K103643 · Ascot Technologies, Inc. · DRG · Oct 6, 2011 · Cardiovascular
Device Facts
Record ID
K103643
Device Name
ASCOT HOME MEDICAL MONITORING SYSTEM (HMMS)
Applicant
Ascot Technologies, Inc.
Product Code
DRG · Cardiovascular
Decision Date
Oct 6, 2011
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.2910
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The HMMS is automation of the existing process of reporting readings to the physician over the phone and receiving verbal instructions for dosage changes. New prescriptions are not handled or ordered in the HMMS. The HMMS is intended for use in the home or at clinical settings as a means to collect physiological readings, measured by FDA approved devices, and transmit to a Physician for review. The readings are accessed by a qualified medical professional (such as Physician or Nurse Practitioner) for review. The qualified medical professional may make medication dosage changes and transmit those changes back to the patient. The patient confirms the receipt of the medication changes are displayed on the patient's Ascot Home Medical Station (HMS). The HMS receives readings measured by various FDA approved medical devices (such as blood pressure with pulse or blood glucose monitor) via standard connections including RS232, Bluetooth, or USB connection. The HMS communicates with the Ascot Central Server (ACS) via standard wireless or wired communication protocols including cellular data and wireless internet. The HMMS does not make any decisions on the data that it conveys. The HMMS does not provide diagnosis of any disease or medical condition. The HMMS is not a substitute for professional healthcare judgment.
Device Story
Software-based system automating patient-to-physician reporting of physiological readings; replaces verbal phone reporting. Inputs: blood pressure, pulse, blood glucose data from FDA-cleared monitors via USB/RS232/Bluetooth. Data transmitted to Ascot Central Server (ACS) via wired/wireless internet or cellular data. Physician/qualified provider reviews data via web interface; may issue medication dosage changes transmitted back to patient's Home Medical Station (HMS). Patient confirms receipt of dosage instructions. Used in home or clinical settings. Benefits: improved communication speed and data accuracy; eliminates manual reporting errors. Does not provide diagnosis, real-time monitoring, or emergency alerts.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Software-based system running on Windows-based PC/netbook. Connectivity via RS232, Bluetooth, USB for device input; wired/wireless internet or cellular data for server communication. Uses standard communication protocols and encryption for privacy. No specific hardware materials or energy sources defined beyond standard computing equipment.
Indications for Use
Indicated for patients with known or suspected hypertension or diabetes requiring home monitoring of blood pressure, pulse, and/or blood glucose. Not for real-time monitoring, emergency calls, or time-critical data. Not a substitute for direct medical supervision or emergency intervention.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
FORA GW 9014 TeleHealth Gateway Telemedicine System (K100427)
Submission Summary (Full Text)
{1}------------------------------------------------
K103643 P1/5
# OCT - 6 2011
# 510(k) Summary
# GENERAL COMPANY INFORMATION COMPANY NAME/ADDRESS/PHONE/FAX:
Ascot Technologies, Inc. 267 Hogans Valley Way Cary, NC 27513 Tel: 919-388-1776 Fax: 919-481-3203
### NAME OF CONTACT:
Mary Ellen Randall President (919) 388-1776
#### DATE PREPARED:
Original December 8, 2010 Updated February 2, 2011 Updated September 18, 2011 Updated October 5, 2011
# TRADE NAME:
Ascot Home Medical Monitoring System (HMMS)
#### COMMON NAME:
Health Management System Software
#### DEVICE NAME:
Ascot Home Medical Monitoring System (HMMS)
## CLASSIFICATION NAME:
The following classifications appear applicable:
| Product<br>Code | Classification Name | Class | CFR<br>Section |
|-----------------|-----------------------------------------------------------------|-------|----------------|
| DRG | Radiofrequency physiological signal transmitter<br>and receiver | II | 870.2910 |
# COMPARISON TO PREDICATE DEVICE
We claim substantial equivalence to the currently legally cleared ACCU-CHEK® Advisor Insulin Guidance Software (K043529) and OneTouch® Zoom™ Diabetes Management Program (K081318), Confidant 2.6 (K083331), and FORA GW 9014 TeleHealth Gateway Telemedicine System (K100427).
