K192608 · Shenzhen Sino-K Medical Technology Co.,Ltd · DQA · Jan 24, 2020 · Cardiovascular
Device Facts
Record ID
K192608
Device Name
Disposable SpO2 Sensor
Applicant
Shenzhen Sino-K Medical Technology Co.,Ltd
Product Code
DQA · Cardiovascular
Decision Date
Jan 24, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Indications for Use
The Disposable SPO2 Sensor is indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients weighing greater than 40 kg at hospital facilities.
Device Story
Disposable SpO2 sensor; accessory to oximeter (e.g., Nellcor N-600). Device uses two LEDs (red 660-666nm, infrared 880-950nm) and a photodiode to measure light absorption through patient's finger. Photodiode detects light intensity changes caused by pulsatile blood flow; electrical signals sent to host oximeter. Host oximeter amplifies signal and calculates SpO2 ratio based on crest/trough absorbance of oxygenated vs. deoxygenated hemoglobin. Used in hospital facilities by clinicians for continuous monitoring. Output displayed on host monitor; assists clinicians in assessing patient oxygenation and heart rate status. Benefits include non-invasive, real-time monitoring of patient physiological status.
Clinical Evidence
Clinical hypoxia testing performed on adult human volunteers per ISO 80601-2-61:2017. Accuracy validated by comparing SpO2 readings against arterial oxygen saturation (SaO2) measured via co-oximetry. Results support accuracy claims within the specified 70%-100% SpO2 range.
Technological Characteristics
Disposable finger-clip sensor. Sensing principle: 2-wavelength relative optical absorption (Red 660-666nm, IR 880-950nm). Biocompatibility: ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization). Electrical safety: IEC 60601-1. EMC: IEC 60601-1-2. Performance: ISO 80601-2-61. Connectivity: Wired connection to host oximeter (e.g., Nellcor N-600).
Indications for Use
Indicated for continuous non-invasive monitoring of functional oxygen saturation (SpO2) and pulse rate (PR) in adult patients weighing >40 kg in hospital settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Shenzhen SINO-K Medical Technology Co., Ltd. % Kevin Wang Consultant Chonconn Medical Device Consulting Co., Ltd. No. A415, Block A, NanShan Medical devices Industrial Park, Nanshan District Shenzhen, 518067 CN
Re: K192608
Trade/Device Name: Disposable SpO2 Sensor Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: December 26, 2019 Received: December 26, 2019
Dear Kevin Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
Device Name Disposable SpO2 Sensor
Indications for Use (Describe)
The Disposable SPO2 Sensor is indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients weighing greater than 40 kg at hospital facilities.
| Type of Use (Select one or both, as applicable) | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <table><tr><td><span> <div style="display:inline-block; border: 1px solid black; width: 10px; height: 10px; text-align: center;"> ✓ </div> Prescription Use (Part 21 CFR 801 Subpart D) </span></td><td><span> <div style="display:inline-block; border: 1px solid black; width: 10px; height: 10px; text-align: center;"> </div> Over-The-Counter Use (21 CFR 801 Subpart C) </span></td></tr></table> | <span> <div style="display:inline-block; border: 1px solid black; width: 10px; height: 10px; text-align: center;"> ✓ </div> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <div style="display:inline-block; border: 1px solid black; width: 10px; height: 10px; text-align: center;"> </div> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
| <span> <div style="display:inline-block; border: 1px solid black; width: 10px; height: 10px; text-align: center;"> ✓ </div> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <div style="display:inline-block; border: 1px solid black; width: 10px; height: 10px; text-align: center;"> </div> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
| FORM FDA 3881 (7/17) | Page 1 of 1 | PSC Publishing Services (301) 443-6740 EF |
|----------------------|-------------|-------------------------------------------|
|----------------------|-------------|-------------------------------------------|
{3}------------------------------------------------
K192608
# 510(K) Summary
### Prepared in accordance with the requirements of 21 CFR Part 807.92
Prepared Date: 2019/09/17
#### Submission sponsor 1.
Name: Shenzhen SINO-K Medical Technology Co., Ltd.
