K082613 · Merit Medical Systems, Inc. · KRA · Dec 30, 2008 · Cardiovascular
Device Facts
Record ID
K082613
Device Name
MERIT MICROCATHETER
Applicant
Merit Medical Systems, Inc.
Product Code
KRA · Cardiovascular
Decision Date
Dec 30, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1210
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Merit Microcatheter is intended for general intravascular use, including peripheral and coronary vasculature. Once the subselective region has been accessed, the microcatheter can be used for the controlled and selective infusion of diagnostic, embolic or therapeutic materials into vessels. The catheter should not be used in the cerebral vessels.
Device Story
Merit Microcatheter is a continuous flush catheter used for intravascular access in peripheral and coronary vasculature. Device features distal radiopaque marker for fluoroscopic visualization and hydrophilic coating on distal shaft to facilitate vessel entry. Available in 2.8/2.8 Fr and 2.8/2.4 Fr sizes with lengths of 110, 130, and 150 cm. Operated by clinicians in clinical settings to deliver diagnostic, embolic, or therapeutic agents to subselective regions. Device provides controlled infusion capabilities; benefits include precise delivery of materials to target vessels.
Clinical Evidence
Bench testing only. Performance verified against ISO 10555-1 (general requirements), ISO 10555-2 (angiographic catheters), ISO 594-2 (luer lock fittings), ISO 11135 (sterilization), and ISO 10993-1 (biocompatibility).
Technological Characteristics
Continuous flush catheter; 2.8/2.8 Fr and 2.8/2.4 Fr diameters; 110-150 cm lengths. Features distal radiopaque marker and hydrophilic coating. Sterilized via ethylene oxide. Complies with ISO 10555-1, ISO 10555-2, ISO 594-2, ISO 11135, and ISO 10993-1 standards.
Indications for Use
Indicated for patients requiring general intravascular access in peripheral and coronary vasculature for controlled, selective infusion of diagnostic, embolic, or therapeutic materials. Contraindicated for use in cerebral vessels.
Regulatory Classification
Identification
A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.
Predicate Devices
RenegadeTM Fiber Braided Microcatheter with Hydro PassTM Hydrophilic Coating (K973645)
Related Devices
K173548 — Merit Pursue Microcatheter · Merit Medical Systems, Inc. · Mar 30, 2018
K172081 — Maestro Microcatheter · Merit Medical Systems, Inc. · Aug 4, 2017
K230636 — Maestro Microcatheter · Merit Medical System, Inc. · Mar 30, 2023
K173430 — SeQure® NF and SeQure® Microcatheters · Accurate Medical Therapeutics, Ltd. · Jan 26, 2018
{0}------------------------------------------------
K082613
DEC 3 0 2008
## Section 5
## 510(k) Summary
| General<br>Provisions | Submitter Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Date of Preparation:<br>Registration Number: | Merit Medical Systems, Inc.<br>1600 West Merit Parkway<br>South Jordan, UT 84095<br>(801) 208-4196<br>(801) 253-6932<br>Michaela Rivkowich<br>September 5, 2008<br>1721504 |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject<br>Device | Trade Name:<br>Common/Usual Name:<br>Classification Name: | To be assigned<br>Merit Microcatheters<br>Continuous Flush Catheter |
| Predicate<br>Device | Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer: | RenegadeTM Fiber Braided Microcatheter with<br>Hydro PassTM Hydrophilic Coating<br>Continuous Flush Catheter<br>K973645<br>Boston Scientific Corporation |
| Classification | Class II<br>21 CFR § 870.1210, 74 KRA<br>Division of Cardiovascular, Respiratory and Neurological Devices | |
| Intended Use | The Merit Microcatheter is intended for general intravascular use, including peripheral and coronary vasculature. Once the subselective region has been accessed, the microcatheter can be used for the controlled and selective infusion of diagnostic, embolic or therapeutic materials into vessels. | |
| Device<br>Description | The Merit Microcatheter is available in 2.8/2.8 Fr and 2.8/2.4 Fr sizes<br>with lengths of 110, 130 and 150 cm. The outer surface of the distal<br>segment of the catheter shaft is coated with a hydrophilic coating<br>designed to facilitate the introduction of the catheter into the vessels.<br>The Microcatheter has a radiopaque marker at the distal tip to<br>facilitate fluoroscopic visualization. | |
| Safety &<br>Performance<br>Tests | No performance standards have been established under Section 514<br>of the Food, Drug and Cosmetic Act for these devices. However, a<br>battery of tests was performed according to protocols based on the<br>requirements of the following documents, and were shown to meet<br>the acceptance criteria that were determined to be applicable to the<br>safety and efficacy of the device:<br>• ISO 10555-1: 1995, Sterile, single-use intravascular catheters,<br>Part 1. General requirements.<br>• ISO 10555-2: 1996, Sterile, single-use intravascular catheters,<br>Part 2. Angiographic catheters<br>• ISO 594-2:1998, Conical fittings with 6% (Luer) taper for syringes,<br>needles and certain other medical equipment - Part 2: Lock<br>fittings<br>• ISO 11135: 1994, Medical Devices - Validation and Routine<br>Control of Ethylene Oxide Sterilization<br>• ISO 10993-1: 2003, Biological Evaluation of medical Devices Part<br>1: Evaluation and Testing, and the FDA Modified ISO 10993 Test<br>Profile | |
| Summary of<br>Substantial | Based on the indications for use, design, and safety and<br>performance testing, the subject Merit Microcatheter meets the<br>minimum requirements that are considered essential for its intended | |
{1}------------------------------------------------
:
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## DEC 3 0 2008
Merit Medical Systems, Inc. c/o Mr. Glenn Norton Director of Regulatory Affairs 1600 West Merit Parkway South Jordan, UT 84095
Re: K082613
> Trade/Device Name: Merit Microcatheter Common Name: catheter, continuous flush Regulation Number: 21 CFR 870.1210 Regulatory Class: II Product Code: KRA Dated: December 19, 2008 Received: December 22, 2008
Dear Mr. Norton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-frec number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
M. A. Killennan
ﻣﺮ Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): K082613
Device Name: Merit Microcatheters
Indications for Use:
The Merit Microcatheter is intended for general intravascular use, including peripheral and coronary vasculature. Once the subselective region has been accessed, the microcatheter can be used for the controlled and selective infusion of diagnostic, embolic or therapeutic materials into vessels. The catheter should not be used in the cerebral vessels.
Prescription Use X
. 1988
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M. G. Hillebert
vascular Devices
**510(k) Number** K082613
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.