CARESCAPE MONITOR B850

K131414 · GE Healthcare Finland Oy · MHX · Aug 28, 2013 · Cardiovascular

Device Facts

Record IDK131414
Device NameCARESCAPE MONITOR B850
ApplicantGE Healthcare Finland Oy
Product CodeMHX · Cardiovascular
Decision DateAug 28, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1025
Device ClassClass 2
AttributesPediatric

Indications for Use

The CARESCAPE Monitor B850 is a multi-parameter high acuity patient monitor intended for use in multiple areas within a professional healthcare facility. The CARESCAPE Monitor B850 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE Monitor B850 is indicated for monitoring of: hemodynamic (including ECG, ST segment, arrhythmia detection, ECG diagnostic analysis and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, cardiac output (thermodilution and pulse contour), temperature, mixed venous oxygen saturation, and central venous oxygen saturation). respiratory (impedance respiration, airway gases (CO2, O2, N2O and anesthetic agents), spirometry, gas exchange) and neurophysiological status (including electroencephalography, Entropy, Bispectral Index (BIS) and neuromuscular transmission. The CARESCAPE Monitor B850 also provides alarms, trends, snapshots and events, and calculations and can be connected to displays, printers and recording devices. The CARESCAPE Monitor B850 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for remote viewing and to data management software devices via a network. The CARESCAPE Monitor B850 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility. The CARESCAPE Monitor B850 is not intended for use during MRI.

Device Story

High-acuity multi-parameter patient monitor; processes physiological signals including ECG, invasive/non-invasive pressure, pulse oximetry, cardiac output, temperature, gas exchange, EEG, and neuromuscular transmission. System comprises host processing unit (ESP V2 software), display, keyboard, and parameter measurement modules. Used in professional healthcare facilities by trained clinicians; provides real-time monitoring, alarms, trends, and diagnostic analysis. Facilitates clinical decision-making through continuous patient status visualization and data integration. Supports standalone operation or network connectivity for remote viewing and data management. Benefits patients through continuous, high-acuity physiological surveillance.

Clinical Evidence

No clinical data required; substantial equivalence supported by bench testing, including module verification, system integration, safety, and performance testing against recognized standards (IEC 60601 series, AAMI EC11/13/57/SP10, ISO 9919/21647).

Technological Characteristics

Multi-parameter monitor; host processing unit with ESP V2 software. Connectivity via network for remote viewing/data management. Standards compliance: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 62304 (software lifecycle), IEC 62366 (usability). Features EK-Pro v13 arrhythmia detection algorithm. Modular design allows various parameter measurement inputs.

