FILTERWIRE EZ EMBOLIC PROTECTION SYSTEM (2.25 MM-3.5 MM)

K061332 · Boston Scientific Corp · NFA · Aug 11, 2006 · Cardiovascular

Device Facts

Record IDK061332
Device NameFILTERWIRE EZ EMBOLIC PROTECTION SYSTEM (2.25 MM-3.5 MM)
ApplicantBoston Scientific Corp
Product CodeNFA · Cardiovascular
Decision DateAug 11, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Intended Use

The FilterWire EZ Embolic Protection System is indicated for use as a guide wire and embolic protection system to contain and remove embolic material (thrombus/debris) while performing angioplasty and stenting procedures in coronary saphenous vein bypass grafts. The diameter of the artery at the site of filter loop placement should be between 2.25 mm and 5.5 mm. The safety and effectiveness of this device as an embolic protection system has not been established in cerebral, carotid and peripheral vasculature or in treating native coronaries, including acute myocardial infarction.

Device Story

FilterWire EZ Embolic Protection System functions as a guide wire and embolic filter during coronary saphenous vein bypass graft interventions. System components include a 0.014-inch protection wire with an integrated circular/oval filter loop assembly, a delivery sheath for filter placement, and a retrieval sheath for filter removal. The filter membrane captures embolic material while maintaining blood flow. Used in clinical settings by physicians; the device is tracked to the target site, deployed to capture debris, and retrieved post-procedure. The system benefits patients by reducing the risk of distal embolization during stenting/angioplasty. The device is a modification of the existing FilterWire EZ family, extending the compatible vessel diameter range.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical performance data, including in-vitro and in-vivo testing conducted in conformance with 21 CFR Part 58 (GLP). Testing confirmed the device meets design-input criteria for its intended use.

Technological Characteristics

Single operator exchange delivery system; filter-based technology. Components: Nitinol filter/basket, Bionate polyurethane filter membrane. Compatible with 0.014-inch guide wires and 6F guide catheters. Features radiopaque guide wire tips and radiopaque markers on the filter loop. Available in 190 cm and 300 cm lengths. Mechanical device; no software or energy source.

Indications for Use

Indicated for patients undergoing angioplasty and stenting in coronary saphenous vein bypass grafts (vessel diameter 2.25-5.5 mm) to contain/remove embolic debris. Contraindicated for cerebral, carotid, peripheral vasculature, and native coronary arteries (including acute myocardial infarction).

