Upper Arm Type Electronic Blood Pressure Monitor Series, Electronic Blood Pressure Monitor: YE650A, YE650D, YE660B, YE670A and YE670D
K170605 · Jiangsu Yuyue Medical Equipment & Supply Co., Ltd. · DXN · Mar 27, 2018 · Cardiovascular
Device Facts
Record ID
K170605
Device Name
Upper Arm Type Electronic Blood Pressure Monitor Series, Electronic Blood Pressure Monitor: YE650A, YE650D, YE660B, YE670A and YE670D
Applicant
Jiangsu Yuyue Medical Equipment & Supply Co., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Mar 27, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
Electronic blood pressure monitor is intended to measure the blood pressure and pulse rate of adult at household or medical center. (Not suitable for neonate, pregnancy or pre-eclampsia).
Device Story
Upper arm electronic blood pressure monitor; uses oscillometric method to measure systolic/diastolic blood pressure and pulse rate. Device consists of cuff, pneumatic flow control, pressure sensor, and single-chip microcomputer. Operation: user places cuff on upper arm; device automatically inflates cuff to exceed arterial pressure; pneumatic module releases pressure while sensor collects pulse pressure signals; signals are amplified, filtered, and processed by microcomputer algorithm to calculate blood pressure and pulse. Results displayed on LCD; stored in internal memory. Used in home or clinical settings by patients or healthcare providers. Benefits include non-invasive, automated blood pressure monitoring for hypertension management.
Clinical Evidence
Bench testing only. Performance validated against ANSI/AAMI/ISO 81060-2 and IEC 80601-2-30 standards. Biocompatibility testing performed per ISO 10993-1. Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2. Home use safety verified per IEC 60601-1-11.
Technological Characteristics
Oscillometric measurement; semiconductor pressure sensor; automatic electronic pump inflation/deflation; LCD display; 4 AA battery power; modular software design; cuff materials tested per ISO 10993; compliant with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and IEC 60601-1-11 (home use).
Indications for Use
Indicated for adult patients in household or medical center settings for the measurement of blood pressure and pulse rate. Contraindicated for neonates, pregnant patients, or patients with pre-eclampsia.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
OMRON Healthcare (China) Co., Ltd. Model HEM-7320 (K133383)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 27, 2018
Jiangsu Yuyue Medical Equipment & Supply Co., Ltd Rong Zhong Quality Assurance Yunyang Industrial Park Danyang, Jiangsu, China 212300
Re: K170605
Trade/Device Name: Upper Arm Type Electronic Blood Pressure Monitor Series, Electronic Blood Pressure Monitor: YE650A, YE650D, YE660B, YE670A and YE670D Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: January 11, 2018 Received: February 26, 2018
Dear Rong Zhong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
M.A. Willhelmme
for Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K170605
#### Device Name
Upper Arm Type Electronic Blood Pressure Monitor Series Electronic Blood Pressure Monitor: YE650A, YE650D, YE660B, YE670A and YE670D
Indications for Use (Describe)
Electronic blood pressure monitor is intended to measure the blood pressure and pulse rate of adult at household or medical center. (Not suitable for neonate, pregnancy or pre-eclampsia).
| Type of Use (Select one or both, as applicable) | Over-The-Counter Use (21 CFR 201.66) |
|-------------------------------------------------|--------------------------------------|
|-------------------------------------------------|--------------------------------------|
_ | Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the word "yuwell" in red font. The font is bold and slightly italicized. The word is written in lowercase letters.
