K051670 · Baylis Medical Co., Inc. · PDU · Mar 30, 2006 · Cardiovascular
Device Facts
Record ID
K051670
Device Name
RF TUNNELER WIRE
Applicant
Baylis Medical Co., Inc.
Product Code
PDU · Cardiovascular
Decision Date
Mar 30, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The RF Tunneler Wire is intended to create a channel in totally occluded peripheral vessels 3 mm or greater.
Device Story
RF Tunneler Wire is a guide wire used in peripheral interventional procedures to cross total occlusions. Device consists of a core wire with polymer insulation and an active distal tip. It connects to an external RF perforation generator via a connector cable. During use, the device delivers RF energy at the active tip to ablate or penetrate target tissue, creating a channel through the occlusion. It is designed for compatibility with standard balloon and stent catheters. The device is operated by a physician in a clinical setting. By facilitating the crossing of occluded vessels, it enables subsequent interventional procedures such as balloon angioplasty or stenting, potentially restoring blood flow in patients with peripheral vascular disease.
Clinical Evidence
Bench testing only. No clinical data provided. Mechanical and electrical bench testing verified safety and performance characteristics against predicate devices.
Technological Characteristics
Core wire with polymer insulation; active distal tip for RF energy delivery; connects to external RF perforation generator. Device is a percutaneous catheter-based wire. Biocompatibility and sterilization validated.
Indications for Use
Indicated for creating channels in totally occluded peripheral vessels 3 mm or greater in diameter.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
K232562 — PowerWire Radiofrequency Guidewire Kit · Baylis Medical Technologies, Inc. · Feb 28, 2024
K101615 — POWERWIRE RADIOFREQUENCY GUIDEWIRE · Baylis Medical Co., Inc. · Jul 28, 2010
K251158 — PowerWire® 14 Radiofrequency Guidewire Kit · Baylis Medical Technologies, Inc. · Sep 11, 2025
K041973 — MODIFICATION TO SAFE-CROSS RADIO FREQUENCY TOTAL OCCLUSION CROSSING SYSTEM · Intra Luminal Therapeutics, Inc. · Aug 16, 2004
K050916 — SAFE-CROSS RADIO FREQUENCY TOTAL OCCLUSION CROSSING SYSTEM [PERIPHERAL] · Intra Luminal Therapeutics, Inc. · Jun 8, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
K051670
## MAR 3 0 2006
## 1. 510(k) Summary
### Submitter Information
- A. Company Name: Baylis Medical Company Inc.
- B. Company Address: 2580 Matheson Blvd. E. Mississauga, Ontario L4W 4J1 Canada
- C. Company Phone: (905) 602-4875 ext 252
- D. Company Facsimile: (905) 602-5671
- E. Contact Person: Meghal Khakhar Regulatory Affairs Manager
- F. Summary Prepared on: 28-Mar-2006
#### Device Identification
- A. Device Trade Name: RF Tunneler Wire
- B. Device Common Name: Tunneler Wire
- C. Classification Name: Wire, guide, catheter
- D. Device Class: Class II (per 21 CFR 870.1330)
- E. Device Code: DQX
#### Identification of Predicate Devices
Predicate devices are the Baylis Medical RF Perforation Probe (Baylis Medical Company Inc.), which is cleared under 510(k) Premarket Notification number K010265; the Safe-Cross Radio Frequency Total Occlusion Crossing System (IntraLuminal Therapeutics Inc.), which is cleared under 510(k) Premarket Notification number K031842; and the CLiRpath Excimer Laser Catheter (Spectranetics Corp.), which is cleared under 510(k) Premarket Notification number K040067.
{1}------------------------------------------------
## Device Description
The RF Tunneler Wire consists of a core wire surrounded with a polymer insulation. The wire connects to a perforation generator at the proximal end via a connector cable, and has an active tip at the distal end to delivery RF energy.
The RF Tunneler Wire is designed to be compatible with most balloon and stent catheters approved for use in peripheral interventional procedures.
### Intended Use
The RF Tunneler Wire is intended to create a channel in totally occluded peripheral vessels 3 mm or greater.
### Comparison to Predicate Devices
The RF Tunneler Wire is similar to the Baylis Medical RF Perforation Probe in design and function. The intended use of the RF Tunneler Wire is substantially equivalent to the Safe-Cross Radio Frequency Total Occlusion Crossing System and the CLiRpath Excimer Laser Catheter in intended use.
The RF Tunneler Wire and the predicate devices are all wires used in interventional procedures. Thev all focus energy at the active tip to create channels in target tissue.
### Biocompatibility. Sterilization. Packaging, and Bench Testing
Biocompatibility of the materials used in the RF Tunneler Wire has been tested and confirmed to verify compliance with safety requirements. Sterilization of the RF Tunneler Wire has been validated. Packaging of the RF Tunneler Wire has been validated.
The RF Tunneler Wire has undergone mechanical and electrical bench testing to verify compliance with safety and performance requirements. These tests include comparing the RF Tunneler Wire to a predicate device.
#### Conclusion
Testing of the RF Tunneler Wire verified the safety and performance characteristics of this device for its intended use. The functions of the RF Tunneler Wire, as well as its performance characteristics as verified in bench testing are substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Baylis Medical Company c/o Meghal Khakhar, MBBS, CerRAP, RAC Regulatory Affairs Manager 2580 Matheson Blvd. E Mississauga, ON L4W 4J1 CANADA
SEP 1 8 2013
Re: K051670
Trade/Device Name: Baylis RF Tunneler Wire Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: PDU Dated: March 22, 2006 Received: March 24, 2006
Dear Ms. Khakhar:
This letter corrects our substantially equivalent letter of March 30, 2006.
We have reviewed your Section 510(k) premarket notification of intent to market the device we nate roviewed your been mined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreations of the enactment date of the Medical Device Amendments or to conninered phortes inly 2011/09/2017 accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of I he general ocharing practice, labeling, and prohibitions against misbranding and adviteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I Dr hose intatutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Z.M.Z.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K051670
Baylis RF Tunneler Wire Device Name:
Indications For Use:
The RF Tunneler Wire is intended to create a channel in totally occluded peripheral vessels 3 imm or greater.
× Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
R. Jachner
(Division Sign-Off)
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number_Ko516 70
Page 1 of _1_
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.