← Product Code [DPS](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DPS) · K223013

# AccurECG Analysis System (K223013)

_Accurkardia, Inc. · DPS · Jun 23, 2023 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DPS/K223013

## Device Facts

- **Applicant:** Accurkardia, Inc.
- **Product Code:** [DPS](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DPS.md)
- **Decision Date:** Jun 23, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2340
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Software as a Medical Device

## Indications for Use

The AccurECG Analysis System is intended for use by qualified healthcare professionals in the assessment of a patient's recorded ambulatory ECG data. Analysis results are provided in a standard report for review and printing. The System provides single-lead analysis on a beat-by-beat basis, Ventricular Ectopic Beat and Supraventricular Ectopic Beat detection, heart rate measurement, and rhythm analysis. The AccurECG Analysis System is compatible with ECG recordings taken with silver/silver chloride (Ag / AgCl) electrodes (wet leads). The AccurECG Analysis System is not for use in life-supporting or sustaining systems or ECG monitor and alarm devices. The System is not intended to be used with cardiac telemetry monitors. The AccurECG Analysis System's automated interpretation results are not intended to be the sole means of diagnosis. The AccurECG Report is an adjunct intended to facilitate health care decision making in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms, and other diagnostic information to arrive at a diagnostic conclusion. The AccurECG Analysis System does not interface with data management systems or hardware and is device independent.

## Device Story

AccurECG Analysis System is a web-based software platform for analyzing ambulatory ECG recordings. Workflow: user uploads ECG files via web interface to secure database; proprietary algorithm performs signal processing, QRS complex/T-wave delineation, and arrhythmia classification; system generates report with original signal and analysis results for physician review. Used by healthcare professionals in clinical settings to support diagnostic decision-making. Device is independent, does not interface with external hardware/data management systems. Benefits include automated beat-by-beat analysis and rhythm interpretation to assist clinicians in identifying cardiac abnormalities.

## Clinical Evidence

Bench testing only. Performance evaluated per AAMI/ANSI EC 57:2012 and IEC 60601-2-47:2012. Software verification and validation conducted against gold standard clinical inputs. Usability testing performed with Certified Cardiovascular Technicians.

## Technological Characteristics

Web-based software platform; client-side web interface; server-side proprietary interpretation algorithm. Compatible with Ag/AgCl wet lead ECG recordings. Software Level of Concern: Major. Algorithm implemented in LabVIEW using G programming language.

## Regulatory Identification

An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.

## Predicate Devices

- Cardiologs Holter Platform ([K212112](/device/K212112.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".

June 23, 2023

AccurKardia, Inc. % Adrienne Lenz Principle Medical Device Regulation Expert Hyman, Phelps & McNamara, PC 700 Thirteenth Street, N.W., Suite 1200 Washington, District of Columbia 20005-5929

Re: K223013

Trade/Device Name: AccurECG® Analysis System Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS, DQK Dated: May 26, 2023 Received: May 26, 2023

Dear Adrienne Lenz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shruti N. Mistry -S

for

Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known)

#### K223013

#### Device Name

AccurECG® Analysis System

#### Indications for Use (Describe)

The AccurECG Analysis System is intended for use by qualified healthcare professionals in the assessment of a patient's recorded ambulatory ECG data. Analysis results are provided in a standard report for review and printing.

The System provides single-lead analysis on a beat-by-beat basis, Ventricular Ectopic Beat detection, heart rate measurement, and rhythm analysis System is compatible with ECG recordings taken with silver/silver chloride (Ag / AgCI) electrodes (wet leads).

The AccurECG Analysis System is not for use in life-supporting or ECG monitor and alarm devices. The System is not intended to be used with cardiac telemetry monitors.

The AccurECG Analysis System's automated interpretation results are not intended to be the sole means of diagnosis. The AccurEGG Report is an adjunct intended to facilitate health care decision with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms, and other diagnostic information to arrive at a diagnostic conclusion.

The AccurECG Analysis System does not interface with data management systems or hardware and is device independent.

Type of Use (Select one or both, as applicable)

区 Prescription Use (Part 21 CFR 801 Subpart D) [ ] Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displaysa currently valid OMB number."

