CONVEY 7F GUIDING CATHETER, CONVEY 8F GUIDING CATHETER

K132197 · Pendracare International, B.V. · DQY · Aug 12, 2013 · Cardiovascular

Device Facts

Record IDK132197
Device NameCONVEY 7F GUIDING CATHETER, CONVEY 8F GUIDING CATHETER
ApplicantPendracare International, B.V.
Product CodeDQY · Cardiovascular
Decision DateAug 12, 2013
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2

Indications for Use

The Convey Guiding Catheter is designed to provide a pathway through which therapeutic and diagnostic devices are introduced. The Convey Guiding Catheter is intended to be used in the coronary or peripheral vascular system.

Device Story

Flexible plastic catheter; features luer hub, strain relief, body, intermediate tip, and soft tip. Body/intermediate tip composed of inner liner and outer jacket reinforced with stainless steel braid. Distal section includes various preformed shapes (e.g., Judkins, Amplatz, hockey stick) for target vessel placement; some models include side holes for perfusion. Used in coronary or peripheral vascular procedures; operated by physicians. Provides pathway for percutaneous, transluminal passage of guidewires and interventional devices. Single-use; discarded after procedure. Benefits patient by facilitating access to target vessels for diagnostic or therapeutic interventions.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical design verification (bench) testing, including visual inspection, shape retention, dimensional analysis, kink resistance, radial stiffness, bending stiffness, pull force, coating integrity, burst pressure, particulate testing, and radiopacity. Biocompatibility testing leveraged from predicate devices per ISO 10993 standards.

Technological Characteristics

Flexible plastic catheter; inner liner/outer jacket reinforced with stainless steel braid. Preformed distal shapes. Single-use. Sterilized via Ethylene Oxide. Materials biocompatibility per ISO 10993 (haemocompatibility, cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity).

