K192239 · C.R. Bard, Inc. · PQK · Oct 17, 2019 · Cardiovascular
Device Facts
Record ID
K192239
Device Name
WavelinQ Plus EndoAVF System
Applicant
C.R. Bard, Inc.
Product Code
PQK · Cardiovascular
Decision Date
Oct 17, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1252
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The WavelinQ™ EndoAVF System is indicated for the creation of an arteriovenous fistula (AVF) using concomitant ulnar artery and ulnar vein or concomitant radial artery and radial vein in patients with minimum artery and vein diameters of 2.0 mm at the fistula creation site who have chronic kidney disease and need hemodialysis.
Device Story
System consists of two single-use magnetic catheters (venous and arterial) for percutaneous creation of arteriovenous fistula (AVF). Venous catheter contains RF cutting electrode; arterial catheter contains backstop. Magnets in both catheters align and appose vessels; RF energy delivered to create fistula. Catheters feature braid-reinforced Pebax shafts, radiopaque tips, and handles for navigation over 0.014" guidewires. Used by clinicians in clinical setting. Provides minimally invasive alternative to surgical AVF creation, potentially reducing trauma and improving access for hemodialysis patients.
Clinical Evidence
No clinical data provided. Evidence consists of non-clinical bench testing, including electrical safety (IEC 60601), EMC testing, magnet array force coaptation, S-turn tortuosity alignment comparison, and a design validation cadaver study.
Technological Characteristics
Braid-reinforced Pebax catheter shafts; radiopaque Pebax tips; magnetic alignment system; RF energy source for tissue cutting. Single-use, disposable. Compatible with 0.014" guidewires. Electrical safety per IEC 60601; EMC compliant.
Indications for Use
Indicated for creation of arteriovenous fistula (AVF) in patients with chronic kidney disease requiring hemodialysis; requires minimum artery and vein diameters of 2.0 mm at creation site; utilizes concomitant ulnar artery/vein or radial artery/vein.
Regulatory Classification
Identification
This device is a single use percutaneous catheter system that creates an arteriovenous fistula in the arm of patients with chronic kidney disease who need hemodialysis.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must evaluate:
(i) The ability to safely deliver, deploy, and remove the device;
(ii) The ability of the device to create an arteriovenous fistula;
(iii) The ability of the arteriovenous fistula to attain a blood flow rate and diameter suitable for hemodialysis;
(iv) The ability of the fistula to be used for vascular access for hemodialysis;
(v) The patency of the fistula; and
(vi) The rates and types of all adverse events.
(2) Animal testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be assessed:
(i) Delivery, deployment, and retrieval of the device;
(ii) Compatibility with other devices labeled for use with the device;
(iii) Patency of the fistula;
(iv) Characterization of blood flow at the time of the fistula creation procedure and at chronic followup; and
(v) Gross pathology and histopathology assessing vascular injury and downstream embolization.
(3) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Simulated-use testing in a clinically relevant bench anatomic model to assess the delivery, deployment, activation, and retrieval of the device;
(ii) Tensile strengths of joints and components;
(iii) Accurate positioning and alignment of the device to achieve fistula creation; and
(iv) Characterization and verification of all dimensions.
(4) Electrical performance, electrical safety, and electromagnetic compatibility (EMC) testing must be performed for devices with electrical components.
(5) Software verification, validation, and hazard analysis must be performed for devices that use software.
(6) All patient-contacting components of the device must be demonstrated to be biocompatible.
(7) Performance data must demonstrate the sterility of the device components intended to be provided sterile.
(8) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
(9) Labeling for the device must include:
(i) Instructions for use;
(ii) Identification of system components and compatible devices;
(iii) Expertise needed for the safe use of the device;
(iv) A detailed summary of the clinical testing conducted and the patient population studied; and
(v) A shelf life and storage conditions.
K182796 — WavelinQ 4F EndoAVF System · C.R. Bard, Inc. · Feb 6, 2019
K191114 — Ellipsys Vascular Access System (Ellipsys System), Power Controller, Ellipsys Vascular Access System (Ellipsys System), Catheter, Ellipsys Vascular Access System (Ellipsys System), Needle · Avenu Medical · Aug 9, 2019
DEN160006 — everlinQ endoAVF System · Tva Medical, Inc. · Jun 22, 2018
DEN170004 — Ellipsys Vascular Access System · Avenu Medical, Inc. · Jun 22, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 17, 2019
Bard Peripheral Vascular Kulveen Dhatt Senior Regulatory Affairs Specialist 1625 West 3rd Street Tempe, AZ 85281
Re: K192239
Trade/Device Name: WavelinQ EndoAVF System Regulation Number: 21 CFR 870.1252 Regulation Name: Percutaneous Catheter for Creation Of An Arteriovenous Fistula Regulatory Class: Class II Product Code: PQK Dated: August 16, 2019 Received: August 19, 2019
Dear Ms. Dhatt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Carmen Gacchina Johnson Acting Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192239
Device Name WavelinQ EndoAVF System
#### Indications for Use (Describe)
The WavelinQ™ EndoAVF System is indicated for the creation of an arteriovenous fistula (AVF) using concomitant ulnar artery and ulnar vein or concomitant radial vein in patients with minimum artery and vein diameters of 2.0 mm at the fistula creation site who have chronic kidney disease and need hemodialysis.
