K201360 · Xinkang Medical Instrument Co. , Ltd. · DQA · Aug 19, 2020 · Cardiovascular
Device Facts
Record ID
K201360
Device Name
Reusable and Disposable SpO2 Sensors
Applicant
Xinkang Medical Instrument Co. , Ltd.
Product Code
DQA · Cardiovascular
Decision Date
Aug 19, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Indications for Use
The Reusable and Disposable SpO2 Sensors are indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients weighing greater than 40kg. The sensors are intended to be used in hospital settings where patient care is offered by qualified healthcare personnel.
Device Story
Device consists of reusable and disposable SpO2 sensors used as accessories to patient monitors (e.g., Nellcor). Sensors attach to patient fingers to measure oxygenation via red and infrared light absorption. Optical components (LED and photodetector) transmit light through tissue; light absorption varies with hemoglobin oxygenation. Data acquisition system calculates SpO2 and pulse rate based on absorbance ratios of oxygenated vs. total hemoglobin. Used in hospital settings by qualified healthcare personnel. Output displayed on compatible monitor to assist clinicians in assessing patient oxygenation and heart rate, facilitating timely clinical intervention.
Clinical Evidence
Clinical hypoxia study conducted on healthy adult volunteers (ages 21-50) with varying skin pigmentations, including three darkly pigmented subjects. Accuracy validated against arterial blood samples measured via CO-oximeter per ISO 80601-2-61. Results support accuracy claims of ±3% for SpO2 (70-100%) and ±3 bpm for pulse rate (30-250 bpm).
Technological Characteristics
Materials: Silicone, non-woven. Sensing: 2-wavelength optical absorption (Red 660nm, Infrared 880-950nm). Connectivity: Wired connection to compatible patient monitors. Standards: IEC 60601-1, IEC 60601-1-2, ISO 80601-2-61, ISO 10993-5/10. Non-sterile.
Indications for Use
Indicated for continuous non-invasive monitoring of functional SpO2 and pulse rate in adult patients weighing >40kg in hospital settings. Contraindications not specified.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Unimed Reusable and Disposable SpO2 Sensors (K142832)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
August 19, 2020
Xinkang Medical Instrument Co. Ltd. % James Tsai Consultant Shenzhen Joyantech Consulting Co., Ltd. 1713A, 17th Floor, Block A, Zhongguan Times Square, Nanshan District Shenzhen, Guangdong 518000 China
Re: K201360
Trade/Device Name: Reusable and Disposable SpO2 Sensors Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: March 18, 2020 Received: May 21, 2020
Dear James Tsai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201360
#### Device Name
Reusable and Disposable SPO2 Sensors
#### Indications for Use (Describe)
The Reusable and Disposable SPO2 Sensors are indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients weighing greater than 40kg. The sensors are intended to be used in hospital settings where patient care is offered by qualified healthcare personnel.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| <span> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# K201360
This summary of 510(K) safety and effectiveness information is submitted as required by requirements of SMDA and 21 CFR §807.92.
## 1 Administrative Information
| Date of Summary prepared | July 13, 2020 |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer information | Company: Xinkang Medical Instrument Co., Ltd<br>Company address:<br>2 Floor, Puhua Science and Technology Park,<br>Tongsheng Community Dalang Street, Longhua<br>District, 518109 Shenzhen, People's Republic of<br>China<br>Contact person: Xu Changsheng<br>Phone: +86-755-23776681<br>Fax: +86-755-23776861<br>E-mail: 751857289@qq.com |
| Submission Correspondent | Shenzhen Joyantech Consulting Co., Ltd.<br>Address: 1713A, 17th Floor, Block A,<br>Zhongguan Times Square, Nanshan District,<br>Shenzhen<br>Contact person: James Tsai<br>E-Mail: james_tsai@cefda.com |
| Establishment registration number | |
| 2 Device Information | |
| Type of 510(k) submission: | Traditional |
| Trade Name: | Reusable and Disposable SpO2 Sensors |
| Classification name: | Oximeter |
| Review Panel: | Cardiovascular devices |
| Product Code: | DQA |
| Device Class: | II |
{4}------------------------------------------------
## Regulation Number: | 870.2700
#### 3 Predicate Device Information
Unimed Medical Supplies, Inc. Sponsor: Unimed Reusable and Disposable SpO₂ Sensors Device: K142832 510(K) Number: Review Panel: Cardiovascular devices DQA Product Code: Device Class: ll 870.2700 Regulation Number:
#### 4 Device Descriptions
The proposed device, Reusable and Disposable SpO2 sensors are the accessory of the patient monitors, which are intended for continuous monitoring of functional arterial oxygen saturation and pulse rate in non-invasive way with legally marketed devices.
The sensor shall be connected to its corresponding monitor (Nellcor, N-600x cleared in K060576), and it is used to attach the patient's finger and measure oxygenation of blood from detecting the infrared-light and red-light absorption characteristics of deoxygenated hemoglobin and oxygenated hemoglobin. The sensor is connected to a data acquisition system which is used to calculate and display oxygen saturation values and heart rate conditions.
The Reusable and Disposable SpO2 sensors consist of compatible connectors, cable, and patient sensor terminal. And the optical components of sensor are designed to a light emitting diode and a light detector. Red and Infrared lights are shone through the terminal tissues. Then parts of the emitting lights are absorbed by blood and tissues. The light absorbed by the blood varies with the oxygen saturation of haemoglobin. The light detector detects the light volume transmitted through the tissues on the basis of blood pulse, then the microprocessor calculates a value for the oxygen saturation by measuring the absorbance of the wave peak and the wave trough. The saturation values are determined by the percentage ratio of the oxygenated hemoglobin (HbO2) to the total amount of hemoglobin (Hb).
