MICROLIFE AUTOMATIC BLOOD PRESSURE MONITOR, MODEL BP3AC1-1PC

K060686 · Microlife Intellectual Property GmbH · DXN · Jun 9, 2006 · Cardiovascular

Device Facts

Record IDK060686
Device NameMICROLIFE AUTOMATIC BLOOD PRESSURE MONITOR, MODEL BP3AC1-1PC
ApplicantMicrolife Intellectual Property GmbH
Product CodeDXN · Cardiovascular
Decision DateJun 9, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1130
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The Microlife Automatic Blood Pressure Monitor, Model BP 3AC1-1 PC, is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive MAM (Microlife Average Mode) technique in which an inflatable cuff is wrapped around the upper arm. The device detects the appearance of irregular heartbeat during measurement, and gives a warning signal with the reading once the irregular heartbeat is detected.

Device Story

Device measures systolic/diastolic blood pressure and pulse rate using oscillometric method; electronic capacitive pressure sensor converts cuff pressure changes to electrical signals. Features MAM (Microlife Average Mode) technique, automatically repeating 3 measurement cycles and calculating average. Detects irregular heartbeats, providing warning signal. Used in home or clinical settings by adults. Includes PC-link functionality via USB cable to transfer memory data to Microlife Blood Pressure Analyzer (BPA) software for display and storage. Healthcare providers or patients use output to monitor blood pressure trends; aids in hypertension management.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing including reliability (storage, operation, vibration, drop, life), EMC testing, and software validation (BPA version 3.1.6).

Technological Characteristics

Non-invasive oscillometric blood pressure monitor; electronic capacitive pressure sensor; upper arm cuff; USB connectivity for PC data transfer; software-based MAM (Average Mode) algorithm; factory-set bleed valve for deflation. Complies with ANSI/AAMI SP10-2002.

