K201288 · Baylis Medical Company, Inc. · DRE · Jun 12, 2020 · Cardiovascular
Device Facts
Record ID
K201288
Device Name
ExpanSure Large Access Transseptal Dilator
Applicant
Baylis Medical Company, Inc.
Product Code
DRE · Cardiovascular
Decision Date
Jun 12, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1310
Device Class
Class 2
Intended Use
The ExpanSure Large Access Transseptal Dilator is indicated for use in procedures where access to the left atrium via the transseptal technique is desired.
Device Story
ExpanSure Large Access Transseptal Dilator facilitates transseptal access to left atrium; used in cardiac catheterization/angiography. Device consists of polyethylene dilator with internal braid for torqueability and metallic marker coil at distal tip for enhanced fluoroscopic visualization; includes stainless steel J-tip guidewire. Operated by physicians in clinical settings. Dilator tracks over guidewire to dilate transseptal puncture site; metallic marker provides visual feedback to clinician during navigation. Single-use, sterile device; compatible with 0.032"/0.035" guidewires and 12.5 Fr introducer sheaths. Silicone lubricant coating reduces friction. Enhances procedural safety and efficiency through improved visualization of dilator tip.
Clinical Evidence
Bench testing only. No clinical data presented. Verification included mechanical testing (tip cap deflection/tensile strength), visual/dimensional analysis, biocompatibility (ISO 10993-1), pyrogen testing (LAL), packaging integrity (ISO 11607), and sterilization validation (ISO 11135) to a SAL of 10^-6.
Technological Characteristics
Polyethylene dilator shaft with internal braid; metallic marker coil at distal tip; stainless steel J-tip guidewire. Silicone lubricant coating. Compatible with 0.032"/0.035" guidewires and 12.5 Fr sheaths. Sterilized via Ethylene Oxide (ISO 11135). Mechanical testing per ISO 11070:2014. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for patients requiring transseptal access to the left atrium during cardiac procedures.
Regulatory Classification
Identification
A vessel dilator for percutaneous catheterization is a device which is placed over the guide wire to enlarge the opening in the vessel, and which is then removed before sliding the catheter over the guide wire.
K182064 — ExpanSure Transseptal Dilation System · Baylis Medical Company, Inc. · Mar 21, 2019
K220414 — VersaCross Connect Transseptal Dilator · Baylis Medical Company, Inc. · May 9, 2022
K241720 — VersaCross Connect Transseptal Dilator · Baylis Medical Company, Inc. · Jul 12, 2024
K251325 — VersaCross Connect Transseptal Dilator · Baylis Medical Company, Inc. · May 29, 2025
K233647 — VersaCross Connect Transseptal Dilator · Baylis Medical Company, Inc. · Dec 14, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
June 12, 2020
Baylis Medical Company Inc. May Tsai Regulatory Affairs Manager 2580 Matheson Blvd. East Mississauga, Ontario L4W 4J1 Canada
Re: K201288
Trade/Device Name: ExpanSure™ Large Access Transseptal Dilator Regulation Number: 21 CFR 870.1310 Regulation Name: Vessel Dilator For Percutaneous Catheterization Regulatory Class: Class II Product Code: DRE Dated: May 13, 2020 Received: May 14, 2020
Dear May Tsai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rachel Neubrander Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201288
Device Name
ExpanSure Large Access Transseptal Dilator
Indications for Use (Describe)
The ExpanSure Large Access Transseptal Dilator is indicated for use in procedures where access to the left atrium via the transseptal technique is desired.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 7. 510(k) Summary
# Submitter Information
| A. Company Name: | Baylis Medical Company Inc. |
|-------------------------|--------------------------------------------------------------------|
| B. Company Address: | 2580 Matheson Blvd. East<br>Mississauga, Ontario L4W 4J1<br>Canada |
| C. Company Phone: | (905) 602-4875 |
| D. Company Facsimile: | (905) 602-5671 |
| E. Contact Person: | May Tsai<br>Manager, Regulatory Affairs |
| F. Summary Prepared on: | 13-May-2020 |
# Device Identification
| A. Device Trade Name: | ExpanSure™ Large Access Transseptal<br>Dilator |
|-------------------------|---------------------------------------------------------------------|
| B. Device Common Name: | Dilator, Vessel, For Percutaneous<br>Catheterization |
| C. Classification Name: | CFR 870.1310 – Dilator, Vessel, For<br>Percutaneous Catheterization |
| D. Product Code: | DRE |
| E. Device Class: | Class II |
{4}------------------------------------------------
## I dentification of Legally Marketed Device
| Predicate Device | Manufacturer | 510(k) | Indications for Use |
|----------------------------------------|-----------------------------|---------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| ExpanSure™ Transseptal Dilation System | Baylis Medical Company Inc. | K182064 | The ExpanSure™ Transseptal Dilation System is indicated for use in procedures where access to the left atrium via the transseptal technique is desired. |
#### Table 7 1: Predicate Device
### Indications for Use
The ExpanSure™ Large Access Transseptal Dilator is indicated for use in procedures where access to the left atrium via the transseptal technique is desired.