| Type of Use (Select one or both, as applicable) |
|-----------------------------------------------------------------------------------|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
| <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## WavelinQ™ EndoAVF System
### 510(k) Summary 21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
### Submitter Information:
| Applicant: | Bard Peripheral Vascular, Inc. |
|------------|-----------------------------------------------------|
| | 1625 West 3rd Street |
| | Tempe, Arizona 85281 |
| Phone: | 480-379-2875 |
| Fax: | 480-449-2546 |
| Contact: | Kulveen Dhatt, Senior Regulatory Affairs Specialist |
| Date: | August 16, 2019 |
### Subject Device Name:
| Device Trade Name: | WavelinQ™ EndoAVF System |
|-----------------------|--------------------------------------------------------------------|
| Common or Usual Name: | Percutaneous catheter for creation of an arteriovenous fistula for |
| | hemodialysis access |
| Classification Name: | Percutaneous catheter for creation of an arteriovenous fistula for |
| | hemodialysis access |
| Product Code: | PQK |
| Regulatory Class: | Class II |
| Review Panel: | Cardiovascular |
| Regulation Number: | 21 CFR 870.1252 |
| Predicate Device: | |
| Device Trade Name: | WavelinQ™ 4F EndoAVF System |
|-----------------------|----------------------------------------------------------------------------------------|
| 510(k) Number: | K182796 |
| Common or Usual Name: | Percutaneous catheter for creation of an arteriovenous fistula for hemodialysis access |
{4}------------------------------------------------
K192239
| Classification Name: | Percutaneous catheter for creation of an arteriovenous fistula for<br>hemodialysis access |
|----------------------|-------------------------------------------------------------------------------------------|
| Product Code: | PQK |
| Regulatory Class: | Class II |
| Review Panel: | Cardiovascular |
| Regulation Number: | 21 CFR 870.1252 |
### Device Description:
The WavelinQ™ EndoAVF System consists of two single-use disposable magnetic catheters: a venous catheter and an arterial catheter. The venous catheter contains an array of magnets positioned on either side of a radiofrequency (RF) cutting electrode contained within an isolative housing. The arterial catheter contains a matching array of magnets positioned on either side of an electrode "backstop". The backstop serves as a mechanical stop for the cutting electrode to contact following the creation of the AVF. The magnets in the two catheters serve to align and appose the arterial backstop of the arterial catheter with the RF electrode of the venous catheter when positioned in the target AVF location. Radiofrequency energy can then be delivered through the electrode for cutting tissue and AVF creation.
The arterial and venous catheters are both comprised of braid reinforced Pebax catheter shafts. These shafts provide flexibility for device delivery and torquability to aid in rotational alignment and positioning. Both catheters include a soft, radiopaque, rapid exchange style Pebax tip for atraumatic device navigation with radiographic visibility. These tips allow the catheters to track over a standard guide wire 0.014" or smaller. The catheters include a handle/hub to facilitate device delivery, positioning and alignment.
### Indications for Use of Device:
The WavelinQ™ EndoAVF System is indicated for the creation of an arteriovenous fistula (AVF) using concomitant ulnar artery and ulnar vein or concomitant radial artery and radial vein in patients with minimum artery and vein diameters of 2.0 mm at the fistula creation site who have chronic kidney disease and need hemodialysis.
{5}------------------------------------------------
K192239
### Technological comparison to Predicate Device:
The subject WavelinQ™ EndoAVF System has the following similarities to the WavelinQ™ 4F EndoAVF System predicate device (K182796 – cleared on February 6, 2018): same intended use, same indications for use, same target population, same operating principle, same fundamental scientific technology, same sterility assurance level and method of sterilization, same performance bench testing with same acceptance criteria, same biocompatibility requirements, same packaging configurations, and same materials.
The WavelinQ™ EndoAVF System includes minor design modifications when compared to the WavelinQ™ 4F EndoAVF System predicate device. The WavelinQ™ EndoAVF System includes one additional radiopaque rotational indicator to both the venous and arterial catheters and the addition of 24 magnets each to both the arterial and venous catheters. The minor design modifications have been verified and validated through appropriate non clinical performance data. There are no changes with respect to manufacturing, materials and their processing compared to the predicate WavelinQ™ 4F EndoAVF System. Magnet Array Force Coaptation Testing and S-Turn Tortuosity Coaptation and Alignment Comparison Testing were introduced as a verification tests in order to compare the subject device to the predicate device and confirm that minor modifications of the WavelinQ™ EndoAVF System maintain or exceed performance requirements previously established with the predicate WavelinQ™ 4F EndoAVF System.
In order to address the performance requirements previously validated by end users which are potentially impacted by the minor design modifications, repeat validation testing was performed on the subject WavelinQ™ EndoAVF System to confirm design outputs continue to satisfy the design inputs and user needs. The repeat design validation provides confirmation that design outputs satisfy those design inputs that may not otherwise be evaluated through design verification testing.
### Performance Data:
#### Non Clinical Performance Data:
To demonstrate that the subject device WavelinQ™ EndoAVF System is as safe and effective as the predicate device WavelinQ™ 4F EndoAVF System, the technological characteristics and performance criteria were evaluated. Using internal risk assessment procedures, tests of the following characteristics and performance criteria were performed on the subject device:
{6}------------------------------------------------
- Electrical Safety per IEC 60601 ●
- Electromagnetic Compatibility (EMC) Testing
- Design Validation Cadaver Study ●
- . Magnet Array Force Coaptation Testing
- S-Turn Tortuosity Coaptation and Alignment Comparison Testing .
### Conclusions:
The subject device, WavelinQ™ EndoAVF System, met all the predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. The subject device and the predicate device share the same characteristics: intended use, indications for use, target populations for use, and fundamental scientific technology. Therefore, Bard Peripheral Vascular, Inc. concludes that the subject device WavelinQ™ EndoAVF System is substantially equivalent to the legally marketed predicate device WavelinQ™ 4F EndoAVF System.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.