K962115 · Cambridge Heart, Inc. · DRX · Aug 29, 1996 · Cardiovascular
Device Facts
Record ID
K962115
Device Name
CAMBRIDGE HEART HI-RES ECG ELECTRODE
Applicant
Cambridge Heart, Inc.
Product Code
DRX · Cardiovascular
Decision Date
Aug 29, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2360
Device Class
Class 2
Indications for Use
The Cambridge Heart Hi-Res™ ECG Electrode is intended for short term use to measure electrocardiogram signals at rest and during ECG stress testing with the CH 2000 Stress Test System and successor systems. This device is supplied non-sterile and is intended for one time use only.
Device Story
Pregelled, single-use, multi-segment Ag/AgCl electrode; designed for use with Cambridge Heart CH 2000 Stress Test System. Input: electrical signals from skin surface. Device comprises four segments: one center segment for standard ECG signal; three ring segments for additional ECG and trans-thoracic (respiratory) impedance measurements. Noise reduction algorithm on CH 2000 uses these multi-segment inputs to perform adaptive cancellation of baseline wander and movement/muscle artifacts. Output: processed ECG signal for display and printout. Used in clinical settings during stress testing; operated by or under direction of physician. Benefits: improved signal quality via artifact reduction; non-polarizing center segment allows for defibrillation if required.
Clinical Evidence
Bench testing only. Device tested for performance per ANSI/AAMI EC12-1991 (AC impedance, DC offset, bias current tolerance) and biocompatibility per ISO 10993-5 (cytotoxicity). Results demonstrate equivalence to predicate electrodes. Respiratory impedance measurement method validated against AAMI ES1 safety standards.
Indicated for short-term measurement of ECG signals at rest and during stress testing in patients undergoing cardiac evaluation with the CH 2000 Stress Test System.
Regulatory Classification
Identification
An electrocardiograph electrode is the electrical conductor which is applied to the surface of the body to transmit the electrical signal at the body surface to a processor that produces an electrocardiogram or vectorcardiogram.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9. The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Electrocardiograph Electrodes.” See § 870.1(e) for availability information of guidance documents.
{0}
8·27-1996 8:00PM
FROM CAMBRIDGE HEART 617 271 9488
P. 5
K962115
AUG 29 1996
# SECTION 2.0 510 (K) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
As required under Section 12, part (a)(i)(3A) of the Safe Medical Device Act of 1990, an adequate summary of any information regarding safety and effectiveness follows:
## 2.1 General Information
**Name and Address of Submitter:**
Cambridge Heart, Inc.
One Oak Park
Bedford, MA 01730
**Contact:**
Robert T. Miragliuolo
V.P. of Regulatory and Clinical Affairs
**Date of Summary:**
May 15, 1996
**Name of Device:**
Cambridge Heart Hi-Res™ ECG Electrode
**Common/Usual Name of Device:**
ECG Electrode
**Device Classification:**
Electrocardiograph Electrode
**Predicate Devices:**
Smart™ Bi-Lectrode (K791006)
Quinton Quik-Prep Electrode (K782079)
3M Red Dot Electrode (K821438)
HP M1176A Cardiac Monitor (K882609)
000010
{1}
8-27-1996 8:00PM FROM CAMBRIDGE HEART 617 271 9488
P. 6
## 2.2 Device Description and Intended Use
The Cambridge Heart Hi-Res™ ECG Electrode is a pregelled, single use, multi-segment Ag/AgCl electrode used in measuring electrocardiogram signals. It was designed specifically for use with the Cambridge Heart, Inc. CH 2000 Stress Test System and successor systems.
The Cambridge Heart Hi-Res™ ECG Electrode is intended for short term use to measure electrocardiogram signals at rest and during ECG stress testing with the CH 2000 Stress Test System and successor systems. This device is supplied non-sterile and is intended for one time use only.
The Hi-Res™ ECG Electrode is constructed of a polyester basepad, four silver-silver chloride ink electrode segments, an adhesive coated die cut foam, a conductive gel, and a silicone coated polystyrene carrier (release liner).
The electrode comprises four closely spaced electrode segments, consisting of three equal sized segments equidistant around a center segment, each capable of receiving ECG signals. By itself, the center segment functions as a standard Ag/AgCl ECG electrode. It is non-polarizing should defibrillation be required when the electrode is in place. The ECG signal received by the center segment is the primary source of the ECG reported on the display screen and on printouts of the CH 2000 Stress Test System.
The three ring segments are used to receive additional ECG signals and are used, along with the center segment, to obtain real time measures of electrode surface-to-skin impedance and trans-thoracic (respiratory) impedance. These additional ECG signals and impedance measurements are used by the noise reduction algorithms on the CH 2000 to reduce baseline wander and movement artifact from the ECG.
The Cambridge Heart Hi-Res™ ECG Electrode can be used to measure electrocardiogram signals at rest and during stress testing with the CH 2000 Stress Test System and successor systems.
