Reprocessed Dynamic Tip and XT Steerable Diagnostic Electrophysiology Catheters
Applicant
Innovative Health, LLC
Product Code
NLH · Cardiovascular
Decision Date
Sep 15, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1220
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Reprocessed Dynamic Tip and Dynamic XT Steerable Diagnostic EP Catheters are intended for temporary intracardiac sensing, recording, stimulation and temporary pacing during the evaluation of cardiac arrhythmias.
Device Story
Radiopaque, flexible, insulated polymer catheters; plunger mechanism enables distal tip curvature for navigation. Used in clinical electrophysiology settings by physicians to access cardiac chambers. Inputs: intracardiac electrical signals; outputs: electrical signals for recording/monitoring or pacing pulses. Device facilitates arrhythmia diagnosis and mapping. Reprocessed devices undergo cleaning, decontamination, inspection, and functional testing to ensure performance equivalent to original devices. Limited to three reprocessing cycles; serialized and tracked.
Clinical Evidence
Bench testing only. Testing included biocompatibility, cleaning and sterilization validation, visual inspection, dimensional verification, electrical continuity and resistance, simulated use, mechanical characteristics, electrical safety (dielectric and current leakage), and shelf-life validation.
Technological Characteristics
Radiopaque, flexible, insulated polymer shaft; plunger-actuated steerable distal tip. Multi-electrode configuration (4-10 electrodes). Dimensions: 6 French size, 110 cm usable length. Reprocessing includes cleaning, decontamination, and functional testing. No changes to materials or design from predicate.
Indications for Use
Indicated for patients undergoing evaluation of cardiac arrhythmias requiring temporary intracardiac sensing, recording, stimulation, and pacing.
Regulatory Classification
Identification
An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.
Predicate Devices
Reprocessed Dynamic Tip Steerable Diagnostic Electrophysiology Catheter (K161464)
K161393 — Reprocessed Orbiter ST Steerable Diagnostic EP Catheter · Innovative Health, LLC · Oct 26, 2016
K252405 — Medline ReNewal Reprocessed Boston Scientific Dynamic Tip Unidirectional Steerable Diagnostic Catheter · Surgical Instrument Service and Savings, Inc. · Nov 26, 2025
K203261 — Reprocessed Inquiry Steerable Diagnostic EP Catheter · Innovative Health, LLC · Mar 5, 2021
K160496 — Reprocessed Inquiry Steerable Diagnostic EP Catheter · Innovative Health, LLC · Jun 10, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 15, 2021
Innovative Health, LLC. Amanda Babcock Regulatory Affairs Manager 1435 North Hayden Road, Suite 100 Scottsdale, Arizona 85257
Re: K211276
Trade/Device Name: Reprocessed Dynamic Tip and XT Steerable Diagnostic Electrophysiology Catheters Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter Or Electrode Recording Probe Regulatory Class: Class II Product Code: NLH Dated: August 13, 2021 Received: August 16, 2021
Dear Amanda Babcock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Aneesh Deoras Assistant Director (Acting) Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| Description | Item<br>Number | Number of<br>Electrodes | French<br>Size | Electrode<br>Spacing<br>(mm) | Distal Reach | Usable Length<br>(cm) |
