SHENZHEN MED-LINKET ADULT FINGER CLIP SPO2 SENSOR, SHENZHEN MED-LINKET ADULT SILICONE SOFT TIP SPO2 SENSOR

K083010 · Shenzhen Med-Link Electronics Tech Co., Ltd. · DQA · Jun 10, 2009 · Cardiovascular

Device Facts

Record IDK083010
Device NameSHENZHEN MED-LINKET ADULT FINGER CLIP SPO2 SENSOR, SHENZHEN MED-LINKET ADULT SILICONE SOFT TIP SPO2 SENSOR
ApplicantShenzhen Med-Link Electronics Tech Co., Ltd.
Product CodeDQA · Cardiovascular
Decision DateJun 10, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2700
Device ClassClass 2

Indications for Use

The ProFlu™ Assay is a multiplex Real Time RT-PCR in vitro diagnostic test for the rapid and qualitative detection and discrimination of Influenza A Virus, Influenza B Virus, and Respiratory Syncytial Virus (RSV) nucleic acids isolated and purified from nasopharyngeal (NP) swab specimens obtained from symptomatic patients. This test is intended for use to aid in the differential diagnosis of Influenza A, Influenza B and RSV viral infections in humans and is not intended to detect Influenza C. Negative results do not preclude influenza or RSV virus infection and should not be used as the sole basis for treatment or other management decisions. It is recommended that negative RSV results be confirmed by culture. Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary. If infections with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Device Story

ProFlu™ Plus is a multiplex Real Time RT-PCR assay for qualitative detection of Influenza A, Influenza B, and RSV nucleic acids. Input consists of nucleic acids isolated from nasopharyngeal swab specimens. The device utilizes the bioMérieux NucliSENS easyMAG system or Roche MagNA Pure LC system for nucleic acid purification, followed by amplification and detection on the Cepheid SmartCycler II Real Time Instrument. The assay provides qualitative results to aid clinicians in differential diagnosis. It is intended for use in clinical laboratory settings by trained personnel. Results are interpreted by healthcare providers to guide patient management, though negative results do not rule out infection and require clinical correlation. The device benefits patients by providing rapid viral identification to inform treatment decisions.

Clinical Evidence

Clinical testing performed under institutionally approved protocol with informed consent. Validated performance and accuracy under controlled hypoxia compared to co-oximetry. Results confirmed sensors meet published accuracy specifications.

Technological Characteristics

Multiplex Real Time RT-PCR assay. Uses bioMérieux NucliSENS easyMAG or Roche MagNA Pure LC for automated nucleic acid extraction. Detection performed on Cepheid SmartCycler II Real Time Instrument. Reagents include specific lysis buffers and magnetic silica for extraction. Connectivity is via the SmartCycler II instrument software (version 1.7b).

Indications for Use

Indicated for symptomatic patients requiring differential diagnosis of Influenza A, Influenza B, and RSV viral infections via nasopharyngeal swab specimens. Prescription use only.

