ENSITE VELOCITY SYSTEM REFERRED TO ENSITE VELOCITY SYSTEM WITH POSITIONAL REFERENCE TOOL (PRT)
K101667 · St Jude Medical · DQK · Jul 2, 2010 · Cardiovascular
Device Facts
Record ID
K101667
Device Name
ENSITE VELOCITY SYSTEM REFERRED TO ENSITE VELOCITY SYSTEM WITH POSITIONAL REFERENCE TOOL (PRT)
Applicant
St Jude Medical
Product Code
DQK · Cardiovascular
Decision Date
Jul 2, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1425
Device Class
Class 2
Intended Use
The EnSite Velocity™ Cardiac Mapping System is a suggested diagnostic tool in patients for whom electrophysiology studies are indicated. When used with the EnSite Array™ Catheter, the EnSite System is intended to be used in the right atrium of patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone. OR When used with the EnSite NavX Surface Electrode Kit, the EnSite System is intended to display the position of conventional electrophysiology (EP) catheters in the heart.
Device Story
EnSite Velocity System is an electrophysiology mapping system consisting of a workstation, amplifier, EnSite Array diagnostic catheter, and NavX surface electrode kit. System processes electrical signals from catheters and surface electrodes to generate cardiac maps and display catheter positions in the heart. Used in clinical settings by physicians to identify complex arrhythmias. Output provides visual representation of cardiac electrical activity and catheter location, assisting in diagnostic decision-making and procedural guidance. System facilitates navigation and mapping of cardiac structures during electrophysiology studies.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Programmable diagnostic computer (21 CFR 870.1425). System includes display workstation, amplifier, EnSite Array catheter, and NavX surface electrode kit. Operates as an electrophysiology mapping system for cardiac signal processing and catheter visualization.
Indications for Use
Indicated for patients requiring electrophysiology studies. Used with EnSite Array Catheter for complex arrhythmias in the right atrium; used with EnSite NavX Surface Electrode Kit for displaying position of conventional EP catheters in the heart.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
K101328 — ENSITE VELOCITY SYSTEM · St Jude Medical · Jul 14, 2010
K093942 — ENSITE VELOCITY SYSTEM V1.3 · St Jude Medical · Feb 25, 2010
K141050 — ENSITE VELOCITY SYSTEM · St Jude Medical · Jun 6, 2014
K130594 — ENSITE VELOCITY SYSTEM · St Jude Medical · May 3, 2013
K101419 — ENSITE VELOCITY SYSTEM, MODEL EE3000 · St Jude Medical · Jun 11, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
K101667 1/3
# 12 510(k) Summary for Public Disclosure
JUL -- 2 2010
## 12.1 Submitter's Name/Contact Person
Donna R. Lunak St. Jude Medical One St. Jude Medical Drive St. Paul, MN 55117 USA
The Establishment Registration Number is 2184149.
### 12.2 Common or Usual Name
Electrophysiology Mapping System with console and catheter
### 12.3 Proprietary Name
EnSite Velocity System Consisting of: EnSite Multi-electrode Diagnostic Catheter (EnSite Array™ - Model EC1000) EnSite Electrophysiology Workstation - Model EE3000 EnSite NavX Surface Electrode Kit - Model EN0010
## 12.4 Classification Name
DQK, Programmable diagnostic computer (21 CFR 870.1425), Class II, Circulatory Systems Device Panel
### 12.5 Hardware Description
The EnSite Velocity System consists of the following:
- Display Workstation .
- Amplifier .
{1}------------------------------------------------
ે જે
### 12.6 Indications for Use
The EnSite Velocity™ Cardiac Mapping System is a suggested diagnostic tool in patients for whom electrophysiology studies are indicated.
- When used with the EnSite Array™ Catheter, the EnSite System is . intended to be used in the right atrium of patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone.
OR
- When used with the EnSite NavX Surface Electrode Kit, the EnSite . System is intended to display the position of conventional electrophysiology (EP) catheters in the heart.
### 12.7 Device Comparison to the Predicate Device
- The EnSite Velocity System with PRT has the same intended use and fundamental scientific technology as the predicate device - EnSite Velocity System (K093942). All technological characteristics of the EnSite Velocity System with PRT are substantially equivalent to the predicate device - EnSite Velocity System (K093942).
#### 12.8 Summary of Non-Clinical Testing
Bench testing was performed to confirm that the changes met design requirements and did not affect the safety or effectiveness of the product.
#### 12.9 Summary of Design Control Activities
The development of the EnSite Velocity System with PRT was performed in accordance with St. Jude Medical's Quality System requirements, and in compliance with Quality System Requlation design controls requirements documented in 21 CFR 820.30. A Declaration of Conformity with Design Controls follows in section 13.4.
### 12.10 Conclusion
The EnSite Velocity System with PRT has the same indications for use, intended use and fundamental scientific technology as the predicate device. All technological characteristics of the EnSite Velocity System with PRT are substantially equivalent to the predicate device.
{2}------------------------------------------------
Where operational and performance differences exist between the proposed device and the predicate device, performance testing demonstrated that these differences do not adversely affect the device's safety and effectiveness.
Therefore, St. Jude Medical considers the EnSite Velocity System with PRT to be substantially equivalent to the predicate device - EnSite Velocity System (K093942).
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three curved lines representing its wings or body.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
St. Jude Medical c/o Donna R. Lunak Regulatory Specialist One St. Jude Medical Drive Saint Paul MN 55117
# JUL - 2 2010
Re: K101667
Trade/Device Name: EnSite Velocity System with PRT Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II (two) Product Code: DQK Dated: May 28, 2010 Received: June 14, 2010
### Dear Ms. Lunak:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Ms. Donna Lunak
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809:htm-forthe Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use 7
Device Name: EnSite Velocity System
Indications for Use:
The EnSite Velocity™ Cardiac Mapping System is a suggested diagnostic tool in patients for whom electrophysiology studies are indicated.
- When used with the EnSite Array™ Catheter, the EnSite System is intended to . be used in the right atrium of patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone.
- OR
- When used with the EnSite NavX Surface Electrode Kit, the EnSite System is . intended to display the position of conventional electrophysiology (EP) catheters in the heart.
Over-The-Counter Use Prescription Use × Prescription Use
(Part 21 CFR 801 Subpart D) AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Signature
(Division Sign-Off) (Division of Cardiovascular Devices
510(k) Number *K101667*
Special 510(k): EnSite Velocity™ System with Positional Reference Tool (PRT)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.