Arm-Type Fully Automatic Digital Blood Pressure Monitor
Applicant
Sejoy Electronics & Instruments Co., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Oct 16, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
Measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents age 12 through 21 years of age with arm circumference of 22- 42cm.
Device Story
Arm-type fully automatic digital blood pressure monitor; uses inflatable cuff wrapped around upper arm. Internal pump inflates cuff; oscillometric method with silicon integrated pressure sensor determines systolic/diastolic blood pressure and pulse rate. Mechanical valve controls deflation rate; automatic pressure release at measurement end. Device displays results on LCD for three minutes; includes WHO blood pressure classification indicators and irregular heartbeat detection. Designed for over-the-counter use by patients. Benefits include non-invasive, automated monitoring of blood pressure and pulse rate in home or clinical settings.
Clinical Evidence
Clinical validation performed per ISO 81060-2:2013 (non-invasive sphygmomanometers - clinical validation of automated measurement type). Results confirmed compliance with accuracy requirements.
Technological Characteristics
Oscillometric measurement; silicon integrated pressure sensor; internal pump; mechanical deflation valve. Max pressure safety setting 300 mmHg. LCD display. Complies with AAMI/ANSI/IEC 80601-2-30:2009. Biocompatibility per ISO 10993-5 and ISO 10993-10. Electromagnetic compatibility per IEC/EN 60601-1-2:2007. Safety per AAMI/ANSI ES 60601-1:2005.
Indications for Use
Indicated for adults and adolescents aged 12-21 years with arm circumference 22-42cm for measurement of systolic/diastolic blood pressure and pulse rate.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is simple and recognizable, representing the department's role in promoting the health and well-being of Americans.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 16, 2015
Sejoy Electronics & Instruments Co., Ltd. Mr. Ren Yunhua General Manager Building 2, No. 202, Zhenzhong Road West Lake Economy & Technology Zone Hangzhou, 310030 CN
Re: K152687
> Trade/Device Name: Arm-Type Fully Automatic Digital Blood Pressure Monitor, BP-35 Regulation Number: 21 CFR 870.1130 Regulation Name: Non-Invasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: September 15, 2015 Received: September 18, 2015
Dear Mr. Ren Yunhua,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Mitchell Stein
forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152687
Device Name
The Arm-Type Fully Automatic Digital Blood Pressure Monitor, BP-35
Indications for Use (Describe)
Measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents age 12 through 21 years of age with arm circumference of 22- 42cm.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for SEJOY. The logo is in blue, with an orange triangle above the "J". The logo is simple and modern, and it is likely used by a company that sells products or services related to health or wellness.
# K152687
# Page 1 of 4
# 510(k) Summary
The assigned 510(k) number_
# 2.1. Date Prepared:
2015.08.31
# 2.2. Submitter's Identification:
Name: Sejoy Electronics & Instruments Co., Ltd.
Add.: Building 2, No.202, Zhenzhong Road, West Lake Economy & Technology Zone,
Hangzhou, China 310030
Contact Person: Yunhua Ren
Phone: +86-571-81957767
Fax: +86-571-81957750
Email: renyh@sejoy.com
# 2.3. Name of the Device:
Trade Name: The Arm-Type Fully Automatic Digital Blood Pressure Monitor
Including the following models: BP-35
Common Name: Blood Pressure Monitor
Classification name: Non-invasive blood pressure measurement System
21 CFR 870-1130, Class II, 74-DXN.
# 2.4. Classification Information:
Regulation Number: 870.1130
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for SEJOY. The logo is blue, with a small orange triangle above the "J". The logo is simple and modern.
K152687
Page 2 of 4
Product Code: DXN Device Class: II
Panel: 74 Cardiovascular
# 2.5. Predicate Device Information:
The Arm-Type Fully Automatic Digital Blood Pressure Monitor are substantially equivalent to the following device: device: Blood Pressure (Model BP-35) , FDA 510(k) number: K141755, manufactured ELECTRONICS & INSTRUMENTS CO., LTD.
# 2.6. Device Description:
The Arm-Type Fully Automatic Digital Blood Pressure Monitor uses an inflatable cuff which is wrapped around the patient's upper arm. The cuff is inflated automatically by an internal pump in the device. The systolic and diastolic blood pressures are determined by oscillometric method and silicon integrate pressure sensor technology. The deflation rate is controlled by a preset mechanical valve at a constant rate. The pressure of the cuff is completely released automatically at the end of the measurement. At the same time, the measurements are displayed on the LCD display for three minute. There is a maximum pressure safety setting at 300 mmHg. The device will not inflate the cuff higher than 300 mmHg.
For BP-35, the blood pressure results are compared with WHO (World Health Organization) Blood Pressure classification, which are severe Hypertension. Moderate Hypertension, Mild Hypertension, High-normal, Normal, and Optimal. The corresponding LCD segment will be turned on along with the systolic, diastolic, and pulse rate information. BP-35 will display an irregular heartbeat symbol " (♥) " if an irregular heartbeat was detected during the measurement process. It can display average results in three ways.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for SEJOY. The logo is in blue, with the letters "SEJOY" in a bold, sans-serif font. There is a small orange triangle above the "J" in the logo. The logo is simple and modern.
K152687
Page 3 of 4
The device is designed and manufactured according to AAMI / ANSI / IEC 80601-2-30:2009, medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers.
### 2.7. Indications for Use:
Measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents age 12 through 21 years of age with arm circumference of 22- 42cm.
### 2.8. Difference of comparison with predicate device:
The difference between the predicate device and submit Arm-type fully automatic blood pressure monitor is one new arm-cuff (Size: 22.0-42cm) added as optional.
# 2.9. Non-clinical Tests:
Electromagnetic Compatibility Test according to IEC/EN 60601-1-2:2007;
General Safety Provisions Test according to AAMI/ANSI ES 60601-1:2005/(R) 2012 and C1:2009/(R)2012 and A2:2010/(R)2012;
Performance Test according to IEC 80601-2-30:2009, medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers. The test result all meet or exceed the requirement of the standards.
Biocompatibility Test according to FDA Bluebook Memorandum G95-1Use of International Standard ISO 10993, ISO 10993-5: 2009 Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity and ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for SEJOY. The logo is blue and has an orange triangle above the "J". The logo is simple and modern.
<152687 age 4 of 4
# 2.10. Discussion of Clinical Tests Performed:
Clinical tests were performed and comply with the accuracy requirements of ISO 81060-2 Second edition 2013-05-01, non-invasive sphygmomanometers - part 2: clinical validation of automated measurement type.
### 2.11. Conclusions:
The submit Arm-type Fully Automatic Blood Pressure Monitor is itself of the predicate device BP-35 manufactured by SEJOY ELECTRONICS & INSTRUMENTS CO., LTD. (FDA 510(k) number: K141755). The addition of new arm-cuff (Size: 22cm-42cm) will not: (1) affect the intended use or (2) alter the fundamental scientific technology of the devices