K051628 · Draeger Medical Systems, Inc. · MHX · Jul 20, 2005 · Cardiovascular
Device Facts
Record ID
K051628
Device Name
INFINITY MONITORS WITH SCIO MODIFICATIONS
Applicant
Draeger Medical Systems, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Jul 20, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Infinity Modular Monitors are intended for multi-parameter patient monitoring. The devices will produce visual and audible alarms if any of the physiological parameters monitored vary beyond preset limits and timed or alarm recordings will be produced. These devices will connect to an R50 Bedside recorder, either directly or via the Infinity Network.
Device Story
Infinity Modular Monitors provide multi-parameter physiological monitoring in clinical settings. The system integrates a Scio gas module for continuous measurement of CO2, N2O, O2, and anesthetic agents (halothane, isoflurane, enflurane, sevoflurane, desflurane). The modification adds a sample gas return kit, allowing gas return to the breathing circuit, identical to the Vamos device. The system processes inputs from various sensors (ECG, pressure, temperature, gas sampling) to display real-time data and derived parameters. It generates visual and audible alarms when physiological values exceed preset limits. Healthcare professionals (physicians, nurses, technicians) operate the device to assess patient status and guide clinical interventions. Data can be recorded via the R50 Bedside recorder or the Infinity Network. The device is not MRI compatible.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Modular patient monitor with gas sampling module. Features sample gas return capability. Connectivity via Infinity Network and MIB protocol converter. Not MRI compatible.
Indications for Use
Indicated for adult, pediatric, and neonatal patients for multi-parameter monitoring including heart rate, respiration, invasive/non-invasive pressure, arrhythmia, temperature, cardiac output, SpO2, pulse rate, apnea, ST segment, tcpO2/tcpCO2, EEG, and FiO2. Gas module (Scio) measures CO2, N2O, O2, and anesthetic agents. Cardiac output, ST segment, and arrhythmia are for adult/pediatric only. tcpO2 is for neonates only (not under gas anesthesia). Used by healthcare professionals in clinical settings.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Special 510(k) Infinity Monitors with Scio Modifications
JUL 2 0 2005
germedical
A Dräger and Siemens Company
K051628
510(k) SUMMARY as required per 807.92(c)
| Submitter's Name and Address: | Draeger Medical Systems, Inc.<br>16 Electronics Avenue<br>Danvers, MA 01923 |
|-------------------------------|-----------------------------------------------------------------------------------------------|
| Contact Person: | Penelope H. Greco<br>Regulatory Affairs Manager<br>Tel: (978) 907-7503<br>Fax: (978) 750-6879 |
| Date submission was prepared: | June 17, 2005 |
| Device Name: | |
| Common Name: | Monitor, Physiological, Patient<br>(with arrhythmia detection or alarms) |
| Classification Name: | MHX |
| Regulation Number: | 21 CFR 870.1025 |
| Class: | 2 |
Legally Marketed Device Identification: Infinity Modular Monitors
# Device Description:
The intent of this 510(k) is to submit a modification specific to Scio, a gas module used exclusively with Draeger Medical Systems' Infinity monitors. The modification is the addition of a sample gas return kit identifical to that of Draeger Medical's Vamos, 510(k) K040847. The modification allows the user the option of returning the sample gas from the Scio gas module to the breathing circuit.
#### Intended Use:
The Infinity Modular Monitors are intended for multi-parameter patient monitoring. The devices will produce visual and audible alarms if any of the physiological parameters monitored vary beyond preset limits and timed or alarm recordings will be produced. These devices will connect to an R50 Bedside recorder, either directly or via the Infinity Network.
### Predicate Devices:
Infinity Monitors with Scio: 510(k) K031340, K033957, K040188 Vamos: K040847
### Substantial Equivalence:
Scio with sample gas return was tested in accordance with internal design control procedures and was determined to be as safe and effective for its intended use as the predicate device.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Page 1 of 1
# COMPANY CONFIDENTIAL
Draeger Medical Systems, Inc. 16 Electronics Avenue Danvers, MA 01923 Tel: (978) 907-7500 Fax: (978) 750-6879
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized eagle or bird-like figure.
JUL 2 0 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Penelope H. Greco Regulatory Affairs Manager DRAEGER Medical Systems, Incorporated 16 Electronics Ave. Danvers, Massachusetts 01923
Re: K051628
Trade/Device Name: Infinity Monitors with SCIO Modifications Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: II Product Code: MHX Dated: June 17, 2005 Received: June 20, 2005
Dear Ms. Greco:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
# Page 2 - Ms. Greco
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Ficase oe advised that I Dris issuation sevice complies with other requirements of the Act that I Dri has made a astatutes and regulations administered by other Federal agencies. You must of ally I coleral statutes and registents, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CTN in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic form in the qualisy by ovisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Section ) vice as described in your Section 510(k) This letter will anow you to ough finding of substantial equivalence of your device to a legally prematics notineation. The a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acrioliary at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain Misoranding by reference to presenters willities under the Act from the Division of Small other general mionnational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Senette y. Michael Ond.
Chiu Lin, PH.D. Director Division of Anesthesiology, General Hospital Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known):_
Device Name: Infinity Modular Monitors with Scio Modifications
Indications for Use:
The INFINITY Modular monitors are capable of monitoring:
- Heart rate . . . . . . . . . . . ●
- Respiration rate .
- Invasive pressure .
- Non-invasive pressure .
- Arrhythmia .
- Temperature .
- Cardiac output .
- Arterial oxygen saturation .
- Pulse rate .
- Apnea .
- . ST Segment Analysis
- 12-Lead ST Segment Analysis .
- tcp02/tcpC02 .
- EEG signals .
- FiO2 .
> Prescription Use_ (Per 21 CFR 801.109)
OR
Over-The-Counter Use_
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sukite Y. Michael Om.d
(Division Sign-Off) Division of Anesthesiology, General Ho Infection Control, Dental
510(k) Number: K051628
Page 1 of 2
{4}------------------------------------------------
### Indications for Use
The SCIO module samples breathing gases from adults and pediatrics. The gas modules continuously measure the content of CO2, N2O, O2 and one of the anesthetic agents, halothane, isoffurane, Enflurane, Sevoflurane and Desflurane in any mixture and communicates real time and derived gas information to the INFINITY monitors.
With etCO2 the monitors can measure end tidal carbon dioxide, inspired carbon dioxide, and respiration rate in either mainstream or side-stream measurement mode; and with etCO2+Respiratory Mechanics, spirometry and carbon dioxide can be monitored. The monitors can interface with specific third party devices via an MIB protocol converter.
The devices are intended to be used in the environment where patient care is provided by Healthcare Professionals, i.e. Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition.
The devices are intended to be used on Adult, Pediatric and Neonatal populations, with the exception of the parameter Cardiac Output, ST Segment Analysis, and arrhythmia which are intended for use in the adult and pediatric populations only; and tcpO2 which is to be used in the neonatal population only when the patient is not under gas anesthesia.
#### MRI Compatibility Statement:
ﻌﺘﻤ ﺴ
The INFINITY Modular Monitors are not compatible for use in a MRI magnetic field.
Page 2 of 2 .
Auy Sulleron
(Division Sign-Off (Division Sign-Off) \
Division of Anesthestelogy General Hospital, Infection Control, Dental D
510(k) Number: