PACEART WRIST ELECTRODES

K960721 · Paceart, Inc. · DRX · Jul 18, 1996 · Cardiovascular

Device Facts

Record IDK960721
Device NamePACEART WRIST ELECTRODES
ApplicantPaceart, Inc.
Product CodeDRX · Cardiovascular
Decision DateJul 18, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2360
Device ClassClass 2

Indications for Use

Paceart Wrist Electrodes are intended to be used for home monitoring of pacemaker patients via telephone monitoring systems.

Device Story

Paceart Wrist Electrodes consist of flat Type 304 stainless steel electrodes mounted on metallic expansion watchbands. Used in pairs, one on each wrist, to capture ECG signals for home telephonic monitoring. Device connects to trans-telephonic ECG transmitters (e.g., Paceart Cardiophone) via shielded cables with oversized 3.5mm pins to prevent power outlet insertion. No gel or skin preparation required. Designed for short-term use (max 5 minutes). Operated by patients in home settings. Output is an ECG signal transmitted via telephone to a healthcare provider for clinical review. Benefits include ease of use for elderly patients requiring pacemaker monitoring without complex skin preparation.

Clinical Evidence

Bench testing performed per AAMI EC12-1991 (paragraphs 4.2.2.2 and 4.2.2.3) for d-c offset and internal noise, showing performance better than AAMI requirements and equivalent to predicate. Clinical comparison conducted on a human subject wearing both subject and predicate electrodes simultaneously; ECG recordings were virtually identical.

Technological Characteristics

Materials: Type 304 stainless steel electrode and expansion band. Sensing: Passive metallic electrode. Connectivity: Wired connection via 3.5mm jack to ECG transmitter. Form factor: Wrist-worn band. Sterilization: Not specified (reusable).

Indications for Use

Indicated for home telephonic ECG monitoring of patients, particularly elderly pacemaker patients. Not for general-purpose or long-term ECG monitoring.

Regulatory Classification

Identification

An electrocardiograph electrode is the electrical conductor which is applied to the surface of the body to transmit the electrical signal at the body surface to a processor that produces an electrocardiogram or vectorcardiogram.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9. The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Electrocardiograph Electrodes.” See § 870.1(e) for availability information of guidance documents.

Predicate Devices

Submission Summary (Full Text)

