U-RIGHT TD-3135 BLOOD PRESSURE MONITORING SYSTEM

K112274 · Taidoc Technology Corporation · DXN · Dec 21, 2011 · Cardiovascular

Device Facts

Record IDK112274
Device NameU-RIGHT TD-3135 BLOOD PRESSURE MONITORING SYSTEM
ApplicantTaidoc Technology Corporation
Product CodeDXN · Cardiovascular
Decision DateDec 21, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1130
Device ClassClass 2

Intended Use

The U-RIGHT TD-3135 Blood Pressure Monitoring System is intended to be used to measure the systolic and diastolic blood pressure and pulse rate by using a non-invasive technique in which an inflatable cuff is wrapped on the upper arm. This system should only be used for the testing on people over 18 years of age and over.

Device Story

The U-RIGHT TD-3135 Blood Pressure Monitoring System is an oscillometric device for measuring systolic/diastolic blood pressure and pulse rate. It consists of a monitor and an inflatable upper arm cuff. The device operates by wrapping the cuff around the arm; the system automatically inflates and deflates to detect pressure oscillations, which are processed to calculate blood pressure and pulse rate. It is intended for over-the-counter use by adults 18 years and older. The device provides digital output of measurements to the user. It is a modified version of the predicate TD-3132, featuring a different physical appearance, reduced memory capacity, and the removal of data transmission functions.

Clinical Evidence

Bench testing only. Software verification and validation, along with design validation, confirmed that the performance, safety, and effectiveness of the device are equivalent to the predicate device.

Technological Characteristics

Oscillometric non-invasive blood pressure measurement system. Consists of a monitor and inflatable upper arm cuff. Operates via electronic circuit design. Class II device. No specific materials or software architecture details provided beyond equivalence to predicate.

Indications for Use

Indicated for adults 18 years of age and older for non-invasive measurement of systolic and diastolic blood pressure and pulse rate using an upper arm cuff.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ DEC 2 1 2011 # Section 11. 510(k) Summary ে এ ব #### 510(K) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. The Assigned 510(k) number is: __ 长川2274 - l . Submitter's Identification: TaiDoc Technology Corporation 3F, 5F, No.127, Wugong 2nd Rd., Wugu District, New Taipei City, 24888, Taiwan Correspondence: Linda Ko Regulatory Affairs Specialist Tel: +886-2-6625-8188 #1158 Fax: +886-2-6625-0288 Email: kolinda@taidoc.com.tw Date of submission: 08/05/2011 - 2. Device name: Proprietary name: U-RIGHT TD-3135 Blood Pressure Monitoring System Regulatory information: | A. Regulation section: 21 CFR §870.1130, Noninvasive blood pressure<br>measurement system | |-------------------------------------------------------------------------------------------| | B. Classification: Class II (Blood Pressure Measurement System) | - B. Classification: Class II (Blood Pressure Measurement System) - ﺰ Product Code: DXN, System, Measurement, Blood-Pressure, Non-Invasive - D. Panel: 74, Cardiovascular - Blood Pressure Measurement System {1}------------------------------------------------ 3. Intended Use: ् ェ The U-RIGHT TD-3135 Blood Pressure Monitoring System is intended to be used to measure the systolic and diastolic blood pressure and pulse rate by using a non-invasive technique in which an inflatable cuff is wrapped on the upper arm. This system should only be used for the testing on people over 18 years of age and over. - 4. Device Description: The kit of U-RIGHT TD-3135 Blood Pressure Monitoring System consists of two main products: the Blood Pressure Monitor and the pressure cuff. The product has been designed and tested to work together as a system to produce accurate blood pressure measurements. - 5. Substantial Equivalence Information: - A. Predicate device name: FORA P20/U-RIGHT TD-3132 Blood Pressure Monitoring System Predicate K number: K092106 B. ﻥ Comparison with predicate: The modified U-RIGHT TD-3135 Blood Pressure Monitoring System has the following similarities to the predicate device: - same operating principle, - . same fundamental scientific technology, - incorporate the same basic circuit design, - . incorporate the same materials, - . same shelf life - packaged using the same materials, and - manufactured by the same process. The modifications encompass: - Modification of the devices physical appearance - 비 Decreased memory storage capacity - 트 Removed the data transmission function - . Labeling change due to the above modifications 11- 2 of 3 {2}------------------------------------------------ - 6. Test Principle: The measurement is by using oscillometric, non-invasive blood pressure (systolic, diastolic blood pressure and pulse rate) measuring technology. ### 7. Performance Characteristics: The U-RIGHT TD-3135 Blood Pressure Monitoring System has the same performance characteristics as the predicate device. Software verification and validation, and design validation confirmed that the performance, safety and effectiveness of the U-RIGHT TD-3135 Blood Pressure Monitoring System are equivalent to the predicate device. ### 8. Conclusion: Based on the information provided in this submission, the U-RIGHT TD-3135 Blood Pressure Monitoring System is substantially equivalent to the predicate FORA P20/U-RIGHT TD-3132 Blood Pressure Monitoring System. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle or bird-like figure with three distinct segments forming its body and wings. Encircling the bird is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA", arranged in a circular fashion around the bird. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 DEC 2 1 2011 TaiDoc Technology Corporation c/o Ms. Linda Ko Regulatory Affairs Specialist 6F, No. 127, Wugong 2nd Rd. Wugu Township, New Taipei City TAIWAN 24888 Re: K112274 > Trade/Device Names: TaiDoc U-RIGHT TD-3135 Blood Pressure Monitoring System Regulatory Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (Two) Product Code: DXN Dated: November 30, 2011 Received: December 1, 2011 Dear Ms. Ko: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may . publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 1 - Ms. Linda Ko Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. Mark Fellman for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Section 8. ## Indications for Use 510(k) Number: K1/2274 Device Name: U-RIGHT TD-3135 Blood Pressure Monitoring System Indications for Use: The U-RIGHT TD-3135 Blood Pressure Monitoring System is intended to be used to measure the systolic and diastolic blood pressure and pulse rate by using a non-invasive technique in which an inflatable cuff is wrapped on the upper arm. This system should only be used for the testing on people over 18 years of age and over. Prescription Use (21 CFR Part 801 Subpart D) And/Or Over the Counter Use _ X (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Mink Keller for Bram fuckerman Division Sign Off Director DCD 510(K Page 1 of __ 1 8-1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...