Indications for Use

Indicated for monitoring hemodynamic, respiratory, and neurophysiological status in adult, pediatric, and neonatal patients in professional healthcare facilities. Contraindicated for use during MRI.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K131414 Page 1 of 5 GE Healthcare Finland Oy Kuortaneenkatu 2, P.O. Box 900 FI-00031 GE Finland T: +358 10 39411 F: +358 9 1463310 # 510(k) Summary In accordance with 21 CFR 807.92 the following summary of information is provided: ## Owner/Contact/Date (807.92(a)(1): | Date:<br>Owner/Submitter: | August 20, 2013<br>GE Healthcare Finland Oy.<br>Kuortaneenkatu 2<br>00510 Helsinki<br>FINLAND<br>Phone: +358 10 39411<br>Fax: +358 9 1463310 | AUG 2 8 2013 | |------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------| | Primary Contact Person: | Joel Kent<br>Manager, Quality and Regulatory Affairs<br>GE Healthcare<br>Phone: 781-449-8685<br>Fax: 781-433-1344<br>E-mail: joel.kent@ge.com | | | Secondary Contact Person: | Rauno Ruoho<br>Regulatory Affairs Manager<br>GE Healthcare Finland Oy<br>Kuortaneenkatu 2<br>00510 Helsinki<br>Finland<br>Phone: + 358 10 394 3624<br>Fax: +358-92726532<br>E-mail: rauno.ruoho@ge.com | | | Device names (807.92(a)(2)): | | | | Trade Name:<br>Common/Usual Name:<br>Classification Names: | CARESCAPE Monitor B850<br>multi-parameter patient monitor<br>21 CFR 870.1025 Arrhythmia detector and alarm (including ST-segment<br>measurement and alarm) | | | Primary Product Code:<br>Secondary Product Codes: | MHX<br>BZK, BZL, BZQ, CAP, CBQ, CBR, CBS, CCK, CCL, DPS, DPZ, DQA,<br>DQK, DRT, DSI, DSJ, DSK, DXG, DXN, FLL, GWJ, GWQ, KOI, KRB,<br>MLD, NHO, NHP, NHQ, OLT, OLW, OMC, ORT | | Image /page/0/Picture/7 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a circular frame. The letters and the frame are in black, contrasting with the white background. The logo is a well-known symbol associated with the General Electric company. {1}------------------------------------------------ Predicate Device(s) (807.92(a)(3):: Device Description (807.92(a)(4)):: #### K092027 CARESCAPE Monitor B850 The CARESCAPE Monitor B850 includes both new and existing subsystems interconnected to form a high acuity, multi-parameter patient monitoring system. A typical configuration would be a CARESCAPE Monitor B850 host processing unit running the CARESCAPE Monitoring platform software (in this submission the ESP V2 software) , a display with integrated keyboard and a frame for the insertion of parameter measurement modules. Many times a secondary display with keyboard is also used. The CARESCAPE Monitor B850 is designed to give the user maximum flexibility to choose the particular options for the particular care environment at the healthcare facility. A variety of options are available to the customer including additional displays, various input devices (keyboard, mouse, bar code reader), additional modules and frames and software options. Some of these major subsystems include non-patient contact accessory items (e.g. cables and mounting hardware). The CARESCAPE Monitor B850 interfaces to a variety of existing physiological parameter measurement modules. In addition, the CARESCAPE Monitor B850 interfaces to a variety of existing OEM medical devices via the existing network infrastructure. There are also various care area specific main software options and software licenses for specific features the customer can choose from for the CARESCAPE Monitor B850. There are also various upgrade programs offered to existing customers via the CARESCAPE Life Upgrade Programs to provide new CARESCAPE Monitor B850 hardware and/or software to the customers, but offer the benefit that the customers can continue to use the majority of their parameter measurement modules. {2}------------------------------------------------ K131414 Page 3 of 5 #### Intended Use: (807.92(a)(5): The indications for use are: The CARESCAPE Monitor B850 is a multi-parameter high acuity patient monitor intended for use in multiple areas within a professional healthcare facility. The CARESCAPE Monitor B850 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time The CARESCAPE Monitor B850 is indicated for monitoring of: · hemodynamic (including ECG, ST segment, arrhythmia detection, ECG diagnostic analysis and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, cardiac output (thermodilution and pulse contour), temperature, mixed venous oxygen saturation, and central venous oxygen saturation). · respiratory (impedance respiration, airway gases (CO2, O2, N2O and anesthetic agents), spirometry, gas exchange) and · neurophysiological status (including electroencephalography, Entropy, Bispectral Index (BIS) and neuromuscular transmission. The CARESCAPE Monitor