TRANSONIC HCM 103 SYSTEM

K134035 · Transonic Systems, Inc. · DXG · Sep 17, 2014 · Cardiovascular

Device Facts

Record IDK134035
Device NameTRANSONIC HCM 103 SYSTEM
ApplicantTransonic Systems, Inc.
Product CodeDXG · Cardiovascular
Decision DateSep 17, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1435
Device ClassClass 2
AttributesPediatric

Intended Use

The Transonic HCM103 system is intended for diagnostic assessment of cardiovascular status and circulatory variables in patients aged 1 month and older. The measurements include continuous cardiac output and associated hemodynamic parameters (such as stroke volume variation; pulse pressure variation, systemic vascular resistance and related index parameters) by continuous arterial pressure waveform analysis. In this stand-alone system, cardiac output (CO) value is entered from such devices as COstatus (HCM101) system or equivalent to calibrate the pressure wave form for measurement of continuous cardiac output. In addition the system also provides heart rate, systolic, diastolic pressures, mean arterial pressure and indexed parameters.

Device Story

System performs continuous hemodynamic monitoring via arterial pressure waveform analysis; inputs include arterial pressure signals from standard hospital pressure transducers; requires manual calibration via external cardiac output measurement (e.g., Transonic COstatus HCM101). Device calculates cardiac output, stroke volume variation, pulse pressure variation, systemic vascular resistance, heart rate, and blood pressures. Used in ICU, OR, or clinical environments by healthcare professionals. Output displayed for clinician assessment of patient cardiovascular status; aids in hemodynamic management and clinical decision-making. Benefits include continuous monitoring of circulatory variables to guide patient care.

Clinical Evidence

No clinical or animal data provided. Substantial equivalence supported by bench testing, including software verification and electrical safety testing per IEC 60601-1. Performance confirmed via bench studies validating accuracy of hemodynamic parameter derivation from arterial pressure waveforms.

Technological Characteristics

System comprises monitor, sensor cable, and data transfer module; utilizes standard FDA-cleared pressure transducers. Operates via arterial pressure waveform analysis. Electrical safety per IEC 60601-1. Software-based processing for hemodynamic parameter derivation.

Indications for Use

Indicated for diagnostic assessment of cardiovascular status and circulatory variables in neonatal, pediatric, and adult patients (aged 1 month and older).

