← Product Code [DRG](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DRG) · K211305

# ANNE One (K211305)

_Sibel, Inc. · DRG · Sep 14, 2021 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DRG/K211305

## Device Facts

- **Applicant:** Sibel, Inc.
- **Product Code:** [DRG](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DRG.md)
- **Decision Date:** Sep 14, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2910
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

ANNE One is a wireless vital signs and physiological data monitoring platform indicated for the measurement of heart rate, respiratory rate, step count, fall temperature, and body temperature by qualified healthcare professionals in healthcare settings. The device is intended for use on general care patients who are 18 years of age or older as a general patient monitor to provide continuous physiological information as an aid to diagnosis and treatment. The device is not intended for use on critical care patients. The device is not intended to monitor or measure respiratory rate on heart rate on ambulatory patients.

## Device Story

ANNE One is a wireless multi-parameter vital signs monitoring system for healthcare settings. It comprises a Samsung Galaxy tablet running the ANNE View mobile application, a chest-worn sensor, a limb-worn sensor, and a wireless charger. Sensors are skin-mounted and battery-operated; they continuously collect physiological data (heart rate, respiratory rate, step count, fall count, skin/body temperature) and transmit encrypted data via Bluetooth to the tablet. Healthcare professionals use the tablet to view real-time physiological information to aid in diagnosis and treatment. The system provides continuous monitoring for general care patients; it does not include alarm notifications or long-term data storage. The device benefits patients by providing continuous, non-invasive physiological tracking without the need for wired connections, facilitating timely clinical assessment.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench and simulated use testing. Simulated use studies included heart rate validation in 35 healthy subjects and respiratory rate validation in 50 healthy subjects against FDA-cleared reference devices. Step and fall count features were validated against manual counts. Skin temperature and body temperature accuracy were validated per ISO 80601-2-56:2017.

## Technological Characteristics

System includes chest and limb sensors, tablet, and charger. Sensors use wireless Bluetooth transmission. Materials are biocompatible per ISO 10993-1. Electrical safety complies with ANSI/AAMI ES60601-1 and IEC 60601-1-2. Non-sterile. Software is Moderate level of concern. Respiratory rate is derived from ECG (EDR).

## Regulatory Identification

A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.

## Predicate Devices

- Patient Status Engine ([K172329](/device/K172329.md))

## Reference Devices

- Fever Scout Continuous Monitoring thermometer ([K162137](/device/K162137.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 14, 2021

Sibel Inc. % Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Blvd Warren, New Jersey 07059

Re: K211305

Trade/Device Name: ANNE One Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency physiological signal transmitter and receiver Regulatory Class: Class II Product Code: DRG, MWI, FLL Dated: July 23, 2021 Received: July 27, 2021

Dear Dave Yungvirt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K211305

Device Name ANNE One

Indications for Use (Describe)

ANNE One is a wireless vital signs and physiological data monitoring platform indicated for the measurement of heart rate, respiratory rate, step count, fall temperature, and body temperature by qualified healthcare professionals in healthcare settings. The device is intended for use on general care patients who are 18 years of age or older as a general patient monitor to provide continuous physiological information as an aid to diagnosis and treatment. The device is not intended for use on critical care patients. The device is not intended to monitor or measure respiratory rate on heart rate on ambulatory patients.

| Type of Use (Select one or both, as applicable)                                       |                                                                                      |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| <div> <span>☑</span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span>☐</span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(K) Summary

Submitter: .

Sibel Inc. 6650 W. Touhy Avenue, Niles, IL 60714 Tel: (224) 251-8859

Official Correspondent: Peter Xu, VP of Quality Assurance and Regulatory Affairs 6650 W. Touhy Avenue, Niles, IL 60714 Tel: (224) 251-8859

Date Prepared: 9/7/2021

#### ll. Device Information

Name of Device: ANNE One 510K Number: K211305 Common or Usual Name: Wireless Remote Monitoring System Classification Name: Radiofrequency Physiological Signal Transmitter and Receiver Review Panel: Cardiovascular Regulation: 21 CFR §870.2910 Regulatory Class: Class II

Product Classification Code: DRG, MWI, FLL

#### . Predicate Device

Predicate - Trade Name: Patient Status Engine Predicate 510(k): K172329 Predicate device manufacturer: Isansys Lifecare Ltd

#### IV. Reference Device

Reference device – Trade Name: Fever Scout Continuous Monitoring thermometer Reference 510(k): K162137 Reference Manufacturer: Vivalnk Inc.

#### V. Device Description

ANNE One is a wireless multi-parameter vital signs monitoring system that consists of the following components, and subsystems:

A. ANNE Tablet (Samsung Galaxy Tablet) - manufactured by Samsung

- B. ANNE View Mobile Application (software) manufactured by Sibel Inc.

{4}------------------------------------------------

- C. ANNE Chest Sensor manufactured by Sibel Inc.
- D. ANNE Limb Sensor manufactured by Sibel Inc.
- E. ANNE Wireless Charger manufactured by Sibel Inc.

The ANNE Chest Sensor is a battery-operated, skin-mounted sensor with wireless transceiver capabilities worn on the upper body to measure heart rate, respiratory rate, step count, fall count, and skin temperature.

The ANNE Limb Sensor is an additional battery-operated skin-mounted sensor with wireless transceiver capabilities worn on the finger to measure skin temperature. When placed underneath the axillae, the ANNE Limb Sensor can measure body temperature.

Both the ANNE Chest and ANNE Limb Sensors continuously gather vital signs data from the person being monitored and then transmit the encrypted data to the ANNE Tablet operating the ANNE View Mobile Application-a mobile device software application.

### VI. Indications for Use

ANNE One is a wireless vital signs and physiological data monitoring platform indicated for the measurement of heart rate, respiratory rate, step count, fall count, skin temperature, and body temperature by qualified healthcare professionals in healthcare settings. The device is intended for use on general care patients who are 18 years of age or older as a general patient monitor to provide continuous physiological information as an aid to diagnosis and treatment. The device is not intended for use on critical care patients. The device is not intended to monitor or measure respiratory rate or heart rate on ambulatory patients.

