RTX3320PSTN WIRELESS TELEHEALTH GATEWAY

K041816 · Rtx Healthcare A/S · DRG · Jul 19, 2004 · Cardiovascular

Device Facts

Record IDK041816
Device NameRTX3320PSTN WIRELESS TELEHEALTH GATEWAY
ApplicantRtx Healthcare A/S
Product CodeDRG · Cardiovascular
Decision DateJul 19, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2910
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The RTX3320 device is for use by patients at home. It is intended to be used in combination with a variety of patient monitors upon the prescription of a licensed physician or other authorized healthcare provider. RTX3320 serves as the remote communication link between compatible patient monitors, and the compatible healthcare facility at another location. The healthcare facility could be at a disease management center, or with the healthcare/wellness provider or other out-of-hospital caregivers. The purpose is to transmit selected medical information (such as weight, blood pressure, blood glucose) over a normal residential telephone line. RTX3320 is a "black box" device – it is not connected to the patient and not operated by the patient. The installation must be performed by or under supervision of an authorized professional healthcare company. This device does not measure, interpret or make any decisions on the vital data that it conveys.

Device Story

RTX3320 Wireless Telehealth Gateway acts as a 'black box' communication hub; receives physiological data (weight, BP, glucose, ECG) from compatible wireless patient monitors via Bluetooth; transmits data over public switched telephone network to remote healthcare servers. Installed by authorized professionals; not operated by patients. Device performs no measurement, interpretation, or clinical decision-making. Healthcare providers receive transmitted data at remote facilities to support clinical management. Benefits include remote monitoring of chronic conditions without requiring patient interaction with the gateway.

Clinical Evidence

Bench testing only. No clinical investigation performed. Verification and validation activities established performance and reliability. Compliance with IEC 60601-1, IEC 60601-1-2, FCC Part 15, and FCC Part 68 demonstrated.

Technological Characteristics

Hub/gateway device; Bluetooth wireless connectivity to patient monitors; PSTN modem for data transmission. Designed for home use. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), FCC Part 15 (radio), FCC Part 68 (telecom). Risk management per ISO 14971:2000.

Indications for Use

Indicated for home-based patients requiring remote transmission of physiological data (e.g., weight, BP, glucose, ECG) from compatible monitors to healthcare facilities. Not for emergency calls, real-time alarms, time-critical data, or patients requiring direct medical supervision/emergency intervention.

