CARTO(R) EP NAVIGATION SYSTEM, VERSION 3.0 (WITH STANDARD LOCATION PAD)
K120550 · Biosense Webster, Inc. · DQK · May 7, 2012 · Cardiovascular
Device Facts
Record ID
K120550
Device Name
CARTO(R) EP NAVIGATION SYSTEM, VERSION 3.0 (WITH STANDARD LOCATION PAD)
Applicant
Biosense Webster, Inc.
Product Code
DQK · Cardiovascular
Decision Date
May 7, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Intended Use
The CARTO® 3 System V3.0 is intended for catheter-based atrial and ventricular mapping. The mapping system allows real-time display of cardiac maps in a number of different formats. Maps may be displayed as anatomical maps, cardiac electrical activation maps, cardiac electrical propagation maps, cardiac electrical potential maps, impedance maps, cardiac chamber geometry maps and ECG fragmentation maps. The acquired patient signals, including body surface ECG and intracardiac electrograms, may also be displayed in real time on the system's display screen. The CARTO® 3 System V3.0 is also intended to support EP procedures, maintaining CARTO® System capabilities in the presence of a high metallic environment and magnetic field strengths up to 0.1 T and provide a data communication channel to the Stereotaxis Niobe Catheter Navigation System. The CARTO® 3 System V3.0 includes CARTOMERGE® PLUS Module functionality to import, register and merge CT or MRI structural images with CARTO map's physiological information and real time catheter navigation. The system includes the Fast Anatomical Mapping (FAM) functionality that allows for the quick creation of cardiac anatomical volumes using catheters with magnetic location sensors. The system's CARTOSOUND® image integration functionality enables integration of intracardiac echo (ICE) to enable visualization of 3D combined maps. A fluoro background capability enables visualization of X-ray images as reference to the CARTO® Maps images and catheters. In addition to the use of specialized navigation catheters with magnetic location sensors, the system is also intended for use with conventional, non-navigational, electrophysiology catheters without magnetic location sensors.
Device Story
CARTO 3 System V3.0 is a catheter-based cardiac mapping system; acquires/analyzes intracardiac electrograms (IECG) and body surface ECG signals; utilizes magnetic sensor technology and Advanced Catheter Location (ACL) technology for real-time 3D anatomical and electroanatomical mapping. Used by physicians in clinical EP procedures; supports high metallic environments (up to 0.1 T). Features include CARTOMERGE PLUS (CT/MRI image registration), Fast Anatomical Mapping (FAM) for rapid volume creation, CARTOSOUND (ICE integration), and fluoroscopic background visualization. Provides data communication to Stereotaxis Niobe system. Output displayed on screen for real-time visualization of cardiac electrical activity and geometry; assists physicians in navigating catheters and identifying target sites for ablation or diagnosis; improves procedural efficiency and anatomical accuracy.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by extensive bench and animal testing to verify modified features and demonstrate that new functionality does not negatively impact existing performance.
Technological Characteristics
Cardiac mapping system; utilizes magnetic sensor technology and Advanced Catheter Location (ACL) technology. Supports integration of CT/MRI images, ICE, and fluoroscopy. Connectivity includes data communication channel to Stereotaxis Niobe system. Operates in magnetic fields up to 0.1 T. Compatible with specialized magnetic navigation catheters and conventional non-navigational EP catheters.
