The SC 7000 and SC 9000XL INFINITY Modular Bedside monitors are capable of monitoring: - Heart rate - Respiration rate - Invasive pressure - Non-invasive pressure - Arrhythmia - Temperature - Cardiac output - Arterial oxygen saturation - Pulse rate - End-tidal carbon dioxide - (central) apnea - ST segment analysis. With the MultiGas™ and MultiGas+™ modules the SC 7000 and SC 9000XL are capable of measuring: - Respiration rate - Inspired and expired Carbon Dioxide (CO2) - Inspired and expired Oxygen (MultiGas+TM only) - Average inspired Oxygen (MultiGas™ only) - Inspired and expired gas concentrations of Enflurane, Isoflurane, Desflurane, Desflurane, Sevollurane, and Nitrous Oxide. The SC 7000 and SC 9000XL can interface with third party devices. The devices are intended to be used in the environment where patient care is provided by Healtheare Professionals, i.e. Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition. The devices are intended to be used on Adult, Pediatric and Neonatal populations, with the exception of the parameter Cardiac Output and ST Segment Analysis which are intended for use in the adult and pediatric populations only; and Arrhythmia which is intended for use in the adult population only.
Device Story
Modular bedside monitors (SC7000/SC9000XL) for continuous physiological monitoring; inputs include ECG, pressure transducers, temperature probes, SpO2 sensors, and gas modules; processes signals to display waveforms and numeric data; provides visual/audible alarms for parameter limit violations; used in clinical settings by physicians, nurses, and technicians; integrates with Infinity network and R50 recorders; supports clinical decision-making by providing real-time patient status and trend data; benefits include early detection of physiological instability.
Clinical Evidence
No clinical data provided; substantial equivalence based on non-clinical performance data and design similarity to the predicate device.
Technological Characteristics
Modular bedside monitor; color TFT 10.4" display; supports up to 8 waveforms; oscillometric NBP; MIB compatible; MultiGas module support; 3/7/8 lead ST monitoring; 8-channel IBP; networked via Infinity network; software-based processing; not MRI compatible.
Indications for Use
Indicated for adult, pediatric, and neonatal patients in healthcare settings for monitoring heart rate, respiration, invasive/non-invasive pressure, temperature, cardiac output, SpO2, pulse rate, EtCO2, apnea, ST segment, and anesthetic gases. Contraindications: Arrhythmia monitoring limited to adults; Cardiac Output and ST segment analysis limited to adult/pediatric populations. Not for use in MRI environments.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Siemens SC9000/SC9015 Patient Monitoring System (K946306)
Related Devices
K982730 — SC 7000 & SC 9000 XL INFINITY MODULAR BEDSIDE MONITOR · Siemens Medical Solutions USA, Inc. · Aug 24, 1998
K043439 — INFINITY MODULAR MONITORS WITH VF5 MODIFICATIONS (DELTA/KAPPA/DELTA XL/VISTA XL/GAMMA X XL AND SC 7000/8000/9000XL · Draeger Medical Systems, Inc. · Feb 18, 2005
K032749 — BEDSIDE MONITOR, MODEL OPV-1500 · Nihon Kohden America, Inc. · Sep 16, 2003
K031433 — SIEMENS INFINITY MODULAR MONITORS, MODELS SC 7000, SC 9000XL AND SC 8000 WITH VF3 MODIFICATIONS · Siemens Medical Solutions USA, Inc. · Aug 19, 2003
K070566 — INFINITY DELTA/DELTA XL/ KAPPA/ GAMMA X XL/ VISTA XL/ SC7000/SC8000/ SC9000XL · Draeger Medical Systems, Inc. · Jun 4, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the alphanumeric string "K980882" on the top line. The second line shows the alphanumeric string "P 1/3". The text appears to be handwritten with a dark pen or marker on a white background.
