← Product Code [DQA](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DQA) · K201887

# VS Newborn Heart Rate Monitor (K201887)

_Surepulse Medical Limited · DQA · Jul 23, 2021 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DQA/K201887

## Device Facts

- **Applicant:** Surepulse Medical Limited
- **Product Code:** [DQA](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DQA.md)
- **Decision Date:** Jul 23, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The VS Newborn Heart Rate Monitor is indicated for the continuous monitoring of heart rate in neonatal patients. The device uses reflectance photoplethysmography to determine heart rate. The sensor is integrated into a cap placed on the head of the patient. The VS Newborn Heart Rate Monitor is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.

## Device Story

Non-invasive heart rate monitor for neonates; used in delivery rooms and NICUs. Device utilizes reflectance photoplethysmography (PPG) via a sensor integrated into a single-use soft cap worn on the newborn's head. PPG signals processed to determine heart rate; output displayed on monitor. Operated by healthcare professionals; provides real-time heart rate data to assist clinical decision-making during neonatal care. Benefits include continuous, non-invasive monitoring without requiring traditional adhesive electrodes.

## Clinical Evidence

Clinical study at Nottingham University Hospitals NHS Trust (n=74 newborns, 13 mothers, 40 healthcare professionals). Assessed thermal properties, device function in NICU and C-section newborns. Primary endpoint: reliability and accuracy (Bland-Altman limits of agreement) compared to ECG electrodes (K024264) and LNCS pulse oximetry sensors (K110723). Secondary endpoints: acquisition time, user feedback. Safety endpoints: skin damage, adverse events. Results: device maintained temperature, demonstrated reliability/accuracy comparable to comparators, and showed no skin irritation or adverse events.

## Technological Characteristics

Reflectance photoplethysmography (PPG) sensor integrated into single-use soft cap. Uses green visible light for HR determination. Biocompatibility per ISO 10993-1. Electrical safety per ISO 60601-1, IEC 60601-1-2, IEC 62133, ISO 80601-2-61. LED safety per IEC 62471. Usability per IEC 62366-1 and IEC 60601-1-6. Algorithm testing per JIS T 1303.

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Predicate Devices

- Masimo Rainbow SET® Rad 87 CO-Oximeter and Accessories ([K091241](/device/K091241.md))

## Reference Devices

- ECG electrodes ([K024264](/device/K024264.md))
- LNCS pulse oximetry sensors ([K110723](/device/K110723.md))
- CARESCAPE Monitor B450 (132533)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

July 23, 2021

SurePulse Medical Limited % Allison Komiyama Prinicipal Consultant AcKnowledge Regulatory Strategies, LLC 2251 San Diego Ave, Suite B-257 San Diego, California 92110

Re: K201887

Trade/Device Name: VS Newborn Heart Rate Monitor Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: July 6, 2020 Received: July 8, 2020

Dear Allison Komiyama:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K201887

#### Device Name VS Newborn Heart Rate Monitor

#### Indications for Use (Describe)

The VS Newborn Heart Rate Monitor is indicated for the continuous monitoring of heart rate in neonatal patients. The device uses reflectance photoplethysmography to determine heart rate. The sensor is integrated into a cap placed on the head of the patient. The VS Newborn Heart Rate Monitor is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.

| Type of Use (Select one or both, as applicable)                                                     |                                                                                                    |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">\u2612</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">\u2610</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Surepulse. The logo consists of a stylized heartbeat symbol in blue, followed by the word "surepulse" in a gradient from blue to green. Below the word "surepulse" is the tagline "Every second counts" in a smaller, light gray font.

510(k) Summary

# 510(k) Summary

#### DATE PREPARED

July 6, 2020

### MANUFACTURER AND 510(k) OWNER

SurePulse Medical Limited Medicity, D6 Thane Road, Nottingham, NG90 6BH, United Kingdom Telephone: Official Contact: James Carpenter, CEO

#### REPRESENTATIVE/CONSULTANT

Allison C. Komiyama, Ph.D., R.A.C. Michelle Rubin-Onur, Ph.D. Lucie Dalet, Ph.D. AcKnowledge Regulatory Strategies, LLC Telephone: Email: akomiyama@acknowledge-rs.com Website: www.AcKnowledge-RS.com

### DEVICE INFORMATION

| Proprietary Name/Trade Name: | VS Newborn Heart Rate Monitor                                                                                                                        |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:                 | Oximeter                                                                                                                                             |
| Regulation Number:           | 21 CFR 870.2700                                                                                                                                      |
| Class:                       | II                                                                                                                                                   |
| Product Code:                | DQA                                                                                                                                                  |
| Premarket Review:            | Anesthesiology                                                                                                                                       |
| Review Panel:                | Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices (OHT1)<br>ENT, Sleep Disordered Breathing, Respiratory and<br>Anesthesia Devices (DHT1C) |

#### PREDICATE DEVICE IDENTIFICATION

The VS Newborn Heart Rate Monitor is substantially equivalent to the following predicate:

| 510(k) Number | Predicate Device Name / Manufacturer                                           | Primary Predicate |
|---------------|--------------------------------------------------------------------------------|-------------------|
| K091241       | Masimo Rainbow SET® Rad 87 CO-Oximeter and<br>Accessories / Masimo Corporation | ✓                 |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Surepulse. The logo consists of a stylized heartbeat symbol in blue, followed by the word "surepulse" in a gradient from blue to green. Below the word "surepulse" is the tagline "Every second counts" in a smaller, light gray font.

