SORIN BIOMEDICA SMART BCD VANGUARD,SURFACE MODIFIED BLOOD CARDIOPLEGIA SYSTEM
K012763 · Cobe Cardiovascular, Inc. · DRT · Nov 15, 2001 · Cardiovascular
Device Facts
Record ID
K012763
Device Name
SORIN BIOMEDICA SMART BCD VANGUARD,SURFACE MODIFIED BLOOD CARDIOPLEGIA SYSTEM
Applicant
Cobe Cardiovascular, Inc.
Product Code
DRT · Cardiovascular
Decision Date
Nov 15, 2001
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sorin Biomedica SMAR T® BCD Vanguard™ Surface Modified Blood Cardioplegia System is intended to mix, cool, warm, and deliver oxygenated blood and cardioplegic solution for periods up to 6 hours. The device allows monitoring of temperature and pressure, traps bubbles, and removes air. Blood and cardioplegic solution are delivered to the patient by a 100% occlusive roller pump, through the extension line and appropriate cannula.
Device Story
Cardioplegia heat exchanger with integral bubble trap; used during cardiopulmonary bypass. Inputs: oxygenated blood and cardioplegic solution. Device mixes, cools, or warms fluids; traps bubbles; removes air via hydrophobic membrane; monitors temperature and pressure via integrated ports. Output: temperature-controlled, air-free cardioplegic solution delivered to patient via 100% occlusive roller pump and cannula. Used in surgical settings by perfusionists/surgeons. Surface-modifying material added to blood contact assembly to improve blood compatibility. Pressure relief valve prevents over-pressurization. Benefits: maintains myocardial protection during surgery; reduces blood trauma.
Cardiopulmonary bypass heat exchanger. Features hydrophobic membrane, one-way valve, pressure relief valve, filter screen, temperature monitoring port, and pressure monitoring port. Includes surface-modifying material on blood contact surfaces to enhance biocompatibility. Single-use, nonpyrogenic fluid pathways. Operates with 100% occlusive roller pump.
Indications for Use
Indicated for patients requiring extracorporeal circulation during surgical procedures where mixing, cooling, warming, and delivery of oxygenated blood and cardioplegic solution are necessary for up to 6 hours.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Predicate Devices
SORIN Biomedica BCD Vanguard™ Blood Cardioplegia System
Submission Summary (Full Text)
{0}------------------------------------------------
510(k) Pre-Market Notification: SMAR,T® BCD Vanguard™ Blood Cardioplegia System COBE Cardiovascular, Inc.
NOV 1 5 2001
K012763
### 510(k) Summary
| SUBMITTER: | COBE Cardiovascular, Inc. |
|------------|---------------------------|
| | 14401 W. 65th Way |
| | Arvada, CO 80004 |
Lynne Leonard CONTACT PERSON: Phone:(303) 467-6586 Fax: (303) 467-6429
August 16, 2001 DATE PREPARED:
SORIN Biomedica SMARŢT® BCD Vanguard™ DEVICE TRADE NAME: Surface Modified Blood Cardioplegia System
COMMON/USUAL NAME: Cardioplegia Heat Exchanger
CLASSIFICATION NAME: Cardiopulmonary Bypass Heat Exchanger
SORIN Biomedica BCD Vanguard™ PREDICATE DEVICE: Blood Cardioplegia System
DEVICE DESCRIPTION AND INTENDED USE:
The SMAR T BCD Vanguard is a cardioplegia heat exchanger with integral bubble trap, r he only സ്റ്റ് ഇയ varigurations connected to the heat exchanger available with various tudifying material has been added to the primary blood contact assembly. 71 Junate mostlying to improve the blood compatibility of the device. The sundocs of the mout oxehanger seide, is for single use only, and has nonpyrogenic fluid pathways.