Regulatory Classification

Identification

A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ KC4816 page 1 of 4 JUL 1 9 2004 Image /page/0/Picture/2 description: The image shows the logo for RTX Healthcare. The logo consists of the letters "RTX" in a bold, sans-serif font. To the right of the letters, there is a stylized arrow pointing to the right. Below the logo, the words "RTX Healthcare" are printed in a smaller, sans-serif font. ### 510(k) Summary This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirement of 21 CFR 807.92 | Submitter: | Bjarne Flou | |------------------|----------------------| | | Managing Director | | | RTX Healthcare | | | Stroemmen 6 | | | DK-9400 Noerresundby | | | Denmark | | | Tel: +45 96322300 | | | Fax: +45 96322310 | | | Email: bf@rtx.dk | | Contact person: | Niels Ole Andersen | | | Engineering Manager | | | Email: noa@rtx.dk | | Date of summary: | 04/15/2004 | Physiological Transmitter and Receiver Common Name: Trade name: RTX3320 Wireless Telehealth Gateway Classification name:21 CFR 870.2910 Physiological Signal Transmitter And Receiver. Classification no: DRG #### Predicate Device: The RTX3320 device is substantially equivalent to the following predicate device: 510(k) number: K023749 Device name: M3810A Philips Telemonitoring System with M3812B TeleStation. Applicant: Philips Medical Systems #### Submission Device Description: The RTX3320 telemedicine device perform transmission of physiological patient information to and from wireless patient monitors, and a remote data server healthcare facility using standard digital communication technologies and protocols. The RTX3320, with its build-in modem, transmits data using the public switched telephone network. It is designed not to interfere with the normal use of phones on the same telephone line. The RTX3320 device is not operated by, or used directly on a patient, and poses no significant risk to the patient or other people within the patient's home. #### Intended use and indications for use: The RTX3320 device is for use by patients at home. It is intended to be used in combination with a variety of patient monitors upon the prescription of a licensed physician or other authorized healthcare provider. RTX3320 serves as the remote communication link between compatible patient monitors, and the compatible healthcare facility at another location. The healthcare facility could be at a disease management center, or with the healthcare/wellness provider or other out-of-hospital caregivers. {2}------------------------------------------------ The purpose is to transmit selected medical information (such as weight, blood pressure, blood glucose) over a normal residential telephone line. RTX3320 is a "black box" device – it is not connected to the patient and not operated by the patient. The installation must be performed by or under supervision of an authorized professional healthcare company. This device does not measure, interpret or make any decisions on the vital data that it conveys. #### Contraindications, precautions and warnings This device is not intended for emergency calls, and may not be used for transmission or indication of any real-time alarms or time-critical data. Clinical judgement and experience are required to check and interpret the measurements collected and transmitted. This device is not for use in systems which substitute for medical care. This device is not intended for patients requiring direct medical supervision or emergency intervention. | Item | Predicate device | Submission device | | |------|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------| | 1 | Intended use / Indication for use | See section 6. | See section 2. | | 2 | Intended users | Home users and healthcare providers. | Home users and healthcare providers. | | 3 | Site of use | Typically for use in patient's home, placed on a normal table. | Typically for use in patient's home, placed on a normal table. | | 4 | Prescription | The device is intended to be used upon prescription of a licensed physician or authorized healthcare provider. | The device is intended to be used upon prescription of a licensed physician or authorized healthcare provider. | | 5 | System description | Telemedicine system consisting of a device that is working as hub/gateway sending data measured by the system patient monitors to a system data server. | Telemedicine device that is working as hub/gateway sending measured data from compatible patient monitors to a compatible data server. | | 6 | Wireless patient monitors | Wireless connection between the patient monitors and the hub/gateway. | Wireless connection between the patient monitors and the hub/gateway. | | 7 | Transmission | Residential telephone lines | Residential telephone lines | | 8 | Patient Interactions | Display and push buttons for collection of patient typed data | No patient interaction. ("Black box") | | 9 | Measurements taken | Blood pressure, weight, ECG and Blood glucose | Blood pressure, weight, ECG, Blood glucose and other measurements provided from compatible monitor devices. | | 10a | Contra indications and warnings | The device does not send any real time alarms. | The device is not for emergency calls, and may not be used to send any real-time alarms or time-critical data | #### Substantial Equivalence Comparison table {3}------------------------------------------------ KC41816 Page 3 of 4 | 10b | Contra indications<br>and warnings | Clinical judgment and<br>experience are required to<br>check and interpret the<br>information delivered. | Clinical judgment and<br>experience are required to<br>check and interpret the<br>measurements that are taken,<br>collected, and delivered by<br>systems using a RTX3320<br>device. | |-----|-----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 10c | Contra indications<br>and warnings | The device is not intended as a<br>substitute for medical care. | The device is not for use in<br>systems which substitute for<br>medical care. | | 10d | Contra indications<br>and warnings | The device is not for use in<br>systems set up for patients who<br>need direct medical supervision | The device is not for use in<br>systems set up for patients who<br>need direct medical supervision<br>or who might need emergency<br>intervention. | | 11 | Wireless link<br>between patient<br>monitors and the<br>gateway | Short range radio system. | Short range radio system using<br>Bluetooth technology. | | 12 | Environmental<br>specifications | Proprietary information | See section 5. | | 11 | Wireless link<br>between patient<br>monitors and the<br>gateway | Short range radio system. | Short range radio system using<br>Bluetooth technology. | | 12 | Environmental<br>specifications | Proprietary information | See section 5. | #### Discussion on differences: ltem 1: The intended use / Indications for use for the predicate device and submission device is generally the same. The exact use is for both devices to be decided and supervised by a licensed physician or an authorized healthcare provider, but limited to be within the intended use and with respect to the contra indications and warnings which are also generally the same for both devices. ltem 5 The predicate device is a system consisting of patient monitor devices, a hub/gateway device and a system server software. The Submission device is a hub/gateway device with a specified generic protocol interface to any compatible patient monitor, and with a specified generic protocol interface to any compatible system server. The protocols are validated against existing compatible patient monitors and servers, and in the labeling it is stated that only compatible