Room401, Bldg2, Veteran Ind. City, Gongle Community, Xixiang Street, Baoan District, Shenzhen, Guangdong, China Contact person: Lao Chengxin Title: General Manager E-mail: boss@sino-k.com Tel: +86 137 15333326
#### 2. Submission correspondent
Name: Chonconn Medical Device Consulting Co., Ltd.
Address: No. A415, Block A, NanShan Medical devices Industrial Park Nanshan District, Shenzhen, Guangdong, P.R. China 518067 Contact person: Kevin Wang
E-mail: kevin(@chonconn.com
Tel: +86-755 33941160
| Trade/Device Name | Disposable SpO2 Sensor |
|-------------------|---------------------------------|
| Common Name | Oximeter (Accessory-sensor) |
| Regulatory Class | Class II |
| Classification | 21CFR 870.2700 / Oximeter / DQA |
| Submission type | Traditional 510(K) |
#### 3. Subject Device Information
#### 4. Predicate Device
Manufacturer: Orantech Inc. 510 (k) #: K181270 Product Name: Disposable SpO2 Sensors
#### 5. Device Description
The proposed device, Disposable SpO2 Sensor is accessory to the oximeter, which is intended for continuous monitoring of functional arterial oxygen saturation and pulse rate in non-invasive with U.S. legally marketed oximeter. It is only intended for adult.
The sensor shall be connected with the corresponding monitor (Nellcor N-600). Oxygenation of blood is measured by detecting the infrared and red-light absorption characteristics of deoxygenated hemoglobin
{4}------------------------------------------------
and oxygenated hemoglobin, which consists of a probe attached to the patient's finger. The sensor is connected to a data acquisition system which is used to calculate and display oxygen saturation levels and heart rate conditions.
Each sensor has two LEDs, emitting both red and infrared light, and a photodiode. Red and infrared light lit alternately according to certain sequence, when the fingertips of capillary repeatedly with the heart pumps blood congestion, light emitting diode after blood vessels and projected onto a photodiode, photodiode can be induced to change with pulse light intensity, the electrical signals in the form of change. Then the received signal is forwarded to the corresponding oximeter that amplifies the signal and an algorithm that calculates the ratio. By measuring the wave crest of the pulse wave and the absorbance of the trough, SpO2 is calculated to obtain the correct oxygen saturation value. The saturation value is determined by the percentage ratio of the oxygenated hemoglobin (HbO2) to the total amount of hemoglobin (Hb).
#### Intended use & Indication for use 6.
The Disposable SPO2 Sensor is indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients weighing greater than 40 kg at hospital facilities.