{2}------------------------------------------------
K103643 P 2/5
# COMPARISON TO PREDICATE DEVICE
| Feature/<br>Claim | ACCU-<br>Chek®<br>Advisor<br>Software | OneTouch®<br>Zoom™<br>Diabetes<br>Management<br>Program | Confidant<br>2.6 | FORA<br>Telehealth<br>Gateway | Ascot HMMS |
|-----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|----------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| Data Upload | Yes | No. Data<br>obtained from<br>DataVault | Yes through<br>standard<br>communicati<br>on<br>technologies. | Yes through<br>standard<br>communicati<br>on<br>technologies. | Yes through<br>standard<br>RS232,<br>Bluetooth or<br>USB<br>Connection |
| Support | Through call<br>center<br>support,<br>labeling &<br>health care<br>professionals | Unknown. | Unknown | Unknown | Through<br>trained health<br>care staff, staff<br>training,<br>written<br>instructions &<br>Ascot support. |
| Data Storage | On computer<br>media | On computer<br>media | On computer<br>media | On Computer<br>media | On computer<br>media |
| Reports &<br>Graphs | Based on<br>readings<br>obtained | Based on<br>readings<br>obtained | Based on<br>readings<br>obtained | Based on<br>readings<br>obtained | Based on<br>readings<br>obtained |
| Review<br>Readings | On Computer | Web Interface | On cell<br>phone. | Web<br>Interface | Web Interface |
| Enables<br>patient &<br>health care<br>professional<br>to share<br>readings | Yes | Yes. | Yes. | Yes. | Yes. |
| Provides<br>feedback to<br>patient | No. | No. | Yes.<br>Automated. | No. | Yes.<br>Authorized By<br>Physician only. |
| Devices<br>Supported | Blood<br>Glucose | Blood Glucose | Blood<br>glucose, non-<br>invasive<br>blood<br>pressure,<br>weight. | Blood<br>glucose,<br>blood<br>pressure,<br>Weight,<br>temp, body<br>fat,<br>hydration. | Blood glucose,<br>non-invasive<br>blood pressure. |
| Feature/<br>Claim | ACCU-<br>Chek®<br>Advisor<br>Software | OneTouch®<br>Zoom™<br>Diabetes<br>Management<br>Program | Confidant<br>2.6 | FORA<br>Telehealth<br>Gateway | Ascot HMMS |
| Intended Use:<br>Home setting | Yes. | Yes. | Yes. | Yes. | Yes. |
| Method of<br>transmitting<br>data to<br>physician | Brought in at<br>appointment,<br>read off<br>patient's<br>machine or<br>called in to<br>Physician. | Transmitted<br>across<br>computer<br>connection to<br>remote<br>database hosted<br>by Microsoft. | Unknown.<br>Readings are<br>transmitted to<br>a central<br>database. | Transmitted<br>across<br>standard<br>computer<br>connection to<br>remote<br>database. | Transmitted<br>across secure<br>computer<br>connection to<br>Ascot server<br>database. |
| WebViewer | No. Patient<br>brings in<br>readings. | Readings read<br>through<br>HealthVault<br>account. | Feedback to<br>patient's cell<br>phone. | Yes. | Yes. Readings<br>transmitted<br>eliminating<br>reporting<br>errors. |
| Provides<br>Diagnosis | No. | No. | No. | No. | No. |
{3}------------------------------------------------
K103643 p3/5
# DEVICE DESCRIPTION
The HMMS is automation of the existing process of reporting readings to the physician over the phone and receiving verbal instructions for dosage changes. New prescriptions are not handled or ordered in the HMMS.
The Ascot HMMS (Home Medical Monitoring System) is a software based device which receives readings in the home or clinical setting from various FDA approved devices (such as home Blood Pressure monitor with pulse or Glucose testing monitor) and transmits readings to the Ascot Central Server (ACS) for review by the Physician. (For the purposes of this submission, Physician refers to a qualified medical professional who can change medication dosages, such as Nurse Practitioners or Physician Assistants.) If the Physician modifies medication dosages, the patient medication dosage instructions ordered by the Physician, are transmitted back to the patient and displayed on the HMS. Receipt of each transmission is confirmed within the system.
The Ascot HMMS (Home Medical Monitoring System) reduces risks to health because it speeds up the communication between the Physician and the patient. It improves accuracy because readings are transmitted directly from the home to server for later review by the Physician.
If the patient prefers, they may still report readings using the former process.