{3}------------------------------------------------

# SECTION 5 : 510(k) Summary

## 5.1 Applicant/Submitter

| Company Name   | : | AccurKardia, Inc.                |
|----------------|---|----------------------------------|
| Street Address | : | 101 Avenue of the Americas, FL 8 |
| City           | : | New York                         |
| State          | : | New York                         |
| Zip Code       | : | 10013                            |
| Phone Number   | : | 646-334-5728                     |

## 5.2 Contact Person

| Full Name | : Juan C. Jiménez |
|-----------|-------------------|
| Phone     | : 646-334-5728    |

## 5.3 Correspondent Information

| Full Name | : Juan C. Jiménez |
|-----------|-------------------|
| Phone     | : 646-334-5728    |

## 5.4 Date of Preparation

Date of Preparation

: 05/26/2023

## 5.5 Device Information

| Device Name            | AccurECG® Analysis System        |
|------------------------|----------------------------------|
| Model Number           |                                  |
| Common Name            | ECG Analysis System              |
| Regulation             | 21 CFR 870.2340, 21 CFR 870.1425 |
| Product Code           | DPS;DQK                          |
| Reviewing Panel/Branch | Cardiovascular                   |
| Submission Number      | K223013                          |

{4}------------------------------------------------

### 5.6 Predicate Device(s)

|  | Table - 5.1  Predicate Device(s) |  |
|--|----------------------------------|--|
|--|----------------------------------|--|

| Predicate Type | 510(k) Number | Device Name                | Manufacturer            | Product Code |
|----------------|---------------|----------------------------|-------------------------|--------------|
| Primary        | K212112       | Cardiologs Holter Platform | Cardiologs Technologies | DPS,DQK      |

### 5.7 Device Description

The AccurECG Analysis System ("AccurECG" or the "System") consists of: (1) A client-side web interface which provides tools to upload and review ECG recordings via a web application programming interface (API), and (2) A server-side automated proprietary ECG interpretation support algorithm which measures and analyzes ECGs to provide physicians supportive information for ECG analysis.

AccurECG is intended to analyze recordings performed on adults over the age of 22 and works in the following sequence:

- i. The web interface allows the user to select files and upload to the AccurECG secure database.
- ii. The proprietary AccurECG algorithm analyzes and labels the ECG:
	- o Delineation detection of ORS complexes and T waves on the ECG signal through advanced signal processing and image-based techniques.
	- · Abnormality labels automated arrhythmia interpretation and statistical classification.
- iii. AccurECG Analysis System displays the resulting analysis and original ECG signal with reviewing tools in the web interface.
- iv. AccurECG Analysis System allows for preliminary comments summarizing the review before generating a report.

## 5.8 Intended Use/Indications for Use

The AccurECG Analysis System is intended for use by qualified healthcare professionals in the assessment of a patient's recorded ambulatory ECG data. Analysis results are provided in a standard report for review and printing.

The System provides single-lead analysis on a beat-by-beat basis. Ventricular Ectopic Beat and Supraventricular Ectopic Beat detection, heart rate measurement, and rhythm analysis. The AccurECG Analysis System is compatible with ECG recordings taken with silver/silver chloride (Ag / AgCl) electrodes (wet leads).

The AccurECG Analysis System is not for use in life-supporting or sustaining systems or ECG monitor and alarm devices. The System is not intended to be used with cardiac telemetry monitors.

The AccurECG Analysis System's automated interpretation results are not intended to be the sole means of diagnosis. The AccurECG Report is an adjunct intended to facilitate health care decision making in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms, and other diagnostic information to arrive at a diagnostic conclusion.

The AccurECG Analysis System does not interface with data management systems or hardware and is device independent.

## 5.9 Comparison of Technological Characteristics with Predicate

The subject device has the same intended use and substantially equivalent technological characteristics that do not raise questions of safety or effectiveness. Table - 5.2 Feature Comparison: AccurECG System to the predicate device illustrates a comparison of detection features between the AccurECG system and the cited predicate.