### 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS
(As Required by 21 CFR 807.92)
- 1. Date Prepared [21 CFR 807.92(a)(1)] 01/11/2018
#### 2. Submitter's Information [21 CFR 807.92(a)(1)]
| Company Name: | Jiangsu Yuyue Medical Equipment& Supply Co., Ltd |
|------------------|--------------------------------------------------|
| Company Address: | Yunyang Industrial Park, Danyang, Jiangsu, China |
| Contact Person: | Rui, Li |
| Phone: | 0511-86900827 |
| Fax: | 0511-86900991 |
| Email: | Lirui@yuyue.com.cn |
#### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
| Trade Name: | Upper Arm Type Electronic Blood Pressure Monitor Series, Electronic Blood<br>Pressure Monitor: YE650A, YE650D, YE660B, YE670A and YE670D |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Electronic Blood Pressure Monitor Series |
| Product Code: | DXN |
| Regulation Number: | 870.1130 |
| Device Class: | II |
#### 4. Identification of Predicate Devices(s) [21 CFR 807.92(a)(3)[
The identification of predicates within this submission is as follow:
| Predicate I | OMRON Healthcare (China) Co., Ltd. |
|----------------------|----------------------------------------------|
| Manufacturer: | |
| Trade Name: | Model HEM-7320 |
| Common Name: | Noninvasive Blood PressureMeasurement System |
| Product Code: | DXN |
| Classification Name: | Noninvasive blood pressure measurement |
| Regulation Number: | 21CFR 870.1130 |
| Classification: | Class II |
| FDA 510 (k) #: | K133383 |
- 5. Description of the Device [21 CFR 807.92(a)(4)]
## SECTION 7 - 2 of 5
{4}------------------------------------------------
#### Jiangsu Yuyue Medical Equipment & Supply Co., Ltd.
Image /page/4/Picture/2 description: The image shows the word "yuwell" in red font. The font is sans-serif and appears to be bolded. The word is slightly blurred, suggesting the image may be a close-up or a lower-resolution file.
The measuring method for the Upper Arm Type Electronic Blood Pressure Monitor Series is oscillation mensuration. The series will automatically starting to take measurements after the inflation of the cuff finished, the results will show the systolic pressure and diastolic pressure and pulse rate. The monitor will store the measurements automatically, which including the time, date, blood pressure and pulse. The record could be revisited. Total 74 sets of data could be stored for YE670A, 60 sets of data for YE670D, 60 sets of data for YE650A, 80 sets of data for YE650D and 74 sets of data for YE660B.
### 6. Intended Use [21 CFR 807.92(a)(5)]
Electronic blood pressure monitor is intended to measure the blood pressure and pulse rate of adult at household or medical center. (Not suitable for neonate, pregnancy or pre-eclampsia).
#### 7. Technological Characteristic [21 CFR 807.92(a)(6)]
YE670series, YE650series and YE660B utilize modular design method, It consists of nine main modules: power-on self-test module, system initialization module, sampling data processing and pressure, pulse rate calculation module, display processing module, power detection processing module, data storage module, key scanning processing module, sampling processing module, voice broadcast processing module, and each module communicates through a message queue.
The blood pressure monitor controls the pneumatic flow control module through single-chipped microcomputer to pressurize the cuff module in order to exceed the lower pressure of patients, the blood being pushed against the artery walls; Pneumatic Flow Control Module being directed to release the pressure, while the pressure detection module collect pulse pressure signal and amplify filter; amplified filter signal being read by single-chipped microcomputer for pressure and pulse signal, through unique algorithm to obtain the systolic and diastolic pressure with pulse; Single-chipped microcomputer will control the inflation/deflation module to release the pressure after receive measurements; in the meanwhile, display the measurements results then stored the values with memory module.
### SECTION 7 - 3 of 5
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the text "K170605" in the upper left corner. Below this text is the word "yuwell" in a stylized, red font. The word is slightly blurred, giving it a sense of motion or depth. A black line is at the bottom of the image.
