NETKONNECT REMOTE NETWORK EXTENSION

K112637 · Nihon Kohden · MHX · Dec 9, 2011 · Cardiovascular

Device Facts

Record IDK112637
Device NameNETKONNECT REMOTE NETWORK EXTENSION
ApplicantNihon Kohden
Product CodeMHX · Cardiovascular
Decision DateDec 9, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1025
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

The NetKonnect Remote Network Extension is intended to interface the Nihon Kohden monitoring network with networked client PCs to annunciate and display patient monitoring information to healthcare providers. The device is intended for use in real-time monitoring of routine patient status and alarm events. NetKonnect Remote Network Extension supplements the primary patient monitoring system by providing a forwarding mechanism for annunciating and displaying patient alarm events and event related information including vital signs values and waveforms. The system is intended for use by qualified medical personnel within a hospital or clinical environment. The stimulator is available for use on any patients as determined by the qualified medical personnel.

Device Story

NetKonnect is a software-only product that interfaces Nihon Kohden patient monitoring networks with networked client PCs. It automatically detects Nihon Kohden devices on the network and establishes communication to forward vital signs, waveforms, and alarm events. The software operates in the background without a user interface, except for administrator functions. Clinicians use client PCs to remotely view patient data and receive alarm notifications. It serves as a secondary notification system and does not replace primary bedside monitoring. The device benefits patients by extending the reach of monitoring information to clinicians inside or outside the hospital via network, dial-up, or internet connections, facilitating timely response to patient status changes.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Software-only product; runs on Microsoft Windows (Win32 service processes). Communicates via Ethernet using proprietary "NET9" protocol. Functions as a network interface/gateway for patient monitoring data. No physical accessories.