Indications for Use

Indicated for adult individuals to measure systolic/diastolic blood pressure and pulse rate via non-invasive upper arm cuff. Detects irregular heartbeats.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(K) SUMMARY This summary of 5l0(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. The assigned 5l0(k) number is: K ص8666068 ### Submitter's Identification: 1. JUN - 9 2006 Microlife Intellectual Property GmbH, Switzerland Max Schmidheiny-Strasse 201 9435 Heerbrugg / Switzerland Date Summary Prepared: January 11, 2006 ### 2. Name of the Device: Microlife Automatic Blood Pressure Monitor, Model BP3AC1-1 PC. ### 3. Information for the 510(k) Cleared Device (Predicate Device): Microlife Automatic Blood Pressure Monitor, Model BP3BT0-AP, K#041411. #### 4. Device Description: Microlife Automatic Blood Pressure Monitor, Model BP3AC1-1 PC, is designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a non-invasive MAM (Microlife Average Mode) technique in which an inflatable cuff is wrapped around the Upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic capacitive pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well - known technique in the market called the "oscillometric method". The device detects the appearance of irregular heartbeat during measurement, and the arrhythmia symbol " device can be used in connection with your personal computer (PC) running the Microlife Blood Pressure Analyzer (BPA) software. The memory data can be transferred to the PC by connecting the monitor via cable with the PC. Exhibit #1-1 page 1 of 4 {1}------------------------------------------------ ### 5. Intended Use: The Microlife Automatic Blood Pressure Monitor, Model BP 3AC1-1 PC, is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive MAM (Microlife Average Mode) technique in which an inflatable cuff is wrapped around the upper arm. The device detects the appearance of irregular heartbeat during measurement, and gives a warning signal with the reading once the irregular heartbeat is detected. ### Comparison to the 510(k) Cleared Device (Predicate Device): 6. Both devices use the well-known oscillometric method within the software algorithm to determine the systolic and diastolic blood pressure and pulse rate. An upper arm cuff is inflated automatically; deflate rate is controlled by a factory set bleed valve and the deflation pressures are transferred via tubing to a sensor in both units. Moreover, both devices have irregular heart beat detection function. The differences between BP3AC1-1 PC and the predicate device are described as the following: # 1.1. Average Mode function (MAM function): BP3AC1-1 PC contains a switchable average mode in which the device automatically repeats 3 individual measurements cycles. After that, the average of these 3 individual measurements is calculated and also be shown on the display. ## 1.2. PC-link function: BP3AC1-1 PC can be used in connection with your personal computer (PC) running the Microlife Blood Pressure Analyzer (BPA) software. The memory data can be transferred to the PC by connecting the blood pressure monitor to personal computer via USB cable. All the memory data can be transferred to the connected computer through USB cable and be shown on the computer Exhibit #1-2 Page 2 of 4 {2}------------------------------------------------ monitor. After transferred to computer, the memory data can then be saved in the personal computer memory. ### Discussion of Non-Clinical Tests Performed for Determination of 7. Substantial Equivalence are as follows: Testing information demonstrating safety and effectiveness of the Microlife Automatic Blood Pressure Monitor, Model BP3AC1-1 PC in the intended environment of use is supported by testing that was conducted in accordance with the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND, which outlines Electrical, Mechanical and Environmental Performance Requirements. The following tests were conducted accordingly: - a. Reliability Test - Storage test - Reliability Test Operation test b. - Reliability Test Vibration test C. - d. Reliability Test - Drop test - Reliability Test Life test e. - f. EMC Test - PC-link software BPA version 3.1.6 test report ದ. None of the tests demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards. It is then concluded that the Microlife Automatic Blood Pressure Monitor, Model BP3AC1-1 PC, meets all relevant requirements of the aforementioned tests. ### 8. Discussion of Clinical Tests Performed: Clinical performance of the modified device will remain unchanged; therefore anther clinical test for the modified device, BP3AC1-1PC is not required. ### 9. Conclusions: We have demonstrated that the Microlife Automatic Blood Pressure Monitor, Model BP3AC1-1 PC, is as safe and effective as the predicate device, the Microlife Automatic Blood Pressure Monitor, Model BP 3BT0-AP based on electrical, Exhibit #1-3 page 3 of 4 {3}------------------------------------------------ mechanical and environmental testing results as well as the FDA DCRND November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", and, the ANSI/AAMI Voluntary Standard, SP10-2002 testing results. page 404 Exhibit #1-4 and and the country of the county of : . : {4}------------------------------------------------ Image /page/4/Picture/2 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all capital letters and is evenly spaced around the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN - 9 2006 Microlife Intellectual Property c/o Mr. Tzu-Wei Li Center for Measurement Standards/ Industrial Technology Research Institute Bldg 16, 321 Kuang Fu Rd. Sec. 2 Hsinchu, Taiwan 30042 R.O.C. R.O.C. Re: K060686 Trade/ Name: Microlife Automatic Blood Pressure Monitor, Mode BP3AC1-PC Regulation Number: 21 CFR 870.1130 Regulation Name: Blood Pressure Measurement System, Non-invasive Regulatory Class: II (two) Product Code: DXN Dated: March 6 and May 17, 2006 Received: March 15 and May 23, 2006 Dear Mr. Li: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ ## Page 2 - Mr. Tzu-Wei Li Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) rms leket notification. The FDA finding of substantial equivalence of your device to a legally premated predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at 240-276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, B. Gimmerman for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## Indications For Use Form Page 1 of 1 510(k) Number (if known):_ Device Name: Microlife Automatic Blood Pressure Monitor , Model BP3AC1-1PC Indications for Use: The Microlife Automatic Blood Pressure Monitor, Model BP 3 AC1-1 P C i s a d evice intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive MAM (Microlife Average Mode) technique in which an inflatable cuff is wrapped around the upper arm. The device detects the appearance of irregular heartbeat during measurement, and gives a warning signal with the reading once the irregular heartbeat is detected. Prescription Use (Part 21 CFR 801 Subpart D) AND/ OR Over-The-Counter Use _ X (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) B.B. Commemor
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...