### Device Description
The subject ExpanSure™ Large Access Transseptal Dilator represents a modification to the dilator component of the legally marketed ExpanSure™ Transseptal Dilation System (K182064). The modification was for the addition of a metallic marker coil to the dilator distal tip for increased visibility under imaging during procedures.
The subject device is comprised of a dilator (ExpanSure Dilator) and J-tip guidewire, which are single-use and supplied sterile to the user. The ExpanSure Dilator is designed for safe and easy catheterization and angiography of specific heart chambers and locations. The dilator consists of an inner and outer polyethylene layer containing radio-pacifier for visibility under imaging during procedures. The inner and outer shaft layers are separated by a braid material for torqueability. The dilator distal tip includes a metallic marker coil to facilitate visualization under imaging during procedures. The outer surface of the dilator is coated with a silicone lubricant. The ExpanSure Dilator is compatible with 0.032″ and 0.035″ transseptal devices/quidewires and 12.5 Fr introducer sheaths. The J-tip quidewire is comprised of stainless steel.
### Comparison to Predicate Device
The intended use and indications for use of the ExpanSure™ Large Access Transseptal Dilator remains unchanged from the predicate ExpanSure™ Transseptal Dilation System (K182064).
{5}------------------------------------------------
The subject and predicate device share the same fundamental scientific technology, principles of operation, mechanism of action, packaging configuration, and sterilization method (Table 7.2). Differences in design and technological characteristics between the subject and predicate devices do not raise any new concerns or different questions of safety and effectiveness. The results of verification and validation testing provide reasonable assurance of substantial equivalence of the ExpanSure™ Large Access Transseptal Dilator with the predicate device.
| Characteristic | Subject Device Compared to<br>Predicate ExpanSure™<br>Transseptal Dilation System<br>(K182064) | |
|-----------------------------------|------------------------------------------------------------------------------------------------|-----------|
| Intended Use | Identical | |
| Indications for Use | Identical | |
| Fundamental scientific technology | Identical | |
| Operating principles | Identical | |
| Mechanism of action | Identical | |
| Technological characteristics | Similar | |
| Materials | Patient contacting | Identical |
| | Non-patient contacting | Similar |
| Packaging configuration | Identical | |
| Sterilization method | Identical | |
| Accessory devices | Identical | |
Table 7 2: Comnarison of Subject and Predicate Device
# Performance Testing
Non-clinical performance testing was completed for the subject device to support substantial equivalence to the predicate device. The following verification and validation activities were completed to support the device modifications:
### Mechanical Testing
Mechanical verification was conducted for the subject device to verify compliance with the applicable requirements of ISO 11070:2014 and Baylis self-enforced requirements. The following mechanical tests were performed:
- 1. Strength of Union Tip Cap Deflection
- 2. Strength of Union Tip Cap Tensile
# General Physical Testing
General physical verification was conducted for the subject device to verify compliance with the applicable requirements of ISO 11070:2014 and Baylis self-enforced requirements. The following general physical tests were performed:
{6}------------------------------------------------
- 1. Visual Analysis
- 2. Measurements
### Pyrogen Testing
The subject device is supplied non-pyrogenic. Limulus Amoebocyte Lysate (LAL) testing was performed using the Kinetic Chromogenic method, as per ANSI/AAMI ST72:2011 and the FDA guidance document, " Guidance for Industry - Pyrogens and Endotoxins Testing: Questions and Answers," to verify the subject device meets current FDA and USP pyrogen limit specifications.
### Benchtop Validation
Customer requirements were validated through benchtop validation activities. Benchtop validation testing was performed to validate the performance of the subject device during normal intended use as per current applicable requirements of ISO 11070:2014 and Baylis selfenforced requirements.
In addition, the below evaluations were completed to confirm the subject device meets the following performance requirements:
## Biocompatibility Verification
Biological safety was evaluated for the subject device to verify compliance with the current applicable requirements of ISO 10993-1:2009/Cor.1:2010 and the FDA quidance document, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process."
### Packaging Verification
Ship testing was evaluated to verify the integrity of the subject device packaging through the rigors of shipping and handling as well as storage over time. The seal strength and sterile barrier integrity was also evaluated to verify compliance with the current applicable requirements of ANSI/AAMI/ISO 11607:2006/(R)2010 (Parts 1 and 2) over the proposed intended shelf life of the subject device.
### Sterilization Verification
Sterilization and residual limit verification were evaluated for the subject device to verify compliance with the current applicable requirements of ISO 11135:2014 and ISO 10993-7:2008/Cor.1:2009, respectively. Sterilization was performed with Ethylene Oxide to a Sterility Assurance Level (SAL) of 10-6.
The ExpanSure™ Large Access Transseptal Dilator met all test requirements as specified by applicable standards and test protocols. The verification and validation activities demonstrated the subject device meets its intended use
{7}------------------------------------------------
and is as safe, as effective, and performs in a manner that is substantially equivalent to the predicate device.
### Conclusions
The subject and predicate device share the same intended use and indications for use, fundamental scientific technology, principles of operation, mechanism of action, packaging configuration, and sterilization method. Design differences between the subject and predicate device do not raise any new concerns or different questions of safety and effectiveness. The results of verification and validation activities demonstrate the substantial equivalence of the ExpanSure™ Large Access Transseptal Dilator to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.