{5}------------------------------------------------

| Feature                                                   | AccurECG | CardioLogs ECG Analysis Platform |
|-----------------------------------------------------------|----------|----------------------------------|
| Heart rate determination for non-paced<br>adult           | Yes      | Yes                              |
| QRS Detection                                             | Yes      | Yes                              |
| Non-paced arrhythmia interpretation for<br>adult patients | Yes      | Yes                              |
| Non-paced ventricular arrhythmia calls                    | Yes      | Yes                              |
| Intervals measurements                                    | Yes      | Yes                              |
| Ventricular ectopic beat detection                        | Yes      | Yes                              |
| Patient Population                                        | Adult    | Adult & Pediatric                |

#### Table - 5.2 Feature Comparison: AccurECG to the predicate device

Table - 5.3 Comparison of technological characteristics: AccurECG to the predicate device provides a comparison of the technological characteristics of the AccurECG System with the cited predicate.

### Table - 5.3 Comparison of technological characteristics: AccurECG System to the predicate device

|                              | AccurECG<br>(this submission)      | CardioLogs ECG Analysis Platform<br>(K170568) | Comparison to<br>predicate device |
|------------------------------|------------------------------------|-----------------------------------------------|-----------------------------------|
| Device Trade<br>Name         | AccurECG® Analysis System          | CardioLogs ECG Analysis Platform              | N/A                               |
| Manufacturer                 | AccurKardia, Inc.                  | CardioLogs Technologies                       | N/A                               |
| 510(K) No.                   | (this submission)                  | K212112                                       | N/A                               |
| Regulation<br>Number         | 21 CFR 870.2340<br>21 CFR 870.1425 | 21 CFR 870.2340<br>21 CFR 870.1425            | Same                              |
| Product Code                 | DQK, DPS                           | DQK, DPS                                      | Same                              |
| Class                        | 2                                  | 2                                             | Same                              |
| Device<br>Class/Name         | Electrocardiograph                 | Electrocardiograph                            | Same                              |
| Software Level of<br>Concern | Major                              | Moderate                                      | Same                              |