#### 5 Intended Use/ Indications for Use
The Reusable and Disposable SpO2 Sensors are indicated for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpOz) and pulse rate (PR) for adult patients weighing greater than 40kg. The sensors are intended to be used in hospital settings where patient care is offered by qualified healthcare personnel.
#### 6 Comparison to predicate device
{5}------------------------------------------------
| Comparison<br>item | Proposed Device | Predicate Device (K142832) |
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product name | Reusable and<br>disposable SpO2 sensor | Unimed Reusable and Disposable<br>SpO2 Sensors |
| Model | Adult/reusable: XSAE3<br>Adult/disposable: XDAN | Adult/reusable: U403S-01<br>Adult/disposable: U503-01 |
| Product Code | DQA | DQA |
| Regulation<br>Number | 870.2700 | 870.2700 |
| Classification | II | II |
| Intended<br>use &<br>Indication<br>s for Use | The Reusable and Disposable<br>SpO2 Sensors are indicated for<br>continuous non-invasive monitoring<br>of functional oxygen saturation of<br>arterial hemoglobin (SpO2) and<br>pulse rate (PR) for adult patients<br>weighing greater than 40kg. The<br>sensors are intended to be used in<br>hospital settings where patient care<br>is offered by qualified healthcare<br>personnel. | Unimed Disposable and Reusable<br>SpO2 Sensors are indicated for<br>continuous non-invasive monitoring of<br>functional oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse rate<br>(PR) for adult patients weighing<br>greater than 40kg, pediatric patients<br>weighing 10-50 kg, and neonatal<br>patients weighing no less than 3 Kg. |
| Measurement<br>Method | 2-wavelength Relative<br>Optical Absorption | 2-wavelength Relative<br>Optical Absorption |
| Light<br>Emitting | Red: 660nm±3nm<br>Infrared: 880-950nm | Red: 660-666nm<br>Infrared: 880-950nm |
| Signal<br>Detection<br>Method | Photodetector | Photodetector |
| SPO2<br>Accuracy | ±3% (70-100%) | ±3% (70-100%) |
| Pulse<br>Rate<br>Accuracy | ±3(30-250bpm) | ±3 (30-250bpm) |
| Applied<br>population | Adult (≥40Kg) | Adult (≥40Kg) |
| Measurement<br>part | Fingers | Fingers |
| Distal<br>connector<br>design | Soft tip, non-woven | Soft tip, textile adhesive |
| Compatible<br>monitor | Nellcor | Nellcor |
| Sterile | No | No |
| Material | Silicone, Non-woven | Silicone, 3M |
| Biocompatibility | Cytotoxicity<br>Irritation<br>Sensitization | Cytotoxicity<br>Irritation<br>Sensitization |
| Electrical<br>Performance<br>and Safety | IEC60601-1,IEC60601-1-2,<br>ISO80601-2-61,ISO10993-<br>5/10 | IEC60601-1,IEC60601-1-2,<br>ISO80601-2-61,ISO10993-<br>5/10 |
{6}------------------------------------------------
The proposed device and the predicate device have the same intended use and similar technological characteristics; they both measure SpO2 values for the patients.
#### 7 Performance data
The following performance data of Reusable and Disposable SpO2 Sensors were provided in support of the substantial equivalence determination:
#### Biocompatibility testing
The biocompatibility evaluation for the Reusable and Disposable SpO2 Sensors were conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process," as recognized by FDA. The terminal sensor and cable are considered to be contacted with patient's intact skin for duration of less than 24 hours. The biocompatibility testing includes the following:
- Cytotoxicity .
- Skin Sensitization ●
- Skin Irritation .
#### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted, and the results show that the proposed device complies with the following standards:
- IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) + AM1 (2012) Medical . electrical equipment - Part 1: General requirements for basic safety and essential performance for safety
- . IEC 60601-1-2: 2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests standard for EMC
- . ISO 80601-2-61: 2017 + COR1:2018 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of Pulse Oximeter Equipment standard for performance effectiveness
#### Clinical study
Clinical studies were conducted to verify the accuracy of proposed device. The clinical studies were conducted per following standards:
- ISO 80601-2-61: 2017 Medical Electrical Equipment Part 2-61: Particular Requirements for Basic Safety and Essential Performance of Pulse Oximeter Equipment.
- · Pulse Oximeters-Premarket Notification Submissions: Guidance for Industry and Food and Drug Administration Staff Clinical testing has been performed under an approved protocol with subject informed consent.
{7}------------------------------------------------
Clinical hypoxia test results were obtained in human adult volunteers to validate the accuracy of Reusable and Disposable SpO2 Sensors versus arterial oxygen saturation (SaOz) as determined by CO-oximeter. Clinical test results support device accuracy claims for the specified saturation range. The SpO2 accuracy has been validated in human studies against arterial blood sample reference measured with a CO-oximeter. Pulse oximeter measurements are statistically distributed, the volunteer population in the studies composed of healthy men and women from age 21 to 50, with variations of skin pigmentations and per FDA's guidance for Pulse Oximeters, three darkly pigmented subjects are included in the clinical study.
### Summary
Based on the non-clinical performance and clinical data as documented in the device development, the proposed device has a safety and effectiveness profile that is similar to the predicate device.
### 8 Conclusions
Based on device comparison information and performance data, the proposed device has the same intended use as the predicate device, and the differences in technological characteristics does not raise different questions of safety and effectiveness. Therefore, the proposed device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.