The HealthView ECG Manager is a computer system, used within a clinical network, which is intended to be used by trained professionals. It provides the ability to retrieve, store, edit, send and print ECGs and related digitized clinical documents through the use of on-screen measurement and editing tools. The system receives files (such as ECG, stress, or Holter) from any compatible device, displaying such data to the clinician for analysis and review. The product does not modify the original ECG waveform and does not provide an automated ECG analysis.
Device Story
HealthView ECG Manager is a software-only programmable diagnostic computer system; installed on clinical servers/workstations. It receives digitized ECG, stress, and Holter files from compatible devices; provides tools for retrieval, storage, editing, sending, and printing. Clinicians use on-screen measurement and editing tools to review data; device does not modify original waveforms or provide automated analysis. Used in clinical settings by trained professionals to assist in diagnosis and treatment planning. Device does not contact patients or control life-sustaining equipment; relies on human intervention for interpretation.
Clinical Evidence
Bench testing only. Verification and validation included: i) accuracy of measurement tools compared to cleared devices, ii) performance speed in simulated network environment, iii) retrieval speed validation, and iv) tool accuracy validation at a clinical site. No clinical data required.
Technological Characteristics
Software-only device; operates on standard server/workstation hardware. Functions include acquisition, archiving, viewing, and editing of ECG data. No patient contact. Risk management per ISO 14971 (2007).
Indications for Use
Indicated for use by trained professionals in a clinical network to retrieve, store, edit, send, and print ECGs and related digitized clinical documents (ECG, stress, or Holter) for analysis and review.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
# K123466
## 5. HealthView ECG Manager 510(k) Summary
# DEC 2 0 2012
The following safety and effectiveness summary has been prepared pursuant to 21 CFR 807.87 and 21 CFR 807.92 (for 510(k) summaries).
### Submitter Information
#### LUMEDX
| 110 110th Ave Ne (Suite 475) | |
|------------------------------|--------------|
| Bellevue, WA 98004 | |
| Phone: | 425 818 2326 |
| Fax: | 425 450 0962 |
| Contact Person: | Chris Pearce |
| Date prepared: | Nov 5th 2012 |
ባ
#### Device name and Classification
Trade name:
HealthView ECG Manager (version 1.0)
ECG Management System Common name: Classification Name: Programmable Diagnostic Computer 21 CFR Part 870.1425 Classification number: Device Class: Class II DQK (Cardiology Panel) Product Code:
### Substantial Equivalence
The HealthView ECG Manager (version 1.0) device, addressed in this premarket notification, is substantially equivalent to both of the following commercially available devices:
| 510(k) # | Trade Name | Manufacturer |
|----------|---------------|---------------|
| K052883 | Cardio Server | Epiphany |
| K083639 | MUSE | GE Healthcare |
#### Device description
HealthView ECG Manager (version 1.0), herein after referred to as ECG Manager, is a Programmable Diagnostic Computer system. It is a "software only" medical device, to be
Lumedx HealthView ECG Manager
Page 1 of 3
{1}------------------------------------------------
installed on a server and workstation(s) that meet the minimum hardware requirements noted in the documentation. The hardware itself is not considered a medical device and is not part of this 510(k) submission. The device provides a trained user with the ability to find, retrieve, view and integrate ECGs into a single patient record, to assist in the diagnosis and treatment planning of patients. The device does not contact the patient and does not control any life sustaining devices.
#### Intended Use
The HealthView ECG Manager is a computer system, used within a clinical network, which is intended to be used by trained professionals. It provides the ability to retrieve, store, edit, send and print ECGs and related digitized clinical documents through the use of on-screen measurement and editing tools. The system receives files (such as ECG, stress, or Holter) from any compatible device, displaying such data to the clinician for analysis and review.
The product does not modify the original ECG waveform and does not provide an automated ECG analysis.
## Technological characteristics of the device
ECG Manager is a software-only device that can be used on multiple hardware platforms (provided that the minimum hardware requirements are met) that allows acquiring, archiving, viewing, and editing. This device is used by trained and qualified professionals who have ample opportunity for competent human intervention in interpreting the waveforms and information presented to them. These technological characteristics are the same as those in the predicate devices, in terms of hardware needs, operating system requirements, overall functional characteristics, storage methodology, and waveform generation. This device does not physically come in contact with a patient, nor does it control any life-sustaining devices.
#### Performance Test Data
Every identified requirement has been tested and confirmed to be performing as expected (see Section 16, and provided screen shots of test files). Additionally, performance of the device has been substantiated in multiple ways: i) verifying accuracy of measurement tools using other cleared devices, ii) verifying the speed of performance in a simulated network environment, iii) validating the retrieval speed at a validation site, and iv) validating tools accuracy at a validation site.
Lumedx HealthView ECG Manager
Page 2 of 3
{2}------------------------------------------------
K123466
## General Safety and Effectiveness Concerns
Risk management is ensured by use of the ISO14971 (2007) standard, which has been used to identify and mitigate potential hazards. These potential hazards are also individually confirmed to be controlled via verification and validation testing, and any necessary cautions and warnings are included in user documentation.
Data created or modified on this device are evaluated by medical professionals, thus allowing for intervention in the event of a malfunction.
Lumedx therefore believes this device to be as safe and effective as the predicate devices referenced above; it does not introduce new technology or new indications for use.
## Conclusions - Substantial Equivalence
This submission includes the results of a hazard analysis, and shows that the potential hazards have been controlled. The Level of Concern of the device has been demonstrated to be "Moderate". All verification and validation testing has successfully concluded. The 510(k) premarket notification for the HealthView ECG Manager (version 1.0) contains adequate information to show substantial equivalence to the listed predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a circular logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol consists of three stylized, curved lines that converge at the bottom, resembling a stylized human form or a flame. The text is arranged around the circle's perimeter, with "DEPARTMENT OF HEALTH & HUMAN SERVICES" at the top and "USA" at the top right.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
# DEC 2 0 2012
LUMEDX. c/o Mr. Chris Pearce Director, Regulatory Affairs 110, 110th Ave NE Bellevue, WA 98004
#### Re: K123466
Trade/Device Name: HealthView ECG Manager Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable diagnostic computer Regulatory Class: Class II (two) Product Code: DQK Dated: November 8, 2012 Received: November 9, 2012
Dear Mr. Pearce:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
## Page 2 – Mr. Chris Pearce
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Owen P. Faris -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## 4. Indications for Use Statement
The HealthView ECG Manager is a computer system, used within a clinical network, which is intended to be used by trained professionals. It provides the ability to retrieve, store, edit, send and print ECGs and related digitized clinical documents through the use of on-screen measurement and editing tools. The system receives files (such as ECG, stress, or Holter) from any compatible device, displaying such data to the clinician for analysis and review.
The product does not modify the original ECG waveform and does not provide an automated ECG analysis.
Prescription Use × AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1
> (Division Sign-Off) Division of Cardiovascular Devices
**510(k) Number** k123466
Lumedx HealthView ECG Manager
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.