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K061332 AUG 11 2006 # Section 5 # 510(k) Summary # "510(k) Summary" as required by section 807.92(c) #### 5.1 General Provisions | Submitter's Name<br>and Address | Boston Scientific Corporation<br>2011 Stierlin Court<br>Mountain View, CA 94043-4655, U.S.A. | |---------------------------------|-------------------------------------------------------------------------------------------------------| | Contact Person | Debbie McIntire<br>Senior Regulatory Affairs Specialist<br>Tel: (650) 623-1703<br>Fax: (650) 623-1610 | | Classification Name | 21CFR870.1250, Percutaneous Catheter | | Common or Usual Name | Distal embolic protection guide wire | | Proprietary Name | Boston Scientific FilterWire EZ Embolic Protection<br>System | | Manufacturing Facilities | Boston Scientific Corporation<br>2011 Stierlin Court<br>Mountain View, CA 94043-4655, U.S.A. | ### 5.2 Name of Predicate Device The Boston Scientific FilterWire EZ Embolic Protection System (2.25 mm - 3.5 mm) is substantially equivalent in design and intended use to the FilterWire EZ Embolic Protection System (2.25 mm - 3.5 mm) cleared under K051984. ### 5.3 Device Description The subject FilterWire EZ Embolic Protection System (2.25 mm - 3.5 mm) consists of the same four main components as the FilterWire EZ Embolic Protection System (2.25 mm - 3.5 mm) cleared under K051984. These components are: {1}------------------------------------------------ - · A Protection Wire (0.014 in. [0.36 mm] guide wire and integrated filter loop) that incorporates a circular/oval-shaped structure (filter loop assembly), which supports a thin, porous, filter membrane designed to contain and remove embolic material while maintaining blood flow. - · An EZ Delivery Sheath, a low-profile sheath that constrains the filter on the protection wire for delivery to the site of filter loop placement. It is prepackaged with the protection wire for coaxial delivery and has a slit for easy removal in a peel-away fashion. - · An EZ Retrieval Sheath is intended to negotiate through the anatomy over the protection wire to retrieve the filter. The sheath is designed to track easily to the filter without catching on obstacles such as stents. The sheath has a radiopaque marker and is silicone coated. - · A tool kit containing one peel-away introducer, one wire torquer and one hemostasis valve dilator ### 5.4 Intended Use The FilterWire EZ Embolic Protection System is indicated for use as a guide wire and embolic protection system to contain and remove embolic material (thrombus/debris) while performing angioplasty and stenting procedures in coronary saphenous vein bypass grafts. The diameter of the artery at the site of filter loop placement should be between 2.25 mm and 5.5 mm. The safety and effectiveness of this device as an embolic protection system has not been established in cerebral, carotid and peripheral vasculature or in treating native coronaries, including acute myocardial infarction. ### ર્સ્ડ Summary of Technological Characteristics The Filter Wire EZ Embolic Protection System (2.25 mm - 3.5 mm) indicated for use in saphenous vein bypass grafts is equivalent to the Filter Wire EZ Embolic Protection System (2.25 mm - 3.5 mm) cleared under K051984 in that the intended use, principles of operation, design and materials are similar. The following design attributes are the same or similar for the subject device and the predicate device: ## Design Attributes Single Operator Exchange delivery systems Filter-based technology Nitinol® filter/basket component Bionate™ polyurethane filter membrane <sup>&</sup>quot;The FilterWire EZ Embolic Protection System (2.25 mm - 3.5 mm) is an extension of the FilterWire EZ Embolic Protection System family (sizes 3.5 mm - 5.5 mm) cleared under K032884 and K052880; the indications for use in the combined directions for use (for the FilterWire EZ family) will reference the entire vessel diameter range of 2.25 mm to 5.5 mm. {2}------------------------------------------------ Compatibility with .014 in. guide wires Compatibility with 6F guide catheters Available in 2 wire lengths; 190 cm and 300 cm Radiopaque guide wire tips Radiopaque markers on filter loop #### Non-Clinical Performance Data ર્સ્ડ Where appropriate, testing conformed to the requirements of 21 CFR Part 58, Good Laboratory Practices (GLP). Specifically, non-clinical tests conducted for the FilterWire EZ Embolic Protection System (2.25 mm - 3.5 mm) showed the device met its design-input criteria, and is safe and effective for its intended use. #### 5.7 Statement of Substantial Equivalence Based on a comparison of design, fundamental technology, intended use, results of in-vitro testing, in-vivo testing, biocompatibility, and packaging, Boston Scientific considers the FilterWire EZ Embolic Protection System (2.25 mm - 3.5 mm) to be substantially equivalent to the FilterWire EZ Embolic Protection System (2.25 mm - 3.5 mm) (K051984). {3}------------------------------------------------ Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like symbol with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 1 1 2006 Boston Scientific Corporation C/O Ms. Debbie McIntire Senior Regulatory Affairs Specialist 2011 Stierlin Court Mountain View, CA 94043-4655 Re: K061332 > Trade/Device Name: FilterWire EZ Embolic Protection System (2.25mm - 3.5mm) Regulation Number: 21 CFR 870.1250 Regulation Name: Distal Embolic Protection Guidewire Regulatory Class: Class II Product Code: NFA Dated: August 4, 2006 Received: August 4, 2006 Dear Ms. McIntire: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ # Page 2 – Ms. Debbie McIntire Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease oc advised that I Dr is lossants over device complies with other requirements of the Act that i DA has made a develations administered by other Federal agencies. You must bir any I cuttlar statures und regulations and limited to: registration and listing (21 Comply with an the 11ct 821 CFR Part 801); good manufacturing practice requirements as set CI K ratt 607); adoning (21 CFR Part 820); and if applicable, the electronic forth in the quant) by events (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Section (Section of the as described in your Section 510(k) This letter will and will your of substantial equivalence of your device to a legally premailer notificated. " ceresults in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n you don't e ffice of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general mional and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Special 510(k): Device Modification FilterWire EZ™ Embolic Protection System (2.25 mm - 3.5 mm) ### Section 3 Indications For Use FilterWire EZ™ Embolic Protection System Device Name: Indications for Use: The indications for use for the FilterWire EZ Embolic Protection System are as follows: The FilterWire EZ Embolic Protection System is indicated for use as a guide wire and embolic protection system to contain and remove embolic material (thrombus/debris) while performing angioplasty and stenting procedures in coronary saphenous vein bypass grafts. The diameter of the artery at the site of filter loop placement should be between 2.25 mm and 5.5 mm. The safety and effectiveness of this device as an embolic protection system has not been established in cerebral, carotid and peripheral vasculature or in treating native coronaries, including acute myocardial infarction. | Prescription Use | X | |----------------------------|---| | (Per 21 CFR 801 Subpart D) | | AND/OR Over-The-Counter Use_ (Per 21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRH Office of Device Evaluation (ODE) | | |-------------------------------------------------------|---------| | (Division Sign-Off) | | | Division of Cardiovascular Devices | | | 510(k) Number | K061332 |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...