangular/square); disposable; bulk packaged in OPP/PE laminated pouches.
Indications for Use
Indicated for short-term (< 24 hours) ECG monitoring in adults (Model 4110) and infants (Model 4140).
Regulatory Classification
Identification
An electrocardiograph electrode is the electrical conductor which is applied to the surface of the body to transmit the electrical signal at the body surface to a processor that produces an electrocardiogram or vectorcardiogram.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9. The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Electrocardiograph Electrodes.” See § 870.1(e) for availability information of guidance documents.
Predicate Devices
Disposable ECG Monitoring Electrode Models 4110 and 4140 (K950472)
Related Devices
K990113 — DISPOSABLE ECG MONITORING ELECTRODES, MODELS 4500,4530, 4533, 4539, 4540, AND 4560 · Maersk Medical A/S · Dec 6, 1999
K990115 — DISPOSABLE ECG MONITORING ELECTRODES, MODELS, 4420, 4440 · Maersk Medical A/S · Dec 6, 1999
K990112 — DISPOSABLE ECG MONITORING ELECTRODES, MODELS 4060, 4610 · Maersk Medical A/S · Dec 6, 1999
K990111 — DISPOSABLE ECG MONITORING ELECTRODES, MODELS 4520, 4535, 4570 · Maersk Medical A/S · Dec 6, 1999
K091856 — CONMED ECG MONITORING ELECTRODES , MODEL 1700 ( INCLUDING 2500), 1750, 2700, 2710, 1800, SERIES · Conmed Corporation · Aug 7, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
K990110
# 510 (k) SUMMARY
## l. ADMINISTRATIVE
- Submitter: Maersk Medical A/S Niko Business Unit Enqmosen 1 DK-3540, Lynge Denmark Phone No .: 011 45 48 16 70 30
Contact Person: Mr. Christian Pelch
Date of Preparation: October 15, 1999
### II. DEVICE NAME
Proprietary Name: Niko ECG Monitoring Electrodes Common Name: ECG Electrode Classification Name: Electrocardiograph Electrode
#### III. PREDICATE DEVICES
Disposable ECG Monitoring Electrode Models 4110 and 4140 (K950472); Nikomed USA, Inc.
# IV. DEVICE DESCRIPTION
Pregelled electrodes are of Ag/AgCl construction with a sensor element area between 10 and 20 mm in diameter, and an adhesive part between 20 and 70 mm in diameter or rectangular/square in shape. Electrodes are bulk packaged in OPP/PE laminated pouches; 60/pouch; 300/box.
# V. INTENDED USE
ECG monitoring electrodes for short-term use (< 24 hours) in adults (Model 4110) and infants (Model 4140).
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC - 6 1999
Maersk Medical A/S c/o Mr. Richard A. Hamer Consultant to Maersk Medical A/S Richard Hamer Associates, Inc. P.O. Box 16598 Ft. Worth, TX 76132
K990110 Re: Disposable ECG Monitoring Electrodes Models 4110 and 4140 Requlatory Class: II (two) Product Code: DRX October 15, 1999 Dated: October 18, 1999 Received:
Dear Mr. Hamer:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act
{2}------------------------------------------------
Page 2 - Mr. Richard A. Hamer
This letter will allow you to beqin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at Also, please note the regulation entitled, (301) 594-4639. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
L. Ruth Tiller
lia M. Witten, Ph.D., M.D. Acting Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Page of 1 1
510(k) Number (if known): K990110
Niko ECG Monitoring Electrodes Model 4110 and 4140 Device Name:
Indications for Use:
ECG monitoring electrodes for short-term use (< 24 hours) in adults (Model 4110) and infants (Model 4140).
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dan K. Miller
Cardiovascular, Respiratory, and Neurological Devices 510(k) Number
Prescription Use
(Per 21 CFR 801.109)
ાર
Over-the-Counter Use _________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2 96)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.