000011
{2}
8-27-1996 8:01PM
FROM CAMBRIDGE HEART 617 271 9488
P. 7
The Hi-Res™ ECG Electrode reduces baseline wander, movement artifact and muscle artifact from the ECG, when used with the CH 2000 Stress Test system or successor systems.
The center segment of the electrode is non-polarizing should defibrillation be required while the electrode is in place.
The electrode is flexible, has a low profile, is capable of adhering to the skin under the variety of conditions particular to exercise stress testing, and is pregelled as required for lowering skin-electrode impedance.
The Hi-Res™ ECG Electrode was designed to specifications, confirmed by testing, that meet or exceed the following voluntary guidelines and standards as they apply to disposable ECG electrodes used for exercise testing: ISO 10933-5 and ANSI/AAMI EC12-1991.
The Hi-Res™ ECG Electrode is used at LL, V2, V4, V5, V6, I and H. It interfaces with a high resolution patient lead cable supplied by Cambridge Heart, Inc., for use with the CH 2000 Stress Test System. This high resolution patient lead cable is similar to the standard 14 electrode patient lead cable included in the 510(k), K950018, for the CH 2000 Stress Test System with the exception that the leadwires and electrode connectors have been changed so as to be used with the Hi-Res™ ECG Electrode. Testing has confirmed that the high resolution patient cable meets the AAMI EC-53 standard.
A noise reduction algorithm has been added to the CH 2000 specifically for recognizing and processing the additional ECG and impedance signals from the Hi-Res™ ECG Electrode. This algorithm creates a weighted average of the ECG signals from the segments of the each electrode to reduce noise through adaptive cancellation of artifact in the ECG. In creating the average, the algorithm also uses the impedance signals to cancel baseline noise artifact. This noise reduction process creates an ECG signal that can then be processed by the CH 2000 like a standard ECG signal from a conventional single segment electrode.
000012
{3}
8-27-1996 8:01PM
FROM CAMBRIDGE HEART 617 271 9488
P. 8
The CH 2000 Stress Test System and the HP M1176A device, K882609, measure respiratory impedance in a similar manner. Both devices inject a low level high-frequency current through the torso, using existing electrodes, and measure the resulting voltage drop across the body.
The Hi-Res™ ECG Electrode is substantially equivalent to the Smart™ Bi-Lectrode (K791006), the Quinton Quik-Prep Electrode (K782079), and the 3M Red Dot Electrode (K821438). Refer to Exhibit 2.1 for a summary of similarities and differences.
## 2.3 Summary of Similarities and Differences
The table below shows the frequency and current used by the CH 2000 and the predicate device for the purpose of measuring respiratory impedance. The CH 2000 current is significantly lower than the maximum of 280 µA allowed by AAMI standard ES1, Section 3.5 for a frequency of 28 kHz.
| | CH 2000 | HP M1176A |
| --- | --- | --- |
| Frequency | 28 kHz | 39 kHz |
| Maximum RMS current | 100 µA | 68 µA |
Please refer to the exhibit 2.1 on the following page for a summary of the similarities and differences.
000013
{4}
8-27-1996 8:02PM
FROM CAMBRIDGE HEART 617 271 9488
P. 9
EXHIBIT 2.1
SUMMARY OF SIMILARITIES AND DIFFERENCES
| Characteristics | Cambridge Heart Hi-Res™ ECG Electrode | Quinton Quik Prep Electrode | Red Dot Electrode | Smart™ Bi Electrode |
| --- | --- | --- | --- | --- |
| Indications for Use | The Cambridge Heart Hi-Res™ ECG Electrode is intended for short term use. In measure electrocardiogram signals at rest and during atres testing with the CH 2000 stress Test System or successor systems. | The Quinton Quik Prep electrode represents an addition to Quinton's line of disposable ECG electrodes. The electrode is designed for stress or exercise test as well as longer term monitoring. | Red Dot Electrodes are designed for obtaining electrocardiographic signals from subjects who are being monitored in all areas of a hospital or in a clinical setting. | Intended for patient ECG monitoring |
| Number of Electrode Samples | Four | One | One | Two |
| Conductive System Electrode Conductive Gel | Silver-Silver Chloride YES | Silver-Silver Chloride YES | Silver-Silver Chloride YES | Silver-Silver Chloride Unknown |
| SAFETY STANDARDS Biological Response Evaluation | Protocol- ANSYAAMEC12-1991 Cytotoxicity Kligman Sensitivity Primary Examination PASSED | Not Specified | Protocol- Unknown Cytotoxicity Human Drize Compatible Inhibitory Protocols PASSED | Unknown |
{5}
8-27-1996 8:05PM
FROM CAMBRIDGE HEART 617 271 9488
P. 10
EXHIBIT 2.1 (continued)
SUMMARY OF SIMILARITIES AND DIFFERENCES
| Performance Standard ANGRAM (EC 12.1991) | Cambridge HeartRI-Res® ECG Electrode | Quinton Quik Prep Electrode | 3M Red Dot Electrodes 90.2120 | Smart 76RI Electrode |
| --- | --- | --- | --- | --- |