|                                             | Subject device<br>Sibel Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate device<br>Isansys Lifecare Ltd                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Variances / Equivalence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name                                  | ANNE One                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Patient Status Engine                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| 510(k) Number                               | K211305                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | K172329                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Class                                       | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Equivalent                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Product Code                                | DRG, MWI, FLL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | DRG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Equivalent                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Regulation<br>Number and<br>Regulation Name | 870.2910<br>Transmitters and<br>Receivers, Physiological<br>Signal, Radiofrequency                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 870.2910<br>Transmitters and Receivers,<br>Physiological Signal,<br>Radiofrequency                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Equivalent                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                             | Subject device<br>Sibel Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate device<br>Isansys Lifecare Ltd                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Variances / Equivalence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Indications for<br>Use                      | ANNE One is a wireless<br>vital signs and<br>physiological data<br>monitoring platform<br>indicated for the<br>measurement of heart<br>rate, respiratory rate,<br>step count, fall count,<br>skin temperature, and<br>body temperature by<br>qualified healthcare<br>professionals in<br>healthcare settings. The<br>device is intended for<br>use on general care<br>patients who are 18<br>years of age or older as<br>a general patient monitor<br>to provide continuous<br>physiological information<br>as an aid to diagnosis<br>and treatment. The<br>device is not intended for<br>use on critical care<br>patients. The device is<br>not intended to monitor<br>or measure respiratory<br>rate or heart rate on<br>ambulatory patients. | The device is a wireless<br>remote monitoring system<br>intended for use by<br>healthcare professionals for<br>continuous collection of<br>physiological data in home<br>and healthcare settings.<br>This includes heart rate,<br>heart rate variability (R-R<br>interval), ECG derived<br>respiration rate data (EDR),<br>skin temperature, activity,<br>posture and optional SpO2<br>and noninvasive Blood<br>Pressure (BP). Data is<br>transmitted wirelessly from<br>the sensors to the Patient<br>Gateways and from the<br>Patient Gateways to a<br>central Server where it is<br>stored for analysis. The<br>Patient Status Engine can<br>include the ability to notify<br>healthcare professionals<br>when physiological data fall<br>outside selected<br>parameters. | Similar<br>ANNE One does not include<br>optional SpO2 and<br>non-invasive BP<br>measurements. ANNE One<br>does not measure heart rate<br>variability. ANNE One does not<br>include the ability to notify<br>healthcare professionals when<br>physiological data fall outside<br>selected parameters. ANNE<br>One does not store data for<br>future analysis. ANNE One is<br>not intended to monitor or<br>measure respiratory rate or<br>heart rate on ambulatory<br>patients. The differences do<br>not affect safety or<br>effectiveness. |
| Intended Use                                | ANNE One is intended<br>for use on general care<br>patients who are 18<br>years of age or older as<br>a general patient monitor,<br>to provide continuous<br>physiological information.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | The device is intended for<br>use on general care patients<br>who are 18 years of age or<br>older as a general patient<br>monitor, to provide<br>physiological information.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Equivalent                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Use Environment                             | Healthcare Settings                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Home and Healthcare<br>Settings                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Different<br>ANNE One does not offer<br>Home use. The differences do<br>not affect safety or<br>effectiveness.                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Heart Rate                                  | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Equivalent                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                             | Subject device<br>Sibel Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate device<br>Isansys Lifecare Ltd                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Variances / Equivalence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| R-R Interval                                | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Different<br>ANNE One is not indicated for<br>the display of R-R interval.<br>However, it does calculate the<br>R-R interval to determine HR.<br>This difference does not affect<br>safety or effectiveness.                                                                                                                                                                                                                                                                                                                                |
| Respiration Rate                            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Equivalent<br>Both the ANNE One and<br>Patient Status Engine derive<br>respiratory rate from ECG<br>alone (EDR).                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Skin Temperature                            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Equivalent                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Body<br>Temperature                         | Yes on axilla/armpit                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Yes*                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Equivalent<br>The differences do not affect<br>safety or effectiveness.                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| SpO2                                        | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes - optional with a FDA<br>cleared SpO2 oximeter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Different<br>Without the optional SpO2<br>oximeter, the devices are<br>equivalent. The difference<br>does not affect safety or<br>effectiveness.                                                                                                                                                                                                                                                                                                                                                                                            |
| Activity                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Equivalent<br>The ANNE One specifically<br>measures activity as step<br>count. The differences do not<br>affect safety or effectiveness.                                                                                                                                                                                                                                                                                                                                                                                                    |
| Posture                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Equivalent<br>The ANNE One specifically<br>measures posture only as it<br>relates to fall count. The<br>differences do not affect safety<br>or effectiveness                                                                                                                                                                                                                                                                                                                                                                                |
| Non-Invasive<br>Blood Pressure<br>(NIBP)    | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes - optional with a FDA<br>cleared NIBP device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Different<br>Without the optional NIBP<br>device, the devices are<br>equivalent. The difference<br>does not affect safety or<br>effectiveness.                                                                                                                                                                                                                                                                                                                                                                                              |
|                                             | Subject device<br>Sibel Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate device<br>Isansys Lifecare Ltd                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Variances / Equivalence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Data                                        | Data transmitted<br>wirelessly via Bluetooth<br>from sensors to mobile<br>device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Data is transmitted<br>wirelessly via Bluetooth from<br>the sensors to the Patient<br>Gateways and from the<br>Patient Gateways to a<br>central Server where it is<br>stored for analysis.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Equivalent<br>ANNE One does not offer data<br>transmission to a central<br>Server or data download. The<br>differences do not affect safety<br>or effectiveness.                                                                                                                                                                                                                                                                                                                                                                            |
| Notification                                | No notification ability                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | The Patient Status Engine<br>can include the ability to<br>notify healthcare<br>professionals when<br>physiological data fall<br>outside selected<br>parameters.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Different<br>ANNE One does not offer a<br>notification but provides a<br>real-time display. This<br>difference does not affect<br>safety of the device.<br>Performance data<br>demonstrates that this change<br>does not affect effectiveness.                                                                                                                                                                                                                                                                                              |

Table 5-1: Substantial Equivalence for Indication for Use

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

*The labeling for the Patient Status Engine states that the LifeTemp Sensor can be used for body temperature measurement if placed under the axilla, but this feature is not included in the device's indications for use.