{0} Page A 2 K960721 Paceart Wrist Electrodes Paceart Associates LP 510(k) Submission Paceart Wrist Electrodes 510(K) Summary JUL 18 1996 (1) Submitter Information: Name: Paceart Associates LP. Address: 22 Riverview Drive, Wayne New Jersey 07470 Telephone Number: 201-696-1122 Contact Person: Dr. George Myers, 201-438-2310 (2) Names: Trade: Paceart Wrist Electrodes Common Usual Name: Electrocardiograph Wrist Electrodes Classification Name: Electrode, Electrocardiograph (3) Classification, Panel Class II, 74DRX (4) Predicate Device: Medtronic 9427 Wrist Electrode (5) Description Paceart Wrist Electrodes are metallic electrodes with elastic wristbands intended for electrocardiographic monitoring. Physically the device consists of a flat metallic electrode fastened to a familiar metallic expansion band of the type commonly used with wrist watches. Ordinarily, the patient would have one electrode and band on each wrist. Each electrode has attached to it a single cable, attached to a connector with an oversized pin. The cable is used to provide the connection to the monitor. No gel or other skin preparation is indicated, and in general is not necessary: the patient slips on the bracelet so that the electrode is on the inside of the wrist, and connects the cable connector to a jack on the monitor. The plug is of sufficiently large diameter so that it will not fit into a power outlet. The wrist electrode is used for home telephonic monitoring purposes, and in general will be in contact with the skin for a maximum of five minutes; it is not intended for use as a general-purpose electrocardiographic electrode or for long-term monitoring. The major patient population who is likely to use such an electrode is elderly (for example, pacemaker patients). {1} Paceart Wrist Electrodes Page A 3 (6) Intended Use Paceart Wrist Electrodes are intended to be used for home monitoring of pacemaker patients via telephone monitoring systems. (7a) Predicate Devices The predicate device for the Paceart Wrist Electrode is the Medtronic 9427 Wrist Electrode (7b) Testing The Paceart Wrist Electrode has been compared in tests to the predicate device with respect to noise and offset voltage. The values for the two devices are equivalent. Test reports show that the Paceart Electrode and strap are manufactured from a biologically-compatible stainless steel, Type 304. These tests all show that the Paceart Wrist Electrode is safe and effective for its intended use. {2} Paceart Wrist Electrodes Page A 4 ## Photograph ## Paceart Wrist Electrode ![img-0.jpeg](img-0.jpeg) ![img-1.jpeg](img-1.jpeg) {3} Paceart Wrist Electrodes Page A 5 Paceart Associates LP 510(k) Submission Paceart Wrist Electrodes Overview December 4, 1995 I. Introduction and Intended Use Paceart Wrist Electrodes are metallic electrodes with elastic wristbands intended for electrocardiographic monitoring. Physically the device consists of a flat metallic electrode fastened to a familiar metallic expansion band of the type commonly used with wrist watches. Ordinarily, the patient would have one electrode and band on each wrist. Each electrode has attached to it a single cable attached to a connector with an oversized pin. The cable is used to provide the connection to the monitor. No gel or other skin preparation is indicated, and in general is not necessary: the patient slips on the bracelet so that the electrode is on the inside of the wrist, and connects the cable to a jack on the monitor by menas of the connector. The plug is of sufficiently large diameter so that it will not fit into a power outlet. The wrist electrode is used for home telephonic monitoring purposes, and in general will be in contact with the skin for a maximum of five minutes; it is not intended for use as a general-purpose electrocardiographic electrode or for long-term monitoring. Since the major patient population who is likely to use such an electrode is quite elderly (for example, pacemaker patients), it is important that the electrode can be easily attached with a minimum of adjustment and preparation. II. Construction The device is shown in Figure 1, which shows both the electrode plate and the strap. The electrode material is Type 304 stainless steel, which is an accepted material for implants, a far more stringent application. The materials section has a certificate from the manufacturer showing the composition of this steel. The watch band is a commercial watch band which has been in commercial distribution in the United States for this purpose for some time, with no reported cases of any skin irritations. The cables are shielded, and the shield is connected to a segment of the connector. One end of each cable (one from each electrode) is permanently attached to one of the electrodes, and cannot be removed by the user. The other ends are permanently {4} Paceart Wrist Electrodes Page A 6 connected to a small jack with a pin which is 3.5 mm in diameter, which is too large to fit into a power outlet. The connections are shown in Figure 2, and the jack is shown schematically in Figure 3. The jack is specifically made to mate with the Paceart Cardiophone, a trans-telephonic ECG transmitter. However, the electrodes can be used with any other transmitter with a compatible socket. Figure 4 shows more detail of the construction of the electrode. ## III. Predicate Device The predicate device for the Paceart wrist electrode is the Medtronic 9427 wrist electrode, which has the same intended use and virtually the same construction. The characteristics of the two electrodes are compared in the appropriate section of this submission. ## IV. Biocompatibility The only metal in contact with the patient's skin is Type 304 Stainless Steel (the material of both the bracelet and the electrode). Type 304 stainless steel is generally considered to be biologically inert and, in fact, is used for commercial metal watchbands. Since the electrode itself is worn for only short periods of time and is considered inert, there is no risk of biological reaction. Metals are not covered by the tri-partite agreement. No plastic materials are in contact with the patient. ## V. Bench Tests No AAMI standard exists for this type of electrodes. Therefore, the electrodes have been tested for d-c offset and internal noise, according to AAMI standard EC12-1991, Paragraph 4.2.2.3, and AAMI standard EC12-1991, Paragraph 4.2.2.2. The tests were also run on the predicate device, and the results for the two devices are compared. Results and protocols are in the bench test section. Since the electrodes are in general used without gel, the offset voltages were measured both with gel (as specified by the standard) and without gel. The Paceart electrodes had a lower offset voltage without gel than did the Medtronic electrodes. {5} Paceart Wrist Electrodes Page A 7 For all tests, both devices had comparable results, and both are better than the AAMI requirements for reusable electrodes. ## VI. Clinical Comparison The Paceart and Medtronic electrodes were compared with a human subject, who wore both sets of electrodes simultaneously, and recorded an ECG using both electrodes at the same time. No gel was used in these tests. The results, shown in the Clinical Section, are virtually identical. ## Figures 1. Electrode plate and strap (drawing) 2. Connections 3. Drawing of connecting jack 4. Detail of electrode construction {6} ![img-2.jpeg](img-2.jpeg) Figure 1 Page 1 {7} ![img-3.jpeg](img-3.jpeg) Figure 2 {8} Cable to Electrodes ![img-4.jpeg](img-4.jpeg) Figure 3 Body of Connector Pin 3.5 mm diameter Paceart Wrist Electrode Connector Schematic Drawing {9} ![img-5.jpeg](img-5.jpeg)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...