Shockwave Intravascular Lithotripsy (IVL) System with the Javelin Peripheral Intravascular Lithotripsy (IVL) Catheter
Applicant
Shockwave Medical, Inc.
Product Code
PPN · Cardiovascular
Decision Date
Sep 27, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Shockwave Medical IVL System with the Javelin Peripheral IVL Catheter is intended for lithotripsy-enabled modification and crossing of calcified lesions in the peripheral vasculature, including the iliac, femoral, popliteal, and infra-popliteal arteries, prior to final treatment. Not for use in the coronary, carotid, cerebral, or pulmonary vasculature.
Device Story
System consists of IVL catheter, generator, and connector cable. Catheter features integrated emitter near distal end, fluid-filled IVL window, and guidewire lumen. Used in peripheral arterial system to treat calcified stenosis. Generator delivers electrical energy to emitter, creating acoustic pressure pulses within saline-filled window; pulses disrupt calcium in lesion, facilitating vessel dilation. Operated by physicians in clinical settings under fluoroscopic guidance. Output is mechanical disruption of plaque; enables subsequent angioplasty or stenting. Benefits include improved luminal gain and lesion crossing in heavily calcified peripheral arteries.
Clinical Evidence
Pooled analysis of 90 subjects from FORWARD PAD IDE and New Zealand/Australia Mini-S Feasibility studies. Prospective, multi-center, single-arm. Primary safety endpoint: 30-day MAE rate 1.1% (upper 95% CI 6.0%), meeting 11.2% PG (p=0.0012). Primary effectiveness endpoint: Technical Success (residual stenosis ≤50% without flow-limiting dissection) 99.0% (lower 95% CI 94.4%), meeting 85.0% PG (p<0.0001). Secondary endpoints included angiographic complications (1.0%) and IVL Device Success (93.0%).
Technological Characteristics
Catheter with integrated emitter, non-expandable polymer IVL window, and saline-filled lumen. Energy source: electrical-to-acoustic pressure pulses. Features: hydrophilic coating, radiopaque emitter. Connectivity: IVL generator and connector cable exchange data during treatment. Single-use. Sterilization: not specified.
Indications for Use
Indicated for lithotripsy-enabled modification and crossing of calcified lesions in peripheral vasculature (iliac, femoral, popliteal, infra-popliteal arteries) in patients with peripheral artery disease (PAD). Contraindicated for use in coronary, carotid, cerebral, or pulmonary vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Shockwave Medical Intravascular Lithotripsy System (K191840)
Related Devices
K240954 — Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave L6 Peripheral IVL Catheter · Shockwave Medical, Inc. · May 7, 2024
K221852 — Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave L6 Peripheral IVL Catheter · Shockwave Medical, Inc. · Aug 16, 2022
K203365 — Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave M5+ Peripheral IVL Catheter · Shockwave Medical, Inc. · Apr 22, 2021
K240225 — Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave E8 Peripheral IVL Catheter · Shockwave Medical, Inc. · Mar 22, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health and Human Services logo is a stylized depiction of a human figure, while the FDA acronym and full name are written in blue.
Sepember 27, 2024
Shockwave Medical, Inc. Shira Stone Senior Regulatory Affairs Specialist 5403 Betsy Ross Drive Santa Clara, California 95054
## Re: K242213
Trade/Device Name: Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave Javelin Peripheral Intravascular Lithotripsy (IVL) Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: PPN Dated: July 26, 2024 Received: July 29, 2024
Dear Shira Stone:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Ariel G. Ashshakoor -S
Digitally signed by Ariel G. Ash-shakoor -S Date: 2024.09.27 11:42:11 -04'00'
# For
Greg O'Connell Assistant Director DHT2C: Division of Intervetional Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242213
Device Name
Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave Javelin Peripheral Intravascular Lithotripsy (IVL) Catheter
Indications for Use (Describe)
The Shockwave Medical IVL System with the Javelin Peripheral IVL Catheter is intended for lithotripsy-enabled modification and crossing of calcified lesions in the peripheral vasculature, including the iliac, femoral, popliteal, and infra-popliteal arteries, prior to final treatment.
Not for use in the coronary, carotid, cerebral, or pulmonary vasculature.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Shockwave Medical. The word "SHOCKWAVE" is in large, bold, gray letters, with a design of concentric circles and a cracked globe above the "O". Below "SHOCKWAVE" is the word "MEDICAL" in smaller, light gray letters.
Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave Javelin Peripheral IVL Catheter K242213 510(k) Summary Page 1 of 5
#### K242213 510(k) Summary
This 510(k) Summary is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92.
#### Name, Address, and Phone Number of Applicant
Shockwave Medical, Inc. 5403 Betsy Ross Drive Santa Clara, CA 95054 Phone: (510) 279-4262
#### Contact Person
Shira Stone
### Date Prepared
July 26, 2024
| Trade Name: | Shockwave Intravascular Lithotripsy (IVL) System with the<br>Shockwave Javelin Peripheral Intravascular Lithotripsy (IVL)<br>Catheter |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Catheter, lithotripsy, peripheral, transluminal |
| CFR Classification: | 21 CFR 870.1250 |
| Classification<br>Name: | Percutaneous catheter |
| Product Code: | PPN |
#### Device Name and Classification
#### Predicate Device
The predicate device is the Shockwave Medical Intravascular Lithotripsy System, K191840, cleared by FDA on August 7, 2019.
#### Indications for Use / Intended Use
The Shockwave Medical IVL System with the Javelin Peripheral IVL Catheter is intended for lithotripsy-enabled modification and crossing of calcified lesions in the peripheral vasculature, including the iliac, femoral, ilio-femoral, popliteal, and infra-popliteal arteries, prior to final treatment.
Not for use in the coronary, carotid, cerebral, or pulmonary vasculature.
#### Device Description
The IVL Catheter is a proprietary lithotripsy device delivered through the peripheral arterial system of the lower extremities to the site of an otherwise difficult to treat calcified stenosis. Intravascular Lithotripsy (IVL) is an interventional procedure that utilizes a fluid-filled catheter connected to power sources that generate acoustic shock waves modify calcified
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Shockwave Medical. The word "SHOCKWAVE" is in large, bold, gray letters, with a design of concentric circles and a web-like pattern incorporated into the letters. Below "SHOCKWAVE" is the word "MEDICAL" in smaller, lighter gray letters.
Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave Javelin Peripheral IVL Catheter K242213 510(k) Summary Page 2 of 5
plaque in peripheral arteries. Energizing the intravascular lithotripsy device will generate acoustic pressure pulses within the target treatment site, disrupting calcium within the lesion and allowing dilation of peripheral artery stenosis.
The Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave Javelin Peripheral Intravascular Lithotripsy (IVL) Catheter consists of three main components: the IVL Catheter, the IVL Generator, and the IVL Connector Cable. The Shockwave Javelin IVL Catheter is comprised of a catheter with an integrated emitter located near the distal end to enable the localized delivery of acoustic pressure pulses in the peripheral vasculature. The emitter is racilitate catheter visibility under fluoroscopy and is surrounded by a fluid-filled space (IVL window) that allows for the transmission of acoustic pressure pulses.
The Shockwave Javelin Peripheral IVL Catheter shaft contains a lumen to pressurize and a lumen to de-pressurize the catheter with saline (the medium to create IVL), a guidewire lumen, and a lithotripsy emitter. The emitter is enclosed within a non-expandable polymer fluid-filled space (i.e., IVL window) containing saline that is connected to the proximal shaft, inlet and outlet ports, and is tapered down to the distal tip of the catheter. The IVL window is located near the distal tip of the catheter. The emitter is radiopaque to facilitate catheter visibility under fluoroscopy and is surrounded by the IVL window that allows for the transmission of acoustic pressure pulses. The IVL window is designed to provide a static catheter profile.
The IVL Generator and Connector Cable are used with the Shockwave Javelin Peripheral IVL Catheter to deliver localized, lithotripsy-enabled modification and crossing of calcified, stenotic arteries. The IVL Generator, IVL Connector Cable and IVL Catheters are designed to exchange data during patient treatment.
#### Technological Comparison
This Traditional 510(k) Premarket Notification describes dimensional modifications to the predicate Shockwave S4 Peripheral IVL Catheter including fewer emitters, increased catheter length, distally located emitter, an IVL window in lieu of a balloon, and extended hydrophilic coating length.
