SURESIGNS VM4, SURESIGNS VM6, SURESIGNS VM8 MODEL: 863063, 863064, 863065, 863066, 863068

K101067 · Philips Medical Systems · MHX · May 7, 2010 · Cardiovascular

Device Facts

Record IDK101067
Device NameSURESIGNS VM4, SURESIGNS VM6, SURESIGNS VM8 MODEL: 863063, 863064, 863065, 863066, 863068
ApplicantPhilips Medical Systems
Product CodeMHX · Cardiovascular
Decision DateMay 7, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1025
Device ClassClass 2
AttributesPediatric

Intended Use

The SureSigns VM Series Patient Monitors are for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. For monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics and neonates in healthcare environments. Additionally, the monitors may be used in transport situations within a healthcare facility.

Device Story

Multi-parameter patient monitors (VM4, VM6, VM8) for healthcare environments; inputs include physiological signals (ECG, impedance respiration, SpO2, NBP, CO2, temperature). Device processes signals to provide real-time monitoring, recording, and alarming; includes apnea alarm functionality and SARA respiratory algorithm. Operated by healthcare professionals; output displayed on monitor; supports trend database clock synchronization with external systems (e.g., EMR). Benefits include continuous patient status tracking and timely clinical alerts. Modifications include new Oridion miniMedi CO2 module, updated main board with speaker malfunction detection, and software enhancements (technical alarms, patient ID fields, battery reconditioning).

Clinical Evidence

Bench testing only. Verification and validation activities included system-level tests, performance tests, and safety testing based on hazard analysis. Pass/fail criteria were based on predicate device specifications and subject device requirements. No clinical data provided.

Technological Characteristics

Multi-parameter monitor; includes Oridion miniMedi CO2 module, NBP module, and impedance respiration. Features speaker malfunction detection, optional barcode reader, and RS232 serial port. Connectivity includes EMR clock synchronization. Software includes SARA respiratory algorithm. Complies with China SFDA dielectric production test requirements. IPX1 rated.

