MORTARA SURVEYOR PATIENT MONITOR

K123556 · Mortara Instrument, Inc. · MHX · Mar 22, 2013 · Cardiovascular

Device Facts

Record IDK123556
Device NameMORTARA SURVEYOR PATIENT MONITOR
ApplicantMortara Instrument, Inc.
Product CodeMHX · Cardiovascular
Decision DateMar 22, 2013
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 870.1025
Device ClassClass 2
AttributesPediatric

Intended Use

The Mortara Surveyor Patient Monitor is a prescription device intended to be used by knowledgeable healthcare professionals within a healthcare facility. It is designed for continuous monitoring in either stationary or portable applications. It is intended to be used by clinicians and medical qualified personnel for monitoring ECG, Respiration, NIBP, Temperature, SPO2, Invasive Blood Pressure, End-Tidal & Inspired CO2, 12-lead resting ECG and cardiac output.

Device Story

Mortara Surveyor Patient Monitor (S12/S19) is a portable, multi-parameter physiological monitor for use by clinicians in healthcare facilities. It accepts inputs from various sensors for ECG, respiration, NIBP, temperature, SpO2, invasive blood pressure, CO2, and cardiac output. The device processes these signals to provide real-time patient monitoring and arrhythmia/ST-segment detection. Data can be viewed on 11.6" or 18.5" displays and transmitted via secure Ethernet protocol to a Surveyor Central Station for remote monitoring. The system assists clinicians in continuous patient assessment, enabling timely clinical decision-making through visual and alarm-based feedback. It is intended for both stationary and portable clinical applications.

Clinical Evidence

No clinical data was performed or deemed necessary to demonstrate safety and effectiveness. Evidence is based on bench testing, including performance, functional, environmental, and vibration/shock testing, alongside electrical safety and EMC verification.

Technological Characteristics

Portable patient monitor with 11.6" or 18.5" display. Parameters: ECG, NIBP, SpO2, CO2, IBP, Temp, Respiration, Cardiac Output. Connectivity: Ethernet for central station integration. Software-based monitoring and alarm system. Not sterilizable. Biocompatible patient-contacting materials.