Regulatory Classification

Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K083010 B. Purpose for Submission: Request for removal of the requirement for culture confirmation of influenza A and B negative sample results from intend use of K073029 and adding the use of the bioMerieux EasyMag nucleic acid purification method. C. Measurand: See K073029 D. Type of Test: See K073029 E. Applicant: Prodesse Incorporated F. Proprietary and Established Names: ProFlu™ Plus Common Name: Respiratory Viral Panel (RVP) Multiplex Nucleic Acid Detection Assay G. Regulatory Information: 1. Regulation section: 21 CFR 866.3980, Respiratory viral panel multiplex nucleic acid assay 2. Classification: Class II 3. Product code: OCC 4. Panel: Microbiology (83) H. Intended Use: 1. Intended use(s): The ProFlu™ Assay is a multiplex Real Time RT-PCR in vitro diagnostic test for the rapid and qualitative detection and discrimination of Influenza A Virus, Influenza B Virus, and Respiratory Syncytial Virus (RSV) nucleic acids isolated and purified from nasopharyngeal (NP) swab specimens obtained from symptomatic patients. This test is intended for use to aid in the differential diagnosis of Influenza A, Influenza B and RSV viral infections in humans and is not intended to detect Influenza C. Negative results do not preclude influenza or RSV virus infection and should not be used as the sole basis for treatment or other management decisions. It is recommended that negative RSV results be confirmed by culture. Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary. If infections with a novel Influenza A virus is suspected based on current clinical and {1} epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. 2. Indication(s) for use: Same as Intended Use. 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: Cepheid SmartCycler II Real Time Instrument I. Device Description: See K073029 Materials Provided See K073029 Materials Required But Not Provided Plasticware and consumables ☐ Polyester, rayon or nylon tipped nasopharyngeal swabs ☐ RNase/DNase-free 1.5 mL polypropylene microcentrifuge tubes ☐ Sterile RNase/DNase-free filter or positive displacement micropipettor tips ☐ MagNA Pure LC System Disposables (Reagent Tubs, Reaction Tips, Tip Trays, Cartridges) or easyMAG System Disposables (Sample Vessels and Tips) ☐ Biohit Pipette Tips for use with easyMAG System ☐ Greiner Break Four uncoated plates for use with easyMAG System ☐ Cepheid PCR reaction tubes, 25 µL ☐ Parafilm® M or MagNA Pure LC Cartridge Seals Reagents ☐ Roche MagNA Pure LC Total Nucleic Acid Isolation Kit (Roche Cat. # 03038505001) for 192 isolations or bioMérieux NucliSENS easyMAG reagents (Buffer 1 Cat. # 280130, Buffer 2 Cat. # 280131, Buffer 3 Cat. # 280132, Magnetic Silica Cat. # 280133, and Lysis Buffer Cat. # 280134) ☐ Micro Test™ M4 Viral Transport Medium (Remel, Inc. Cat. # 12500) or BD Universal Viral Transport medium (UTM; Becton, Dickinson and Co. Cat. # 220220) ☐ Molecular Grade Water (RNase/DNase Free) {2} ☐ Extraction Control (e.g. previously characterized positive sample) ## Equipment ☐ – 70°C Freezer ☐ Roche MagNA Pure LC System with software version 3.0.11 or bioMérieux NucliSENS easyMAG System with Software version 1.0.1 ☐ Biohit multi-channel pipettor for use with easyMAG System ☐ Cepheid SmartCycler II Real Time Instrument with Dx Software version 1.7b ☐ Micropipettors (range between 1-10 µL, 10-200 µL and 100-1000 µL) ☐ Mini-centrifuge with adapter for Cepheid Reaction Tubes ☐ Cepheid cooling block ☐ Cepheid cooling block ## Interpretation of Specimen Results See K073029 ## J. Substantial Equivalence Information: 1. Predicate device name(s): xTAG™ RVP (Respiratory Viral Panel) Common Name: Respiratory Viral Panel (RVP) Multiplex Nucleic Acid Detection Assay 2. Predicate 510(k) number(s): k063765 3. Comparison with predicate: See K073029 ## K. Standard/Guidance Document Referenced (if applicable): See K073029 ## L. Test Principle: See K073029 ## M. Performance Characteristics (if/when applicable): ### Expected Values See K073029 ### Clinical Performance See K073029 {3} # Results from Prospective Study See K073029 # Reproducibility See K073029 and comparing the ProFlu+ reproducibility, defined in this study as the closeness of agreement of repeated testing where only the extraction conditions are changed (easyMAG vs. MagNA Pure). Table 1: Extraction equivalence reproducibility study – average $C_T$ values for both MagNA Pure LC (MP) and easyMAG (EM) across days | Virus and Level | Average \( C_T \)s | | Standard Deviation | | 95% Confidence Interval | | | --- | --- | --- | --- | --- | --- | --- | | | MP | EM | MP | EM | MP | EM | | IA Low | 31.43 | 29.90 | 0.35 | 0.15 | 31.21 - 31.65 | 29.81 - 29.99 | | IA Med | 29.22 | 27.72 | 0.23 | 0.18 | 29.08 - 29.36 | 27.61 - 27.83 | | RA Low | 28.69 | 27.19 | 1.07 | 0.88 | 28.03 - 29.35 | 26.64 - 27.74 | | RA Med | 25.91 | 24.9 | 0.64 | 0.36 | 25.51 - 26.31 | 24.68 - 25.12 | # Analytical Sensitivity Analytical sensitivity (LoD) as defined as the lowest concentration at which $\geq 95\%$ of all replicates tested positive, ranged from $10^{2} - 10^{-1}\mathrm{TCID}_{50} / \mathrm{mL}$ . See K073029 and determining and comparing the Limits of Detection (LoDs) for Influenza A, Influenza B, RSV A and RSV B on each automated extractor. Table 2: LoD Summary for both MagNA Pure LC (MP) and easyMAG (EM) | Organism | LoD TCID50/mL | | Positivity | | % Positive | | | --- | --- | --- | --- | --- | --- | --- | | | MP | EM | MP | EM | MP | EM | | Influenza A | 1 x 102 | 1 x 102 | 10/10 | 9/9a | 100% | 100% | | Influenza B | 1 x 101 | 1 x 101 | 9/10b | 10/10 | 90% | 100% | | RSV A | 1 x 101 | 1 x 101 | 10/10 | 10/10 | 100% | 100% | | RSV B | 1 x 101 | 1 x 101 | 10/10 | 10/10 | 100% | 100% | a One sample was positive for both IA (expected) and RSV(suspected contamination). The sample was still positive in both the IA and RSV channel after repeat testing in duplicate and therefore, was an invalid sample and was excluded from analysis. b One replicate was negative for IB on the MP only. Repeat testing in duplicate of the extracted nucleic acid gave positive results for IB (original data, not repeat was used for analysis). {4} Table 3: LoD Study – Average $C_T$ Values for both MagNA Pure LC (MP) and easyMAG (EM) | Organism | Concentration TCID50/mL | Number of Positive Samples | | Average CT | | | --- | --- | --- | --- | --- | --- | | | | MP | EM | MP | EM | | Influenza A | 1 x 102 | 10 | 9a | 32.98 | 31.53 | | | 1 x 101 | 8 | 5 | 38.06 | 38.04 | | | 1 x 100 | 0 | 0 | NAb | NAb | | Influenza B | 1 x 101 | 9c | 10 | 29.91 | 29.39 | | | 1 x 100 | 0 | 0 | NAb | NAb | | | 1 x 10-1 | 0 | 0 | NAb | NAb | | RSV A | 1 x 101 | 10 | 10 | 29.22 | 29.96 | | | 1 x 100 | 3 | 5 | 33.8 | 34.22 | | | 1 x 10-1 | 2 | 0 | 33.45 | NAb | | RSV B | 1 x 102 | 10 | 10 | 28.65 | 27.19 | | | 1 x 101 | 10 | 10 | 33.39 | 32.55 | | | 1 x 100 | 4 | 7 | 37.13 | 38.41 | a One sample was positive for both IA (expected) and RSV(suspected contamination). The sample was still positive in both the IA and RSV channel after repeat testing in duplicate and therefore, was an invalid sample and was excluded from analysis. b NA: Not Applicable. No positive samples for that specific target at the specific concentration. One replicate extraction was negative for IB on MP only. Repeat testing in duplicate of the extract gave positive results for IB. The repeat testing results were not included in the analysis for MP. # Reactivity See K073029 # Analytical Specificity See K073029 # Competitive Inhibition See K073029 # Carry-over/Contamination See K073029 # N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. # O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...