#### 8. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92(b)(2)]
| Table 1 | YE650A | YE650D | YE670A | YE670D | YE660B | Predicate Device |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | DXN | DXN | DXN | DXN | DXN | DXN |
| Regulation No. | 870.1130 | 870.1130 | 870.1130 | 870.1130 | 870.1130 | 870.1130 |
| Classification | II | II | II | II | II | II |
| Intended Use | Electronic blood pressure monitor is intended to measure the blood pressure and pulse rate of adult at household or medical center. (Not suitable for neonate, pregnancy or pre-eclampsia) | Electronic blood pressure monitor is intended to measure the blood pressure and pulse rate of adult at household or medical center. (Not suitable for neonate, pregnancy or pre-eclampsia) | Electronic blood pressure monitor is intended to measure the blood pressure and pulse rate of adult at household or medical center. (Not suitable for neonate, pregnancy or pre-eclampsia) | Electronic blood pressure monitor is intended to measure the blood pressure and pulse rate of adult at household or medical center. (Not suitable for neonate, pregnancy or pre-eclampsia) | Electronic blood pressure monitor is intended to measure the blood pressure and pulse rate of adult at household or medical center. (Not suitable for neonate, pregnancy or pre-eclampsia) | The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population. The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings. |
| 510 (k) Number | N/A | N/A | N/A | N/A | N/A | K133383 |
| Design Method | Oscillometric Method | Oscillometric Method | Oscillometric Method | Oscillometric Method | Oscillometric Method | Oscillometric Method |
## SECTION 7 - 1 of 5
{6}------------------------------------------------
| IIIIIe | |
|--------|--|
| | |
| | |
| | |
| | |
| Patient<br>Population | Adult | Adult | Adult | Adult | Adult | Adult |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Measurement<br>Site | Upper Arm | Upper Arm | Upper Arm | Upper Arm | Upper Arm | Upper Arm |
| Cuff<br>Circumference | 220mm ~ 320mm | 220mm ~ 320mm | 220mm ~ 320mm | 220mm ~ 320mm | 220mm ~ 320mm | 220mm ~420mm |
| Inflation<br>Method | Automatic by<br>electronic pump | Automatic by<br>electronic pump | Automatic by<br>electronic pump | Automatic by<br>electronic pump | Automatic by<br>electronic pump | Automatic by<br>electronic pump |
| Deflation<br>Method | Automatic Pressure<br>Release Valve | Automatic Pressure<br>Release Valve | Automatic Pressure<br>Release Valve | Automatic Pressure<br>Release Valve | Automatic Pressure<br>Release Valve | Automatic Pressure<br>Release Valve |
| Display | LCD Digital Display | Backlight LCD<br>Digital Display | LCD Digital Display | LCD Digital Display | LCD Digital Display | LCD Display |
| Patients<br>Contacting<br>Materials | Patient contact<br>materials of the cuff<br>have been tested in<br>accordance with ISO<br>10993 and FDA<br>guidance. | Patient contact<br>materials of the cuff<br>have been tested in<br>accordance with ISO<br>10993 and FDA<br>guidance. | Patient contact<br>materials of the cuff<br>have been tested in<br>accordance with ISO<br>10993 and FDA<br>guidance. | Patient contact<br>materials of the cuff<br>have been tested in<br>accordance with ISO<br>10993 and FDA<br>guidance. | Patient contact<br>materials of the cuff<br>have been tested in<br>accordance with ISO<br>10993 and FDA<br>guidance. | Patient contact<br>material of the cuff<br>have been tested in<br>accordance with ISO<br>10993 and FDA<br>guidance. |