Indications for Use

Indicated for adult and pediatric patient populations in healthcare facilities for monitoring non-invasive blood pressure, impedance respiration, invasive blood pressure, temperature, SpO2, end-tidal/inspired CO2, ECG with arrhythmia and ST-segment analysis, 12-lead resting ECG, and cardiac output.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a heart shape inside of a square. The heart shape is white, and the square is gray. The heart shape is in the center of the square. The heart shape is slightly smaller than the square. K123556 Page 1 of 3 : · 100 - 100 - 100 ### Abbreviated 510(k) Notification . . - : # MAR 2 2 2013 #### 510(k): Surveyor S12 and S19 Device Summary . | Submitter: | Chien Hui (Amy) Yang, Regulatory Affairs Engineer<br>Mortara Instrument, Inc.<br>7865 N. 86th Street<br>Milwaukee, WI 53224 | |------------|-----------------------------------------------------------------------------------------------------------------------------| | Date: | November 6, 2012 | | FAX: | (414) 354-4760 | | Phone: | (414) 354-1600 | | Contact: | Chien Hui (Amy) Yang (see above) | | Trade Name: | Mortara Surveyor Patient Monitor | |----------------------------|----------------------------------------------------------------------| | Common Name: | Patient Physiological Monitor (with Arrhythmia Detection or Alarms) | | Classification Name: | Monitor, Physiological Patient (with Arrhythmia Detection or Alarms) | | Classification Regulation: | 21 CFR §870.1025 | | Product Code: | MHX | Legally marketed devices to which S.E. is claimed: | | Predicate 510(k) Number | Predicate Manufacturer / Model | |---------------------------------------------------------------------------|-------------------------|------------------------------------------------------| | Mortara Surveyor Patient Monitor<br>Non-Invasive Blood Pressure component | K090556 | Spacelabs Healthcare / élance Vital<br>Signs Monitor | | Impedance Respiration component | | | | Invasive Blood Pressure component | | | | Temperature component | | | | Functional Arterial Oxygen<br>Saturation component | | | | End-Tidal & Inspired CO2 | | | #### End-Tidal & Inspired CO2 component | component | | | |---------------------|---------|------------------------------------------------------------------| | ECG Monitoring | K090556 | Spacelabs Healthcare / élance Vita<br>Signs Monitor | | | K060135 | Mortara Instrument, Inc. / Surveyor<br>Telemetry Central Station | | 12-Lead Resting ECG | K100127 | Mortara Instrument, Inc. / ELI 230<br>Electrocardiograph | | Cardiac Output | K012226 | Schiller / ARGUS PB-1000<br>Monitoring System | # © 2012 Mortara Instrument, Inc. © 2012 Mortara Instrument, Inc. . {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows a heart shape inside of a rectangle. The heart shape is white, and the background is gray. The heart shape is not completely enclosed, as the bottom of the heart is open. K123556 Page 2 of 3 #### Abbreviated 510(k) Notification #### Description: The Mortara Instrument, Inc. Surveyor Patient Monitoring System (S12 and S19) is a portable patient monitors intended to be used by clinicians and medical qualified personnel for monitoring ECG, Respiration, NIBP, Temperature, SPO2, Invasive Blood Pressure, End-Tidal & Inspired CO2, 12-lead resting ECG and cardiac output. Models within the Mortara Surveyor family come in two different sized viewing areas (11.6" and 18.5") and offer selected monitoring features. See the Product Specification for identified models and monitoring features. #### Central Station: A secure Mortara protocol over an Ethernet connection allows the Mortara Surveyor Patient Monitor to exchange patient information with the Mortara Surveyor Central Station. This allows monitoring of the Surveyor \$12/S19 at a central workstation. #### Technology Comparison: The Mortara Surveyor utilizes the same or similar technology for each parameter as utilized by the predicate devices. #### Intended Use: The Mortara Surveyor Patient Monitor is a prescription device intended to be used by knowledgeable healthcare professionals within a healthçare facility. It is designed for continuous monitoring in either stationary or portable applications. It is intended to be used by clinicians and medical qualified personnel for monitoring ECG, Respiration, NIBP, Temperature, SPO2, Invasive Blood Pressure, End-Tidal & Inspired CO2, 12-lead resting ECG and cardiac output. #### Indications for Use: The Mortara Surveyor Patient Monitor is indicated for use in adult & pediatric patient populations. The Mortara Surveyor Patient Monitor facilitates the monitoring of: - Non-invasive blood pressure . - Impedance respiration - Invasive blood pressure - Temperature ● - Functional arterial oxygen saturation (SpO2) - End-tidal & inspired CO2 - ECG monitoring with arrhythmia & ST-segment - 12-Lead resting ECG - Cardiac output The Mortara Surveyor Patient Monitor is a prescription device intended to be used by healthcare professionals in all areas of a healthcare facility. #### Performance Testing: #### Sterilization Validation: The Mortara Surveyor Patient Monitor is not sterilizable, and therefore this section does not apply to the monitor itself. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows a white heart shape cut out of a dark square. The heart shape has a pointed bottom and two rounded humps at the top. The square is dark and has a rough texture. The heart shape is centered in the square. 3 of 3 #### Abbreviated 510(k) Notification #### Shelf Life Testing: The Mortara Surveyor Patient Monitor is not sterilizable, and therefore this section does not apply to the monitor itself. #### Biocompatibility Testing: The S12 and S19 patient contacting components use the same materials used on other products that Mortara has used historically and/or has been cleared as substantially equivalent previously. #### Software Testing: Software for the Mortara Surveyor Patient Monitor was designed according to a robust software development process, and was rigorously verified and validated. Test results indicated that the Mortara Surveyor Patient Monitor complies with its predetermined specification. #### Electrical Safety: The Mortara Surveyor Patient Monitor was tested for patient safety in accordance with applicable Standards. Test results indicated that the Mortara Surveyor Patient Monitor complies with its predetermined specification. #### Electromagnetic Compatibility Testing: The Mortara Surveyor Patient Monitor was tested for EMC in accordance with applicable Standards. Test results indicated that the Mortara Surveyor Patient Monitor complies with its predetermined specification. #### Performance Testing - Bench: The Mortara Surveyor Patient Monitor was tested in accordance with internal requirements and procedures, and test results indicated that the device complies with the predetermined requirements. This testing includes performance and functional, environmental, and vibration/ shock testing. #### Performance Testing - Animal: Animal performance testing was not performed and is not necessary to demonstrate safety and effectiveness of the Mortara Surveyor Patient Monitor. #### Performance Testing - Clinical: Clinical performance testing was not performed and is not necessary to demonstrate safety and effectiveness of the Mortara Surveyor Patient Monitor. #### Conclusion: Based upon a comparison of devices and performance testing results, the Mortara Surveyor Patient Monitor is substantially equivalent to the predicate device. © 2012 Mortara Instrument, Inc. . . . . . . {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an eagle or bird in flight, composed of three curved lines. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 22, 2013 Mortara Instrument, Inc. c/o Ms. Amy Yang 7865 North 86th St. Milwaukee. WI 53224 , Re: K123556 Trade/Device Name: Surveyor Patient Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: Class II (two) Product Code: MHX Dated: February 27, 2013 Received: February 28, 2013 Dear Ms. Yang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ #### Page 2 - Ms. Amy Yang Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Owen P. Faris -S Bram D. Zuckerman, M.D. for · . Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health #### Enclosure {5}------------------------------------------------ ## Indications for Use K123556 510(k) Number (if known): Mortara Surveyor Patient Monitor · Device Name: Indications for Use: The Mortara Surveyor Patient Monitor is indicated for use in adult & pediatric patient populations. The Mortara Surveyor Patient Monitor facilitates the monitoring of: - Non-invasive blood pressure ● - Impedance respiration . - Invasive blood pressure . - Temperature ● - Functional arterial oxygen saturation (SpO2) ● - End-tidal & inspired CO2 . - ECG monitoring with arrhythmia & ST-segment ● - 12-Lead resting ECG . - Cardiac output . The Mortara Surveyor Patient Monitor is a prescription device intended to be used by healthcare professionals in all areas of a healthcare facility. Prescription Use AND/OR × (21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Owen P. Faris -S 2013.03.22 10:42:26'-04'00' Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...