| AC Impedance (10Hz) Avg <2 KΩ Max <3 KΩ | Shell Life Labeling - 9 months
CENTER DOT SEGMENT
AC Impedance (10Hz) Average: 0.25 KΩ Max: <3 KΩ | Shell Life Labeling - 9 months
Impedance (60Hz) - 1.5 KΩ | Shell Life Labeling - 5 year
AC Impedance (10Hz) - Average: 488 KΩ Max: 3 KΩ | No Information Available |
| DC Offset Voltage <100mV after 1 min
Combined Offset Instability and Noise <150μV | DC Offset Voltage - Average: 0.3 mV Max: <100 mV | DC Offset Voltage - <20mV | DC Offset Voltage - Average: 1.72mV Max: <100 mV | |
| SDR <100mV @5 sec
(Bias Current Tolerance <100 mV over 6 hrs) | DC Offset Instability and Internal Noise - Average: 0.4 μV/sec Max: <100mV @5 sec
Bias Current Tolerance - Average: 0.4 mV over 6 hours Max: <100mV over 6 hours
ANSIAMM EC-12-1991 | DC Offset Dist. Recovery - <100mV @5 sec
Bias Current - Not Specified
(Not specified) | DC Offset Instability and Internal Noise - 54 mV over 1 sec
Defibrillator Recovery - 23.8mV @5 sec
Bias Current Tolerance - 11.4 mV over eight hours
[Proposed AAMI Standard May 1991] | |
{6}
8-27-1996 8:08PM
FROM CAMBRIDGE HEART 617 271 9488
P. 11
## 2.4 Substantial Equivalence Decision Tree
The 510(k) "Substantial Equivalence" Decision Making Process in ODE Guidance Memo #K86-3, was used in making the determination of substantial equivalence. The answer to these questions lead to a determination of substantial equivalence.
1. Does New Device Have the Same Indication Statements?
Yes. The Hi-Res™ ECG Electrode and the predicate electrode devices are all indicated for the same basic purpose of conducting the ECG signal from the surface of the skin to the electrocardiograph. The Hi-Res™ ECG Electrode is indicated for short time use, at rest and during stress testing; the predicate devices are intended for the same use and also for long term monitoring. The HP M1176A is a cardiac device for obtaining ECG and respiratory measurements. The CH 2000 Stress Test System is intended for the recording of ECGs and vector cardiograms by a physician, or under the direction of a physician, at rest and during ECG stress testing.
2. Does the New Device Have the Same Technology Characteristics, e.g., Design, Materials, Etc.?
Yes. The Hi-Res™ ECG Electrode functions in the same manner as the predicate devices in that it is a transducer of the electrical signals emanating from the heart by converting an ionic signal in tissue to a metal conductive electrical signal. It has a similar conductive system consisting of silver-silver chloride elements to increase the rate of depolarization; a conductive gel to enhance skin contact and conductivity; and an adhesive on a flexible material, made from a relatively inert polymer, to hold the electrode on to the skin. The multi-segment design of the Hi-Res™ ECG is an extension of the bi-segment design of the Smart™ Bi-Lectrode.
000016
{7}
8-27-1996 8:08PM
FROM CAMBRIDGE HEART 617 271 3488
P. 12
The CH 2000 Stress Test System and the HP M1176A device, K882609, measure respiratory impedance in a similar manner. Both devices inject a low level high-frequency current through the torso, using existing electrodes, and measure the resulting voltage drop across the body.
3. Are the Descriptive Characteristics Precise Enough to Ensure Equivalence?
NO. The descriptive characteristics are not precise enough to ensure safety of effectiveness equivalence.
4. Are Performance Data Available to Assess Equivalence
The Hi-Res™ ECG Electrode was tested and met the performance requirements of ANSI/AAMI EC12-1991 and the biocompatibility requirements of ISO 10993-5. In their respective premarket applications, the Quik Prep Electrode and the 3M Red Dot Electrode were tested to demonstrate compliance with certain voluntary standards as they applied to ECG electrodes. The requirements of these standards are similar to the requirements of ANSI/AAMI EC12-1991 and ISO10933-5 for performance and biocompatibility. The CH 2000 Stress Test System's method of measuring respiratory impedance meets the AAMI Standard For Safe Current Limits for Electromedical Apparatus ES1, Section 3.5.
5. Performance Data Demonstrates Equivalence?
Yes. The results of biocompatibility testing on the Hi-Res™ ECG Electrode and on the predicate devices were substantially equivalent. The results of the performance testing on the Hi-Res™ ECG Electrode and the predicate devices were substantially equivalent. The results of the testing demonstrate
000017
{8}
8-27-1996 8:09PM FROM CAMBRIDGE HEART 6'7 27' 9435
that the manner in which CH 2000 Stress Test System measures respiratory impedance is substantially equivalent to the predicate device.
Based on the FDA's decision tree, the Hi-Res™ ECG Electrode is substantially equivalent to the Smart™ Bi-Lectrode, the Quinton Quik-Prep Electrode, and the 3M Red Dot Electrode, and the HP M1176A and CH 2000 Stress Test System are substantially equivalent.
000018