| Memory Size | Up to 60sets of data | Up to 80 sets of data | Up to 74 sets of data | Up to 60 sets of data | Up to 74 sets of data | Up to 100 sets of<br>data |
| Measurement<br>Pressure Range | 0 ~ 280 mmHg (0<br>kPa ~ 37.3 kPa) | 0~280 mmHg (0<br>kPa ~ 37.3 kPa) | 0 ~ 280 mmHg (0<br>kPa ~ 37.3 kPa) | 0 ~ 280 mmHg (0<br>kPa ~ 37.3 kPa) | 0~280mmHg (0-<br>37.3kPa) | 0 mmHg ~ 299<br>mmHg(0 kPa~ 39.9<br>kPa) |
| Range Accuracy | 士3 mmHg<br>(±0.4kPa) | 士3 mmHg<br>(±0.4kPa) | ±3 mmHg<br>(±0.4kPa) | 士3 mmHg<br>(±0.4kPa) | 士3mmHg<br>(±0.4kPa) | 土3mmHg<br>(±0.4kPa)or 2% of<br>reading |
| Measurement<br>Pulse Range | 40 ~ 200 beats/min | 40 ~ 200 beats/min | 40 ~ 200 beats/min | 40 ~ 200 beats/min | 40 ~ 200 beats/min | 40 ~ 180 beats/min |
| Pulse Accuracy | ±5% of reading<br>value | 土5% of reading<br>value | ±5% of reading<br>value | ±5% of reading<br>value | 土5% of reading<br>value | Within ±5%of<br>reading |
| Pressurization<br>Source | Automatic<br>Pressurize | Automatic<br>Pressurize | Automatic<br>Pressurize | Automatic<br>Pressurize | Automatic<br>Pressurize | Automatic Internal<br>Pump |
| Pressure Sensor | Semiconductor<br>Pressure Sensor | Semiconductor<br>Pressure Sensor | Semiconductor<br>Pressure Sensor | Semiconductor<br>Pressure Sensor | Semiconductor<br>Pressure Sensor | Semiconductor<br>Pressure Sensor |
| Operating<br>Environment | Temperature: +10 °C<br>~ +40 °C<br>Humidity: 15% RH<br>~ 90% RH(no<br>condensation) | Temperature: +10 °C<br>~ +40 °C<br>Humidity: 15% RH<br>~ 90% RH(no<br>condensation) | Temperature: +10 °C<br>~ +40 °C<br>Humidity: 15% RH<br>~ 90% RH(no<br>condensation) | Temperature: +10 °C<br>~ +40 °C<br>Humidity: 15% RH<br>~ 90% RH(no<br>condensation) | Temperature: +10 °C<br>~ +40 °C<br>Humidity: 15% RH<br>~ 90% RH(no<br>condensation) | Temperature: +10 °C (50-104 °F)<br>Humidity: 30% RH<br>~ 85% RH(no<br>condensation) |
| Storage<br>Environment | Temperature: -20 °C<br>~ +55 °C<br>Humidity: 15% RH<br>~ 90% RH(no<br>condensation) | Temperature: -20 °C<br>~ +55 °C<br>Humidity: 15% RH<br>~ 90% RH(no<br>condensation) | Temperature: -20 °C<br>~ +55 °C<br>Humidity: 15% RH<br>~ 90% RH(no<br>condensation) | Temperature: -20 °C<br>~ +55 °C<br>Humidity: 15% RH<br>~ 90% RH(no<br>condensation) | Temperature: -20 °C<br>~ +55 °C<br>Humidity: 15% RH<br>~ 90% RH(no<br>condensation) | Temperature: -20 °C<br>~ +60 °C<br>Humidity: 10% RH<br>~ 95% RH(no<br>condensation) |
| Energy Source | 4 AA batteries | 4 AA batteries | 4 AA batteries | 4 AA batteries | 4 AA batteries | 4 AA batteries or<br>AC Adapter |
| Display Content | Cuff Pressure, Pulse,<br>Date,<br>Time,<br>Systolic/Diastolic<br>Pressure, error<br>message,<br>measurements<br>results in memory,<br>Irregular Heart Beat<br>Feature, Body<br>movement detection<br>Cuff Wrapping<br>Detection | Cuff Pressure, Pulse,<br>Date,<br>Time,<br>Systolic/Diastolic<br>Pressure, error<br>message,<br>measurements<br>results in memory,<br>Irregular Heart Beat<br>Feature, Body<br>movement detection<br>Cuff Wrapping<br>Detection | Cuff Pressure, Pulse,<br>Date,<br>Time,<br>Systolic/Diastolic<br>Pressure, error<br>message,<br>measurements<br>results in memory,<br>Irregular Heart Beat<br>Feature, Body<br>movement detection<br>Cuff Wrapping<br>Detection | Cuff Pressure, Pulse,<br>Date,<br>Time,<br>Systolic/Diastolic<br>Pressure, error<br>message,<br>measurements<br>results in memory,<br>Irregular Heart Beat<br>Feature, Body<br>movement detection<br>Cuff Wrapping<br>Detection | Cuff Pressure, pulse,<br>Systolic/Diastolic,<br>error message,<br>measurement results<br>in the memory, Date,<br>Time. | Cuff Pressure,<br>Systolic/Diastolic<br>Pressure, pulse rate,<br>error message,<br>measurement results<br>in the memory |