|----------------------------------------------------------------------|----------------|-------------------------|----------------|------------------------------|--------------|-----------------------|
| Reprocessed<br>Dynamic Tip<br>Steerable<br>Diagnostic EP<br>Catheter | 200131 | 4 | 6 | 10 | Large 4.0 | 110 |
| | 200344 | 4 | 6 | 5 | Large 4.0 | 110 |
| | 6DYNTP002 | 4 | 6 | 2,5,2 | Large 4.0 | 110 |
| | 6DYNTP006 | 8 | 6 | 2 | Large 4.0 | 110 |
| | 6DYNTP001 | 10 | 6 | 2,5,2 | Large 4.0 | 110 |
| Reprocessed<br>Dynamic XT<br>Steerable<br>Diagnostic EP<br>Catheters | 201103 | 4 | 6 | 2,5,2 | Large 4.0 | 110 |
| | 201104 | 4 | 6 | 5 | Large 4.0 | 110 |
| | 201110 | 4 | 6 | 10 | Large 4.0 | 110 |
| | 201112 | 4 | 6 | 2 | Large 4.0 | 110 |
| | 201115 | 4 | 6 | 1 | Large 4.0 | 110 |
| | 201109 | 6 | 6 | 10 | Large 4.0 | 110 |
| | 201113 | 6 | 6 | 5 | Large 4.0 | 110 |
| | 201114 | 6 | 6 | 1 | Large 4.0 | 110 |
| | 201105 | 8 | 6 | 2,5,2 | Large 4.0 | 110 |
| | 201106 | 8 | 6 | 2 | Large 4.0 | 110 |
| | 201107 | 8 | 6 | 5 | Large 4.0 | 110 |
| | 201108 | 8 | 6 | 2,10,2 | Large 4.0 | 110 |
| | 201101 | 10 | 6 | 2,5,2 | Large 4.0 | 110 |
| | 201102 | 10 | 6 | 2,6,2 | Large 4.0 | 110 |
| | 6DYNXT009 | 4 | 6 | 2,5,2 | Large 4.0 | 110 |
| | 6DYNXT011 | 4 | 6 | 5 | Large 4.0 | 110 |
| | 6DYNXT002 | 10 | 6 | 5 | Large 4.0 | 110 |
| | 6DYNXT004 | 10 | 6 | 2 | Large 4.0 | 110 |
The item numbers included in the scope of this submission are as follows:
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211276
#### Device Name
Reprocessed Dynamic Tip and XT Steerable Diagnostic Electrophysiology (EP) Catheters
Indications for Use (Describe)
The Reprocessed Dynamic Tip and Dynamic XT Steerable Diagnostic EP Catheters are intended for temporary intracardiac sensing, recording, stimulation and temporary pacing during the evaluation of cardiac arrhythmias.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# SECTION 5: 510(k) SUMMARY
As required by 21 CFR 807.92(c)
#### Submitter's Name and Address:
Innovative Health, LLC. 1435 N. Hayden Road, Suite 100 Scottsdale, AZ 85257
#### Contact Name and Information:
Amanda Babcock Requlatory Affairs Manager Innovative Health, LLC. (480) 525-5911 (office) (888) 965-7705 (fax) ababcock@innovative-health.com
#### Date prepared:
April 26, 2021
#### Device Information:
Trade/Proprietary Name:
Common Name: Classification Name: Classification Number: Product Code:
Reprocessed Dynamic Tip and XT Steerable Diagnostic Electrophysiology Catheters Diagnostic Electrophysiology Catheter Catheter, Recording, Electrode, Reprocessed Class II, 21 CFR 870.1220 NLH
#### Predicate Device:
| 510(k) Number | Device | Manufacturer |
|---------------|----------------------------------------------------------------------------|-------------------------|
| K161464 | Reprocessed Dynamic Tip Steerable<br>Diagnostic Electrophysiology Catheter | Innovative Health, LLC. |
| K891908 | Bard Tip Deflecting Electrode Catheter | C.R. Bard, Inc. |
#### Device Description:
The Reprocessed Dynamic Tip and Dynamic XT Steerable Diagnostic Electrophysiology (EP) Catheters are radiopaque, flexible, insulated catheters with a polymer shaft. The catheters have a plunger mechanism, which, when moved forward or back, results in curvature of the distal tip.