# 510(k) NOTIFICATION 510(k) NOTIFICATION
SIEMENS SC7000 and SC9000XL INFINITY Modular BedsideMonitors
# 510(k) SUMMARY
as required per 807.92(c)
Submitters Name, Address:
Siemens Medical Systems, Inc. Stement Systems Group, PCS Danvers, MA 01923 Tel: (978) 907-7500 Fax: (978) 750-6879 Fax: (978) 750 00:dent: David Simard, Director David Juality Assurance & Regulatory Affairs Contact person for this submission: Jacqueline Emery Contact person for this ents entruary 27, 1998
Trade Name, Common Name and Classification Name:
- A. Trade Name:
Trade Name:
Siemens SC7000 & SC9000 XL INFINITY Modular Bedside Monitor Series
- B. Common Name, Classification Name, Class and Regulation Number:
| Common Name | Classification<br>Number | Class | Regulation Number |
|---------------------------------------------|--------------------------|-------|-------------------|
| Cardiac monitor | 74DRT | II | 21 CFR 870.2300 |
| Pulse rate monitor | 74BWS | II | 21 CFR 870.2300 |
| Pulse oximeter | 74DQA | II | 21 CFR 870.2700 |
| Breathing Frequency Monitor | 73BZQ | II | 21 CFR 868.2375 |
| Clinical Electronic Thermometer | 80BWX | II | 21 CFR 880.2910 |
| Indwelling Blood Pressure Monitor | 74CAA | II | 21 CFR 870.1110 |
| Heart Rate Monitor, Neonatal | 74FLO | II | 21 CFR 870.2300 |
| Ventilatory Effort Monitor (Apnea Detector) | 73FLS | II | 21 CFR 868.2375 |
| Monitor Blood Pressure, Neonatal, Invasive | 74FLP | II | 21 CFR 870.1110 |
| Arrhythmia detector & Alarm | 74DSI | III | 21 CFR 870.1025 |
| Medical Cathode-Ray Tube Display | 74DXJ | II | 21 CFR 870.2450 |
| ST Segment Monitor with Alarm | 74MLD | III | 21 CFR 870.1025 |
| Non-indwelling Blood Pressure Monitor | 74DXN | II | 21 CFR 870.1130 |
| End-tidal Carbon-Dioxide Monitor | 73CCK | II | 21 CFR 868.1400 |
Predicate Device Identification:
licate Device Identification:
Siemens SC9000/SC9015 Patient Monitoring System, original 510(k) submission K946306.
ﻧﺴﻤﺎ
## COMPANY CONFIDENTIAL
## Siemens Medical Systems, Inc.
Electromedical Systems Group, PCS
16 Electronics Avenue Danvers, MA 01923 USA
Tel: (978) 907-7500 Fax: (978) 750-6879 Telex: 511958 (Siemensm SD)
{1}------------------------------------------------
### 510(k) NOTIFICATION SIEMENS SC7000 and SC9000XL INFINITY Modular BedsideMonitors
#### Device Description:
The SC 7000 and SC 9000XL modular bedside monitors are enhanced versions of the SC 9000 (predicate device - original submission K946306) and use the same front panel display and user interface as the SC 9000. In addition, the SC 9000 docking station is compatible with the SC 7000 and SC 9000XL for power and communication capabilities.
The SC 7000 (mid-level monitor) and the SC 9000XL (high-end monitor) are additions to the Siemens INFINITY Modular Portable Bedside Monitoring Series. Both the SC7000 and SC 9000 XL utilize the same electronics and software, but with different base configurations and available options.
#### Intended Use:
The intended use of the SC 7000 and SC 9000XL Bedside Monitoring Series is to measure heart rate, respiration rate, invasive pressure, non-invasive pressure, arrhythmia, temperature, arterial oxygen saturation and pulse rate, cardiac output, central apnea, and ST segment analysis. This device will produce visual and audible alarms if any of these parameters vary beyond preset limits and produce timed or alarm recordings. This device will connect to a Siemens R50 Bedside Recorder, either directly or via the Infinity network.