| Controls | SET Button,<br>Plus/Minus Button,<br>START/PULSE<br>Button | SET Button,<br>Plus/Minus Button,<br>START/PULSE<br>Button | SET Button,<br>Memory Button,<br>START/PULSE<br>Button | SET Button,<br>Memory Button,<br>START/PULSE<br>Button | Memory Button,<br>Start Button | START/STOP<br>button, Date/Tim<br>Setting Button,<br>UP/DOWN Button,<br>USER ID Selections<br>Button |
| Performance | ANSI/AAMI/ISO<br>81060-2<br>IEC<br>80601-2-30 | ANSI/AAMI/ISO<br>81060-2<br>IEC<br>80601-2-30 | ANSI/AAMI/ISO<br>81060-2<br>IEC<br>80601-2-30 | ANSI/AAMI/ISO<br>81060-2<br>IEC<br>80601-2-30 | ANSI/AAMI/ISO<br>81060-2<br>IEC<br>80601-2-30 | ANSI/AAMI/ISO<br>81060-2<br>IEC<br>80601-2-30 |
SECTION 7 - 2 of 5
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the text "K170605" in the upper left corner. Below this text is the word "yuwell" in a stylized, red font. The word is slightly blurred, giving it a sense of motion or depth. A black line is at the bottom of the image.
# SECTION 7 - 3 of 5
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the text 'K170605' in the upper left corner. Below this text is the word 'yuwell' in a stylized, red font. A black line is at the bottom of the image.
| Biocompatibility | ISO 10993-1, FDA<br>Guidance, Tests<br>included<br>Cytotoxicity,<br>Sensitization and<br>Intracutaneous<br>Reactivity | ISO 10993-1, FDA<br>Guidance, Tests<br>included<br>Cytotoxicity,<br>Sensitization and<br>Intracutaneous<br>Reactivity | ISO 10993-1, FDA<br>Guidance, Tests<br>included<br>Cytotoxicity,<br>Sensitization and<br>Intracutaneous<br>Reactivity | ISO 10993-1, FDA<br>Guidance, Tests<br>included<br>Cytotoxicity,<br>Sensitization and<br>Intracutaneous<br>Reactivity | ISO 10993-1, FDA<br>Guidance, Tests<br>included<br>Cytotoxicity,<br>Sensitization and<br>Intracutaneous<br>Reactivity | ISO 10993-1, FDA<br>Guidance, Tests<br>included<br>Cytotoxicity,<br>Sensitization and<br>Intracutaneous<br>Reactivity |
|-------------------|-----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Electrical Safety | IEC<br>60601-1 | IEC<br>60601-1 | IEC<br>60601-1 | IEC<br>60601-1 | IEC<br>60601-1 | IEC<br>60601-1 |
| EMC | IEC<br>60601-1-2 | IEC<br>60601-1-2 | IEC<br>60601-1-2 | IEC<br>60601-1-2 | IEC<br>60601-1-2 | IEC<br>60601-1-2 |
| Home Use | IEC 60601-1-11 | IEC 60601-1-11 | IEC 60601-1-11 | IEC 60601-1-11 | IEC 60601-1-11 | IEC 60601-1-11 |
{9}------------------------------------------------
### Jiangsu Yuyue Medical Equipment & Supply Co., Ltd.
Image /page/9/Picture/2 description: The image shows the word "yuwell" in red font. The font is bold and slightly italicized. The word is on a white background.
### 9. Conclusion [21 CFR 807.92(b)(3)]
The Upper Arm Type Electronic Blood Pressure Monitor650A, 650D, 670A, 670D and 660B are substantially equivalent to predicate device because both devices use identical technology and same intended use, also testing standards are identical with the predicates. The differences between both devices are insignificant in terms of safety and effectiveness.
## SECTION 7 - 5 of 5