CorPath GRX System

K173288 · Corindus, Inc. · DXX · Feb 15, 2018 · Cardiovascular

Device Facts

Record IDK173288
Device NameCorPath GRX System
ApplicantCorindus, Inc.
Product CodeDXX · Cardiovascular
Decision DateFeb 15, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1290
Device ClassClass 2
AttributesTherapeutic

Intended Use

The CorPath GRX System is intended for use in the remote delivery and manipulation of guidewires and rapid exchange catheters, and remote manipulation of guide catheters during percutaneous coronary and vascular procedures.

Device Story

Robotic system for remote interventional procedures; enables physician to manipulate guidewires, rapid exchange catheters, and guide catheters from a seated position away from radiation source. System comprises bedside unit (extended reach arm, robotic drive, single-use cassette) and remote workspace (control console, monitors, foot pedal). Physician uses joysticks and touchscreen to control robotic drive, which advances, retracts, and rotates interventional devices. System facilitates precise, computer-controlled movements under fluoroscopic guidance. Benefits include reduced radiation exposure for clinicians and enhanced procedural control. Used in clinical settings (e.g., cath lab) by interventional physicians.

Clinical Evidence

No new clinical data provided. Substantial equivalence is supported by non-clinical bench testing (device compatibility and simulated use) and prior clinical evaluations of the predicate CorPath 200 System (K152999), which established safety for use in clinical settings.

Technological Characteristics

Robotic catheter control system; consists of bedside unit with robotic drive and remote control console. Connectivity via communication cable. Operates via computer-controlled mechanical manipulation of guidewires and catheters. Single-use cassette interface. Class II device.

Indications for Use

Indicated for patients undergoing percutaneous coronary and vascular procedures requiring the delivery and manipulation of guidewires, rapid exchange catheters, and guide catheters.

Regulatory Classification

Identification

A steerable catheter control system is a device that is connected to the proximal end of a steerable guide wire that controls the motion of the steerable catheter.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that. February 15, 2018 Corindus, Inc. Robert Lavado Manager, Regulatory Affairs 309 Waverley Oaks Road Waltham, Massachusetts 02452 Re: K173288 Trade/Device Name: CorPath GRX System Regulation Number: 21 CFR 870.1290 Regulation Name: Steerable Catheter Control System Regulatory Class: Class II Product Code: DXX Dated: January 12, 2018 Received: January 16, 2018 Dear Robert Lavado: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Kenneth J. Cavanaugh -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K173288 Device Name CorPath GRX System Indications for Use (Describe) The CorPath GRX System is intended for use in the remote delivery and manipulation of guidewires and rapid exchange catheters, and remote manipulation of guide catheters during percutaneous coronary and vascular procedures. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 6.0 510(K) SUMMARY | Submitter's Name and<br>Address: | Corindus, Inc.<br>309 Waverley Oaks Road<br>Suite 105<br>Waltham, MA 02452 | |----------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Establishment<br>Registration Number: | 3007822508 | | Date of Summary: | February 15, 2018 | | Contact Person:<br>Telephone Number:<br>Fax Number: | Robert Lavado, Manager, Regulatory Affairs<br>(508) 653-3335 x211<br>(508) 653-3355 | | Name of the Device: | CorPath GRX System | | Common Name: | CorPath GRX System | | Regulatory Status and<br>Regulation Number: | Class II<br>21 CFR 870.1290 | | Classification Name: | System, Catheter Control, Steerable | | Device Classification: | Product Code:<br>DXX: Steerable Catheter Control System. | | Indications for Use: | The CorPath GRX System is intended for use in the<br>remote delivery and manipulation of guidewires and rapid<br>exchange catheters, and remote manipulation of guide<br>catheters during percutaneous coronary and vascular<br>procedures. | | Identification of the<br>Legally Marketed<br>Device (Predicate<br>Device): | Primary Predicate:<br>CorPath 200 System<br>Device Class: II<br>Product Code: DXX<br>Regulation Number: 21 CFR 870.1290<br>510(k) number: K152999<br>Reference Predicate:<br>CorPath GRX System | {4}------------------------------------------------ | Device Class: | II | |--------------------|-----------------| | Product Code: | DXX | | Regulation Number: | 21 CFR 870.1290 | | 510(k) Number: | K160121 | | Device Description: | The CorPath GRX System is intended to allow physicians to deliver and manipulate commercially available guidewires, rapid exchange catheters and guide catheters during percutaneous coronary and vascular intervention procedures. During the use of the CorPath GRX System, the physician maneuvers interventional devices using intuitive controls under independent angiographic fluoroscopy visual guidance using computer controlled movements while in a seated position away from the radiation source. | |---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| The CorPath GRX System is composed of the following two functional sub-units: - 1. Bedside Unit Which consists of the Extended Reach Arm, Robotic Drive and Single-use Cassette - 2. Remote Workspace - Which consists of the Control Console, angiographic monitor(s), hemodynamic monitors, X-ray foot pedal, and optional Interventional Cockpit. Commercially available guidewires, rapid exchange catheters, and guide catheters are loaded into the Singleuse Cassette. By using the joysticks or the Control Console touch screen, the physician can control the Robotic Drive to advance, retract, and rotate the guidewire, advance and retract the rapid exchange catheter, and advance, retrace, and rotate the guide catheter. The Robotic Drive and Control Console communicate via a single communication cable. The product subject of this premarket notification is Substantial substantially equivalent in design and functionality to the Equivalence: CorPath GRX System (K160121, cleared October 27, 2016); and the CorPath 200 System (K152999, cleared March 18, 2016). > The modified CorPath GRX System and the predicate CorPath GRX System have the same technological characteristics. There have been no changes to the modified CorPath GRX System with respect to design, materials, packaging, sterilization, or method of action. The CorPath GRX System uses the same method of action for driving guidewires and catheters as the predicate CorPath 200 System. The predicate CorPath 200 System was evaluated in a clinical setting and found to be safe for use in the periphery (reference K152999). {5}------------------------------------------------ Verification/validation testing of the CorPath GRX System has been conducted to demonstrate the modified CorPath GRX Systems is substantially equivalent to the predicate devices. Tests conducted were identified on the basis of risk analysis activities performed to evaluate the impact of the modification on the device/components. Specifically, the following non-clinical laboratory tests were performed to determine substantial equivalence: - Device Compatibility Testing - Simulated Use Testing All testing has demonstrated that the device is substantially equivalent to the predicate devices. | Safety and Performance: | The determination of substantial equivalence for this device was based on a detailed device description and non-clinical laboratory testing. The testing demonstrated that the device is safe for its intended use and can be considered substantially equivalent to the predicate devices. Previously conducted clinical evaluation of the predicate device (reference K152999) demonstrates that the device is safe for use in a clinical setting. | |-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Conclusion: | Based on the bench testing conducted, it is concluded that the CorPath GRX System is substantially equivalent to the predicate devices: the CorPath 200 System (K152999, cleared March 18, 2016) and the CorPath GRX System (K160121, cleared October 27, 2016). |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...