| Features | Subject Device<br>Disposable SpO2 Sensors | Predicate Device K181270<br>Disposable SpO2 Sensors | Comparison |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Applicant | Shenzhen SINO-K Medical<br>Technology Co., Ltd. | Orantech Inc. | / |
| Classification<br>Regulation | 21CFR 870.2700 | 21CFR 870.2700 | Same |
| Classification<br>and Code | Class II, DQA | Class II, DQA | Same |
| Common<br>name | Oximeter (Accessory-sensor) | Oximeter (Accessory-sensor) | Same |
| Intended use | The Disposable SPO2 Sensor is<br>indicated for continuous non-<br>invasive monitoring of<br>functional oxygen saturation of<br>arterial hemoglobin (SpO2) and<br>pulse rate (PR) for adult<br>patients weighing greater than<br>40 kg at hospital facilities. | The Disposable and Reusable<br>SPO2 Sensors are indicated for<br>continuous non-invasive<br>monitoring of functional oxygen<br>saturation of arterial hemoglobin<br>(SpO2) and pulse rate (PR) for<br>adult patients weighing greater<br>than 40 kg at hospital facilities. | Same |
| Principle of<br>operation | 2-wavelength Relative Optical<br>Absorption | 2-wavelength Relative Optical<br>Absorption | Same |
#### 7. Comparison to the Predicate Device
{5}------------------------------------------------
| Features | Subject Device<br>Disposable SpO2 Sensors | Predicate Device K181270<br>Disposable SpO2 Sensors | Comparison |
|-------------------------------|-------------------------------------------|---------------------------------------------------------|--------------|
| Light<br>Emitting | Red: 660-666nm<br>Infrared: 880-950nm | Red: 660, 661 and 663nm<br>Infrared: 890, 904 and 940nm | Different 1) |
| Signal<br>Detection<br>Method | Photodetector | Photodetector | Same |
| SpO2 Range | 70%-100% | 70%-100% | Same |
| SpO2<br>Accuracy | ±3% | ±3% | Same |
| PR Range | 30 bpm - 250 bpm | 35 bpm - 240 bpm | Different 2) |
| PR Accuracy | ±3 bpm | ±2 bpm | |
| Sterile | No | No | Same |
| Usage | Disposable | Disposable | Same |
| Electrical<br>Safety | Complied with IEC 60601-1 | Complied with IEC 60601-1 | Same |
| EMC | Complied with IEC 60601-1-2 | Complied with IEC 60601-1-2 | Same |
| Performance | Complied with ISO 80601-2-61 | Complied with ISO 80601-2-61 | Same |
| Biocompatibility | | | |
| Cytotoxicity | Complied with ISO 10993-5 | Complied with ISO 10993-5 | Same |
| Skin Irritation | Complied with ISO 10993-10 | Complied with ISO 10993-10 | Same |
| Sensitization | Complied with ISO 10993-10 | Complied with ISO 10993-10 | Same |
Justifications for differences between proposed device and the predicate device are shown as below:
Different (1): The light emitting between proposed device and predicate device is slightly different. The light emitting affects the SpO2 measurement. The SpO2 and PR measurement has been verified and validated according to ISO 80601-2-61: 2017 clause 201.12.1.101 SpO2 accuracy of pulse oximeter equipment. This difference does not raise any different questions of safety or effectiveness.
Different (2): The PR measurement between proposed device and predicate device is different. This specification has been verified and validated according to ISO 80601-2-61: 2017. This difference does not raise any different questions of safety or effectiveness.
#### 8. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility testing
The biocompatibility evaluation for the subject device was conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The battery of testing included the following tests:
- Cytotoxicity .
{6}------------------------------------------------
- . Sensitization
- Irritation .
The subject device is considered surface contacting for a duration of not exceed 24 hours.
### Non-clinical data
The disposable SpO2 Sensor has been tested according to the following standards:
- IEC 60601-1-2005+CORR.1:2006+CORR.2:2007+A1:2012, Medical Electrical Equipment-. Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic . safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests
- . ISO 80601-2-61: 2017 Medical Electrical Equipment - Part 2-61: Particular Requirements for Basic Safety and Essential Performance of Pulse Oximeter Equipment.
## Clinical data
Clinical studies were conducted to verify the accuracy of proposed device. The clinical studies were conducted per following standards:
- ISO 80601-2-61: 2017 Medical Electrical Equipment Part 2-61: Particular Requirements for . Basic Safety and Essential Performance of Pulse Oximeter Equipment.
- . Pulse Oximeters-Premarket Notification Submissions: Guidance for Industry and Food and Drug Administration Staff
Clinical testing has been performed under an approved protocol with subject informed consent. Clinical hypoxia test results were obtained in human adult volunteers to validate the accuracy of disposable SpO2 Sensor versus arterial oxygen saturation (SaO2) as determined by co-oximetry. Clinical test results support device accuracy claims for the specified saturation range.
#### 9. Conclusion
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate devices.