and system validated patient monitors and servers must be used with the hub/gateway. Verfication and validation done on the compatible patient monitors ensure that no extra safety or performance risks are added when using the submission device compared with the predicate device. The server side and database management must always be validated by the responsible system operator or provider, and using the generic and properly documented protocol adds no extra safety or performance risks to a system using the submission device compared to the predicate device. {4}------------------------------------------------ Kc41816 page 4 of 4 It is very important to state that both the predicate and the submission device may only be n is very important to other that both the presence is used to check and interpret the measurements that are taken, collected and transmitted by the devices. that are taken, bonoots and is fitted with a display and some push buttons for patient interaction. The purpose is to collect some additional information from the patient. The interaction device is not fitted with a display, and no user interaction is possible or necessary to perform the intended use. hecossary to portom the internable assess physiological parameters as the predicate device r to oubmibolon as now over otheraction on the device. If additional information is needed this may be achieved by a normal voice phone call. may be arrithe predicate device the measurements taken are defined by the patient momitors that are a part of the system. For the submission device the measurements taken mention and by the specific compatible patient monitors. Because of the similar intended use and indication for use for both the predicate device and the submission device, the measurements will typically be the same. Inoded.omomonik thir typecifications about the wireless link between the patient monitors and the Gateway for the predicate device are proprietary and not available for RTX Healthcare. Based on the knowledge available as a leading company within wireless connectivity RTX Healthcare has evaluated that many different short range radio solutions are suitable for this type of devices without adding any additional risks to the patient. RTX Healthcare has chosen the Bluetooth wireless technology used in other medical devices like the Nonin Pulseoximeter (FDA-K041156 ) developed by RTX Healthcare. The general radio signal safety requirements and FCC Part 15 rules are regulatory requirements that are equal for both the predicate device and the submission device. Since the only purpose using a wireless radio link is to replace a cable, the wireless link in the submission device is evaluated not to add any additional risks within the intended use compared to the predicate device. Item 12: Detailed environmental specifications for the predicate device are proprietary and not available for RTX Healthcare. This is evaluated not to add any additional risks to the patient since both the predicate device and the submission device, according to the intended use, are designed for use by patients at home. The environmental specifications for the submission device are defined according to IEC60601-1 and specifications available for other medical devices for home use. #### Performance data: The RTX3200 device has been tested to meet the requirements of the following standards and regulations used as acceptance criteria: IEC 60601-1, IEC 60601-1-2, FCC part 15 and FCC Part 68. Risk management is performed according to ISO14971:2000. Based on the fact that the performance comparison of the predicate device and the submission device show that the differences are minor and causes no harm to the user, and the fact that the intended use and indication for use is the same, it was early in the project decided to focus on verification and internal validation instead of large scale validation in form of clinical investigation. Verification and validation testing activities is conducted to establish performance and reliability characteristics of the device. {5}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus. The logo is presented in black and white. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 1 9 2004 RTX Healthcare A/S c/o Dr. J.A. van Vugt Certification Manager KEMA Quality B.V. Utrechtseweg 310, NL-6812 AR Arnhem P.O. Box 5185, 6802 ED Arnhem THE NETHERLANDS Re: K041816 Trade Name: RTX3320 Wireless Telehealth Gateway Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency Physiological Signal Transmitter and Receiver Regulatory Class: II (two) Product Code: DRG Dated: July 2, 2004 Received: July 6, 2004 Dear Dr. van Vugt: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your bocurity is the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosuly to regars and ment date of the Medical Device Amendments, or to commence prior to May 20, 1978, the encordance with the provisions of the Federal Food, Drug, devices mat have occh recultion in assee approval of a premarket approval application (PMA). and Cosmetic Act (Act) that do not requent of the general controls provisions of the Act. The I ou may, merclore, maniel the Act include requirements for annual registration, listing of general controls provisions of the rees, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it IT your device is classified (500 abs. 19) als. Existing major regulations affecting your device can may be subject to such additional controls. Existing major regulations may be subject to suer additions, Title 21, Parts 800 to 898. In addition, FDA may be found in the Oode of Peachar Singerning your device in the Federal Register. {6}------------------------------------------------ Page 2 - Dr. J.A. van Vugt Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours. Neil McOglin Bram D. Zuckerman, M. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ #### 3. Indications for use statement # Indication for Use Statement #### 510(k) Number (if known): Device name: RTX3320 Wireless Telehealth Gateway #### Indications for Use: The RTX3320 device is for use by patients at home. It is intended to be used in combination with a variety of patient monitors upon the prescription of a licensed physician or other authorized healthcare provider. RTX3320 serves as the remote communication link between compatible patient monitors, and the compatible healthcare facility at another location. The healthcare facility could be at a disease management center, or with the healthcare/wellness provider or other out-of-hospital caregivers. The purpose is to transmit selected medical information (such as weight, blood pressure, blood glucose) over a normal residential telephone line. RTX3320 is a "black box" device – it is not connected to the patient and not operated by the patient. The installation must be performed by or under supervision of an authorized professional healthcare company. This device does not measure, interpret or make any decisions on the vital data that it conveys. ## Contraindications, precautions and warnings This device is not intended for emergency calls, and may not be used for transmission or indication of any real-time alarms or time-critical data. Clinical judgement and experience are required to check and interpret the measurements collected and transmitted. This device is not for use in systems which substitute for medical care. This device is not intended for patients requiring direct medical supervision or emergency intervention. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Nil R.P. Ogle +N DDZ (Division Sian-Off **Division of Cardiovascular Devices** 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...