NIHON KOHDEN BSM-1101 AND BSM-1102 PATIENT MONITOR AND ACCESSORIES
K973918 · Nihon Kohden America, Inc. · DRT · Jan 13, 1998 · Cardiovascular
Device Facts
Record ID
K973918
Device Name
NIHON KOHDEN BSM-1101 AND BSM-1102 PATIENT MONITOR AND ACCESSORIES
Applicant
Nihon Kohden America, Inc.
Product Code
DRT · Cardiovascular
Decision Date
Jan 13, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The BSM-1101 and BSM-1102 Patient Monitors are intended for medical purposes to measure the heart rate from a signal produced by an electrocardiograph. The devices sound an alarm when the heart rate falls outside preset upper and lower limits. The devices include an oximeter used to transmit radiation at a known wavelength through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. The devices include a noninvasive blood pressure measurement system that provides a signal from which systolic, diastolic, and mean pressures can be derived through the use of pressure cuffs placed on the arm, leg or finger. The devices may be used as a radio frequency physiological signal transmitter to condition physiological signals so that they can be transmitted via radio frequency from one location to another, e.g. a central monitoring station. The BSM-1101 and BSM-1102 Patient Monitors will be available for use by medical personnel on all patient populations within a medical facility.
Device Story
BSM-1101/BSM-1102 are portable patient monitors used by medical personnel in clinical facilities. Inputs include ECG signals for heart rate, reflected/scattered radiation for SpO2, and pressure cuff signals for noninvasive blood pressure. The device processes these signals to derive physiological parameters, triggers alarms for out-of-limit heart rates, and functions as an RF transmitter to send data to central monitoring stations. Output is displayed for clinician review to support clinical decision-making and patient monitoring. The device benefits patients by providing continuous, real-time physiological status updates.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including electromagnetic, environmental, safety, and performance testing, as well as software validation for data acquisition, processing, display, and recording functions.
Technological Characteristics
Portable patient monitor; measures ECG, SpO2, and noninvasive blood pressure. Includes RF transmitter for physiological signal transmission. Software-based acquisition, processing, and display. Non-contacting device per Tripartite Guidance. Accessories include SpO2 probes (TL-101T, TL-120T, TL-121T, TL-051S, TL-052S, TL-061S, TL-062S).
Indications for Use
Indicated for use by medical personnel on all patient populations within a medical facility for monitoring heart rate, blood oxygen saturation (SpO2), and noninvasive blood pressure (systolic, diastolic, mean).
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
K973918
NIHON KOHDEN AMERICA, INC. October 14, 1997
## 510(k) NOTIFICATION BSM-1101/BSM-1102 Patient Monitor
JAN 13 1998
## SECTION 2 - 510(K) SUMMARY
## Name and Address of Applicant
Nihon Kohden America, Inc. Attn: Requiatory Affairs 2601 Campus Drive Irvine, California 92612-1601 (714) 250-3959
These devices have been class II by the Division of Cardiovascular, Respiratory, and Neurological Devices and the Cardiovascular Device Classification Panel under 21 CFR Part 870.2300 Cardiac Monitor (Including cardiotachometer and rate alams) per 74 DRT, 870.2700 Oximeter per 74 DQA, 870.1130 Noninvasive Blood Pressure Measurement System per 74 DXN, and 870.2910 Radio Frequency Physiological Signal Transmitter per 74 DRG
Common names for the BSM-1102 include Patient Monitor, Portable Monitor, Transport Monitor, Cardiac Monitor and Bedside Monitor.
The predicate device is the Nihon Kohden BSM-2101A Patient Monitor per 510(k) #K914092, commercial distribution certification dated May 28, 1992.
Nihon Kohden's BSM-1101 and BSM-1102 Patient Monitors are intended for medical purposes to measure the heart rate from a signal produced by an electrocardiograph. These devices sound an alam when the heart rate falls outside oreset upper and lower limits. The devices include an oximeter used to transmit radiation at a known wavelenath through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. These devices include a noninvasive blood pressure measurement system that provides a signal from which systolic, and mean pressures can be derived through the use of pressure cuffs placed on the arm, leg or finger. The devices may be used as a radio frequency physiological signal transmitter to condition physiological signals so that they can be transmitted via radio frequency from one location to another, e.g. a central monitoring station. These devices will be available for use by medical personnel on all patient populations within a medical facility.
To date, no performance standards or special controls are known or established for these devices as required by Section 514 of the Food, Drug and Cosmetic Act and implemented by 21 CFR Part 861.
The BSM-1101 and BSM-1102 are not intended to be sterile.
The BSM-1101 and BSM-1102 Patient-Monitors were determined to be non-contacting per the Tripartie Guidance. Therefore, good laboratory practice studies were not required per 21 CFR part 58. Accessories manufactured.by .Nihon-Kohden-that .max .contact the .patientinclude-TL-101T, .- TL-120T, TL-121T, TL-051S, TL-052S, TL-061S and TL-062S SpO2 probes. The TL-101T, TL-120T and TL-121T SpO2 problement accessories submitted for review in a separate 510/k) submission. The TL-0515, TL-052S, TL-061S and TL-062S SpO2 probes were previously submitted as accessories to the ZB-831PA Telemetry Transmitter, per 510(k) #K946175, commercial distribution certification dated November 22, 1995.
The BSM-1101 and BSM-1102 Patient Monitors are subjected to electromagnetic, environmental, safety and performance testing procedures to verify the operation of the device. Software validation tests the operation of the software functions of acquiring, processing, displaying and recording of the device.
Based on the above, Nihon Kohden believes that the BSM-1101 and BSM-1102 Patient Monitors are substantially equivalent to the Nihon Kohden BSM-2101A Patient Monitor.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, often used as a symbol of medicine. The caduceus is positioned to the right of the text, which reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement around the logo.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856
JAN 1 3 1998
Mr. Gary Reasoner Director of Product Operations Nihon Kohden America, Inc. 2601 Campus Drive Irvine, CA 92715
K973918 Re : Nihon Kohden BSM-1101 and BSM-1102 Patient Monitor Trade Name: Requlatory Class: II (two) 74 DRT Product Code: Dated: October 14, 1997 Received: October 15, 1997
Dear Mr. Reasoner:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: "General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
Page 2 - Mr. Gary Reasoner
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neuroloqical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
G. Indications for Use Statement
510(k) Number (if known):
Device Name: BSM-1101 and BSM-1102 Patient Monitor
Indications for Use:
The BSM-1101 and BSM-1102 Pattern Monitors are intended for medical purposes to measure the heart rate from a signal produced by an electrocardiograph. The devices sound an alarm when the heart rate falls outside preset upper and lower limits. The devices include an oximeter used to transmit radiation at a known wavelength through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. The devices include a noninvasive blood pressure measurement system that provides a signal from which systolic, diastolic, and mean pressures can be derived through the use of pressure cuffs placed on the arm, leg or finger. The devices may be used as a radio frequency physiological signal transmitter to condition physiological signals so that they can be transmitted via radio frequency from one location to another, e.g. a central monitoring station.
The BSM-1101 and BSM-1102 Patient Monitors will be available for use by medical personnel on all patient populations within a medical facility.
**Prescription Use**
(Per 21 CFR 801.109)
Qouglas Killeen
**(Division Sign-Off)**
Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.