Validation methods and laboratory accuracy data for body temperature measurements from the Fever Scout reference device were further utilized to demonstrate substantial equivalence to ANNE One.

### VII. Performance Data

Verification and Validation tests established the safety and effectiveness of ANNE One. The following performance data have been provided in support of the substantial equivalence determination:

# Sterilization & Shelf-life Testing

ANNE One is non-sterile, and therefore sterilization data is not provided. The shelf-life study under accelerated aging data has been provided to support the labeled shelf-life claim for ANNE One. Additional peel force testing established the safety of the ANNE adhesive compared to standard FDA-cleared medical adhesives.

# Biocompatibility Testing

Biocompatibility testing was performed for all patient-contacting materials used in ANNE One, and the results indicated that all patient-contacting materials in the ANNE One system are biologically compatible and satisfy the requirements defined in ISO 10993 Part 1, entitled 'Use of International Standard ISO-10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.

# Electromagnetic Compatibility and Electrical Safety

Electrical safety and electromagnetic compatibility testing were conducted on ANNE One. The device complies with the ANSI/AAMI ES60601-1:2005/(R)2012 and IEC 60601-1-2:2014

{8}------------------------------------------------

standards. The device also complies with the FCC CFR 47 Part 15 Subpart C standards for wireless communication. The device was tested and demonstrated to be safe in case of defibrillation according to Section 8.5.5 of ANSI/AAMI ES60601-1:2005/(R)2012.

## Software Verification and Validation Testing

Software verification and validation testing demonstrated the safety and effectiveness of ANNE One. Based on the FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, the software level of concern for ANNE One is determined to be Moderate.

## Benchtop Testing and Simulated Use Testing

Bench testing was performed to verify device specifications and to ensure mechanical and electrical requirements were met. Laboratory accuracy testing according to ISO 80601-2-56:2017 Section 201.101.2 validated the effectiveness of the ANNE Chest and Limb sensor for skin temperature measurements and the Limb Sensor for body temperature measurements when placed under the axillae. Heart rate measurements with ANNE One were verified according to IEC 60601-2-27 Sections 201.12.101.15 and 201.12.1.101.17. A simulated use study validated heart rate measurements in 35 healthy subjects against an FDA-cleared reference device. Respiratory rate measurements with ANNE One were validated in 50 healthy subjects against an FDA-cleared End Tidal Carbon Dioxide (EtCO-) reference device. The step and fall count features were demonstrated to conform to their design specifications through comparison to manual counts.

Simulated use testing validated the measurements of skin temperature, step count, and fall detection by the ANNE One system in n=10 subjects. The ANNE Hydrogel adhesive was also assessed for skin adhesion, pain upon removal, irritation, and comfort. The study demonstrated the functionality and accuracy of the system.

## Animal Studv

Animal testing was not required to demonstrate the safety and effectiveness of ANNE One.

## Clinical Study

Clinical testing was not required to demonstrate the safety and effectiveness of ANNE One. Instead, substantial equivalence is supported by Bench and Simulated Use testing.

## Usability Testing

Studies related to human factors engineering and ease of use of the ANNE One were conducted in accordance with the principles of IEC 62366, Medical devices - Applications of usability engineering to medical devices and the document entitled Applying Human Factors and Usability Engineering to Medical Devices.

## VII. Conclusion

The results of the substantial equivalence assessment, taken together with the performance testing data, demonstrate that ANNE One's performance characteristics are substantially equivalent to the predicate device in both design and intended use.

{9}------------------------------------------------

It has been shown through the documents provided in this 510(k) submission that the minor differences between ANNE One and the predicate device of Patient Status Engine, as well as the reference device of the Fever Scout Continuous Monitoring thermometer, do not raise any new questions regarding its safety and performance. Any differences between the subject device and the predicate/reference devices have no significant effect on safety or effectiveness as established through performance testing.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DRG/K211305](https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DRG/K211305)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