{4}------------------------------------------------
K103643 p4/5
| Device Name | File Number | How Connected | Type Device |
|------------------------------------------------------------------|-------------|---------------|---------------------------------------------------------|
| OneTouch Ultra Mini | K061118 | Serial Cable | Blood Glucose<br>Monitor |
| OneTouch Ultra 2 | K053529 | Serial Cable | Blood Glucose<br>Monitor |
| OneTouch UltraLink | K073231 | Serial Cable | Blood Glucose<br>Monitor |
| Omron HEM-790IT<br>(Product Line extension of<br>Model HEM780N3) | K061822 | USB Cable | Automatic Blood<br>Pressure Monitor<br>(includes pulse) |
| HoMedics BPA-260-CB<br>(Model HL868BF) | K092161 | USB Cable | Blood Pressure<br>Monitor (includes<br>pulse) |
Devices which have been tested to operate with the Ascot Home Medical Monitoring System include:
#### INTENDED USE
The HMMS is automation of the existing process of reporting readings to the physician over the phone and receiving verbal instructions for dosage changes. New prescriptions are not handled or ordered in the HMMS.
The HMMS is intended for use in the home or at clinical settings as a means to collect physiological readings, measured by FDA approved devices, and transmit to a Physician for review. The readings are accessed by a qualified medical professional (such as Physician or Nurse Practitioner) for review. The qualified medical professional may make medication dosage changes and transmit those changes back to the patient. The patient confirms the receipt of the medication changes are displayed on the patient's Ascot Home Medical Station (HMS).
The HMS receives readings measured by various FDA approved medical devices (such as blood pressure with pulse or blood glucose monitor) via standard connections including RS232, Bluetooth, or USB connection. The HMS communicates with the Ascot Central Server (ACS) via standard wireless or wired communication protocols including cellular data and wireless internet.
The HMMS does not make any decisions on the data that it conveys.
The HMMS does not provide diagnosis of any disease or medical condition.
The HMMS is not a substitute for professional healthcare judgment.
{5}------------------------------------------------
K103643 p5/5
The HMMS uses standard communication protocols and encryption to protect patient privacy.
The HMMS Viewer is not a primary alarm source.
The full viewer is for use by qualified medical personnel only.
There is a limited viewer interface for use by the patient in reviewing their own readings and staff for adding notes to patient file and follow up call, if instructed by Physician.
End of 510(k) Summary Section
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a bird in flight, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ascot Technologies, Inc. c/o Ms. Mary Ellen Randall President 267 Hogans Valley Way Cary, NC 27513
- 6 2011
Re: K103643
Trade/Device Name: Ascot Home Medical Management System (HMMS) Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency Physiological Signal Transmitter and Receiver Regulatory Class: Class II (two) Product Code: DRG Dated: September 18, 2011 Received: September 20, 2011
Dear Ms. Randall:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 – Ms. Mary Ellen Randall
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman M.D. Director Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## Indications for Use
510(k) Number: K103643
Device Name: Ascot Home Medical Monitoring System (HMMS)
#### Prescription Use
This medical device is indicated for use by patients with known or suspected hypertension or diabetes that a physician wants to monitor blood pressure, pulse, and/or blood glucose at home.
This medical device is a software based device which runs on a Windows based Computer or netbook. This medical device receives readings from supported FDA approved devices in the home and serves as a remote communication link between the home and the physician or qualified medical provider. The physician or qualified medical provider will review the patient information from a web interface and may make medication changes and transmit those changes back to the patient for display on the Home Medical Station(HMS).
This medical device receives blood pressure, heart rate and blood glucose readings from the supported FDA approved medical devices via USB or serial connection to the HMS. The HMS communicates with the server via standard wireless or wired communication protocols such as cellular data, and wired or wireless internet.
HMMS has been tested to work with Homedics BPA-260-CBL, Omron HEM-7901T and OneTouch Ultra Mini, OneTouch Ultra 2 and OneTouch UltraLink.
This medical device is automation of the exiting process of reporting results to the physician over the phone.
This medical device is not intended to provide real time monitoring. It is not intended for emergency calls or for transmission of any time-critical data. This device is not intended as a substitute for direct medical supervision or emergency intervention.
This medical device does not provide diagnosis of any disease or medical condition and is not a substitute for professional healthcare judgment.
This medical device uses standard communication protocols and encryption to protect patient privacy.
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X |
|-------------------------------------------------|-------------|
| AND/OR | Over<br>(21 |
r-The-Counter Use 1 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K103643
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.