← Product Code [DPS](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DPS) · K043380

# MONEBO AUTOMATIC ARRHYTHMIA DETECTION SOFTWARE LIBRARY, VERSION 1 (K043380)

_Monebo Technologies, Inc. · DPS · Apr 26, 2005 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DPS/K043380

## Device Facts

- **Applicant:** Monebo Technologies, Inc.
- **Product Code:** [DPS](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DPS.md)
- **Decision Date:** Apr 26, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2340
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Software as a Medical Device

## Indications for Use

The Automatic Arrhythmia Detection Software Library is intended for use by qualified medical professionals for assessing historic ambulatory ECG data. The product allows downloading and analyzing data recorded in compatible format by Holter or Event monitoring devices. The Automatic Arrhythmia Detection Software Library provides ECG signal processing and analysis on a beat by beat basis, QRS and Ventricular Ectopic Beat detection, and QRS feature extraction. The product can be integrated into computerized ECG monitoring devices. In this case the medical device manufacturer will identify the indication for use depending on the application of their device.

## Device Story

Monebo Automatic Arrhythmia Detection Software Library is an object library (callable functions) integrated by manufacturers into ECG analysis devices. Input: up to 12 leads of captured ECG data. Processing: signal processing, QRS detection, QRS feature extraction, and ventricular ectopic beat detection on a beat-by-beat basis. Output: annotated ECG data for clinical review. Used by medical professionals to assess historic ambulatory ECG data from Holter or Event monitors. Assists clinicians in identifying arrhythmias, supporting diagnostic decision-making. Benefits patients through automated, consistent analysis of cardiac rhythm data.

## Clinical Evidence

Bench testing only. Performance validated using AHA, MIT-BIH, and NST databases. Results reported for QRS sensitivity (Se) and positive predictivity (+P), and Ventricular Ectopic Beat (VEB) Se and +P. AHA: QRS Se 99.56%, VEB Se 82.49%. MIT-BIH: QRS Se 99.45%, VEB Se 87.03%. NST: QRS Se 91.56%, VEB Se 81.79%.

## Technological Characteristics

Software-only object library. Built on Microsoft .Net framework. Provides ECG signal processing, QRS detection, feature extraction, and ventricular ectopic beat detection. Designed for integration into third-party computerized ECG monitoring devices.

## Regulatory Identification

An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.

## Predicate Devices

- Agilent ([K003621](/device/K003621.md))
- Brentwood ST/AR ([K013717](/device/K013717.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) Summary of Safety and Effectiveness

510(k) Notification K043380 Date: April 21, 2005 Submitter: Monebo Technologies, Inc. 1800 Barton Creek Blvd Austin, Texas 78735-1606

Contact Person: Dale J. Misczynski Phone: 512-732-0235 512-732-0285 Fax: Email: dale.misczynski@monebo.com

Trade/Device Name: Monebo Automatic Arrhythmia Detection Software Library, Version 1 Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II (two) Product Code: DPS

### Predicate Devices:

The Monebo Automatic Arrhythmia Detection Software Library, Agilent and Brentwood ST/AR predicate are software only devices that monitor cardiac function. Table 1 compares the features of the Automatic Arrhythmia Detection Software Library to predicate devices.

| Feature                                                         | Automatic<br>Arrhythmia<br>Detection<br>Software<br>Library | Agilent<br>510(k)<br>K003621      | Brentwood<br>510(k)<br>K013717 |
|-----------------------------------------------------------------|-------------------------------------------------------------|-----------------------------------|--------------------------------|
| Heart rate determination for<br>non-paced adult                 | NO                                                          | YES                               | YES                            |
| Non-paced ventricular<br>arrhythmia calls for adult<br>patients | YES                                                         | YES                               | YES                            |
| Ventricular ectopic beat<br>detection                           | YES                                                         | YES                               | YES                            |
| Patient Populations                                             | Adult                                                       | Adults,<br>Pediatric,<br>Neonatal | Adult                          |

{1}------------------------------------------------

2/2

## Test Results

The bench test results of the software using AHA and MIT-BIH databases are show in the table below:

| Database | QRS Se | QRS +P | VEB Se | VEB +P |
|----------|--------|--------|--------|--------|
| AHA      | 99.56  | 99.9   | 82.49  | 95.65  |
| MIT-BIH  | 99.45  | 99.45  | 87.03  | 87.76  |
| NST      | 91.56  | 85.66  | 81.79  | 53.19  |

Summary results of AHA and MIT testing

### Device Description:

The Monebo Automatic Arrhythmia Detection Software Library is an "object library". An object library is a collection of callable functions that have been complied (or assembled) into machine code or IDL code of the computer on which they execute. The Automatic Arrhythmia Detection Software Library consists of a basic application for viewing, analyzing and annotating ECG data and callable object library built on the Microsoft™ .Net framework. An application software program can be written to invoke some or all of the functions in an object library.

The Monebo Automatic Arrhythmia Detection Software Library provides ECG signal processing, QRS detection, QRS feature extraction, and ventricular ectopic beat detection for up to 12 leads of captured ECG data.

Monebo will compile the Automatic Arrhythmia Detection Software Library specified by an ECG analysis device manufacturer. An object library will be created and delivered to the device manufacturer, who can then integrate it into application software for their ECG analysis. The device manufacturer will be required to submit a 510k with the specific requirements for that device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing, positioned to the right of a circular arrangement of text. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" and is arranged around the top and left side of the circle. The eagle is black, and the text is also in black.

APR 2 6 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Monebo Technologies, Inc. c/o Mr. Dale J. Misczynski President 1800 Barton Creek Blvd. Austin, TX 78735

Re: K043380

> Trade Name: Monebo Automatic Arrhythmia Detection Software Library, Version 1 Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II (two) Product Code: DPS Dated: March 18, 2005 Received: March 21, 2005

Dear Mr. Misczynski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of l'ederal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Dale J. Misczynski

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Bhimmarfor

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): K043380

Device Name: Monebo Automatic Arrhythmia Detection Software Library

Indications For Use:

The Automatic Arrhythmia Detection Software Library is intended for use by qualified medical professionals for assessing historic ambulatory ECG data. The product allows downloading and analyzing data recorded in compatible format by Holter or Event monitoring devices. The Automatic Arrhythmia Detection Software Library provides ECG signal processing and analysis on a beat by beat basis, QRS and Ventricular Ectopic Beat detection, and QRS feature extraction.

The product can be integrated into computerized ECG monitoring devices. In this case the medical device manufacturer will identify the indication for use depending on the application of their device.

Prescription Use AND/OR Over-The-Counter Use -----------

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) .

Blummana

on Sign-Off Division of Cardiova 510(k) Number

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DPS/K043380](https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DPS/K043380)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