The catheter labeling was updated to reference the name of the modified device, Shockwave Javelin Peripheral IVL Catheter, and include clinical study data. The device labeling includes a sterile cable sleeve in the packaging to form a convenience kit.
The IVL System has the same intended use, general purpose, similar principles of operation, and has substantially equivalent technological characteristics including the same fundamental scientific technology, design, energy source, shelf life, and sterilization as the 510(k) cleared Predicate IVL System.
The indication for use statement is similar to the predicate, however minor modifications have been made to further clarify the intended benefit of this specific device which includes modification and crossing of calcified lesions in the peripheral arterial vasculature. The indication
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Shockwave Medical. The logo is in two colors, with the word "SHOCKWAVE" in a dark gray color and the word "MEDICAL" in a light gray color. The "O" in "SHOCKWAVE" is replaced with a blue circle with concentric rings emanating from the center.
for use reflects both the steps in the instructions for use (IFU) and the recent IDE clinical study results.
### Summary of Performance Data
Objective evidence demonstrating that the IVL System design output meets the product design input requirements as well as that device performance characteristics conform to user needs and intended uses as defined in the product specification was provided. Testing was conducted in accordance with Shockwave Medical's Risk Analysis procedures, applicable FDA guidance documents and relevant international standards.
- IVL Catheter design verification and validation testing: 0
- Guidewire compatibility O
- Introducer sheath compatibility о
- Hydrophilic coating length O
- O Hydrophilic coating uniformity
- Useable catheter length O
- Catheter crossing profile O
- Catheter lubricity O
- O Catheter flexibility
- Catheter kink resistance O
- O Catheter distal tip
- Distal tip durability O
- Catheter bonds tensile strength O
- Catheter torsional strength O
- Emitter band integrity O
- Emitter band alignment O
- Emitter radiopacity O
- Single use design O
- Minimum system burst O
- Catheter fatigue (multiple pressure cycles) O
- Sonic output O
- Catheter pulse count and pulse rate O
- Temperature rise O
- Catheter particle count O
- Catheter connector length O
- Catheter connection O
- Catheter identification O
- Catheter sterility (visual inspection) O
- Cable sleeve packaging O
- Catheter compatibility with materials and accessories commonly used in Over-the-O Wire (OTW) peripheral balloon angioplasty and IVL procedures
- Simulated use testing o
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Shockwave Medical. The logo is in gray and features the word "SHOCKWAVE" in large, bold letters. To the left of the word "SHOCKWAVE" are three concentric circles. Below the word "SHOCKWAVE" is the word "MEDICAL" in smaller letters.
Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave Javelin Peripheral IVL Catheter K242213 510(k) Summary Page 4 of 5
Results demonstrated that the performance of the IVL System meets its design specifications and demonstrates substantial equivalence for its intended use.
#### Summary of Clinical Data
The 510(k) submission included the 30-day pooled clinical study data of the FORWARD PAD IDE Study (FORWARD Study) and New Zealand/Australia Mini-S Feasibility Study (Feasibility Study). Data for the first 90 consecutively enrolled subjects across both studies comprised the pooled analysis cohort. The FORWARD and Feasibility studies were prospective, multi-center, single-arm, studies evaluating the treatment of heavily calcified, stenotic peripheral arteries with the Shockwave Javelin Peripheral Intravascular Lithotripsy (IVL). The subjects in the pooled analysis were enrolled at 19 clinical sites: 15 sites located in the United States, and 4 sites in Australia & New Zealand. Subjects with moderately calcified peripheral artery disease (PAD), Rutherford Category (RC) of 2, 3, 4, or 5 of the target limb, a target lesion located in a native, de novo superficial femoral, popliteal or infrapopliteal artery that met all other study criteria were enrolled and treated. At baseline, the mean age was 74.5 years, and the majority of subjects had a target limb(s) categorized as Rutherford Category 3 (43.3%) or 5 (41.1%). Pre-procedure lesion characteristics, as determined by the Core Lab, showed a mean reference vessel diameter (RVD) of 4.2mm (1.1, 7.4), mean minimum lumen diameter (MLD) of 0.7 mm (0.0, 2.4) with a corresponding mean percentage diameter stenosis of 82.9% (49.9, 100), and mean lesion length of 76.9 mm (8.9, 335.2). Procedural imaging characteristics were collected at pre-procedure, post-Javelin, post-dilatation, and final (end of procedure) timepoints.