# 510(k) Summary DEVICE DESCRIPTION

The VS Newborn Heart Rate Monitor is a non-invasive device intended to continuously monitor heart rate (HR) in neonates immediately after delivery and in newborn intensive care units (NICU). The VS Newborn Heart Rate Monitor measures HR using reflectance photoplethysmography (PPG). The sensor is integrated into a single-use soft cap placed on the head of the newborn such that the sensor contacts the forehead of the newborn. The PPG signal and HR are displayed on the monitor. The VS Newborn Heart Rate Monitor is designed for use in hospitals.

# INDICATIONS FOR USE

The VS Newborn Heart Rate Monitor is indicated for the continuous monitoring of heart rate in neonatal patients. The device uses reflectance photoplethysmography to determine heart rate. The sensor is integrated into a cap placed on the head of the patient. The VS Newborn Heart Rate Monitor is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.

# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

SurePulse believes that the VS Newborn Heart Rate Monitor is substantially equivalent to the predicate device based on the information summarized here:

The subject device has a similar design and uses similar materials as the device cleared in K091241. The subject device has an equivalent intended use and similar technological characteristics as the device cleared in K091241. Unlike the predicate device that uses both green visible light and infrared light to measures heart rate and pulse oximetry, the VS Newborn Heart Rate Monitor only uses green visible light to determine heart rate. Unlike the predicate which is indicated for neonates, pediatric, infant, and adult populations, the VS Newborn Heart Rate Monitor is indicated for neonates only. To address the differences in technological characteristics, the subject device underwent biocompatibility testing, performance bench testing, and a clinical study.

# SUMMARY OF NON-CLINICAL TESTING

No FDA performance standards have been established for the VS Newborn Heart Rate Monitor.

Patient-contacting material was subjected to biocompatibility testing in compliance with ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing with a risk management process.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Surepulse. The logo features a stylized heartbeat symbol in blue, followed by the word "surepulse" in a gradient from blue to green. Below the word "surepulse" is the tagline "Every second counts" in a smaller, light gray font.

510(k) Summary

The following tests were performed to demonstrate safety based on current industry standards:

- . Basic safety and essential performance (per ISO 60601-1)
- Electromagnetic disturbances (per IEC 60601-1-2) .
- . Usability (per IEC 62366-1 and IEC 60601-1-6)
- Safety requirements for batteries (per IEC 62133)
- Basic safety and essential performance of pulse oximeter equipment (per ISO 80601-2-61)
- LED safety (per IEC 62471)
- o Algorithm (per JIS T 1303)

The results of these tests indicate that the VS Newborn Heart Rate Monitor is substantially equivalent to the predicate devices.

# SUMMARY OF CLINICAL TESTING

The VS Newborn Heart Rate Monitor underwent clinical testing at the Nottingham University Hospitals NHS Trust, in Nottingham, England. Three phases of the study were 1) assessing the thermal properties of the cap component 2) testing device function with NICU newborns, and 3) testing device function with C-section newborns. The study assessed multiple endpoints including primary, secondary, and safety endpoints. The primary endpoint of the study was to demonstrate the reliability and accuracy (i.e., Bland-Altman limits of agreement) of the subject device. The secondary endpoints included acquisition time and to obtain feedback about the subject device from mothers and healthcare professionals. Finally, there were two safety endpoints that were also considered throughout the study: any sign of skin damage due to the device, and any other unanticipated adverse events attributed to the device. The subject device was compared to ECG electrodes (K024264) and the LNCS pulse oximetry sensors (K110723). Those two devices were connected to the CARESCAPE Monitor B450 (132533) which was connected to a laptop computer. The subject device was connected to the same laptop computer and all data recorded synchronously. In total, 74 newborns, 13 mothers, and 40 healthcare professionals participated in the study. The subject device was found to maintain temperature as well as the standard of care and was as reliable and accurate as the devices it was compared to. Furthermore, there were no signs of skin irritation or adverse events therefore demonstrating that the VS Newborn Heart Rate Monitor is safe.

#### CONCLUSION

Based on the testing performed, including biocompatibility, electrical safety, and clinical testing, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate devices. The similar indications for use, technological characteristics, and performance characteristics for the proposed VS Newborn Heart Rate Monitor are assessed to be substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DQA/K201887](https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DQA/K201887)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