Indications for Use

Indicated for real-time monitoring of patient status and alarm events in hospital or clinical environments. Intended for use by qualified medical personnel on any patient population as determined by the clinician.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Nihon Kohden America, INC. 11/15/2011 # Special 510(k) Notification NetKonnect Remote Network Extension DEC - 9 2011 K112637 わける # 510(k) Summary #### 1.1. Intended Use The NetKonnect Remote Network Extension is interface the Nihon Kohden monitoring network with networked client PCs to annunciate and display patient monitoring information to healthcare providers. The device is intended for use in real-time monitoring of routine patient status and alarm events. NetKonnect Remote Network Extension supplements the primary patient monitoring system by providing a forwarding mechanism and displaying patient alarm events and events and event related information including vital signs values and waveforms. ## Device/Package Labels 1.2. The proposed product labels for the device are located in Attachment 13 . ### 1.3. Proposed Packaging Packaging is depicted in Attachment 13. - 1.4. Instructions for Use The proposed instructions for use (user manuals) are provided with each packaged device and are presented in Attachment 13. - Advertisement/Promotional Literature 1.5. To date no advertisement or promotional literature for the new has been created for distribution in the United States. ## 1.6. Contraindications, Precautions & Warnings Warnings and cautions are listed in the Operator's Manual as shown in Attachment 13. ## 1.7. Indications for Use Statement Indication for use statement is presented in Attachment 4. # DEVICE DESCRIPTION ## Photographs/Engineering Drawings 2.1. The proposed drawings are provided in Attachment 13. ### Physical Description 2.2. NetKonnect is a software product that runs on Microsoft .NetKonnect will communicate with Nihon Kohden patient monitoring devices by the direct network connection (NetKonnect-LT) or through the gateway (NetKonnect will allow hospitals to provide full featured remote overview access capability to its clinicians inside the hospital and outside of hospital by using the dial-up connection or Internet connection at minimal cost. #### 2.3. Principles of Operation Netkonnect has no user interface (except administrator functions) and performs all functions automatically without user intervention. When the network hosting NetKonnect is started, NetKonnect runs automatically and detects each of the Nihon Kohden devices on the patient monitoring network and establishes a vital signs and patient status communication connection with each one. Client PCs can communicate with NetKonnect to provide for remote monitoring of single or multiple patients. Remote monitoring function includes waveform and vital signs viewing, alarm notification and review of stored patient information. #### 2.4. Design Features {1}------------------------------------------------ K112637 17/3 # 2.5. Device Specifications The device specifications are provided in Attachment 16. ## 2.6. Accessories NetKonnect is a software only product, there are no accessories. - 2.7. Significant Changes that could affect Safety or Effectiveness or Intend Use. There are no significant changes as compared to the predicate devices that would affect the safety or effectiveness of the device as intended for use. | | Nihon Kohden<br>NetKonnect Extension<br>CGS-9003 | Nihon Kohden<br>CGS-9001 GWY<br>510k # K083271 | Comments | |-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | SPECIFICATION | | | | | Description of<br>communication<br>requirements | Receives and forwards<br>waveform data using<br>Nihon Kohden's<br>proprietary "NET9"<br>protocol language. | Same, but data must<br>be converted | Predicate device also converts<br>(translates) the data into a<br>language that is supported by the<br>3rd party application. We have<br>determined that this difference<br>does not increase the risk of<br>patient safety or affect the device<br>from performing in the manner in<br>which it is intended to operate<br>from the predicate device CGS-<br>9001 | | | Receives and forwards<br>Numerical data using<br>Nihon Kohden's<br>proprietary "NET9"<br>protocol language. | Same, but data must<br>be converted | Predicate device also converts<br>(translates) the data into a<br>language that is supported by the<br>3rd party application. We have<br>determined that this difference<br>does not increase the risk of<br>patient safety or affect the device<br>from performing in the manner in<br>which it is intended to operate<br>from the predicate device CGS-<br>9001 | | | Receives and forwards<br>Alarm notification to other<br>Nihon Kohden approved<br>devices such as Central<br>Nurse station using Nihon<br>Kohden's proprietary<br>"NET9" protocol<br>language.<br>When used with PC or<br>portable device, no alarm<br>notification is available | Same, but data must<br>be converted | Predicate device also converts<br>(translates) the data into a<br>language that is supported by the<br>3rd party application. We have<br>determined that this difference<br>does not increase the risk of<br>patient safety or affect the device<br>from performing in the manner in<br>which it is intended to operate<br>from the predicate device CGS-<br>9001 | | | Alarm notification is a | Same | | # COMPARATIVE INFORMATION {2}------------------------------------------------ | | secondary notification<br>system and does not<br>replace the primary alarm<br>notification of the beside<br>monitor | | | | |----------------------|---------------------------------------------------------------------------------------------------------------------|------|--|--| | | Communicates with<br>bedside monitors and<br>telemetry receivers via<br>Ethernet | Same | | | | Communication design | Real time monitoring | Same | | | | OS requirements | Microsoft windows that<br>support Win32 service<br>processes | Same | | | ## 3.1. Intended Use The NetKonnect Remote Network Extension is intended to interface the Nihon Kohden monitoring network with networked client PCs to annunciate and display patient monitoring information to healthcare providers. The device is intended for use in real-time monitoring of routine patient status and alarm events. NetKonnect Remote Network Extension supplements the primary patient monitoring system by providing a forwarding mechanism for annunciating and displaying patient alarm events and event related information including vital signs values and waveforms. ## 3.2 Physical Characteristics The NetKonnect Remote Network Extension is a software only product available on CD Rom. ### 3.3 . Target Population The NetKonnect Remote Network Extension will not change the target population of the cleared monitoring devices that the patients are connected to. The device is available for use by medical personnel on all patient populations. #### 3.4. Product Labeling # I. Labeling To date no advertisement or promotional literature has been created for distribution in the United States. Labeling for the currently marketed is located in Attachment 13. # II. Instructions for Use Instructions for use are provided with each packaged device and are presented in Attachment 13. # III. Proposed Packaging Packaging is presented in Attachment 13. # IV. Engineering Drawings/Photographs Drawings of the device are provided in Attachment 13. ### 3.5. Substantial Equivalence Substantial equivalence discussion can be found in attachment # 12 {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, with three abstract human profiles facing right, connected by flowing lines. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 DEC - 9 2011 Nihon Kohden America, Inc. c/o Mr. Steve Geerdes Director or Quality Assurance and Regulatory Affairs 90 Icon Foothill Ranch, CA 92610 Re: K112637 > Trade/Device Name: NetKonnect Remote Network Extension Regulatory Number: 21 CFR 870.1025 Regulation Name: Patient Physiological Monitor (with arrhythmia detection or alarms) Regulatory Class: II (two) Product Code: 74 MHX Dated: November 22, 2011 Received: November 30, 2011 Dear Mr. Geerdes: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Steve Geerdes Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. http://www.fda.gov/MedicalDevices/ResourceforYou/Industry/default.htm. Sincerely yours, Bram D. Zuckerman, M.D Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use Form 510(k) Number (if known): Device Name: ___ NetKonnect Remote Network Extension Indications for Use: The NetKonnect Remote Network Extension is intended to interface the Nihon Kohden monitoring network with networked client PCs to annunciate and display patient monitoring information to healthcare providers. The device is intended for use in real-time monitoring of routine patient status and alarm events. NetKonnect Remote Network Extension supplements the primary patient monitoring system by providing a forwarding mechanism for annunciating and displaying patient alarm events and event related information including vital signs values and waveforms The system is intended for use by qualified medical personnel within a hospital or clinical environment. The stimulator is available for use on any patients as determined by the qualified medical personnel. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 12/12637 510(k) Page 1 of
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...