The primary safety endpoint for the FORWARD and Feasibility studies was major adverse events (MAEs) at 30 days post-index procedure, defined as a composite of: cardiovascular death; clinically-driven target lesion revascularization (CD-TLR); and unplanned target limb major amputation (above the ankle). All MAEs were adjudicated by an independent Clinical Events Committee. The primary safety endpoint performance goal (PG) for 30-Day MAE rate was 11.2%. The observed 30-day MAE rate was 1.1% (1/90), with an upper 95% confidence limit of 6.0%, which was lower than the Performance Goal of 11.2% (p=0.0012). The null hypothesis was rejected, and the 30-Day MAE PG was met.
The primary effectiveness endpoint for the FORWARD PAD and Feasibilities studies was Technical Success, defined as final residual stenosis ≤50% without flow-limiting dissection (≥ Grade D) of the target lesion as assessed by the angiographic core laboratory. The PG for the primary effectiveness endpoint of Technical Success was 85.0%. Technical Success was achieved in 99.0% (97/98) of target lesions with a lower 95% confidence limit of 94.4% which was higher than the PG of 85.0% (p<0.0001). The null hypothesis was rejected, and the Technical Success PG was met.
The secondary safety endpoint was serious angiographic complications at the final (end of procedure) timepoint, defined as flow-limiting dissection (≥ Grade D), perforation, distal embolization, or acute vessel closure as assessed by angiographic core lab, which occurred in 1.0% (1/98) of lesions.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Shockwave Medical. The logo is in gray and features the word "SHOCKWAVE" in large, bold letters. There are blue circles and a web-like design behind the letters. Below "SHOCKWAVE" is the word "MEDICAL" in smaller letters.
Additional secondary endpoints included IVL Technical Success (post-dilatation and final) and IVL Device Success.
- o IVL Technical Success, defined as post-dilatation residual stenosis <50% without flowlimiting dissection (≥ Grade D) of the target lesion by angiographic core lab (measured immediately following mandatory post-dilatation), was achieved in 89.7% (87/97) of lesions.
- o IVL Device Success, defined as the ability to deliver, advance across the target lesion. pressurize, pulse, flush, and retrieve the Javelin IVL catheter, was achieved with 93.0% (107/115) of catheters.
- Technical Success (final), defined as final residual stenosis of ≤30% without flow-limiting dissection (> Grade D) of the target lesion by angiographic core lab, was achieved in 78.6% (77/98) of lesions.
Post-Javelin treatment, drug-coated balloons were used in 40.0% (42/105) of lesions and 22.9% (24/105) of the lesions had a stent/Tack implanted. Commercial IVL was used on 25.7% (27/105) of target lesions.
The study collected residual stenosis data post-Javelin, post-dilatation and at the final angiographic time point. Post-Javelin mean residual stenosis was 59.1 ± 18.4% with 36.5% (31/85) of the lesions having a residual stenosis of ≤ 50% and 3.5% (3/85) with a residual stenosis of ≤ 30%. Postdilatation mean residual stenosis was 31.3 ± 13.7% with 93.8% (91/97) of lesions having residual stenosis of ≤ 50%, and 50.5% (49/97) with a residual stenosis of ≤ 30%. Final mean residual stenosis was 23.0 ± 9.1% with 100% (98/98) of lesions reported with a residual stenosis of < 50%, and with 79.6% (78/98) having a residual stenosis of ≤30%.
The Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave Javelin Peripheral Intravascular Lithotripsy (IVL) Catheter demonstrated a low incidence of MAEs and angiographic complications, consistent with prior peripheral IVL studies. Effectiveness results showed acute luminal gain post-Javelin, and low residual stenosis at the final angiographic timepoint.
The clinical outcomes from the analysis demonstrate the substantial equivalence of the Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave Javelin Peripheral IVL Catheter.
## Basis for Substantial Equivalence
The IVL Catheter with dimensional modifications shares the same intended use, similar principles of operation, overall technical and functional capabilities, and similar design and materials as the identified predicate device. Any differences between the IVL Systems were evaluated through design verification and validation testing which demonstrated device performance and confirmed that there are no new questions of safety or effectiveness. The IVL Catheter is therefore substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.