K051450 · Inovise Medical, Inc. · DQD · Nov 10, 2005 · Cardiovascular
Device Facts
Record ID
K051450
Device Name
EXTENDED MEASUREMENTS SYSTEM
Applicant
Inovise Medical, Inc.
Product Code
DQD · Cardiovascular
Decision Date
Nov 10, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1875
Device Class
Class 2
Intended Use
The Extended Measurements System, when used with AUDICOR Sensors on the chest wall, is intended for use in acquiring, analyzing and reporting ECG and heart sound data and to provide interpretation of the data in an integrated COR report for consideration by physicians. In addition the data can be presented in the format of real-time and trended display of cardiac indices derived from simultaneous ECG data and heart sounds data. The Extended Measurements System can be used as an add-on device to work in conjunction with an AUDICOR-enabled device, or as a stand-alone electrocardiograph. The interpretations of ECG and heart sound data offered by the device are only significant when used in conjunction with physician over read as well as consideration of other relevant patient data. The device is intended for use only under the direct supervision of a physician, and is for use on adults (≥ 18 years).
Device Story
System acquires 12-lead ECG and heart sound data via chest-wall sensors; communicates via Bluetooth to compatible computer or functions as standalone electrocardiograph. Software processes inputs to generate integrated COR report; displays real-time/trended cardiac indices including LVST, LVDT, PADT, AAFT, EMAT, QS2, R-R interval, and S3/S4 evidence. Used in clinical settings under physician supervision. Output assists physicians in evaluating MI, LVH, and heart sound conditions; requires physician over-read and integration with other patient data for clinical decision-making. Benefits include automated derivation and trend monitoring of cardiac indices to support cardiac assessment.
Clinical Evidence
No clinical data provided. Device safety and effectiveness established through bench testing against ANSI/AAMI EC-11, EN 60601-1, EN 60601-1-2, and IEC 60601-2-25 standards.
Technological Characteristics
Software-based system; Bluetooth connectivity; 12-lead ECG and heart sound acquisition. Complies with ANSI/AAMI EC-11, IEC 60601-2-25, and EN 60601-1 safety standards. Provides automated calculation of cardiac indices (LVST, LVDT, PADT, AAFT, EMAT, QS2, R-R interval, S1/S2 intensity).
Indications for Use
Indicated for adult patients (≥ 18 years) requiring acquisition, analysis, and reporting of ECG and heart sound data. Used for monitoring cardiac indices under direct physician supervision.
Regulatory Classification
Identification
Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
K043074 — LIBERTY SYSTEM · Inovise Medical, Inc. · Feb 22, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
K051450
Inovise Medical, Inc.
Extended Measurements Modification to AUDICOR® 200 System
## NOV 1 0 2005
## Summary of Safety and Effectiveness
| As required by 21 CFR, part 807.92 | |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted By: | Inovise Medical, Inc.<br>10565 SW Nimbus Ave, Suite 100<br>Portland, OR 97233-4311<br>Phone 503-431-3837<br>Fax 503-431-3801 |
| Contact: | Steve Hesler<br>Director, Quality and Regulatory |
| Date Prepared: | May 30, 2005 |
| Proprietary<br>Name: | Extended Measurements AUDICOR System |
| Common/ Usual<br>Name: | Electrocardiograph |
| Classification: | 870.2340, 74 DPS |
| Performance<br>Standards: | AANSI/AAMI EC 11 |
| Intended Use: | The Extended Measurements System, when used with AUDICOR Sensors on the<br>chest wall, is intended for use in acquiring, analyzing and reporting ECG and heart<br>sound data and to provide interpretation of the data in an integrated COR report for<br>consideration by physicians. In addition the data can be presented in the format of<br>real-time and trended display of cardiac indices derived from simultaneous ECG<br>data and heart sounds data.<br><br>The Extended Measurements System can be used as an add-on device to work in<br>conjunction with an AUDICOR-enabled device, or as a stand-alone<br>electrocardiograph.<br><br>The interpretations of ECG and heart sound data offered by the device are only<br>significant when used in conjunction with physician over read as well as<br>consideration of other relevant patient data.<br><br>The device is intended for use only under the direct supervision of a physician, and<br>is for use on adults (≥ 18 years). |
| Device<br>Description: | The Extended Measurements System is a software-based device that that can be<br>installed into a compatible computer to communicate with an AUDICOR-capable<br>device via a Blue tooth interface. The Extended Measurements system will also<br>be offered as a stand-alone electrocardiograph/ monitoring device. |
| Test Summary &<br>Conclusion: | The Extended Measurements System was tested to the applicable requirements of<br>the following standards, and shown to comply.<br>ANSI/AAMI EC-11 Standard for Diagnostic Electrocardiographic Devices EN 60601-1 (UL 60601-1) Standard for Medical Electrical Equipment: General<br>Requirements for Safety EN 60601-1-2 Standard for Medical Electrical Equipment Part 1-2: General<br>Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -<br>Requirements and Tests IEC 60601-2-25 Medical Electrical Equipment Part 2: Particular requirements<br>for the safety of electrocardiographs |
{1}------------------------------------------------
Substantial Equivalence:
The Inovise Medical Extended Measurements System is substantially equivalent to the AUDICOR® 200 system (K043074, cleared as Liberty).
Technological Characteristics:
The Extended Measurements System and the AUDICOR predicate device are technologically equivalent in that both acquire 12 lead ECG and heart sounds data from adult patients then present the data in the COR report format which includes graphic display of MI and LVH conditions along with detection and display of S3 and S4 heart sounds.
The Extended Measurements System adds the capability of displaying and monitoring cardiac indices derived form the ECG and hearts sounds information. These indices include:
- . Left Ventricular Systolic Time (LVST)
- Left Ventricular Diastolic Time (LVDT) .
- Pre-atrial Diastolic Filling Time (PADT) .
- Accelerated Atrial Filling Time (AAFT) .
- QS1 (EMAT) .
- � QS2
- R-R Interval .
- � S3 Evidence
- S4 Evidence .
- S1 Intensity .
- S2 Intensity t
- S1/S2 Intensity Ratio ↓
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other.
Public Health Service
NOV 1 0 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Steve Hesler Director, Quality and Regulatory Inovise Medical Inc. 10565 SW Nimbus Ave, Suite 100 Portland, OR 97223-4311
Re: K051450
Trade Name: AUDICOR® Extended Measurements System Regulation Number: 21 CFR 870.1875 Regulation Name: Electronic Stethoscope Regulatory Class: Class II (two) Product Code: DQD Dated: May 31, 2005 Received: June 2, 2005
Dear Mr. Hesler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Steve Hesler
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bhimmar for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): K051450
Extended Measurements AUDICOR System Device Name:
Indications For Use:
The Extended Measurements System, when used with Audicor Sensors on the chest wall, is intended for use in acquiring, analyzing and reporting ECG and heart sound data and to provide interpretation of the data in an integrated COR report for consideration by physicians. In addition, the data can be presented in the format of real-time and trended display of cardiac indices derived from simultaneous ECG data and heart sounds data.
The Extended Measurements System can be used as an add-on device to work in conjunction with an Audicor-enabled device, or as a stand-alone electrocardiograph.
The interpretations of ECG and heart sound data offered by the device are only significant when used in conjunction with physician over read as well as consideration of other relevant patient data.
The device is intended for use only under the direct supervision of a physician, and is for use on adults (> 18 years).
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Bhumimmar
Division of Cardlovascular Devices
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.