K070825 · Little Doctor (Shanghai) Electronic Manufacturing · DXN · May 25, 2007 · Cardiovascular
Device Facts
Record ID
K070825
Device Name
LD8 DIGITAL WRIST BLOOD PRESSURE MONITOR
Applicant
Little Doctor (Shanghai) Electronic Manufacturing
Product Code
DXN · Cardiovascular
Decision Date
May 25, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The LD8 Digital Wrist Blood Pressure Monitor is a device intended to measur systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist.
Device Story
The LD8 Digital Wrist Blood Pressure Monitor is a non-invasive device for measuring systolic/diastolic blood pressure and pulse rate in adults. It utilizes an inflatable wrist cuff and an electronic semiconductor sensor to detect pressure alterations; these signals are processed via the oscillometric method to calculate blood pressure and pulse rate. The device is intended for use by individuals; it provides digital output of measurements. It serves as an alternative to traditional auscultatory methods (stethoscope/mercury manometer), offering a simplified, automated measurement process for patient monitoring.
Clinical Evidence
Clinical testing was performed in accordance with ANSI/AAMI-SP-10 Standard, Section 4.4.2. All testing results met the required parameters for safety and effectiveness.
Technological Characteristics
Non-invasive oscillometric blood pressure monitor; uses electronic semiconductor sensor to convert cuff pressure to electrical signals; wrist-worn form factor; software-based algorithm for signal analysis; powered by internal electronics.
Indications for Use
Indicated for adult individuals to measure systolic and diastolic blood pressure and pulse rate using a non-invasive inflatable wrist cuff.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Digital Blood Pressure Monitor, Model UB-328 (K040229)
Related Devices
K123800 — WRIST AUTOMATIC BLOOD PRESSURE MONITOR · Little Doctor Electronic (Nantong)Co.,Ltd · May 15, 2013
K123600 — DIGITAL BLOOD PRESSURE MONITOR · Little Doctor Electronic (Nantong)Co.,Ltd · May 3, 2013
K133618 — DIGITAL AUTOMATIC WRIST BLOOD PRESSURE MONITOR WBPM22 SERIES · Grandway Technology (Shenzhen) Limited · Mar 5, 2014
K983616 — DAIWA MODEL 180 · Daiwa Products, Inc. · Oct 15, 1999
Submission Summary (Full Text)
{0}------------------------------------------------
EXHIBIT 1
#### 510(k) SUMMARY
This summary of 510(k) and effectiveness information is being submitted in accordance with and requirement of SMDA1990 and 21CFR&807.92.
KO 70825 The assigned 510(k) number is:
#### Submitter's Identification 1.
Little Doctor (Shanghai) Electronic Manufacture Co., Ltd Floor 3dd. 1st Bldg, No 4514 Caoan Road Shanghai China
Phone: +86 021 63056696
Date summary prepared: 2006.9.15
Contact: Mr. David Zhang
#### Name of the Device: 2.
LD8 Digital Wrist Blood Pressure Monitor
#### 3. Common or Usual Name:
Non- Invasive Blood Pressure Measurement System
#### Predicate Device Information 4.
The LD8 Digital Wrist Blood Pressure Monitor is substantially equivalent to the Digital Blood Pressure Monitor, Model UB-328, K040229.
### 5. Device Description:
The LD8 Digital Wrist Blood Pressure Monitor, Model LD8 is designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a non-invasive technique in which an inflatable wrist cuff is wrapped around the wrist. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic semiconductor sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well-known technique in the market called the "oscillometric method".
MAY 25 7000
pg 1 of 3
{1}------------------------------------------------
# 6. Intended Use:
The LD8 Digital Wrist Blood Pressure Monitor is a device intended to measur systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable wrist cuff is wrapped around the wrist.
# 7. Comparison to the Predicate Device:
The LD8 Digital Wrist Blood Pressure and predicate device (Digital Blood Pressure Monitor, Model UB-328, K040229) are identical in functionality and performance with the difference being the external shape of the device, dimensional specification. These differences have no impact on safety or performance of the device. The blood pressure measurement algorithm and its functional technology are identical.
Both the subject and predicate devices use the well-known oscillometric method within the software algorithm to determine the systolic and diastolic blood pressure and pulse rate.
# 8. Discussions of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
Testing information demonstrating safety and effectiveness of the LD8 Digital Wrist Blood Pressure Monitor in the intended environment of use is supported by testing that was conducted in accordance with the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND, which outlines Electrical, Mechanical And Environment Performance Requirement.
The following testing was conducted:
- a. General Functions Test
- b. Reliability Test Operation Conditions
- c. Reliability Test Drop Testing
- d. Reliability Test Storage
- e. Reliability Test Vibration Testing
- EMC Testing f.
- g. IEC 60601-1 Safety Testing
- h. FDA required Unit Intravariability Testing
None of the testing demonstrated any design characteristics that violated the requirement of the Reviewer Guidance or resulted in any safety hazards. It was our conclusion that the LD8 Digital Wrist Blood Pressure Monitor tested met all relevant requirements of the aforementioned tests.
{2}------------------------------------------------
#### 9. Discussion of Clinical Tests Performed:
We have performed clinical testing on the LD8 Digital Wrist Blood Pressure Monitor according to " Clinical Data and Analysis ANSI/AAMI-SP-10 Standard, Section 4.4.2". All testing results met required parameters.
# 10. Conclusion:
We have demonstrated that the LD8 Digital Wrist Blood Pressure Monitor is as safe and effective as the predicate device based on electrical, mechanical and environmental testing results, and SP-10 standard requirements. FDA guidance document requirement were also met. Therefore, our subject device is as safe and effective as our predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 5 2007
Little Doctor (Shanghai) Electronic Manufacture Co. Ltd c/o Mr. David Zhang QA Manager Floor 22, Zhi Yuan Bldg. No 768 Xie Tue Road Luwan District, Shanghai, China
Re: K070825
Trade/Device Name: LD8 Digital Wrist Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Non-Invasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: March 9, 2007 Received: March 26, 2007
Dear Mr. Zhang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. David Zhang
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at 240-276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours,
B.Jimmaster
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Kc70825 pg lof I
Exhibit B
# Indication for use
Page 1 of of 1
770825 510(k) Number (if known): Device Name: LD8 Digital Wrist Blood Pressure Monitor Indication For Use:
The Shanghai Little Doctor LD8 Digital Wrist Blood Pressure Monitor is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist.
escription Use
21 CFR 801 Subpart D)
Over-The Counter Use _________________________________________________________________________________________________________________________________________________________
OR
(21 CFR 807 Subpart C)
SE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE DED)
oncurrence of CDRH, Office of Device Evaluation (ODE)
ision-Off
Cardiovascular Devices
Number 6082.5
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.