Indications for Use
Indicated for catheter-based atrial and ventricular mapping in patients undergoing electrophysiology (EP) procedures. Supports use with specialized magnetic navigation catheters and conventional non-navigational EP catheters.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
CARTO® XP EP Navigation System 2007, Version 9 (K070240)
Stereotaxis NIOBE® Magnetic Navigation System (K060967)
Related Devices
K103746 — CARTO 3 EP NAVIGATION SYSTEM, VERSION 2.0 (W/STANDARD LOCATION PAD),CARTO3 EPNAVIGATION SYSTEM,VERSION 2.0 W/RMT PAD) · Biosense Webster, Inc. · Jan 21, 2011
K112007 — CARTO 3 EP NAVIGATION SYSTEM, VERSION 2.2 · Biosense Webster, Inc. · Oct 5, 2011
K090017 — CARTO 3 EP NAVIGATION SYSTEM, V1.0 AND ACCESSORIES · Biosense Webster, Inc. · Oct 13, 2009
K133916 — CARTO 3 EP NAVIGATION SYSTEM, VERSION 4.2 · Biosense Webster, Inc. · Jul 1, 2014
K093455 — CARTO 3 VERSION 1.05 EP NAVIGATION SYSTEM AND ACCESSORIES · Biosense Webster, Inc. · Dec 4, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
:
## MAY - 7 2012
### 14 510(K) SUMMARY
#### General Information 14.1
| Applicant: | Biosense Webster, Inc.<br>3333 Diamond Canyon Road<br>Diamond Bar, CA 91765<br>USA<br>Phone: 909-839-8597<br>Fax: 909-839-8804 |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Date: | January 15, 2012 |
| Contact Person: | Wayne R. Hohman<br>Project Manager Regulatory Affairs |
| Trade/Proprietary Device<br>Name: | CARTO® 3 System V3.0 and Accessories |
| Manufacturing Part<br>Numbers: | FG-5400-00 (with standard location pad)<br>FG-5600-00 (with RMT location pad) |
| Common Device Name: | Cardiac mapping system |
| Classification Name: | Programmable diagnostic computer<br>Class II, 21 CFR 870.1425 Product Code DQK |
| Predicate Device: | CARTO® 3 SYSTEM V2.2 and Accessories<br>510(k) K112007 (October 5, 2011) |
| Reference Devices: | Reference Device 1:<br>CARTO® XP EP Navigation System 2007, Version 9,<br>510(k) K070240 (May 4, 2007) |
| | Reference Device 2:<br>Stereotaxis NIOBE® Magnetic Navigation System<br>510(k) K060967 (October 19, 2006) |
| Manufacturing Facilities: | Biosense Webster (Israel), Ltd.<br>a Johnson & Johnson Company<br>4 Hatnufa Street, POB 275<br>Yokneam 20692<br>Israel |
.
#### 14.2 Substantial Equivalence
CONFIDENTIAL
{1}------------------------------------------------
The CARTO® 3 EP Navigation System, Version 3.0, is substantially equivalent to the legally marketed CARTO® 3 EP Navigation System, Version 2.2and two Reference Devices as shown in the following Table:
| Table 6. Reference Devices for CARTO® 3 System, Version 3.0 | | |
|-------------------------------------------------------------|---------------|----------------------------------|
| Predicate/Reference Name | 510(k) Number | Equivalence Criteria |
| CARTO® 3 EP Navigation System, Version 2.2 | K112007 | Legacy Functionality and WiseTag |
| CARTO® XP EP Navigation System 2007, Version 9 | K070240 | Multi Electrode Mapping |
| Stereotaxis Niobe® Magnetic Navigation System | K060967 | Fluoroscopic background view |
#### 14.3 Description of Device
The CARTO® 3 System V3.0 is a catheter-based atrial and ventricular mapping system designed to acquire and analyze data points, and use this information to display 3D anatomical and electroanatomical maps of the human heart in real-time. The location information needed to create the cardiac maps and the local electrograms are acquired using a specialized mapping catheter and reference device. The system allows realtime display of electrograms and cardiac maps based on the received intra cardiac signals from the catheters in a number of different formats. The acquired patient signals, including body surface ECG and intracardiac electrograms (IECG) may also be displayed on the display screen. The CARTO® 3 System V3.0 uses two distinct types of location technology - magnetic sensor technology and Advanced Catheter Location (ACL) technology.