The item numbers in scope of this submission are as follows:
| Description | Item<br>Number | Number of<br>Electrodes | French<br>Size | Electrode<br>Spacing<br>(mm) | Distal Reach | Usable Length<br>(cm) |
|----------------------------------------------------------------------|----------------|-------------------------|----------------|------------------------------|--------------|-----------------------|
| Reprocessed<br>Dynamic Tip<br>Steerable<br>Diagnostic EP<br>Catheter | 200131 | 4 | 6 | 10 | Large 4.0 | 110 |
| | 200344 | 4 | 6 | 5 | Large 4.0 | 110 |
| | 6DYNTP002 | 4 | 6 | 2,5,2 | Large 4.0 | 110 |
| | 6DYNTP006 | 8 | 6 | 2 | Large 4.0 | 110 |
| | 6DYNTP001 | 10 | 6 | 2,5,2 | Large 4.0 | 110 |
{5}------------------------------------------------
| Reprocessed<br>Dynamic XT<br>Steerable<br>Diagnostic EP<br>Catheters | 201103 | 4 | 6 | 2,5,2 | Large 4.0 | 110 |
|----------------------------------------------------------------------|-----------|----|---|--------|-----------|-----|
| | 201104 | 4 | 6 | 5 | Large 4.0 | 110 |
| | 201110 | 4 | 6 | 10 | Large 4.0 | 110 |
| | 201112 | 4 | 6 | 2 | Large 4.0 | 110 |
| | 201115 | 4 | 6 | 1 | Large 4.0 | 110 |
| | 201109 | 6 | 6 | 10 | Large 4.0 | 110 |
| | 201113 | 6 | 6 | 5 | Large 4.0 | 110 |
| | 201114 | 6 | 6 | 1 | Large 4.0 | 110 |
| | 201105 | 8 | 6 | 2,5,2 | Large 4.0 | 110 |
| | 201106 | 8 | 6 | 2 | Large 4.0 | 110 |
| | 201107 | 8 | 6 | 5 | Large 4.0 | 110 |
| | 201108 | 8 | 6 | 2,10,2 | Large 4.0 | 110 |
| | 201101 | 10 | 6 | 2,5,2 | Large 4.0 | 110 |
| | 201102 | 10 | 6 | 2,6,2 | Large 4.0 | 110 |
| | 6DYNXT009 | 4 | 6 | 2,5,2 | Large 4.0 | 110 |
| | 6DYNXT011 | 4 | 6 | 5 | Large 4.0 | 110 |
| | 6DYNXT002 | 10 | 6 | 5 | Large 4.0 | 110 |
| | 6DYNXT004 | 10 | 6 | 2 | Large 4.0 | 110 |
# Indications for Use:
The Reprocessed Dynamic Tip and Dynamic XT Steerable Diagnostic EP Catheters are intended for temporary intracardiac sensing, recording, stimulation and temporary pacing during the evaluation of cardiac arrhythmias.
#### Technological Characteristics:
The purpose, design, materials, function, and intended use of the Reprocessed Dynamic Tip and XT Steerable Diagnostic Electrophysiology (EP) Catheters are identical to the predicate devices. There are no changes to the claims, clinical applications, patient population, performance specifications, or method of operation. In addition, Innovative Health's reprocessing of these devices includes removal of visible soil and decontamination. Each device is inspected and function tested prior to packaging and labeling.
# Functional and Safety Testing:
Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of the Reprocessed Dynamic Tip and XT Steerable Diagnostic EP Catheters. This included the following:
- Biocompatibilitv ●
- Cleaning Validation
- Sterilization Validation ●
- Functional Testing ●
- Visual Inspection .
- Dimensional Verification ●
- Electrical Continuity and Resistance ●
- . Simulated Use
- Mechanical Characteristics ●
- . Electrical Safety Testing
- . Dielectric and Current Leakage
- Shelf-life Validation
The Reprocessed Dynamic Tip and XT Steerable Diagnostic EP Catheters are reprocessed no more than three (3) times. Each device is marked, serialized and
{6}------------------------------------------------
tracked. After the device has reached the maximum number of reprocessing cycles, the device is rejected from further reprocessing. Reprocessing is performed only by Innovative Health. Innovative Health restricts its reprocessing to exclude devices previously reprocessed by other reprocessors.
#### Conclusion:
Innovative Health concludes that the Reprocessed Dynamic Tip and XT Steerable Diagnostic EP Catheters are as safe and effective as the predicate devices described herein.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.