B850 also provides alarms, trends, snapshots and events, and calculations and can be connected to displays, printers and recording devices. The CARESCAPE Monitor B850 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for remote viewing and to data management software devices via a network. The CARESCAPE Monitor B850 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility, The CARESCAPE Monitor B850 is not intended for use during MRI. The CARESCAPE Monitor B850 with ESP V2 software is a new version of an existing monitor, introducing new features and improvements to existing features. Hardware used is essentially the same hardware as in the predicate. The predicate device used in this comparison is the CARESCAPE Monitor B850 with ESP VI software (K092027). The CARESCAPE Monitor B850 has the EK-Pro Arrhythmia Detection Algorithm (EK-Pro v13) that employs the same functional technology as the predicate device in the monitoring of ECG parameter data. The predicate arrhythmia algorithms are EK-Pro arrhythmia detection algorithm (EK-Pro v11, K031320) and Datex-Ohmeda bedside arrhythmia detection algorithm, both used with the predicate CARESCAPE Monitor B850 (K092027). #### Summary of Non-Clinical Tests: The CARESCAPE Monitor B850 and its applications comply with voluntary standards as detailed below. The following quality assurance measures were applied to the development of the system: - � Risk Analysis - . Requirements Reviews - . Design Reviews - . Testing on unit level (Module verification) Technology (807.92(a)(6)):: Determination of Substantial Equivalence (807.92(b)(1)):: {3}------------------------------------------------ - . Integration testing (System verification) - . Final acceptance testing (Validation) - . Performance testing (Verification) - . Safety testing (Verification) The CARESCAPE Monitor B850 was designed and tested for compliance to the following standards: IEC 60601-1:1988, A1:1991, A2:1995, Corr1:1995, Medical 1. Electrical Equipment Part 1: General Requirements for Safety - Second Edition 2. IEC 60601-1-1:2000, Medical Electrical Equipment - Part 1-1: General Requirements for Safety - Collateral Standard: Safety Requirements for Medical Electrical Systems - Edition 2.0 IEC 60601-1-2:2001 + A1:2004, Medical electrical equipment - Part 3. 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests - Edition 2.1 4. IEC 60601-1-4:1996 + A1:1999 (AKA ed 1.1:2000), Medical electrical equipment - Part 1: General requirements for safety - 4 - Collateral standard: Programmable electrical medical systems, Edition 1.1 IEC60601-1-6:2006, Medical electrical equipment - Part 1-6: న్. General requirements for basic safety and essential performance - collateral Standard: Usability - Edition 2 IEC 60601-1-8:2006, Medical electrical equipment - Part 1-8: ર્ણ રાજ્યના વિદ્યાર તાલુકામાં આવેલું એક ગામનાં મુખ્યત્વે આ ગામનાં મુખ્યત્વે આ ગામનાં મુખ્યત્વે આ ગામનાં મુખ્યત્વે આ ગામનાં મુખ્યત્વે આ ગામનાં મુખ્યત્વે આ ગામનાં મુખ્યત્વે આ General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests, and guidance for alarm systems in medical electrical equipment and medical electrical systems, Second Edition 7. IEC 60601-2-10:1987 + A1:2001, Medical Electrical Equipment Part 2: Particular Requirements for the Safety of Nerve and Muscle Stimulators - First Edition 8. IEC 60601-2-25:1993 + A1:1999, Medical Electrical Equipment Part 2: Particular requirements for the safety of electrocardiographs - First edition ல். IEC 60601-2-26:2002, Medical electrical equipment - Particular requirements for the safety of electroencephalographs 10. IEC 60601-2-27:2005, Medical electrical equipment Part 2-27: Particular requirements for the safety, including essential performance, of electrocardiographic monitoring equipment IEC 60601-2-30:1999, Medical electrical equipment - Part 2-30: 11. Particular requirements for the safety, including essential performance, of automatic cycling non-invasive blood pressure monitoring equipment-Second Edition IEC 60601-2-34:2000, Medical electrical equipment - Part 2-34: l 2. Particular requirements for the safety, including essential performance, of invasive blood pressure monitoring equipment-Edition 2 IEC 60601-2-40:1998, Medical electrical equipment - Particular 13. requirements for the safety of electromyographs and evoked response equipment 14. IEC 60601-2-49:2001, Medical electrical equipment - Part 2-49: Particular requirements for the safety of multifunction patient monitoring equipment-Edition 1 lEC 60601-2-51:2003, Medical electrical equipment Part 2-51: 15. Particular requirements for safety, including essential performance, of {4}------------------------------------------------ # K131414 ## Page 5 of 5 recording and analyzing single channel and multichannel electrocardiographs-Edition 1 AAMI EC11:1991/(R)2001/(R)2007, Diagnostic 16. Electrocardiographic Devices 17. AAMI EC13: 2002/(R)2007, Cardiac monitors, heart rate meters, and alarms, AAMI EC-57:1998, A1:2003, Testing and reporting performance 18. results of cardiac rhythm and ST segment measurement algorithms AAMI SP10:2002 + A1:2003 + A2:2006, Manual, electronic, or 19. automated sphygmomanometers 20. EN1041:2008, Information supplied by the manufacturer with medical devices 21. EN1060-1:1995 +A1:2002, Non-invasive sphygmomanometers- Part 1: General requirements EN1060-3:1997 +A1:2005, Non-invasive sphygmomanometers- Part 22. 