#### 14.4 Indications for Use
The Indications for Use for the modified device are identical to the predicate device:
The CARTO® 3 System V3.0 is intended for catheter-based atrial and ventricular mapping. The mapping system allows real-time display of cardiac maps in a number of different formats. Maps may be displayed as anatomical maps, cardiac electrical activation maps, cardiac electrical propagation maps, cardiac electrical potential maps, impedance maps, cardiac chamber geometry maps and ECG fragmentation maps. The acquired patient signals, including body surface ECG and intracardiac electrograms may also be displayed in real time on the System's display screen. The CARTO® 3 System V3.0 is also intended to support EP procedures, maintaining CARTO® System capabilities, in the presence of a high metallic environment and magnetic field strengths up to 0.1 T and provide a data communication channel to the Stereotaxis Niobe Catheter Navigation System. The CARTO® 3 System V3.0 includes CARTOMERGE® PLUS Module functionality to import, register and merge CT or MRI structural images with CARTO® Map's physiological information and real time catheter navigation. The system includes the Fast Anatomical Mapping (FAM) functionality that allows for the quick creation of cardiac anatomical volumes using
{2}------------------------------------------------
catheters with magnetic location sensors. The system's CARTOSOUND® Image Integration Module functionality enables integration of intracardiac echo (ICE) to enable visualization of 3D combined maps. In addition to the use of specialized navigation catheters with magnetic location sensors, the system is also intended for use with conventional, non-navigational, electrophysiology catheters without magnetic location sensors.
#### 14.5 Summary of Non-Clinical Performance Testing
The CARTO® 3 System V3.0 underwent extensive Bench and Animal Testing to verify the modified features and to demonstrate with regression testing that the new features did not negatively affect existing features. The CARTO® 3 System V3.0 passed all tests in accordance with appropriate test criteria and standards, and the modified device did not raise new questions of safety or effectiveness.
#### 14.6 Conclusions
The Bench and Animal Non-Clinical Testing demonstrated that the CARTO® 3 System V3.0 is as safe, as effective, and performs as well as or better than the predicate device and two Reference Devices. This testing program supports the determination of substantial equivalence to the predicate devices.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of two main elements: a circular seal on the left and a stylized symbol on the right. The circular seal contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter, with the department's emblem in the center. The stylized symbol on the right resembles an abstract representation of a human figure or a bird in flight, composed of curved lines.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002
MAY - 7 2012
Biosense Webster, Inc. c/o Mr. Wayne R. Hohman Project Manager Regulatory Affairs Biosense Webster, Inc 3333 Diamond Canyon Road Diamond Bar, CA 91765
Re: K120550
> Trade/Device Name: CARTO® 3 EP Navigation System Version 3.0 Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II (two) Product Codes: DQK Dated: February 20, 2012 Received: February 23, 2012
Dear Mr. Hohman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
Page 2 - Mr. Wayne R. Hohman
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
R. D. Zimbawe, M.P.
(Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE
5
# 510(k) No (if known): K 120550
Device Name: CARTO® 3 EP Navigation System, Version 3.0
### Indications for Use:
The CARTO® 3 System V3.0 is intended for catheter-based atrial and ventricular mapping. The mapping system allows real-time display of cardiac maps in a number of different formats. Maps may be displayed as anatomical maps, cardiac electrical activation maps, cardiac electrical propagation maps, cardiac electrical potential maps, impedance maps, cardiac chamber geometry maps and ECG fragmentation maps. The acquired patient signals, including body surface ECG and intracardiac electrograms, may also be displayed in real time on the system's display screen. The CARTO® 3 System V3.0 is also intended to support EP procedures, maintaining CARTO® System capabilities in the presence of a high metallic environment and magnetic field strengths up to 0.1 T and provide a data communication channel to the Stereotaxis Niobe Catheter Navigation System. The CARTO® 3 System V3.0 includes CARTOMERGE® PLUS Module functionality to import, register and merge CT or MRI structural images with CARTO map's physiological information and real time catheter navigation. The system includes the Fast Anatomical Mapping (FAM) functionality that allows for the quick creation of cardiac anatomical volumes using catheters with magnetic location sensors. The system's CARTOSOUND® image integration functionality enables integration of intracardiac echo (ICE) to enable visualization of 3D combined maps. A fluoro background capability enables visualization of X-ray images as reference to the CARTO® Maps images and catheters. In addition to the use of specialized navigation catheters with magnetic location sensors, the system is also intended for use with conventional, non-navigational, electrophysiology catheters without magnetic location sensors.
V-Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ADEA
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number K100550
Biosense Webster, Inc. CARTO® 3 V3.0 EP NAVIGATION CONFIDENTIAL
Traditional 510(k) Page 44 of 1219
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.