WRIST BLOOD PRESSURE METER MODEL SE-330

K973078 · Sein Electronics Co., Ltd. · DPW · May 20, 1999 · Cardiovascular

Device Facts

Record IDK973078
Device NameWRIST BLOOD PRESSURE METER MODEL SE-330
ApplicantSein Electronics Co., Ltd.
Product CodeDPW · Cardiovascular
Decision DateMay 20, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2100
Device ClassClass 2

Indications for Use

For the measurement of systolic and diastolic blood pressure and pulse rate in adult patients.

Device Story

The SE-330 is a semi-automatic digital wrist blood pressure meter. It uses an oscillometric method to measure blood pressure and pulse rate. The device consists of a compact wrist cuff, a manual inflation bulb, and a solenoid valve for controlled deflation. It is intended for use by adults. The user manually inflates the cuff; the device then automatically measures pressure and pulse, displaying results on a liquid crystal display (LCD). It includes a memory function to store and display the previous measurement. The device features an automatic power cut-off after one minute of inactivity to conserve battery life. It is powered by two CR2032 lithium batteries. The device provides a non-invasive method for patients to monitor their blood pressure and pulse, aiding in health management.

Clinical Evidence

Clinical trial conducted on 108 patients following the ANSI/AAMI SP-10 1992 Guidance Document for Electronic or Automatic Sphygmomanometers. The study demonstrated that the SE-330 is equivalent to the cuff-stethoscope auscultatory method for determining blood pressure and pulse.

Technological Characteristics

Oscillometric measurement principle; manual inflation bulb with solenoid valve for deflation; liquid crystal display (LCD); powered by two CR2032 lithium batteries; dimensions 58mm x 70mm x 23.8mm; weight approx 110g; wrist cuff dimensions 314.7mm x 70mm. No specific software architecture or connectivity mentioned.

Indications for Use

Indicated for the measurement of systolic and diastolic blood pressure and pulse rate in adult patients.

Regulatory Classification

Identification

A cardiovascular blood flowmeter is a device that is connected to a flow transducer that energizes the transducer and processes and displays the blood flow signal.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ MAY 20 1995 SEIn ELECTRONICS CO.,LTD. 133 3. PYUNGCHON DONG. ANYANG SHI, KYUNGKI DO, KORFA 11 L . ANYANG (0343) 21-3201, 21 0389 JAX (0343)21 5639 ## 510(k) SUMMARY "This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR -807.92." "The assigned 510(k) number is: _ . " Applicant: SEIN Electronics Co., Ltd. 133--3, Pyung Chon -- Dong, Dong An-Gu Anyang -- City, Kyungki--Do, Korea Tel: Anyang (0343) 21-3201 FAX: (0343) 21 - 5639 Brenda M. Kelly, M.D. Contact Person: Executive V.P., Medical & Regulatory Affairs The Lahr Consulting Group, Inc. One Lethbridge Plaza, Suite #4 Mahwah, N.J. 07430 Tel: (201) 529-3111 FAX: (201) 529-0059 Date August , 1997 Name of Device: | Proprietary Name: | Digital Wrist Blood Pressure Meter, Model SE-330 | |-----------------------|---------------------------------------------------------------------------------| | Common or Usual Name: | Digital Wrist Blood Pressure Meter | | Classification Name: | System, Measurement, Blood Pressure,<br>Non-Invasive, Systolic and/or Diastolic | The Digital Wrist Blood Pressure Meter, Model SE-330, is equivalent to the OMRON compact wrist blood pressure meter, Model HEM-605. The SEIN SE-330 Wrist Blood Pressure Meter is a semi-automatic electronic device designed to measure blood pressure and pulse, and in which a compact wrist cuff is used in place of the standard wrist cuff. Inflation and deflation are accomplished through the use of a manual inflation bulb and solenoid valve respectively. {2}------------------------------------------------ 510(k) SUMMARY Page Two -- ﻟﺪ ﻟﻠﺘﻌ ﻤﺪﻳﻨﺔ Blood pressure and pulse are measured using the oscillometric method. A liquid crystal display of up to six digits is employed to record blood pressure and pulse. Power is supplied by two lithium batteries (size CR2032) Built--in memory allows the display of the previously measured blood pressure and pulse values for comparison to current measurements. An automatic power cut-off feature after one minute of non-use is built into the device to permit savinq of energy. The measuring range of the device is 20-280 mmHg (pressure) and 40-200 pulses per minute with an accuracy of +3mmHg or 2% of the reading, whichever is greater, for the pressure and t5% of the reading for pulse. The dimensions of the SE-330 Blood Pressure Meter are 58 mm (W) and 70 mm (D) x 23.8 mm (H). The weight of the meter is approximately 110 g including two batteries. The wrist cuff dimensions are 314.7 mm (1) x 70 mm (w). Battery life for the device is approximately 150 measurements. A clinical trial based on the ANSI/AAMI SP-10 1992 Guidance Document for Electronic or Automatic Sphygmomanometers and conducted on 108 patients shows the SE-330 Wrist Blood Pressure Meter to be equivalent to the cuff-stethoscope auscultatory method for determining blood pressure and pulse. The SEIN SE-330 and OMRON HEM 605 blood pressure meters are similar in that they employ the oscillometric method of measuring, use batteries as their power source, have liquid crystal display of the measurements, and have similar measuring ranges and accuracy. Both devices employ a compact wrist cuff. Both devices exhibit automatic cut-off power when the instruments are not in use. Operating and storage environment requirements also are similar. Differences between the two devices include the following: The SE-330 uses 2 Lithium batteries, Type CR2032, whereas the HEM-605 uses 2 Alkaline batteries, Type LR03. The SE-330 has built-in memory enabling display of the previously measured value as well as current measurements. The HEM-605 does not have this feature. {3}------------------------------------------------ 510(k) SUMMARY Page Three … > In the SE-330 deflation is controlled by a solenoid valve which keeps velocity constant; the HEM-605 employs a constant release valve system. Differences between the SE-330 and HEM-605 also include manua1 repressurization in the SE-330 and automatic repressurization in the HEM--605 . In addition, the weights and dimensions of the two devices are somewhat different. August 13, 1997 ![]() Tae Young Choi, President - , # {4}------------------------------------------------ Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is an abstract image of an eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY 20 1999 Brenda M. Kelly, M.D. Executive Vice President, Medical and Regulatory Affairs The LAHR Consulting Group, Inc. One Lethbridge Plaza Mahwah, NJ 07430-2113 Re: K973078 Automatic Digital Wrist Blood Pressure Meter Model SE-330 Requlatory Class: II (Two) Product Code: 74 DPW Dated: March 8, 1999 Received: March 11 1999 Dear Dr. Kelly: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in In addition, FDA may publish further regulatory action. announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. {5}------------------------------------------------ Page 2 - Brenda M. Kelly, M.D. This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Thomas Flanagan Carlson Director Division of Cardiovascular, Respiratory and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ 510(k) Number (if known): K923078 Device Name: Automatic Digital Wrist Blood Pressure Meter - Model SE-330 Indications For Use: For the measurement of systo1ic and diastolic blood s of the measurements on adult patients. canons For Osc. For the meaborement. pressure and pulse rate in adult patients. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Bure L. Campbell Division of Curdlovascular, Respiratory, and Neurological Devices 510(k) Number. Prescription Use (Per 21 CFR 801.109) ે જિલ્લાઓ પૈકીના એક એવા ગામનાં એક ગામના લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં મુખ્યત્વે ખેત ## Over-The-Counter Use_x (Üpılonal Format 1-2-96)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...