The heat exchanger assembly (heat exchanger with bubble trap) contains a hydrophobic membrane, a one-way valve, a pressure relief valve, a filter screen, a nyarophobio monitoring port and a pressure monitoring port. The housing is designed such that air entering the device rises to the membrane and is vented to the atmosphere. The one-way valve keeps air from entering the device. The pressure relief valve is designed to relieve excess pressure in the event of outlet line clamping, and has tubing attached to it to vent the solution to the cardiotomy reservoir. The temperature monitoring port is used to measure the temperature of the blood cardioplegia mixture, and the pressure monitoring line is connected to either a pressure manometer or a transducer to measure pressure.
{1}------------------------------------------------
#### INDICATIONS FOR USE
The SMAR, T® BCD Vanguard™ Surface Modified Blood Cardioplegia System is intended to mix, cool, warm, and deliver oxygenated blood and cardioplegic solution for intended to mix, 600, warm, and avice also allows monitoring of temperature and penous of up to ox hes, and removes air. Blood and cardioplegic solution are pressure, traps bubbloo, and rene 100% occlusive roller pump, through the extension line and appropriate cannula.
# STATEMENT OF TECHNICAL CHARACTERISTICS COMPARISON
The SMAR T® BCD Vanguard™ Surface Modified Blood Cardioplegia System has the The OMARY I - BOD Vanguard - Sanaco Vanguard™ Blood Cardioplegia System. The Same intended use as the OOTIN - BCD Vanguard" contains a surface-modifying two uction in that 17 the Stibility; and 2) SMAR, T Surface Modified Tubing and matehal to improve blood compatibled for various tubing and connectors within the circuit configurations. Otherwise, materials, components, design, sterilization and manufacturing processes for the two devices are the same.
Biocompatibility testing and in-vitro testing were performed to demonstrate that the Dioompanbilly tooling and Surface Modified Blood Cardioplegia System is substantially equivalent to the currently marketed BCD Vanguard™ Blood Cardioplegia System.
In-vitro testing consisted of:
- 1. Stability testing, surface-modifying material
- Surface-modifying material elution 2
- Surface-modifying material blood compatibility 3.
- Heat exchanger efficiency 4.
- Priming volume 5.
- Blood pathway pressure drop 6.
- 7. Blood pathway integrity
- Ease of prime 8.
- Blood trauma (including platelet depletion, white blood cell depletion, and 9. plasma free hemoglobin generation)
- 10. Connection integrity
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized graphic of three wavy lines, resembling an abstract human figure or a bird in flight. The logo is black and white.
Food and Drug Administratio 9200 Corporate Boulevard Rockville MD 20850
## NOV 1 5 2001
Mr. Lynne Leonard Sr. Regulatory Affairs Manager COBE Cardiovascular, Inc. 14401 W. 65th Way Arvada, CO 80004-3599
Re: K012763
Trade Name: Sorin Biomedica SMAR T BCD Vanguard Surface Modified Blood Cardioplegia System Regulation Number: 21 CFR 870.4240 Regulation Name: Cardiopulmonary bypass heat exchanger. Regulatory Class: Class II (two) Product Code: DRT Dated: August 16, 2001 Received: August 17, 2001
Dear Mr. Leonard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Lynne Leonard
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
James F. Dillard III
James E. Dillard III Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## NOV 1 5 2001
Indications For Use
510(k) Number (If known):
K012763
Sorin Biomedica SMAR T® BCD Vanguard™ Surface Modified Blood Cardioplegia System Device Name:
Indications For Use:
The Sorin Biomedica SMAR T® BCD Vanguard" Surface Modified Blood Cardioplegia System is The Sorin Biomedica SMAR, I - BCD Vanguard blood and cardioplegic solution for periods up
intended to mix, cool, warm, and deliver oxygenated blood areas bubbles, and removes intended to mix, cool, warm, and deliver oxygenator bround transmont of the bubbles, and removes
to 6 hours. The device allows monitoring of temperature, a 100% occursive ro to 6 hours. The device anows momoring of temporature and processor of 100% occlusive roller pump,
air. Blood and cardioplegic solution are delivered to the patient by a 100 an . Drood and extension line and appropriate cannula.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division of Cardiovascular & Re
510(k) Number 2012763
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.