Regulatory Classification

Identification

A single-function, preprogrammed diagnostic computer is a hard-wired computer that calculates a specific physiological or blood-flow parameter based on information obtained from one or more electrodes, transducers, or measuring devices.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of three faces in profile, stacked on top of each other. Below the faces are three wavy lines. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 17, 2014 Transonic Systems, Inc. Naveen Thuramalla VP, Engineering and Clinical Studies 34 Dutch Mill Rd Ithaca, New York 14850 Re: K134035 Trade/Device Name: Transonic HCM 103 System Regulation Number: 21 CFR 870.1435 Regulation Name: Single-Function, Preprogrammed Diagnostic Computer Regulatory Class: Class II Product Code: DXG Dated: June 18, 2013 Received: June 19, 2014 Dear Naveen Thuramalla, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR {1}------------------------------------------------ Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely your Image /page/1/Picture/8 description: The image shows a logo with the letters "K" and "I" in a stylized font. The letters are connected, with the "K" on the left and the "I" on the right. Below the logo, there is text that reads "Ken Skodacek for". The background of the image is white. Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for "transonic". The word "transonic" is written in blue, with the words "THE MEASURE OF" written in a smaller font below the left side of the word, and the words "BETTER RESULTS." written in a smaller font below the right side of the word. Above the center of the word "transonic" are three gray curved lines. Transonic Systems Inc. 34 Dutch Mill Rd., Ithaca, NY 14850 p: 800.353.3569 f: 607.257.7256 # Indications for Use 510(k) Number (if known): New Submission Device Name: Transonic HCM103 System Indications for Use: (Prescription Device) The Transonic HCM103 system is intended for diagnostic assessment of cardiovascular status and circulatory variables in patients aged 1 month and older. The measurements include continuous cardiac output and associated hemodynamic parameters (such as stroke volume variation; pulse pressure variation, systemic vascular resistance and related index parameters) by continuous arterial pressure waveform analysis. In this stand-alone system, cardiac output (CO) value is entered from such devices as COstatus (HCM101) system or equivalent to calibrate the pressure wave form for measurement of continuous cardiac output. In addition the system also provides heart rate, systolic, diastolic pressures, mean arterial pressure and indexed parameters. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for Transonic. The word "transonic" is written in blue, with the words "THE MEASURE OF" and "BETTER RESULTS." written in gray below it. In the middle of the logo, there are three gray curved lines that resemble sound waves. Transonic Systems Inc. 34 Dutch Mill Rd., Ithaca, NY 14850 p: 800.353.3569 f: 607.257.7256 ## 510(k) SUMMARY ### Summary of Safety & Effectiveness | Submitter's Name & Address: | Transonic Systems Inc.<br>34 Dutch Mill Road,<br>Ithaca, NY 14850 | |-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person & Telephone: | Naveen Thuramalla<br>607-257-5300 (*326) | | Date Summary Prepared: | Dec 16, 2013 (Revised June 18, 2014) | | Device Name: | Classification Name: Computer, Diagnostic, Pre-<br>programmed, Single-function, 21 CFR 870.1435.<br>Product Class and Code: DXG and Class II<br>Classification Panel: Cardiovascular<br>Common/Usual Name: Extracorporeal, Diagnostic<br>monitor<br>Proprietary Name: Transonic HCM103 System | | Predicate Devices: | K082308: MODIFICATION TO VIGILEO ARTERIAL<br>PRESSURE CARDIAC OUTPUT/OXIMETRY<br>MONITOR; EDWARDS LIFESCIENCES, LLC | | Predicate Devices: | K010049: PULSECO HEMODYNAMIC MONITOR<br>CM71, LiDCO Ltd. [K023960: LIDCOPLUS<br>HEMODYNAMIC MONITOR SYSTEM is also similar]. | #### Device Description: In accordance with section 510(k) for the Federal Food, Drug, and Cosmetic Act, Transonic Systems Inc. intends to introduce into interstate commerce the Transonic HCM103 system, which is an apparatus based on arterial pressure waveform analysis to provide diagnostic assessment of cardiovascular status including continuous cardiac output and associated hemodynamic parameters (such as stroke volume variation; pulse pressure variation). In this stand-alone system, calibration value is entered from {4}------------------------------------------------ K134035 such devices as COstatus (HCM101) system or equivalent. In addition the system also measures heart rate, systolic, and diastolic pressures and derives mean arterial pressure. These patients could be in the intensive care units (ICU), operating room (OR) or other such environments. ### Components: Transonic HCM103 system consists of the following components: | Model/Part # | Description | |--------------|---------------------------------------------------------------------------------| | HCM103 | Monitor | | HCPR | Sensor cable with connector to mate with standard hospital pressure transducer. | | HCS40XX | MX950 TranStar Disposable Transducer (K061573) or equivalent. | | HCED03 | Data Transfer Module | #### Indication for Use (Prescription Device): The Transonic HCM103 system is intended for diagnostic assessment of cardiovascular status and circulatory variables in neonatal, pediatric and adult patients. The measurements include continuous cardiac output and associated hemodynamic parameters (such as stroke volume variation; pulse pressure variation, systemic vascular resistance and related index parameters) by continuous arterial pressure waveform analysis. In this stand-alone system, cardiac output (CO) value is entered from such devices as COstatus (HCM101) system or equivalent to calibrate the pressure wave form for measurement of continuous cardiac output. In addition the system also provides heart rate, systolic, diastolic pressures, mean arterial pressure and indexed parameters. ### Substantial Equivalence: The Transonic HCM103 System for use with patients to provide diagnostic assessment cardiovascular status including continuous cardiac output and associated of hemodynamic parameters (such as stroke volume variation; pulse pressure variation, etc) is similar to the VIGILEO arterial pressure cardiac output (K082308) and PULSECO hemodynamic monitor (K010049) which are used to measure continuous cardiac output and related parameters using arterial pressure waveforms. Substantial equivalence determination is based on the following information and was not based on the assessment of non-clinical performance data. {5}------------------------------------------------ Proposed device and predicate devices analyze the arterial pressure waveform using pre-existing arterial access in patients to continuously measure cardiac output and other related hemodynamic parameters. PulseCO software which is used with LiDCOplus hemodynamic monitor requires calibration with cardiac output measurement. HCM103 also works in the same manner. For HCM103. cardiac output measurement to calibrate the pressure waveform, which is used to measure continuous cardiac output, is entered from such systems as COstatus (HCM101). The difference between Vigileo system and the proposed HCM103 system is that the Vigileo system requires use of FloTrac sensor, which is connected to the patient's bedside arterial pressure cable. Transonic HCM103 system requires use of any standard FDA cleared pressure transducer with suitable mating end to be placed in line with the patient's bedside arterial pressure transducer. Vigileo system in conjunction with FloTrac sensor does not require calibration whereas HCM103 system allows for calibration of pressure waveform. The difference between PulseCO (LiDCO) monitor and the proposed HCM103 system is that the PulseCO monitor connects to the bedside blood pressure monitor whereas the HCM103 system connects to the pre-existing patient's arterial pressure transducer using any standard FDA cleared pressure transducer with mating ends. These differences do not raise any new issues of safety or effectiveness regarding the use of Transonic HCM103 System. #### Performance Testing: Bench studies confirmed the accuracy of the HCM103 system based on arterial pressure waveform analysis to accurately measure continuous cardiac output. stroke volume variation, pulse pressure variation, systemic vascular resistance index and other related parameters. Testing consisted of software testing and electrical safety testing according to IEC 60601-1. Prior to shipment, the finished product will be tested and must meet all required release specifications before distribution. The array of testing required for release includes, but are not limited to; physical testing and visual examination (inprocess and finished product). The physical testing is defined by Quality Control Test Procedure documents. These tests are established testing procedures that ensure the product's performance parameters conform to the product design specifications. The testing instruction records for each of the individually required procedures are approved. released, distributed, and revised in accordance with document control cGMP's. Animal and clinical testing was not required to support substantial equivalence. {6}------------------------------------------------ ### Safety and Effectiveness: The successful testing demonstrates the safety and effectiveness of the HCM103 when used for the defined indications for use
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...