2
#### COMPANY CONFIDENTIAL
Siemens Medical Systems, Inc. Electromedical Systems Group, PCS
16 Electronies Avenue Danvers, MA 01923 USA
Tel: (978) 907-7500 Fax: (978) 750-6879 Telex: 511958 (Siemensm SD)
{2}------------------------------------------------
K980882
P3/3
## 510(k) NOTIFICATION 510(K) NOTTITICATION
SIEMENS SC7000 and SC9000XL INFINITY Modular Bedside Monitors
| | Substantial Equivalent Device | Applicant<br>Siemens Medical Systems | |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|-------------------|
| | Siemens SC9000 | Siemens SC 7000 | Siemens SC9000 XL |
| Manufacturer | Siemens | Same | Same |
| 510(k) Number | K946306 (original submission) | To be assigned | Same |
| Intended Use | The intended use of this device<br>is to measure heart rate,<br>respiration rate, invasive<br>pressure, non-invasive pressure,<br>arrhythmia, temperature, cardiac<br>output, arterial oxygen<br>saturation, pulse rate, cardiac<br>output, end-tidal carbon dioxide,<br>(central) apnea, and ST Segment<br>Analysis. This device will<br>produce visual and aural alarms<br>if any of these parameters vary<br>beyond preset limits and produce<br>timed or alarm recordings. This<br>device will connect to a Siemens<br>R50 Bedside recorder, either<br>directly or via the Infinity<br>Network. | Same | Same |
| Intended Population | Adult/Pediatric/Neonatal | Same | Same |
| Intended Environment | In a healthcare environment<br>where patient care is provided<br>by healthcare professionals. | Same | Same |
| Display type | Color TFT, 10.4" | Same | Same |
| Waveforms | Up to 8 | Same | Same |
| Arrhythmia Detection | Basic<br>Enhanced Optional | Same | Same |
| Modular | Yes | Same | Same |
| Networking | Standard | Same | Same |
| NBP | Oscillometric | Same | Same |
| MIB | Compatible | Same | Same |
| MultiGas | Compatible | Same | Same |
| ST | 3/7/8 leads | Same | Same |
| IBP | Up to 8 channels | Same | Same |
# Table of Device Similarities and differences to predicate device
Assessment of non-clinical performance data for equivalence: See Section U
Assessment of clinical performance data for equivalence: See Section V
Biocompatability: Not applicable
Sterilization: Not applicable
Standards and Guidances: See Section W
3
### COMPANY CONFIDENTIAL
### Siemens Medical Systems, Inc.
Electromedical Systems Group, PCS
16 Electronics Avenue Danvers, MA 01923 USA
Tel: (978) 907-7500 Fax: (978) 750-6879 Telex: 511958 (Siemensm SD)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird with outstretched wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
IIIN
Ms. Penelope H. Cameron Regulatory Submissions Manager Siemens Medical Systems, Inc. 16 Electronics Avenue Danvers, MA 01923
K980882 Re: Siemens SC 7000/SC 9000XL INFINITY Modular Bedside Monitors Requlatory Class: III (three) Product Code: DSI Dated: March 4, 1998 Received: March 9, 1998
Dear Ms. Cameron:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Psge 2 - Mr. Joel C. Kent
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
510(k) Number (if known):
Siemens SC 7000 and SC 9000 XL INFINITY Modular Bedside Monitors Device Name:
Indications for Use:
The SC 7000 and SC 9000XL INFINITY Modular Bedside monitors are capable of monitoring:
- Heart rate t
- Respiration rate .
- . Invasive pressure
- Non-invasive pressure 후
- . Arrhythmia
- Temperature .
- Cardiac output .
- Arterial oxygen saturation .
- Pulse rate ♥
- End-tidal carbon dioxide .
- (central) apnea ♥
- ST segment analysis .
With the MultiGas™ and MultiGas+™ modules the SC 7000 and SC 9000XL are capable of measuring:
- Respiration rate .
- Inspired and expired Carbon Dioxide (CO2) .
- Inspired and expired Oxygen (MultiGas+TM only) .
- Average inspired Oxygen (MultiGas™ only) .
- Inspired and expired gas concentrations of Enflurane, Isoflurane, Desflurane, Desflurane, Sevollurane, ● and Nitrous Oxide.
The SC 7000 and SC 9000XL can interface with third party devices.
The devices are intended to be used in the environment where patient care is provided by Healtheare. Professionals, i.e. Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition.
The devices are intended to be used on Adult, Pediatric and Neonatal populations, with the exception of the parameter Cardiac Output and ST Segment Analysis which are intended for use in the adult and pediatric populations only; and Arrhythmia which is intended for use in the adult population only.
#### MRI Compatibility Statement:
The Siemens SC 7000 and SC 9000XL INFINITY Modular Bedside Monitors are not compatible for use in a MRI magnetic field.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format
-
(Optional Format 1-2-96)
Division Sign-Off) ivision of Cardiovascular, Respiratory, and Neurological D
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.