3: Supplementary requirements for electro-mechanical blood pressure measuring systems Except for the following clause: o 7.9 for PDM module: Testing performed in accordance with EN 1060-4 EN 12470-4:2000, A1:2009, Clinical Thermometers - Part 4: 23. Performance of Electrical Thermometers for Continuous Measurement Except for the following clauses: 6.3 b) Temperature measurement error with single use probes 0 exceeded maximum permissible error. 6.4: The response time of the Esophageal stethoscope with O temperature probe exceeds 150s for the probe sizes 18F and 24F. 24. ISO 21647:2004 + C1:2005, Medical electrical equipment -Particular requirements for the basic safety and essential performance of respiratory gas monitors ISO9919:2005, Medical electrical equipment Particular requirements 25. for the safety and essential performance of pulse oximeter equipment for medical use - Second Edition 26. IEC62304:2006, Medical device software - Software life cycle processes IEC62366:2007, Medical Devices - Application of usability 27. engineering to medical devices (General) Clinical (807.92(b)(2)): Summary of Clinical Tests: The subject of this premarket submission, CARESCAPE Monitor B850 did not require clinical studies to support substantial equivalence. Conclusion (807.92(b)(3)):: GE Healthcare considers the CARESCAPE Monitor B850 to be as safe, as effective, and performance is substantially equivalent to the predicate device. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized caduceus, a symbol often associated with healthcare, with a staff entwined by a serpent. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 August 28, 2013 GE Healthcare Finland Oy Mr. Joel Kent Manager, Quality and Regulatory Affairs Kuortaneenkatu 2 Helsinki, FIN-00510 FI Re: K131414 Trade/Device Name: Carescape Monitor B850 Regulation Number: 21 CFR 870.1025 Regulation Name: Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms) Regulatory Class: Class II Product Code: MHX, BZK, BZL, BZQ, CAP, CBQ, CBR, CBS, CCK, CCL, DPS, DPZ, DQA, DQK, DRT, DSI, DSJ, DSK, DXG, DXN, FLL, GWJ, GWQ, KOI, KRB, MLD, NHO, NHP, NHQ, OLT, OLW, OMC, ORT Dated: August 8, 2013 Received: August 9, 2013 Dear Mr. Joel Kent: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. {6}------------------------------------------------ If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Owen P. Faris -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ ## Indications for Use 510(k) Number (if known): Device Name: CARESCAPE Monitor B850 Indications for use: The CARESCAPE Monitor B850 is a multi-parameter high acuity patient monitor intended for use in multiple areas within a professional healthcare facility. The CARESCAPE Monitor B850 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time The CARESCAPE Monitor B850 is indicated for monitoring of: · hemodynamic (including ECG, ST segment, arrhythmia detection, ECG diagnostic analysis and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, cardiac output (thermodilution and pulse contour), temperature, mixed venous oxygen saturation, and central venous oxygen saturation). · respiratory (impedance respiration, airway gases (CO2, O2, N2O and anesthetic agents), spirometry, gas exchange) and · neurophysiological status (including electroencephalography, Entropy, Bispectral Index (BIS) and neuromuscular transmission. The CARESCAPE Monitor B850 also provides alarms, trends, snapshots and events, and calculations and can be connected to displays, printers and recording devices. The CARESCAPE Monitor B850 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for remote viewing and to data management software devices via a network. The CARESCAPE Monitor B850 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility. The CARESCAPE Monitor B850 is not intended for use during MRI. Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEFDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page *_ of _* Digitally signed by Digitally signed by Owen P. Faris -S Date: 2013.08.28 15:18:36-04'00'
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...