Indications for Use

Indicated for use in the coronary or peripheral vascular system to provide a pathway for the introduction of therapeutic and diagnostic devices.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) Summary ## per 21 CFR §807.92 | Submitter's<br>Name<br>Address | PendraCare International B.V.<br>Van der Waalspark 22<br>9351 VC Leek<br>The Netherlands | AUG 1 2 2013 | |--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------| | | FDA Establishment Registration Number 3005635992<br>Owner / Operator Number: 10041705 | | | Contact Name &<br>Information | Bert Roossien, MSc, PhD.<br>Manager Quality Assurance & Regulatory Affairs<br>PendraCare International B.V.<br>Van der Waalspark 22<br>9351 VC Leek<br>The Netherlands | | | | Phone: +31 594 554 077 (direct)<br>Phone: +31 594 510 755 (general)<br>Fax: +31 594 510 769<br>e-mail: broossien@pendracare.com | | | Date prepared | 11 July, 2013 | | | Proprietary Name | Convey™ Guiding Catheter | | | Common Name | Percutaneous Catheter | | | Classification | Class II, 21 CFR 870.1250 | | | Review Panel | Cardiovascular | | | Product Code | DQY | | | Subject Devices | Convey™ 7F Guiding Catheter<br>Convey™ 8F Guiding Catheter | | | Predicate Devices | Convey™ 5F Guiding Catheter (K120585; August 8, 2012)<br>Convey™ 6F Guiding Catheter (K120585; August 8, 2012) | | {1}------------------------------------------------ The guiding catheter is a flexible plastic tube featuring a luer hub, a strain Device relief, a body, an intermediate tip, and a soft tip. The body and the Description intermediate tip exist of an inner liner (basecoat) and an outer jacket (topcoat) reinforced with a tightly wound stainless steel braid wire in between the layers. The central lumen of the catheter is used for the percutaneous, transluminal passage and placement of guidewires, diagnostic and / or therapeutic (interventional) devices within the vascular system. The distal section of the catheter has a variety of preformed shapes (e.g., Judkins Left (abbreviated as JL), Judkins Right (JR), Amplatz (AL), Multipurpose, hockey stick) to facilitate placement of the catheter tip in the desired target vessel. Some catheter models feature two (2) small "inline" side holes in the intermediate tip section to maintain perfusion of the target vessel. This device is a single-use device (i.e., single patient, single procedure, single purpose use). After finalizing the procedure, the catheter is withdrawn, removed and discarded. The Convey Guiding Catheter is designed to provide a pathway through Intended Use of which therapeutic and diagnostic devices are introduced. The Convey Device Guiding Catheter is intended to be used in the coronary or peripheral vascular system. The Convey Guiding Catheter is designed to provide a pathway through Indications for which therapeutic and diagnostic devices are introduced. The Convey Use Guiding Catheter is intended to be used in the coronary or peripheral vascular system. The Convey™ 7F Guiding Catheter and Convey™ 8F Guiding Catheter Substantial incorporate substantially equivalent device materials and design, Equivalence packaging materials and design, intended use, fundamental technology (operating principle & mechanism of action), labeling and manufacturing processes and sterilization process as those featured in the legally marketed predicate devices, the Convey™ 5F Guiding Catheter and Convey™ 6F Guiding Catheter (K120585, clearance: August 8, 2012). {2}------------------------------------------------ | Summary of<br>Non-clinical<br>Testing | To support a determination of substantial equivalence, data was collected<br>from non-clinical design verification (bench) tests. The results provide<br>reasonable assurance that the proposed device has been designed and<br>tested to assure conformance to the requirements for its intended use. | | |-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | | The following in-vitro performance tests (following simulated use, if<br>applicable, were completed of the 7F & 8F Convey Guiding Catheter:<br>Visual Inspection,<br>●<br>Shape Conformance & Shape Retention,<br>●<br>Outer Diameter,<br>�<br>Inner Diameter,<br>Usable Catheter Length,<br>●<br>Side Hole diameter & positioning,<br>�<br>C-Kink (Bending Kink Diameter),<br>�<br>Euler Kink (Axial Kink Displacement),<br>●<br>Radial Stiffness (Collapse),<br>●<br>Three Point Bending Test (Bending Stiffness Body),<br>●<br>Pull Force Test (after simulated use),<br>Coating location & length<br>●<br>Coating Integrity - Outer Friction & Wear functional test,<br>Coating Integrity (after simulated use),<br>Torque Strength (after simulated use),<br>�<br>Body Strength / Burst Pressure test,<br>Particulate testing,<br>Radiopacity / Visibility testing.<br>� | | | Summary of<br>Sterility Testing | Bacterial Endotoxin-Mediated Pyrogenicity Testing,<br>Bioburden testing,<br>●<br>Ethylene Oxide Residual Testing.<br>● | | | Summary of<br>Biocompatibility<br>Testing | The Biocompatibility for the 7F & 8F Convey Guiding Catheter was<br>leveraged from the biocompatibility tests completed on the 5F & 6F<br>Convey™ Guiding Catheter. The following tests were conducted:<br>ISO 10993-4: Haemocompatibility:<br>�<br>o Hemolysis<br>o In vitro Haemocompatibility<br>Coagulation Tests Prothrombin Time Assay-PT<br>o Coagulation Unactivated Partial Thromboplastin Time Assay-UP<br>ISO 10993-5: Cytotoxicity - MEM-elution (USP<87>)<br>ISO 10993-7: Ethylene Oxide Sterilization Residuals<br>●<br>ISO 10993-10: Sensitization (USP<1184>)<br>●<br>ISO 10993-10: Irritation / Intracutaneous Reactivity (USP <88>)<br>●<br>ISO 10993-11: Acute Systemic Toxicity (USP <88>)<br>●<br>ISO 10993-11: Material Mediated Pyrogenicity (USP <151>)<br>◆<br>USP <661>: Packaging Plastic Containers Leachables | | {3}------------------------------------------------ . | Summary of<br>Clinical Testing | Since the modifications do not have an impact on the indications for use<br>and on the safety and performance of the device, a Clinical Evaluation<br>Study was not regarded for these devices. | |--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Conclusion | In summary, the modifications in the subject devices Convey™ 7F<br>Guiding Catheter and Convey™ 8F Guiding Catheter do not affect the<br>intended use and does not alter the fundamental scientific technology of<br>the device. The subject devices Convey™ 7F Guiding Catheter and<br>Convey™ 8F Guiding Catheter are substantial equivalent to the predicate<br>devices Convey™ 5F Guiding Catheter (K120585, dated 8 August 2012)<br>and Convey™ 6F Guiding Catheter (K120585, dated 8 August 2012) with<br>respect to the intended use, fundamental technology, i.e., operating<br>principle, device materials, design, packaging materials & configuration,<br>labeling, manufacturing & sterilization processes and sterility assurance<br>level. | {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features the department's emblem, which is a stylized representation of a human figure embracing a globe. The emblem is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the emblem. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 12, 2013 Pendracare International, B.V. % Mr. Tom Nolan Managing Director Corvitex Corporation 7205 Laketree Dr Raleigh, NC 27615 K132197 Re: Trade/Device Name: Convey™ 7F Guiding Catheter, Convey™ 8F Guiding Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: July 11, 2013 Received: July 16, 2013 Dear Mr. Nolan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. 1f your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {5}------------------------------------------------ Page 2- Mr. Tom Nolan forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, M.A. Hillerman for for Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## Indications for Use Statement 510(k) Number (if known): K132197 Convey™ 7F Guiding Catheter, Device Name: Convey™ 8F Guiding Catheter Indications for Use: The Convey Guiding Catheter is designed to provide a pathway through which therapeutic and diagnostic devices are introduced. The Convey Guiding Catheter is intended to be used in the coronary or peripheral vascular system. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) M.A. Hillemann (Division Sign-Off) Division of Cardiovascular Devices K132197 510(k) Number: _______________________________________________________________________________________________________________________________________________________________ .
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...