{6}------------------------------------------------

| Indications for<br>Use                  | The AccurECG Analysis System is                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | The Cardiologs Platform (Also known<br>as Cardiologs ECG Analysis Platform)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | The predicate device is                                                                                                                                                                                                                                                                                                                           |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                         | intended for use by qualified healthcare<br>professionals<br>("HCP") in the assessment of a patient's<br>recorded ambulatory ECG data.<br>Analysis results are provided in a<br>standard report for review and printing.<br><br>The System provides single-lead analysis<br>on a beat-by-beat basis, Ventricular<br>Ectopic Beat and Supraventricular<br>Ectopic Beat detection, heart rate<br>measurement, and rhythm analysis. The<br>AccurECG Analysis System is<br>compatible with ECG recordings taken<br>with silver/silver chloride (Ag / AgCl)<br>electrodes (wet leads).<br><br>The AccurECG Analysis System is not<br>for use in life-supporting or sustaining<br>systems or ECG monitor and alarm<br>devices. The System is not intended to<br>be<br>used with cardiac telemetry monitors.<br>The AccurECG Analysis System's<br>automated interpretation results are not<br>intended to be the sole means of<br>diagnosis. The AccurECG Report is an<br>adjunct intended to facilitate health care<br>decision making in conjunction with the<br>physician's knowledge of ECG patterns,<br>patient background, clinical history,<br>symptoms, and other diagnostic<br>information to arrive at a diagnostic<br>conclusion.<br><br>The AccurECG Analysis System does<br>not interface with data management<br>systems or hardware and is device<br>independent. | is intended for use by qualified<br>healthcare professionals for the<br>assessment of arrhythmias using ECG<br>data in the adult and pediatric<br>population. The product supports<br>downloading and analyzing data<br>recorded in compatible formats from<br>any device used for the arrhythmia<br>diagnostics such as Holter, event<br>recorder, 12 lead ambulatory ECG<br>devices, or other similar devices when<br>assessment of the rhythm is necessary.<br>The Cardiologs Platform can also be<br>electronically interfaced and perform<br>analysis with data transferred from<br>other computer-based ECG systems,<br>such as an ECG management system.<br>The Cardiologs Platform provides ECG<br>signal processing and analysis, QRS<br>and Ventricular Ectopic Beat<br>detection, QRS feature extraction,<br>interval measurement, heart rate<br>measurement and rhythm<br>analysis.<br><br>The Cardiologs Platform is not for use<br>in life supporting or sustaining systems<br>or ECG monitor and Alarm devices.<br><br>The product can be integrated into<br>computerized ECG monitoring devices.<br>In this case, the medical device<br>manufacturer will identify the<br>indication for use depending on the<br>application of their device.<br>The product can be integrated into<br>computerized ECG monitoring devices.<br>In this case, the medical device<br>manufacturer will identify the<br>indication for use depending on the<br>application of their device.<br><br>Cardiologs Platform interpretation<br>results are not<br>intended to be the sole means of<br>diagnosis. It is offered to physicians<br>and clinicians on an advisory basis only<br>in conjunction with the physician's<br>knowledge of ECG patterns, patient<br>background, clinical history, | indicated for pediatric as<br>well as adult patients.                                                                                                                                                                                                                                                                                             |
| Fundamental<br>scientific<br>technology | The AccurECG Analysis System<br>consists of:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | The Cardiologs Holter Platform<br>consists of:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | The AccurECG Analysis<br>System does not present<br>any major technological<br>differences compared to<br>the<br>predicate device.                                                                                                                                                                                                                |
|                                         | An interface which provides tools to<br>upload, analyze, review ECG data coded<br>in Java.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | An interface which provides tools to<br>measure, analyze and review numerous<br>ECGs coded in java language under the<br>Angular and D3.js frameworks;                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Substantially equivalent:<br>Both<br>devices provide a web<br>interface which provides<br>tools to upload, analyze,<br>review ECG data, but are<br>coded in a different<br>programming language,<br>which should not have<br>any effect on intended us<br>or performance and does<br>not introduce new issues<br>of safety or effectiveness       |
|                                         | An automated proprietary ECG<br>interpretation support<br>algorithm which measures and analyses<br>ECGs to provide<br>supportive information for ECG<br>diagnosis, written in LabVIEW using<br>the G Programming language.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | An automated proprietary ECG<br>interpretation support<br>algorithm which measures and analyzes<br>ECGs to provide supportive<br>information for ECG diagnosis, written<br>in Python language.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Both are accessed throug<br>an<br>internet connection and a<br>web browser, but the<br>predicate device can also<br>be accessed through a<br>direct connection to the<br>API. This connection<br>difference should not<br>have any effect on<br>intended use or<br>performance and does no<br>introduce new issues of<br>safety or effectiveness. |
|                                         | This application is accessed through an<br>Internet connection and a web browser<br>connected to the AccurECG Application<br>Programming Interface (API).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | This application can be accessed<br>through an Internet<br>connection and a web browser, or is<br>directly connected to the Cardiologs'<br>Application Programming Interface<br>(API).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                   |

{7}------------------------------------------------

### 5.10 Summary of Performance Data

Performance tests have been conducted in conformance with the following FDA-recognized consensus standards.

- · AAMI/ANSI EC 57:2012 Testing and Reporting Performance Results of Cardiac Rhythm and ST Segment Measurement Algorithms
- · IEC 60601-2-47:2012 Medical Electrical Equipment, Part 2-47: Particular Requirements For The Basic Safety And Essential Performance Of Ambulatory Electrocardiographic Systems (Section 201.12.1.101 -Algorithm Testing).

### 5.11 Software verification and validation

Testing described in this premarket notification consisted of verification of all design input requirements and product specifications. All clinical input requirements were validated against a gold standard. There are 2 residual anomalies appeared during verification and software validation tests. The Anomalies can be found in the APPX 44 PRJ-21-059 v05 Software Anomalies Listing, AccurECG.

Usability of the AccurECG Analysis System was tested throughout development, validating the effectiveness of risk control measures associated with the user interface. AccurKardia conducted usability tests of the AccurECG user interface with Certified Cardiovascular Technicians to evaluate the effectiveness of the design, labeling and risk. The protocols cover the testable technical requirements relevant to the user interface as described in the User Interface Specification.

### 5.12 Conclusion

Based upon the Intended Use, Indications for Use, product technical information, performance evaluation, and standards compliance provided in this premarket notification, the AccurECG® Analysis System has been shown to be substantially equivalent to the cited predicate.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DPS/K223013](https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DPS/K223013)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
