← Product Code [DRX](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DRX) · K090180

# TENDER-TRODE PREWIRED PEDIATRIC ECG ELECTRODE (K090180)

_Vermed, Inc. · DRX · Jun 17, 2009 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DRX/K090180

## Device Facts

- **Applicant:** Vermed, Inc.
- **Product Code:** [DRX](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DRX.md)
- **Decision Date:** Jun 17, 2009
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 870.2360
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The Tender-Trode® Prewire is a disposable, noninvasive electrode intended to be used for resting ECG monitoring on neonatal and pediatric patients in a hospital setting. These electrodes can be used for shortterm or long-term applications, up to 24 hours. The Tender-Trode® prewired electrodes are single-use, nonsterile, disposable and are to be used on intact (uninjured) skin.

## Device Story

Tender-Trode® Prewired ECG Electrodes are disposable, non-invasive sensors for neonatal and pediatric ECG monitoring. Device consists of conductive solid gel, silver/silver chloride plated ABS eyelet, adhesive cloth substrate, vinyl label, and 24-inch wire with female socket. Used in hospital settings for resting ECG; applied to intact skin for up to 24 hours. Electrodes capture electrical signals from patient skin, transmitting them via integrated wire to monitoring equipment. Healthcare providers use output to monitor cardiac activity. Benefits include reliable signal acquisition for patient monitoring.

## Clinical Evidence

Bench testing only. Device demonstrated conformance with AAMI EC12:2000 standards for disposable ECG electrodes, including labeling, shelf life, packaging, and safety. Biocompatibility testing (cytotoxicity, sensitization, primary skin irritation) per ISO 10993-1 confirmed materials are non-irritating, non-cytotoxic, and non-sensitizing.

## Technological Characteristics

Conductive solid gel; silver/silver chloride plated ABS eyelet; adhesive cloth substrate; vinyl label; 24" wire with .060" dia. female socket; polystyrene release liner. Conforms to AAMI EC12:2000. Biocompatibility per ISO 10993-1. Non-sterile, single-use.

## Regulatory Identification

An electrocardiograph electrode is the electrical conductor which is applied to the surface of the body to transmit the electrical signal at the body surface to a processor that produces an electrocardiogram or vectorcardiogram.

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9. The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Electrocardiograph Electrodes.” See § 870.1(e) for availability information of guidance documents.

## Predicate Devices

- Bunny Electrode/Stealth Electrode Pre Wired ECG, Lead-Lok, Inc. ([K911529](/device/K911529.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# JUN 17 2009

#### 510(k) SUMMARY 16.

## 510(K) SUMMARY of Safety and Effectiveness (as required by 21 CFR 807.92)

Date:

January 12, 2009

(800) 245-4025

Manufacturer:

Vermed, Inc. 9 Lovell Drive Bellows Falls, VT 05101 Registration Number 1219288

Telephone:

Contact Person:

Marc Fillion VP of Quality & Regulatory Affairs (800) 245-4025, Extension 1205 Fax Number: (802) 463-9228

Device Trade Name:

Tender-Trode® Prewired ECG Electrodes

Common Name:

Disposable ECG Monitoring Electrode

Electrocardiograph Electrode Classification Name:

74 DRX Regulatory Reference:

Predicate Device:

Bunny Electrode/Stealth Electrode Pre Wired ECG, Lead-Lok, Inc., K911529

A conductive solid gel electrode consisting of a Description: conductive adhesive gel, a silver/silver chloride plated ABS eyelet, an adhesive cloth substrate, vinyl label, 24" wire w/.060 dia. female socket and polystyrene release liner.

{1}------------------------------------------------

Intended Use:

The Tender-Trode® Prewire is a disposable, noninvasive electrode intended to be used for resting ECG monitoring on neonatal and pediatric patients in a hospital setting. These electrodes can be used for shortterm or long-term applications, up to 24 hours. The Tender-Trode® prewired electrodes are single-use, nonsterile, disposable and are to be used on intact (uninjured) skin.

# Physical/Technical Comparison:

Tender-Trode® Prewired ECG Electrodes are technologically equivalent to the predicate device. Physical and technical characteristics, including materials used in construction, size, intended use and conductive gel are comparable.

The device was tested and demonstrated conformance Performance Summary: with internationally recognized consensus standards. Test results indicate that this device meets or exceeds the performance specifications as established in EC12:2000 for Disposable ECG Electrodes. In addition, the device was found to meet EC12:2000 AAMI standard for labeling, shelf life, packaging and safety. A certification to conformance with this standard has been provided.

The solid conductive gel underwent skin sensitivity Biocompatibility Testing: testing as specified in the Tripartite Biocompatibility Guidance for Medical Devices and ISO 10993-1 requirements for skin contact. These tests include Cytotoxicity, Sensitization, and Primary Skin Irritation (test results attached). The materials were found to be non-irritating, non-cytotoxic, and non-sensitizing.

Shelf Life:

Data obtained in accelerated shelf life studies was reviewed and found to substantiate our claimed shelf life.

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

JUN 17 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Public Health Service

Vermed, Inc. c/o Mr. Marc Fillion VP of Quality & Regulatory Affairs 9 Lovell Drive Bellows Falls, VT 05101

Re: K090180

> Tender-Trode® Prewired ECG Electrodes Regulation Number: 21 CFR 870.2360 Regulation Name: Electrocardiograph Electrode Regulatory Class: Class II (two) Product Code: DRX Dated: May 7, 2009 Received: May 8, 2009

#### Dear Mr. Fillion:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

### Page 2 - Mr. Marc Fillion

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### INDICATIONS FOR USE 5.

510(k) Number (if known):

Device Name: Tender-Trode® Prewired ECG Electrodes

### Indications for Use:

The Tender-Trode® Prewired electrode is a disposable, noninvasive electrode that is intended to be used for resting ECG monitoring on neonatal and pediatric patients in a hospital setting. These electrodes can be used for short-term or long-term applications, up to 24 hours. The Tender-Trode of prewired electrodes are single-use, non-sterile, disposable and are to be used on intact (uninjured) skin.

X Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

for B Zuckermo
Division Sign-Off
6/17/09
Division of Cardiovascular Devices

510(k) Number K090180

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DRX/K090180](https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DRX/K090180)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
