K162230 · Adon Health Co., Ltd. · DXN · Nov 18, 2016 · Cardiovascular
Device Facts
Record ID
K162230
Device Name
Fully Automatic Electronic Blood Pressure Monitor
Applicant
Adon Health Co., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Nov 18, 2016
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
KD-5920, KD-5920L and KD-5923 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
Device Story
Fully automatic electronic blood pressure monitor; uses oscillometric method with silicon pressure sensor; measures systolic/diastolic blood pressure and pulse rate; detects irregular heartbeats. Device consists of main unit and inflatable upper arm cuff (22-48cm). Operated by medical professionals or patients at home. User wraps cuff around arm; device inflates/deflates automatically; calculates values; displays results on LCD. Includes blood pressure classification indicator. Provides objective data for hypertension monitoring and cardiovascular assessment. Benefits include convenient, non-invasive, self-administered blood pressure tracking.
Clinical Evidence
Bench testing only. Conformance to IEC 60601-1, IEC 60601-1-2, IEC 80601-2-30, and ANSI/AAMI/ISO 81060-2:2009 for clinical validation of automated measurement.
Technological Characteristics
Oscillometric measurement principle using silicon pressure sensor. Powered by batteries. Form factor: upper arm cuff (22-48cm) with LCD display. Standards: IEC 60601-1, IEC 60601-1-2, IEC 80601-2-30, ISO 81060-2. Software-based processing for pressure calculation and irregular heartbeat detection.
Indications for Use
Indicated for adult individuals requiring non-invasive measurement of systolic/diastolic blood pressure and pulse rate. For use by medical professionals or at home. Cuff circumference 22cm-48cm.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol representing the human form, with three stylized profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 18, 2016
Andon Health Co., Ltd. Ms. Liu Yi President No.3 Jin Ping St, Ya An Rd, Nankai District Tianjin, 300190 CN
Re: K162230
> Trade/Device Name: KD-5920, KD-5920L and KD-5923 Fully Automatic Electronic Blood Pressure Monitors Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: August 5. 2016 Received: August 11, 2016
Dear Ms. Liu Yi,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mada Jellman
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162230
Device Name
KD-5920, 5920L and KD-5923 Fully Automatic Electronic Blood Pressure Monitor
Indications for Use (Describe)
KD-5920, KD-5920L and KD-5923 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
Counter Use (21 CFR 801 Suppart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary - K162230
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.
# 1.0 submitter's information
| Name: | Andon Health Co., Ltd. |
|----------------------|------------------------------------------------------------|
| Address: | No 3, Jinping Street Ya An Road, Nankai District, Tianjin, |
| | P.R. China |
| Phone number: | 86-22-6052 6161 |
| Fax number: | 86-22-6052 6162 |
| Contact: | Liu Yi |
| Date of Preparation: | 8/3/2016 |
# 2.0 Device information
| Trade name: | KD-5920, KD-5920L and KD-5923 Fully Automatic<br>Electronic Blood Pressure Monitor |
|--------------|------------------------------------------------------------------------------------|
| Common name: | Noninvasive blood pressure measurement system |
Classification name: Noninvasive blood pressure measurement system
## 3.0 Classification
Production code: DXN- Noninvasive blood pressure measurement system. Regulation number: 870.1130 Classification: II Panel: Cardiovascular
# 4.0 Predicate device information
Manufacturer: Andon Health Co., Ltd. Device: KD-5964 Fully Automatic Electronic Blood Pressure Monitor 510(k) number: K102906
### K162230
{4}------------------------------------------------
# 5.0 Intended use
KD-5920, KD-5920L and KD-5923 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
The intended use and the indication for use of KD-5920, KD-5920L and KD-5923, as described in its labeling are the same as the predicate device KD-5964 (K102906).
# 6.0 Device description
KD-5920, KD-5920L and KD-5923 Fully Automatic Electronic Blood Pressure Monitor is designed and manufactured according to IEC 80601-2-30.
The operational principle is based on oscillometric and silicon integrates pressure sensor technology. It can calculate the systolic and diastolic blood pressure, and display the result on the LCD. The measurements results can also be classified by the function of blood pressure classification indicator. If any irregular heartbeat is detected, it can be shown to the user.
# 7.0 Summary comparing technological characteristics with predicate device
| Technological Characteristics | Comparison result |
|-------------------------------|-------------------|
| Design principle | Identical |
| Appearance | Similar |
| Patients contact Materials | Similar |
| Performance | Similar |
| Biocompatibility | Identical |
| Mechanical safety | Identical |
| Standards met | Identical |
| Electrical safety | Identical |
| EMC | Identical |
| Function | Similar |
{5}------------------------------------------------
# 8.0 Discussion of non-clinical and clinical test performed
# Non-clinical Tests have been done as follows:
a. Electromagnetic compatibility test according to IEC 60601-1-2;
b. Electrical safety according test to IEC 60601-1;
c. Safety and performance characteristics of the test according to IEC 80601-2-30
None of the test demonstrates that KD-5920, KD-5920L and KD-5923 Fully Automatic Electronic Blood Pressure Monitor bring new questions of safety and effectiveness.
# 9.0 Performance summary
KD-5920, KD-5920L and KD-5923 Fully Automatic Electronic Blood Pressure Monitor conforms to the following standards:
- IEC 60601-1:2005/(R)2012 And A1:2012,C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD)
- · IEC 60601-1-2:2007,Medical Electrical Equipment -- Part 1-2: General Requirements For Basic Safety And Essential Performance -- Collateral Standard: Electromagnetic Disturbances -- Requirements And Tests
- IEC 80601-2-30:2009 & A1:2013.Medical Electrical Equipment Part 2-30: Particular Requirements For The Basic Safety And Essential Performance Of Automated Noninvasive Sphygmomanometers
- · ANSI/AAMI/ISO 81060-2:2009, Non-invasive sphygmomanometers part 2: Clinical validation of automated measurement type
#### 10.0 Comparison to the predicate device and the conclusion
Our device KD-5920, KD-5920L and KD-5923 Fully Automatic Electronic are substantially equivalent to the Fully Automatic Electronic Blood Pressure Monitor KD-5964 whose 510(k) number is K102906.
KD-5920, KD-5920L and KD-5923 are very similar with its predicate device in the intended use, the design principle, the material, the performance, memory average function, average measure function and the applicable standards. Only their appearance, the memory capacity, the battery and the MCU are different.
However, the test in this submission provides demonstrates that these small differences do not raise any new questions of safety and effectiveness.