Indications for Use

Indicated for use by health care professionals for monitoring, recording, and alarming of physiological parameters in adult, pediatric, and neonatal patients in healthcare environments, including transport within a facility.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K 101067 ## 510K Summary This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807-92(c). - The submitter of this pre-market notification is: 1. Peng Cui Philips Medical Systems 3000 Minuteman Road Andover, MA 01810 United States MAY - 7 2010 Tel: 978-659-7966 Fax: 978-659-3819 Email: peng.cui@philips.com This summary was prepared on April 9, 2010. - 2. The names of the subject devices are the Philips SureSigns Series Patient Monitors, SureSigns VM4, VM6, and VM8 Patient Monitors - The trade names of the devices are the SureSigns VM4, SureSigns VM6, SureSigns VM8 3. Patient Monitors. - 4. The common usual name is multi-parameter patient monitor - ട. The Classification names are as follows: | Device Panel | Classification | ProCode | Description | Applicable<br>Subject<br>Devices | |-----------------------------------------|----------------|---------|-----------------------------------------------------------------------------|----------------------------------| | Circulatory System<br>Devices | 870.1025, II | MHX | Monitor, Physiological,<br>Patient (with arrhythmia<br>detection or alarms) | VM4, VM6,<br>VM8 | | | 870.1110, II | DSJ | Alarm, Blood Pressure | VM4, VM6,<br>VM8 | | | 870.1110, II | DSK | Computer, Blood Pressure | VM4, VM6,<br>VM8 | | | 870.1130, II | DXN | System, Measurement,<br>Blood Pressure, Non-<br>Invasive | VM4, VM6,<br>VM8 | | | 870.2300, II | DRT | Monitor, Cardiac (incl.<br>Cardiotachometer & Rate<br>Alarm | VM4, VM6,<br>VM8 | | | 870.2700, II | DQA | Oximeter | VM4, VM6,<br>VM8 | | | 870.2900, I | DSA | Cable, Transducer and<br>Electrode, incl. Patient<br>connector | VM4, VM6,<br>VM8 | | General Hospital and<br>Personal Use | 880.2910, II | FLL | Thermometer, Electronic,<br>Clinical | VM4 | | Anesthesiology &<br>Respiratory Therapy | 868.1400, II | CCK | Analyzer, Gas, | VM8 | page 1 OF 3 {1}------------------------------------------------ - 6. The modified devices are substantially equivalent to previously cleared Philips device, SureSigns VM Series Patient Monitors marketed pursuant to K052707, K080495 and K090483. - 7. The modifications are as follows: - . The OEM CO2 module, from Oridion Medical Ltd. is being changed to a new version of the Oridion miniMedi module. This module was cleared under FDA K060065. The subject device will include the SARA respiratory algorithm cleared in the predicate device SureSigns VM1 (K082280). - An apnea alarm is being added to the existing impedance respiration and CO2 measurements; . however, the device is not an apnea monitor. - A new main board will be used that is host to the previously cleared Philips NBP module . (previously cleared under K090483) and includes the addition of speaker malfunction detection capability. - . The Front end board has been modified to comply with China SFDA dielectric production test requirements. - . An alternate internal component speaker will be used. - An additional optional bar code reader has been made available. . - Optional RS232 serial port adaptor has been made available. - Several new accessories are being listed for use with the device. . - Software enhancements include: . - the addition of two SpO2 technical alarms । - a speaker malfunction technical alarm - - a new CO2 technical alarm - - additional Patient ID fields have been added - - roll-over tool tips for quick software icon identification - an additional 7 second recording option has been added - - battery reconditioning has been improved । - when connected to an external device the monitors can synchronize its trend database clock to a master device time, such as an EMR system - lmpedance respiration is added to SureSigns VM4 patient monitor, which is already in the . predicate device of SureSigns VM6 and SureSigns VM8 cleared under K052707. - . The Instructions for Use will include some additional warnings and cleaning instruction clarifications for the pTemp measurement. - . ج The Instructions for Use will include some additional information and warnings for the CO measurement. - . The device label will include the IPX1 symbol. - 8. The subject devices have the same intended use as the legally marketed predicate device. The SureSigns VM Series Patient Monitors are for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. For monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics and neonates in healthcare environments. Additionally, the monitors may be used in transport situations within a healthcare facility. - 9. The subject devices have the same fundamental technological characteristics as the legally marketed predicate devices. The subject devices use the same design as the predicate devices. The addition of the apnea alarms, the three new technical alarms and the use of the CO2 miniMedi board does not significantly change the fundamental characteristics over the predicate devices - 10. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the subject devices with respect to the predicates. Testing involved system level tests, performance tests, and safety testing from hazard analysis. page 2 OF 3 Page 72 of 112 April 9, 2010 {2}------------------------------------------------ Pass/Fail criteria were based on the specifications cleared for the predicate device, the specifications of the subject device and test results showed substantial equivalence. The results demonstrate that the Philips SureSigns VM4, VM6 and VM8 Patient Moritors meet all reliability requirements and performance claims and supports a determination of substantial equivalence. page 3 of 3 {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the left side of the logo. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002 MAY - 7 2010 Philips Medical Systems c/o Mr. Peng Cui Senior Manager, Regulatory Affairs 3000 Minuteman Road Andover, MA 01810 Re: K101067 > Trade/Device Name: SureSigns VM4, SureSigns VM6, SureSigns VM8 (reference numbers 863063, 863064, 863065, 863066, and 863068) Regulatory Number: 21 CFR 870.1025 Regulation Name: Patient Physiological Monitor (with arrhythmia detection or alarms) Regulatory Class: II (two) Product Code: 74 MHX Dated: April 9, 2010 Received: April 16, 2010 Dear Mr. Cui: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ ## Page 2 - Mr. Peng Cui Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510 (k) Number (if known): Device Name: SureSigns VM4 (reference number: 863063) SureSigns VM6 (reference numbers: 863064, 863065) SureSigns VM8 (reference numbers: 863066, 863068) Indications for Use: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. KID 1067 AND/OR over-the-counter Use: Prescription Use: YES NO (Part 21 CFFR 801 Subpart D) (21 CFR 807 Subpart C) PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE) W.N. (Division Sign-Off) Division of Cardiovascular Devices Page of 510(k) Number_
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...